6 July 1979 Selection Of Screen-Film Combinations
K. W. Taylor, B. P. McLeish, B. B. Hobbs, T. J. Hughes
Author Affiliations +
Proceedings Volume 0173, Application of Optical Instrumentation in Medicine VII; (1979) https://doi.org/10.1117/12.957157
Event: Application of Optical Instrumentation in Medicine VII, 1979, Toronto, Canada
Abstract
The speeds of screen-film combinations available in Canada have been measured at 75 kVp with appropriately filtered beams. The influence on speeds of kVp, attenuation of the front of the cassette, age and use of screens, reciprocity failure and film variations from batch to batch is surveyed. The significance of the observed variations in speed (in selecting screen-film combinations) is considered. A set of radiographs of head and body phantoms has been used to examine the criteria which radiologists use in defining a 'good' radiograph and from this a qualitative assessment has been made of the priorities in selection of screen-film combinations.
© (1979) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
K. W. Taylor, B. P. McLeish, B. B. Hobbs, and T. J. Hughes "Selection Of Screen-Film Combinations", Proc. SPIE 0173, Application of Optical Instrumentation in Medicine VII, (6 July 1979); https://doi.org/10.1117/12.957157
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Radiography

Image quality

X-rays

Signal attenuation

Medicine

Optical instrument design

Calcium

RELATED CONTENT

A New High Heat Load X-Ray Tube
Proceedings of SPIE (June 15 1984)
Quantitative Digital Fluorography
Proceedings of SPIE (December 29 1982)
Quantum Noise Measurements For X-Ray Image Intensifiers
Proceedings of SPIE (July 06 1979)
Exposures To Patients From Radiological Procedures
Proceedings of SPIE (July 06 1979)
Effects Of kVp On DSA Images
Proceedings of SPIE (June 15 1984)

Back to Top