25 March 1976 Intensifying Screen Speed . . . Its Impact On A Cardiovascular System
Carl C. Scheid
Author Affiliations +
Proceedings Volume 0070, Application of Optical Instrumentation in Medicine IV; (1976) https://doi.org/10.1117/12.954601
Event: Application of Optical Instrumentation in Medicine, IV, 1975, Atlanta, United States
Abstract
Radiographic screens with improved phosphors were evaluated in terms of their impact on image quality and system design. Using tungstate screens as a reference it was shown that system performance can be improved by reducing one or more of the following parameters: MAS, KVP, Focal Spot Size, and Exposure Time. In the case of cardiovascular angiography for instance, maximum benefit was obtained by reductions in exposure time and focal spot size.
© (1976) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Carl C. Scheid "Intensifying Screen Speed . . . Its Impact On A Cardiovascular System", Proc. SPIE 0070, Application of Optical Instrumentation in Medicine IV, (25 March 1976); https://doi.org/10.1117/12.954601
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Absorption

Lanthanum

Image quality

Gadolinium

Calcium

Spectral resolution

Aluminum

RELATED CONTENT

Mixed garnet laser crystals for water vapour DIAL transmitter
Proceedings of SPIE (November 21 2017)
Dual Energy Imaging In Projection Radiography
Proceedings of SPIE (June 12 1986)
Considerations For Selecting Screen - Film Systems
Proceedings of SPIE (March 25 1976)
Selection Of Screen-Film CombinationS
Proceedings of SPIE (July 06 1979)
Large Area Contrast Prediction In Screen-Film Systems
Proceedings of SPIE (August 18 1980)
X-Ray Automatic Exposure Control
Proceedings of SPIE (March 25 1976)

Back to Top