This work presents the prototype design and the status of the project.
Next activities are oriented to realize a Digital Pulse Processor readout electronics implemented in such a way that, in the future, it can be built with space-qualified components. The system shall be able to manage flux variations of several orders of magnitude to deal with the extreme Sun conditions: from quiet to the most energetic class-X flares. Beside this, an activity to identify which are the possible signatures of solar events is on-going.
This paper reports the current and the planned activities to implement the sensor’s readout functions in an FPGA-based space-ready electronics.
Metis features two channels to image the solar corona in two different spectral bands: in the HI Lyman ∝ at 121.6 nm, and in the polarized visible light band (580 – 640 nm). Metis is a solar coronagraph adopting an “inverted occulted” configuration. The inverted external occulter (IEO) is a circular aperture followed by a spherical mirror which back rejects the disk light. The reflected disk light exits the instrument through the IEO aperture itself, while the passing coronal light is collected by the Metis telescope. Common to both channels, the Gregorian on-axis telescope is centrally occulted and both the primary and the secondary mirror have annular shape.
Classic alignment methods adopted for on-axis telescope cannot be used, since the on-axis field is not available. A novel and ad hoc alignment set-up has been developed for the telescope alignment.
An auxiliary visible optical ground support equipment source has been conceived for the telescope alignment. It is made up by four collimated beams inclined and dimensioned to illuminate different sections of the annular primary mirror without being vignetted by other optical or mechanical elements of the instrument.
The stray light calibration was performed in a clean environment in front of the OPSys solar disk divergence simulator (at ALTEC, in Torino, Italy), which is able to emulate different heliocentric distances. Ground calibrations were a unique opportunity to map the Metis stray light level thanks to a pure solar disk simulator without the solar corona. The stray light calibration was limited to the visible light case, being the most stringent. This work is focused on the description of the laboratory facility that was used to perform the stray light calibration and on the calibration results.
The entire alignment and verification phase has been performed by the Metis team in collaboration with Thales Alenia Space Torino and took place in ALTEC (Turin) at the Optical Payload System Facility using the Space Optics Calibration Chamber infrastructure, a vacuum chamber especially built and tested for the alignment and calibration of the Metis coronagraph, and suitable for tests of future payloads.
The goal of the alignment, integration, verification and calibration processes is to measure the parameters of the telescope, and the characteristics of the two Metis channels: visible and ultraviolet. They work in parallel thanks to the peculiar optical layout. The focusing and alignment performance of the two channels must be well understood, and the results need to be easily compared to the requirements. For this, a dedicated illumination method, with both channels fed by the same source, has been developed; and a procedure to perform a simultaneous through focus analysis has been adopted.
In this paper the final optical performance achieved by Metis is reported and commented.
This paper provides a description of the overall manufactured system and its performance and shows the additional resources available at the XUVLab laboratory in Florence that make SCOUT exploitable by whatever compact (within 1 m) optical experiment that investigates the UV band of the spectrum.
ASPIICS is distributed on the two PROBA 3 spacecrafts (S/C) separated by 150 m. The coronagraph optical assembly is hosted by the “coronagraph S/C” protected from direct solar disk light by the occulting disk on the “occulter S/C”.
The most critical issue in the design of a solar coronagraph is the reduction of the stray light due to the diffraction and scattering of the solar disk light by the occulter, the aperture and the optics. In the present article, we deal with two of these issues:
- The analysis of the stray light inside the telescope.
- The optimization of the external occulter edge, in order to eliminate the Poisson spot behind the occulter and to lower the stray light level going through the entrance pupil of the telescope.
This work was performed in the framework of the ESA STARTIGER program which took place at the Laboratoire d’Astrophysique de Marseille (LAM) during a 6-month period from September 2009 to March 2010.
In general, it is a very complicated task to combine the above two stray light issues together in the simulation and design phase as it requires to consider the propagation inside the telescope of the light diffracted by the external occulter. Actually, the present literature only reports diffraction calculations performed for simple occulting systems (i.e., two disks and serrated disk). A more pragmatic approach, also driven by the tight schedule of the STARTIGER program, is to separate the two contributions, and perform two different stray light analyses. This paper is dedicated to the description of both analyses: in particular, the first part is dedicated to the evaluation of the stray light inside the telescope, assuming a simple disk as occulter, and a preliminary baffle design is presented; the second part describes the investigation on the geometry of the external occulter, with a detailed description of the laboratory setup that has been designed and implemented to compare together several types of occulting systems.
The solar corona will be observed thanks to the presence on the first satellite, facing the Sun, of an external occulter producing an artificial eclipse of the Sun disk. The second satellite will carry on the coronagraph telescope and the digital camera system in order to perform imaging of the inner part of the corona in visible polarized light, from 1.08 R⦿ up to about 3 R⦿.
One of the main metrological subsystems used to control and to maintain the relative (i.e. between the two satellites) and absolute (i.e. with respect to the Sun) FF attitude is the Shadow Position Sensor (SPS) assembly. It is composed of eight micro arrays of silicon photomultipliers (SiPMs) able to measure with the required sensitivity and dynamic range the penumbral light intensity on the Coronagraph entrance pupil.
In the following of the present paper we describe the overall SPS subsystem and its readout electronics with respect to the capability to satisfy the mission requirements, from the light conversion process on board the silicon-based SPS devices up to the digital signal readout and sampling.