Significance: The critical closing pressure (CrCP) of cerebral circulation, as measured by diffuse correlation spectroscopy (DCS), is a promising biomarker of intracranial hypertension. However, CrCP techniques using DCS have not been assessed in gold standard experiments. Aim: CrCP is typically calculated by examining the variation of cerebral blood flow (CBF) during the cardiac cycle (with normal sinus rhythm). We compare this typical CrCP measurement with a gold standard obtained during the drops in arterial blood pressure (ABP) caused by rapid ventricular pacing (RVP) in patients undergoing invasive electrophysiologic procedures. Approach: Adults receiving electrophysiology procedures with planned ablation were enrolled for DCS CBF monitoring. CrCP was calculated from CBF and ABP data by three methods: (1) linear extrapolation of data during RVP (CrCPRVP; the gold standard); (2) linear extrapolation of data during regular heartbeats (CrCPLinear); and (3) fundamental harmonic Fourier filtering of data during regular heartbeats (CrCPFourier). Results: CBF monitoring was performed prior to and during 55 episodes of RVP in five adults. CrCPRVP and CrCPFourier demonstrated agreement (R = 0.66, slope = 1.05 (95%CI, 0.72 to 1.38). Agreement between CrCPRVP and CrCPLinear was worse; CrCPLinear was 8.2 ± 5.9 mmHg higher than CrCPRVP (mean ± SD; p < 0.001). Conclusions: Our results suggest that DCS-measured CrCP can be accurately acquired during normal sinus rhythm. |
1.IntroductionAcute brain injury is a leading cause of death and disability in children and adults.1–3 Intracranial hypertension often follows acute injury and is a major cause of secondary brain injury.4–7 The gold standard for its detection is invasive monitoring of intracranial pressure (ICP).4,8,9 Invasive ICP monitors, however, carry risks of intracranial hemorrhage and infection, and they are not always readily available.9,10 Because invasive cerebral monitoring is restricted to critically ill patients and excludes patients on anticoagulation, non-invasive measurement of ICP could extend improved diagnosis and prognosis to a broader population. Numerous techniques have been proposed to this end; they are broadly categorized as electrophysiologic (e.g., electroencephalogram power spectrum analysis), ophthalmic (e.g., optical coherence tomography of the optic nerve), otic (e.g., otoacoustic emissions), and fluid dynamic [e.g., analysis of transcranial Dopper, diffuse correlation spectroscopy (DCS), and near-infrared (NIR) spectroscopy perfusion signals].11–19 Some of these techniques use non-invasive critical closing pressure (CrCP) measurements of the cerebral circulation as a biomarker of intracranial hypertension and ICP.14,18–23 CrCP is the isotropic pressure compressing the cerebral arterioles;23,24 it depends on both ICP and vasomotor tone. To date, most studies of CrCP employ transcranial Doppler ultrasound (TCD) to determine cardiac blood velocity waveforms in major arteries such as the middle cerebral artery.14,24,25 The optical technique of DCS was recently demonstrated as an alternative method for estimating CrCP based on the measurement of arteriolar cerebral blood flow (CBF) through the cardiac cycle.19,23 Importantly, DCS is well-suited for prolonged monitoring at the bedside.26,27 The difference of mean arterial pressure and CrCP is the cerebral perfusion pressure that drives CBF.24,28 Thus, a gold standard procedure for measurement of CrCP would rapidly decrease arterial blood pressure (ABP) to levels for which CBF is zero. This approach is not possible for routine clinical use, but such a scenario does occur during rapid ventricular pacing (RVP) of patients during electrophysiology studies wherein an induced ventricular arrhythmia is performed to identify cardiac tissue causing heart rhythm abnormalities; in this case, CBF approaches zero. This gold standard method of determining CrCP is similar to the method used in a prior TCD study to validate their CrCP measurements.29 Herein, using patients undergoing electrophysiology studies, we compared DCS CrCP measurements acquired during normal sinus rhythm (the typical method) against measurements acquired during RVP (the gold standard). During sinus rhythm, CrCP was obtained from (a) linear extrapolation of CBF and ABP data and (b) fundamental-frequency harmonic Fourier filtering of CBF and ABP data. Based on the previous TCD study,29 we hypothesize that the CrCP measures obtained during regular heartbeats would correlate with CrCP obtained during RVP and that the use of Fourier filtering would improve the agreement. 2.Methods2.1.Experimental ProceduresSeven adult patients undergoing electrophysiology procedures with planned cardiac ablation were enrolled in our study. All subjects provided written consent, and all protocols/procedures were approved by the institutional review board of the University of Pennsylvania, which adheres to the guidelines of the Common Rule and the Food and Drug Administration’s Institutional Review Board and human subject regulations. Electrophysiology studies are performed on patients with ventricular arrhythmia to minimally invasively identify and eliminate tissue in the heart that is causing abnormalities in heart rhythm. During the procedure, at least one catheter is placed into the heart; as part of the procedure, small electric impulses are used to probe the heart wall for abnormalities. During ventricular arrhythmia caused by RVP, cardiac output falls, and ABP approaches CrCP. At the time of the electrophysiology study, after induction of anesthesia, a DCS optical probe was placed on the forehead to monitor CBF in brain supplied by the anterior middle cerebral artery for the entire cardiac ablation procedure [Fig. 1(a)]. CBF was measured continuously at 20 Hz using a custom-built DCS instrument with a software correlator that is described elsewhere.30,31 Specifically, DCS intensity autocorrelation measurements at 2.5 cm source-detector separation and 785 nm wavelength were averaged across four detection channels. Using standard techniques,32 CBF was obtained from a semi-infinite fit to the autocorrelation measurement using an assumed tissue optical absorption coefficient of and reduced scattering coefficient of . Fractional blood flow changes obtained with DCS, which are used to calculate CrCP (Sec. 2.2), are robust to errors in these assumed tissue optical properties.33 ABP was simultaneously measured by a radial artery catheter attached to a FloTrac sensor (Edwards Lifesciences). The FloTrac sensor provides two separate analog outputs of the same ABP signal. One output was connected to the standard vital signs monitor used by the clinical team, while the other output was connected to an analog input of our DCS instrument’s DAQ board (PCIe6323, National Instruments). The ABP and CBF measurements were synchronized and recorded at 20 Hz sampling. 2.2.Data ProcessingThe ABP waveform measurement temporally lagged the CBF waveform measurement because of the differences in distance between the arm arterial tree (where the ABP measurements were made) and heart versus cerebral arterial tree and heart; i.e., it takes longer for ABP cardiac waveforms to reach the arm than the brain. For each patient, we used a cross-correlation technique to measure this lag and align the ABP and CBF time series (xcorr, MATLAB R2018a, Mathworks). The key data of interest for computing CrCP are the continuous high-time-resolution CBF and ABP waveform data (20 Hz sampling) before and during episodes of RVP. We only considered episodes of RVP that resulted in a gap between typical systolic ABP waves of greater than 1.5 s. For each included episode, CrCP was computed during the drop in ABP (the gold standard method). This measurement was compared with calculations of CrCP obtained from data during the 15 s of sinus rhythm data just prior to RVP. ABP and CBF data during an exemplary episode are shown in Fig. 1(b). Our use of a linear resistive model for the cerebral arteriole compartment between the large arteries and capillaries (i.e., the compartment measured by DCS) underlies the computation of CrCP with the gold standard and normal sinus rhythm methods. Specifically, in this model, the in-flow blood pressure at the entrance to the arteriole compartment is ( was assumed34), the out-flow blood pressure at the distal end of the arteriole compartment is CrCP, and the arteriolar resistance is . Then, using an Ohms’ law relation for blood flow, the connection between measured ABP and DCS-CBF at time is given by23 Equation (1) assumes that CrCP and remain constant over the time-scale of the measured fluctuations in ABP and CBF (i.e., CrCP and remain constant during RVP and during the 15 s of sinus rhythm data prior to RVP). For the gold standard method to compute CrCP during RVP (i.e., ), we performed a linear regression of the continuous CBF versus data acquired during the long diastole accompanying RVP [see Fig. 1(c)]; in Fig. 1(c), the continuous synchronized CBF and data during RVP are plotted on the vertical and horizontal axes, respectively. This procedure enabled us to find the extrapolated value of for which CBF is zero; this is the CrCP [Eq. (1)]. We denote CrCP computed this way as (i.e., the -intercept of the linear regression fit). Although is extrapolated, we consider it to be the gold standard measurement because CBF approaches close to zero under these conditions. We used two methods to compute CrCP during the 15 s of regular heartbeat data (normal sinus rhythm) just prior to RVP. In one method, similar to the computation of , we performed a linear regression of CBF versus to find the (extrapolated) blood pressure at which CBF is zero [Fig. 1(d)]. We denote CrCP computed this way as . In the second method, we employed fundamental harmonic Fourier filtering of the 15 s of ABP and CBF data to compute CrCP (i.e., denoted as ). As discussed in detail elsewhere,23,24,29,35 in our vascular model, the ratio of the Fourier amplitudes of the fundamental harmonics of the ABP and CBF waveforms multiplied by , i.e., , is the arteriolar resistance . ( is the heart rate frequency.) Substituting this into Eq. (1), we obtain where the angular brackets, , represent the means across the 15 s of heartbeat data. Exemplary measurements of and are shown in Fig. 1(e). The Fourier amplitudes were obtained from the discrete Fourier transforms of the time-series data (fft, MATLAB R2018a).2.3.Statistical AnalysisSummary statistics are presented using means and standard deviations for , , , mean ABP prior to RVP, DCS detected photon count rate prior to RVP, and the temporal lag of the ABP waveform relative to the CBF waveform. We carried out both linear regression and Bland–Altman analyses to assess agreement between (a) and and (b) and . We further used paired -tests to assess whether differed from and whether differed from . For all statistical tests, a value of was deemed to represent statistical significance. All calculations were carried out using MATLAB R2018a. The use of Fourier filtering requires data spanning multiple cardiac cycles. Therefore, when longer time windows are used to compute the Fourier amplitudes, the frequency resolution for the heart rate is higher. As a check to assess the sensitivity to the 15 s window length, we computed , as described earlier, using different window lengths prior to RVP (i.e., 10, 15, 30, 60, 120, and 180 s). We examined agreement between the computations of with different window lengths using an intraclass correlation coefficient (ICC). ICC was computed based on a single-measurement, absolute-agreement, two-way mixed-effects model.36 In addition, we were interested in whether there was better agreement between the computations of with subsets of longer window lengths. To this end, we used ICCs to assess agreement between the computations of using the following subsets of longer window lengths: (a) 15, 30, 60, 120, and 180 s and (b) 30, 60, 120, and 180 s. In another secondary analysis, we used a Pearson correlation coefficient () to quantify the strength of the linear correlation between and , and we used a paired -test to assess whether differed from . In our final secondary analysis, we quantified the strength of the linear correlations between the CBF and ABP time series during RVP and during the 15 s prior to RVP. Strong linear correlations support the validity of our assumption of constant arteriolar resistance for computing CrCP [Eq. (1)]. For each time-period, we computed the mean (95% CI) of across all RVP episodes. Because Pearson correlation coefficients are not normally distributed and are bound to be between and 1, we first calculated the mean and standard deviation of the mean after transformation of the coefficients using Fisher transforms (i.e., and , where is the Fisher-transformed value, is the number of RVP episodes, and arctanh is the hyperbolic arctangent).37,38 The resulting averaged values for each time period were then transformed back to the correlation space with the hyperbolic tangent (tanh). Specifically, the Pearson correlation mean (95% CI) is . 3.ResultsSeven patients were enrolled (four males and three females), and their average age was years. Fifty-five episodes of RVP across five patients were included. The RVP episodes in the other two enrolled patients were all less than 1.5 s and thus were excluded. Across the 55 episodes, the average mean ABP and DCS detected photon count rate prior to RVP were and , respectively. The ABP waveform lagged the CBF waveform by . The average , , and measurements across the included RVP episodes were , , and , respectively. A linear regression and Bland–Altman analysis showed agreement between and [Figs. 2(a) and 2(b), Table 1]. The two parameters were significantly correlated (, ), the slope and intercept of the linear regression were not significantly different from unity and zero, respectively, and their mean difference was not significantly different from zero. Agreement was worse between and [Figs. 2(c) and 2(d) and Table 1]. Although the two parameters were significantly correlated (, ), overestimated by (). Table 1Linear relationships between the normal sinus rhythm CrCP (i.e., CrCPFourier or CrCPLinear) and gold standard CrCP (i.e., CrCPRVP) measurements and the normal sinus rhythm minus gold standard CrCP difference.
In a secondary analysis, we varied the window length used to compute to values different by 15 s (i.e., 10, 30, 60, 120, and 180 s). The ICC between the six sets of data (i.e., one data set for each window length) was 0.43 (95% CI, 0.23 to 0.61). This poor overall agreement in computed using these different time windows is driven largely by data from one window length, i.e., the 10-s window. The ICC between the five sets of data computed with the longer windows (i.e., 15, 30, 60, 120, and 180 s) was substantially higher, i.e., 0.76 (95% CI, 0.68 to 0.84). This higher ICC was, in turn, not different from the ICC between the four sets of data computed with even longer windows (i.e., 30, 60, 120, and 180 s), i.e., 0.85 (95% CI, 0.81 to 0.89). Together, these results suggest that a 15-s window is sufficiently long to compute in adults and that windows of 10-s or less are too short. In another secondary analysis, a significant linear correlation of between and was observed (), and the difference of (mean ± SD) was greater than zero (). In our third secondary analysis, significant linear correlations between ABP and CBF during RVP and prior to RVP were observed. The mean (95% CI) linear correlation between the ABP and CBF data during RVP was (0.89, 0.92). The corresponding mean linear correlation between the ABP and CBF data acquired during the 15-s interval prior to RVP was (0.78, 0.85). The strong linear correlations support our assumption of constant arteriolar resistance for computing CrCP [see Eq. (1)]. 4.DiscussionThe optical method for deriving CrCP relies on measurements of CBF and ABP variations over a time scale for which arteriolar resistance is constant. Herein, we compared CrCP estimates obtained from two distinct causes of fast CBF/ABP variations: normal sinus rhythm and RVP. We treated the latter estimate () as the gold standard because, during RVP, ABP approaches the true CrCP. Importantly, we found that, during the regular heartbeats, CrCP estimated from fundamental harmonic Fourier filtering (i.e., ) agreed with . Agreement was worse for CrCP estimates based on linear extrapolation during regular heartbeats (i.e., ). The advantages of the Fourier method include its insensitivity to errors in temporal alignment between the CBF and ABP waveforms and its removal of higher order harmonics from the waveforms (in contrast to the linear method where all harmonics are present). During normal sinus rhythm, the CBF and ABP waveforms are repetitive periodic oscillations. Accordingly, the waveforms can be represented by a Fourier series with frequencies (or harmonics) that are integral multiples of the frequency of repetition (i.e., the fundamental harmonic). The fundamental harmonic is the heart rate (). The removal of higher harmonics (i.e., ) is advantageous because vascular compliance effects are more pronounced at higher signal frequencies, i.e., the single-resistor vascular model used to calculate CrCP is less accurate for larger harmonics.19 In addition, both the DCS and TCD CrCP measurement techniques make the approximation that the ABP waveforms in the arm (where the ABP measurements were made) and cerebral arterial trees are equivalent. In reality, there are subtle differences between the shapes of the ABP waveforms in the arm and cerebral arterial trees. These subtle differences primarily affect the higher harmonics in the ABP waveform.39 Thus, fundamental harmonic Fourier filtering renders the CrCP measurement less sensitive to errors that arise from ABP waveform estimation. These reasons could explain why overestimated the gold standard. Accordingly (especially because the gold standard method is not usually available), we recommend use of the fundamental harmonic Fourier filtering method to measure CrCP during routine use. We note that another recently proposed DCS-based method to estimate CrCP has also shown promise. This method, employing an innovative algorithm to resolve pulsatile CBF waveforms at 100 Hz sampling with DCS, used linear extrapolation solely during diastole between regular heartbeats to determine CrCP (i.e., CrCP was the -intercept of the best-fit line to CBF versus ABP data after ABP drops below the dicrotic notch).19 The CBF sampling rate in our study was too low to estimate CrCP in this manner. Future work is needed to assess whether this diastole-only method for calculating CrCP improves agreement with the gold standard CrCP estimate. Our results indicate that DCS CrCP measurements during normal sinus rhythm are accurate in adults with presumed normal ICP levels. Future work is needed, however, to assess the impact of ICP pulsatility on CrCP measurements in adults with intracranial hypertension. Both the DCS and TCD methods for deriving CrCP assume that CrCP is constant on the time scale of CBF and ABP fluctuations [Eq. (1)]. In practice, is greater than zero because of the pulsatility in ICP caused by the transfer of systolic ABP increases into brain tissue.40 From Eq. (1), the arteriolar resistance is more precisely given as . The neglect of results in overestimation of and underestimation of CrCP. At normal ICP levels, typical ICP pulsatility is small (),40 but it can become more substantial during intracranial hypertension.40 On a related note, ICP pulsatility can also induce subtle oscillations in intracranial tissue volume (i.e., the ICP-volume compliance curve),40 which may cause motion artifacts in DCS measurements. The observed agreement between and suggests that these artifacts do not significantly impact measurements at normal ICP levels, but further investigations at higher ICP pulsatility levels are warranted. The DCS CrCP measurement method has notable advantages compared with the TCD method.19,23 DCS is well-suited for prolonged monitoring at the bedside,26,27 less susceptible to the confound of turbulent flow in the vasculature41 (the Reynolds number of flow in arterioles is substantially lower than the Reynolds number in large arteries42), and more sensitive to localized injuries.19,23 One limitation is the need to assume the coefficient in Eq. (1). This accounts for the blood pressure drop across the large cerebral arteries. Our assumed of 0.6 is based on systemic and arteriolar blood pressure measurements in rats.34 If the same is used for every measurement, then the specific assumed will not affect the diagnostic accuracy of detecting intracranial hypertension based on CrCP measurements. It would only alter the specific threshold used for detection. (Note that, in the recent study that introduced the diastole-only method to estimate CrCP from DCS measurements, was used.19) The analysis is more problematic, however, if varies substantially between subjects or within subjects at different timepoints. Future work is needed to investigate the relationship between systemic and arteriolar blood pressure waveforms in humans and large animal models. 5.ConclusionThe observed agreement between DCS measurements of cerebral CrCP acquired prior to and during RVP suggests that CrCP can be accurately measured with DCS during regular heartbeats. Further, agreement was better for CrCP estimated from fundamental harmonic Fourier filtering than from linear extrapolation. To further investigate the use of CrCP as a biomarker for intracranial hypertension, future work is needed to study the relation between systemic and arteriolar blood pressure waveforms and to assess the impact of ICP pulsatility on the CrCP calculation during intracranial hypertension. DisclosuresThe authors disclose partial ownership of the following patents. Pending: WO 2021/091961 [T.S.K., W.B.B., A.G.Y., and D.J.L.], 63/257685 [T.S.K., W.B.B., D.J.L., and R.M.F.]. Granted: US8082015 [A.G.Y.], US10064554 [A.G.Y.], US10342488 [W.B.B. and A.G.Y.], US10827976 [W.B.B., D.J.L., and A.G.Y.]. No author currently receives royalties or payments from these patents. AcknowledgmentsThis work was funded by the National Institutes of Health Grant Nos. R01NS113945 [W.B.B. and D.J.L.], K08NS117897 [B.R.W.], R01NS082309 [W.A.K.], T32GM112596 [E.M.G.], T32HL007915 [T.S.K.], and P41EB015893 [A.G.Y.] and the Children’s Hospital of Philadelphia Frontier Program. We gratefully acknowledge Eileen Maloney-Wilensky and Olivia Amendolia for their regulatory support of the study; Ramani Balu, Lian He, Lin Wang, and Christopher Favilla for useful discussions; and the Electrophysiology Section of the Division of Cardiology at the Hospital of the University of Pennsylvania for supporting data acquisition. Code, Data, and Materials AvailabilityThe data and code presented in this study are available on request from the corresponding author. ReferencesV. L. Feigin et al.,
“Global, regional, and national burden of neurological disorders, 1990-2016: a systematic analysis for the Global Burden of Disease Study 2016,”
Lancet Neurol., 18
(5), 459
–480
(2019). https://doi.org/10.1016/S1474-4422(18)30499-X Google Scholar
C. A. Taylor et al.,
“Traumatic brain injury-related emergency department visits, hospitalizations, and deaths—United States, 2007 and 2013,”
Morb. Mortal. Wkly. Rep., 66
(9), 1
–16
(2017). https://doi.org/10.15585/mmwr.ss6609a1 Google Scholar
M. C. Dewan et al.,
“Epidemiology of global pediatric traumatic brain injury: qualitative review,”
World Neurosurg., 91 497
–509.e1
(2016). https://doi.org/10.1016/j.wneu.2016.03.045 Google Scholar
N. Carney et al.,
“Guidelines for the management of severe traumatic brain injury, fourth edition,”
Neurosurgery, 80
(1), 6
–15
(2017). https://doi.org/10.1227/NEU.0000000000001432 NEQUEB Google Scholar
Jr. E. S. Connolly et al.,
“Guidelines for the management of aneurysmal subarachnoid hemorrhage: a guideline for healthcare professionals from the American Heart Association/American Stroke Association,”
Stroke, 43
(6), 1711
–1737
(2012). https://doi.org/10.1161/STR.0b013e3182587839 SJCCA7 0039-2499 Google Scholar
K. T. Kahle et al.,
“Hydrocephalus in children,”
Lancet, 387
(10020), 788
–799
(2016). https://doi.org/10.1016/S0140-6736(15)60694-8 LANCAO 0140-6736 Google Scholar
M. J. Bell and P. M. Kochanek,
“Pediatric traumatic brain injury in 2012: the year with new guidelines and common data elements,”
Crit. Care Clin., 29
(2), 223
–238
(2013). https://doi.org/10.1016/j.ccc.2012.11.004 CCCLEH 0749-0704 Google Scholar
C. Hawthorne and I. Piper,
“Monitoring of intracranial pressure in patients with traumatic brain injury,”
Front. Neurol., 5 121
(2014). https://doi.org/10.3389/fneur.2014.00121 Google Scholar
J. B. Rosenberg et al.,
“Non-invasive methods of estimating intracranial pressure,”
Neurocrit. Care, 15
(3), 599
–608
(2011). https://doi.org/10.1007/s12028-011-9545-4 Google Scholar
T. Zeng and L. Gao,
“Management of patients with severe traumatic brain injury guided by intraventricular intracranial pressure monitoring: a report of 136 cases,”
Chin. J. Traumatol., 13
(3), 146
–151
(2010). https://doi.org/10.3760/cma.j.issn.1008-1275.2010.03.003 Google Scholar
X. Zhang et al.,
“Invasive and noninvasive means of measuring intracranial pressure: a review,”
Physiol. Meas., 38
(8), R143
–R182
(2017). https://doi.org/10.1088/1361-6579/aa7256 PMEAE3 0967-3334 Google Scholar
L. C. Padayachy,
“Non-invasive intracranial pressure assessment,”
Child’s Nervous Syst.: ChNS, 32
(9), 1587
–1597
(2016). https://doi.org/10.1007/s00381-016-3159-2 CNSYE9 1433-0350 Google Scholar
J. W. Swanson et al.,
“Evaluation of optical coherence tomography to detect elevated intracranial pressure in children,”
JAMA Ophthalmol., 135
(4), 320
–328
(2017). https://doi.org/10.1001/jamaophthalmol.2017.0025 Google Scholar
D. Cardim et al.,
“Non-invasive monitoring of intracranial pressure using transcranial Doppler ultrasonography: is it possible?,”
Neurocrit. Care, 25
(3), 473
–491
(2016). https://doi.org/10.1007/s12028-016-0258-6 Google Scholar
J. B. Fischer et al.,
“Non-invasive estimation of intracranial pressure by diffuse optics: a proof-of-concept study,”
J. Neurotrauma, 37
(23), 2569
–2579
(2020). https://doi.org/10.1089/neu.2019.6965 JNEUE4 0897-7151 Google Scholar
A. Ruesch et al.,
“Estimating intracranial pressure using pulsatile cerebral blood flow measured with diffuse correlation spectroscopy,”
Biomed. Opt. Express, 11
(3), 1462
–1476
(2020). https://doi.org/10.1364/BOE.386612 BOEICL 2156-7085 Google Scholar
A. Ruesch et al.,
“Fluctuations in intracranial pressure can be estimated non-invasively using near-infrared spectroscopy in non-human primates,”
J. Cereb. Blood Flow Metab., 40
(11), 2304
–2314
(2020). https://doi.org/10.1177/0271678X19891359 Google Scholar
T. M. Flanders et al.,
“Optical detection of intracranial pressure and perfusion changes in neonates with hydrocephalus,”
J. Pediatr., 236 54
–61.e1
(2021). https://doi.org/10.1016/j.jpeds.2021.05.024 Google Scholar
K. C. Wu et al.,
“Validation of diffuse correlation spectroscopy measures of critical closing pressure against transcranial Doppler ultrasound in stroke patients,”
J. Biomed. Opt., 26
(3), 036008
(2021). https://doi.org/10.1117/1.JBO.26.3.036008 JBOPFO 1083-3668 Google Scholar
M. Czosnyka et al.,
“Critical closing pressure in cerebrovascular circulation,”
J. Neurol. Neurosurg. Psychiatr., 66
(5), 606
–611
(1999). https://doi.org/10.1136/jnnp.66.5.606 JNNPAU 0022-3050 Google Scholar
D. Cardim et al.,
“Prospective study on noninvasive assessment of intracranial pressure in traumatic brain-injured patients: comparison of four methods,”
J. Neurotrauma, 33
(8), 792
–802
(2016). https://doi.org/10.1089/neu.2015.4134 JNEUE4 0897-7151 Google Scholar
C. Thees et al.,
“Relationship between intracranial pressure and critical closing pressure in patients with neurotrauma,”
Anesthesiology, 96
(3), 595
–599
(2002). https://doi.org/10.1097/00000542-200203000-00014 ANESAV 0003-3022 Google Scholar
W. B. Baker et al.,
“Noninvasive optical monitoring of critical closing pressure and arteriole compliance in human subjects,”
J. Cereb. Blood Flow Metab., 37
(8), 2691
–2705
(2017). https://doi.org/10.1177/0271678X17709166 Google Scholar
R. B. Panerai,
“The critical closing pressure of the cerebral circulation,”
Med. Eng. Phys., 25
(8), 621
–632
(2003). https://doi.org/10.1016/S1350-4533(03)00027-4 MEPHEO 1350-4533 Google Scholar
G. V. Varsos et al.,
“A noninvasive estimation of cerebral perfusion pressure using critical closing pressure,”
J. Neurosurg., 123
(3), 638
–648
(2015). https://doi.org/10.3171/2014.10.JNS14613 JONSAC 0022-3085 Google Scholar
J. Selb et al.,
“Prolonged monitoring of cerebral blood flow and autoregulation with diffuse correlation spectroscopy in neurocritical care patients,”
Neurophotonics, 5
(4), 045005
(2018). https://doi.org/10.1117/1.NPh.5.4.045005 Google Scholar
W. B. Baker et al.,
“Continuous non-invasive optical monitoring of cerebral blood flow and oxidative metabolism after acute brain injury,”
J. Cereb. Blood Flow Metab., 39
(8), 1469
–1485
(2019). https://doi.org/10.1177/0271678X19846657 Google Scholar
R. Aaslid et al.,
“Estimation of cerebral perfusion pressure from arterial blood pressure and transcranial Doppler recordings,”
Intracranial Pressure VI, 226
–229 Springer, Berlin, Heidelberg
(1986). Google Scholar
R. Aaslid et al.,
“Dynamic pressure-flow velocity relationships in the human cerebral circulation,”
Stroke, 34
(7), 1645
–1649
(2003). https://doi.org/10.1161/01.STR.0000077927.63758.B6 SJCCA7 0039-2499 Google Scholar
R. M. Forti et al.,
“Transcranial optical monitoring of cerebral hemodynamics in acute stroke patients during mechanical thrombectomy,”
J. Stroke Cerebrovasc. Dis., 28
(6), 1483
–1494
(2019). https://doi.org/10.1016/j.jstrokecerebrovasdis.2019.03.019 Google Scholar
D. Wang et al.,
“Fast blood flow monitoring in deep tissues with real-time software correlators,”
Biomed. Opt. Express, 7
(3), 776
–797
(2016). https://doi.org/10.1364/BOE.7.000776 BOEICL 2156-7085 Google Scholar
T. Durduran et al.,
“Diffuse optics for tissue monitoring and tomography,”
Rep. Progr. Phys. Phys. Soc., 73
(7), 076701
(2010). https://doi.org/10.1088/0034-4885/73/7/076701 Google Scholar
W. B. Baker et al.,
“Modified Beer-Lambert law for blood flow,”
Biomed. Opt. Express, 5
(11), 4053
–4075
(2014). https://doi.org/10.1364/BOE.5.004053 BOEICL 2156-7085 Google Scholar
G. L. Baumbach and D. D. Heistad,
“Remodeling of cerebral arterioles in chronic hypertension,”
Hypertension, 13
(6 Pt 2), 968
–972
(1989). https://doi.org/10.1161/01.HYP.13.6.968 Google Scholar
E. Michel et al.,
“Frequency dependence of cerebrovascular impedance in preterm neonates: a different view on critical closing pressure,”
J. Cerebral Blood Flow Metab., 17
(10), 1127
–1131
(1997). https://doi.org/10.1097/00004647-199710000-00015 Google Scholar
T. K. Koo and M. Y. Li,
“A guideline of selecting and reporting intraclass correlation coefficients for reliability research,”
J. Chiropractic Med., 15
(2), 155
–163
(2016). https://doi.org/10.1016/j.jcm.2016.02.012 Google Scholar
D. M. Corey, W. P. Dunlap and M. J. Burke,
“Averaging correlations: expected values and bias in combined Pearson rs and Fisher’s transformations,”
J. General Psychol., 125
(3), 245
–261
(1998). https://doi.org/10.1080/00221309809595548 Google Scholar
N. C. Silver and W. P. Dunlap,
“Averaging correlation coefficients: should Fisher’s transformation be used?,”
J. Appl. Psychol., 72
(1), 146
(1987). https://doi.org/10.1037/0021-9010.72.1.146 Google Scholar
C. Vlachopoulos, M. O’Rourke and W. W. Nichols, McDonald’s Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles, CRC Press, Boca Raton, Florida
(2011). Google Scholar
M. E. Wagshul, P. K. Eide and J. R. Madsen,
“The pulsating brain: a review of experimental and clinical studies of intracranial pulsatility,”
Fluids Barriers CNS, 8
(1), 5
(2011). https://doi.org/10.1186/2045-8118-8-5 Google Scholar
M. Soehle et al.,
“Critical closing pressure in subarachnoid hemorrhage: effect of cerebral vasospasm and limitations of a transcranial Doppler-derived estimation,”
Stroke, 35
(6), 1393
–1398
(2004). https://doi.org/10.1161/01.STR.0000128411.07036.a9 SJCCA7 0039-2499 Google Scholar
D. N. Ku,
“Blood flow in arteries,”
Annu. Rev. Fluid Mech., 29
(1), 399
–434
(1997). https://doi.org/10.1146/annurev.fluid.29.1.399 ARVFA3 0066-4189 Google Scholar
BiographyAlec Lafontant received his PhD in biomedical engineering from Drexel University in 2019. He is now a senior R&D development engineer at Medtronic. Prior to Medtronic, he was a postdoctoral fellow in the Wolfson Laboratory for Clinical and Biomedical Optics at the Children’s Hospital of Philadelphia. His research interests focus on the development of diffuse optics technologies. Elizabeth Mahanna Gabrielli is an assistant professor in the Department of Anesthesiology, Perioperative Medicine and Pain Management at the University of Miami Miller School of Medicine. She received her MD from the UT Southwestern Medical Center in 2007. Her research interests are in neuromonitoring and postoperative delirium. Rodrigo M. Forti is a postdoctoral fellow in the Wolfson Laboratory for Clinical and Biomedical Optics at the Children’s Hospital of Philadelphia. He received his PhD in physics from the University of Campinas in 2020. His interests focus on developing solutions to the technical challenges associated with incorporating diffuse optical methods into practical bedside devices. Tiffany S. Ko is a research scientist in the Department of Anesthesiology and Critical Care Medicine at the Children’s Hospital of Philadelphia. She received her PhD in bioengineering from the University of Pennsylvania in 2018. The focus of her research is the development of non-invasive diffuse optical tools to identify critical periods of neurological vulnerability and to facilitate individualized care pathways to mitigate subsequent neurological injury in critically-ill children. Daniel J. Licht is a professor of pediatrics at the University of Pennsylvania and a pediatric neurologist at the Children’s Hospital of Philadelphia. He is also the director of the Wolfson Laboratory for Clinical and Biomedical Optics. He is an expert in cerebral complications of congenital heart disease and has extensive experience in implementing non-invasive neuroimaging techniques for clinical use. Arjun G. Yodh is the James M. Skinner Professor of Science at the University of Pennsylvania. His current interests include fundamental and applied questions in condensed matter physics, medical, and biophysics and optical sciences. He has extensive experience utilizing diffuse optics techniques for monitoring and imaging of hemodynamics in living tissues. W. Andrew Kofke is an emeritus professor in the Department of Anesthesiology and Critical Care, Perelman School of Medicine at the University of Pennsylvania. He received his MD from the University of Pittsburgh in 1978 and his MBA in health sciences emphasis from West Virginia University. His research interests are in neurointensive care management. Brian R. White is an assistant professor of pediatrics at the University of Pennsylvania and an attending physician in the division of pediatric cardiology at the Children’s Hospital of Philadelphia. He received his MD and PhD in physics from Washington University in St. Louis in 2012. His research interests focus on the development of imaging biomarkers of neurologic injury in pediatric illness using optical functional neuroimaging. Wesley B. Baker is an assistant professor in the Division of Neurology at the Children’s Hospital of Philadelphia and Perelman School of Medicine at the University of Pennsylvania. He received his PhD in physics from the University of Pennsylvania in 2015. His interests include the translation of diffuse optics technologies, as well as using optical techniques to study the interplay between the delivery of critical care and the patient’s long-term brain health. |