The objective of this study is to design and characterize a piezoelectric composite and evaluate its suitability for
viscosity-measuring applications, i.e., monitoring the coagulation rate of blood. The composite is manufactured of a
platinum-core lead zirconate titanate (PZT) fiber inserted into an aluminum matrix. This study characterizes the
described composite by testing its impedance, capacitance, voltage sensitivity response to vibrational inputs, and
deformation due to electrical input. As actuators, different voltage inputs are fed into the probes and displacement is
measured with results on the range of nanometers. As sensors, the devices are used to monitor cantilever beam
vibrations. The probe's response is in the mV range and follows the same pattern as an accelerometer. Additional
tests in air, water, and deionized water are carried out to evaluate the sensor's suitability for measuring viscosity
using two probes: one as an actuator and the other as a sensor. Results of the gain and phase between the two probes
indicate that the phase shift may be used as an indicator of viscosity changes. The first significant phase shift was
measured as 2.45, 2.77, and 4.065x107Hz, for water, air, and oil, respectively, which is directly proportional to the
kinematic viscosity of each fluid.
Access to the requested content is limited to institutions that have purchased or subscribe to SPIE eBooks.
You are receiving this notice because your organization may not have SPIE eBooks access.*
*Shibboleth/Open Athens users─please
sign in
to access your institution's subscriptions.
To obtain this item, you may purchase the complete book in print or electronic format on
SPIE.org.
INSTITUTIONAL Select your institution to access the SPIE Digital Library.
PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.