The effect of cycling on charge-storage, actuation and sensing behavior of a polypyrrole is studied, having
its application for an electroactive catheter in mind. It is shown that the electrochemical capacitance of a
polypyrrole film decreases by about 15 % over the course of 100 cycles, while the per cycle rate of this
decrease drops by 75 % between the first and the last ten cycles, implying that a steady-state value may
exist. The decrease in capacitance is shown to have a significant effect on actuation strain. In order to
achieve a more constant capacitance and more robust actuation performance, it is proposed to pre-cycle the
potential of the film to exhaust the effect of processes that contribute to the decrease in capacitance and
allow it to reach a more constant value. The ability of a polypyrrole film to generate currents corresponding
to applied external load during actuation is verified and the cycle life time of such a sensor is studied. It is
shown that after an initial decrease, the sensor current reaches a steady-state value as well, and maintains
that value at least over 5600 cycles.
Access to the requested content is limited to institutions that have purchased or subscribe to SPIE eBooks.
You are receiving this notice because your organization may not have SPIE eBooks access.*
*Shibboleth/Open Athens users─please
sign in
to access your institution's subscriptions.
To obtain this item, you may purchase the complete book in print or electronic format on
SPIE.org.
INSTITUTIONAL Select your institution to access the SPIE Digital Library.
PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.