PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.
Commonly, microfluidic devices are based on the movement of fluids. For molecular diagnostics assays which often include steps like PCR, this practically always involves a more or less complicated set of external pumps, valves and liquid controls. In the presented paper, we follow a different approach in which the fluid after sample introduction remains stationary and the main bioactive sample molecules are moved through a chain of reaction compartments which contain the different reagents necessary for the assay. The big advantage of this concept is the lack of any external fluid actuation/control. Results on sample carry-over experiments and complete assays will be given.