1 March 1994 Scanned-fringe, spatial-harmonic-distortion test for detector nonlinearity
Alberto J. Varela, Glenn D. Boreman
Author Affiliations +
Abstract
Results are presented for a modification of the spatial-harmonic-distortion test, suitable for nonlinearity measurement of single-element detectors, using a scanned-fringe technique and an electronic spectrum analyzer. Detector nonlinearity of less than 1% was measured at a wavelength of 10.6 μm. Previously, the smallest measured nonlinearity for the spatial-harmonic-distortion test was of the order of 3%, using a stationary-fringe technique and an 8-bit frame grabber.
Alberto J. Varela and Glenn D. Boreman "Scanned-fringe, spatial-harmonic-distortion test for detector nonlinearity," Optical Engineering 33(3), (1 March 1994). https://doi.org/10.1117/12.159335
Published: 1 March 1994
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Pyroelectric detectors

Information operations

Spectrum analysis

Mirrors

Carbon dioxide lasers

Frame grabbers