1.IntroductionBoth morphology (depth, location, and shape) and the wavelength-dependent optical properties of the underlying tissue affect the observed signals of diffuse optics (DO).1,2 These are often modeled to be homogeneous or as simplified layered signals when fitting them with a theoretical model to obtain the quantities of interest.3–8 This is done not because the researchers do not appreciate the complex relationship between the two, but rather because the details of that relationship are too complex to know and utilize for practical measurements. Nevertheless, some more advanced methods utilizing a large number of source–detector pairs to reconstruct tomographic images use the underlying tissue morphology derived from brain scans as a priori information.9–11 This comes at the cost of being less practical. The availability of a complex model is not the only issue. Some of the atypical, heterogeneous tissues reduce the signal-to-noise ratio (SNR) to levels that would require unpractical solutions with the current technologies to obtain a reliable number of photons for data processing. This consequence is not a surprise because the physics of the photon propagation is well understood,3,5,12,13 and it is important to note that these heterogeneities also affect other modalities. The lack of a universal agreement over the analysis methods and a reliable assessment of the limitations of the technology led to the failures of several clinical trials14–17 using commercial near-infrared spectroscopy (NIRS) oximeters for neuro-monitoring. This has been particularly impactful in the neurocritical care of conditions such as traumatic brain injury (TBI) and led the clinical community to debate over the utility of NIRS due to the observed limitations in reproducibility, reliability, and accuracy which has been reflected in both the literature18–22 and elsewhere (e.g., in various discussions in congresses and forums). Many early adopters were discouraged by these results and abandoned NIRS-based cerebral oximetry. It can be argued that this could have been addressed better by the manufacturers through a careful study of the relationship between the tissue morphology and the optical signals and with more advanced online quality control measures. As expected, similar problems arose in studies emerging from research groups that are the developers of the technology and were addressed on an ad hoc basis during the data analysis, mainly by discarding unreliable data.18–22 Detailed studies of this relationship were also carried out using simulations and with limited in vivo data.23–32 In this article, an analysis of the data obtained by time-resolved spectroscopy (TRS)33 and diffuse correlation spectroscopy (DCS)3 over several years using the patient brain scans as the guiding data is reported. We opted not to use this information to build a complex model (would have been unstable due to the limited number of measurements) but rather to use them as a set of case studies. It will be argued here that this is an important contributor to a controversial issue in the field. Our goal is to provide anecdotal examples and speculate on different actions that could be taken to minimize the impact of these complexities on both the clinical science and also the clinical trial levels. From our presentation, it will be evident that this population is very heterogeneous, and each case should be inspected with respect to the quality of the optical signals as well as the potential ability to probe the brain through overlaying, complex tissue structures. The open question is whether these instructions or suggestions on how to improve the measurements and to develop indicators as to when to trust the results and/or to discard them could lead to a more successful future for near-infrared spectroscopic methods. Our goal is to offer an illustrated guide to assist end-users—such as developers of emerging commercial technologies, young researchers, and clinical collaborators—in applying diffuse optical methods for neuromonitoring in patients with critical care pathologies. 2.Materials and Methods2.1.Diffuse Optical InstrumentationTwo similar hybrid DO devices were used in this study, both previously described in Refs. 3435.–36. Both devices are based on NIRS principles, and they combine TRS and DCS. TRS allows the retrieval of the wavelength-dependent absorption coefficient () and the reduced scattering coefficient () to derive the microvascular oxy- and deoxyhemoglobin concentrations and the tissue/blood oxygen saturation ().3,33 TRS is based on the measurement of the distribution of the time-of-flight (DTOF) of photons, which allows for the accurate separation of the absorption and scattering effects at each wavelength.33 TRS data were acquired every 1 s with a 3- to 5-mW light injection at 687 and 830 nm. DCS uses a 785-nm continuous wave (CW) and coherent light source (27.5 mW) to quantify the statistics of the diffuse speckles to derive a blood-flow index (BFI) that is well known to be proportional to absolute cerebral blood flow.37 During the data acquisition, the optical probes were maintained fixed to the skin by self-adherent elastic bandages, i.e., kinesiology tape (Classic, Kintex, Talheim, Germany) and Coban wrap (Coban™, 3M Science Applied to Life, St. Paul, Minnesota, United States). Hand-held measurements without bandages were performed whenever the condition of the subject did not allow for the use of fixed probes, for example, due to the presence of decompressive craniectomy. 2.2.Brain ScansNon-contrast 3D X-ray computed tomography (CT) scans of the head by a standard CT scanner (SOMATOM Definition AS or AS+, SIEMENS Healthcare, Erlangen, Germany) were performed by the expert personnel of Vall D’Hebron University Hospital (VHUH) whenever required for routine diagnosis. The available CT scans closest in time to the DO acquisitions were then used for the purposes of this study. CT scans were performed with a maximum voltage () of 120 kV, a current of 35 mA, and either 1- or 4-mm slice thickness (meaning the distance between a scanned plane and the subsequent). When present, the 1-mm slice resolution was always preferred to the 4-mm one for the following evaluation. The acquired CT dataset for each instance was reconstructed with the hospital’s proprietary software, anonymized, and exported as a Digital Imaging and Communications in Medicine (DICOM) dataset for analysis using third-party software. A standard viewing software (RadiAnt DICOM Viewer, trial version 2021.1, Medixant, Poznań, Poland) was utilized outside the hospital to display DICOM data and with default settings for conventional CT images. Further information was removed using another user interface free tool (DicomCleaner™, PixelMed™ Toolkit). 2.3.Subject Inclusion Criteria and Study ProtocolsThree different groups of patients present in the intensive care unit (ICU) of the VHUH were selected for enrollment in this study: a cohort of malignant middle cerebral artery (MCA) infarction patients, a cohort of subjects with different severity of TBI and with subarachnoid hemorrhage (SAH). All cohorts obtained clearance from the local ethical committee within VHUH (ACU-AT-203/2012-3531) and conducted according to the Declaration of Helsinki.38 Informed consent was collected prior to the measurements and signed by either the patients or their legal representative. In the case of the malignant MCA infarction cohort, patients with large hemispheric infarction were recruited for a prospective interventional study. The inclusion criteria were (1) the presence of malignant MCA infarction and a hypodense lesion involving at least 50% of the middle cerebral artery territory with or without decompressive craniotomy and (2) age between 18 and 65 years. For the SAH cohort, patients were included in case of ischemic lesions related to subarachnoid hemorrhage and/or conditions following SAH [i.e., aneurysmal rupture with low Glasgow Coma Scale (GCS) score, bleeding, vasospasm, mild ischemia, and ischemic infarction]. These patients, aged more than 18 years, were included in a prospective interventional study. In the TBI cohort, the inclusion criteria were the same as reported in Ref. 34. In summary, patients over 16 years old with a TBI, GCS score of 3 to 15 upon admission, who required ICP monitoring, and mechanical ventilation from onset or after clinical decline were enrolled. The eligibility criteria included normocapnia [partial pressure of arterial carbon dioxide () of 35 to 40 mmHg] and stable hemodynamics allowing for a short hypocapnic challenge, with informed consent from a legal representative. The exclusion criteria were patients with contraindications for invasive neuromonitoring (e.g., anticoagulant use, scalp infection, and bleeding disorders) and clinical signs of brain death. Despite the populations being different, the study protocol has been uniform: it consisted of of steady-state acquisition on selected positions on the head (Fig. 1). The choice of the position for the measurements was made in collaboration with the expert medical staff who identified the different areas of interest according to the CT scans available prior to each measurement. Up to four positions were chosen on the same subject, and few subjects were measured multiple times during different acquisition sessions. After the acquisition, the position of the probes on the head was recorded. First, the location of the TRS detection fiber tip was marked with a sterile surgical staple, which was fixed on the intact skin by standard transparent medical tape. Subsequently, the radial distance between the nasion and the staple was measured, as well as the angle spanning between this radial distance and the straight line passing through the nasion and the eye line, as visible in Fig. 2a. It was decided to use the TRS tip to represent the whole probe for convenience. To measure the radial distance and the angle, a flexible measuring tape and a protractor were used. Figure 1(b) shows an illustration to better understand these parameters. In cases where CT scans were taken after measurement, the staple position was then captured and visible in CT scans for post-processing reference, as shown in Fig. 2. Additional demographic and clinical information such as presence of injury, presence of decompressive craniectomy, sedation state, presence of hematoma, gender, presence of bone, and age were also collected. 2.4.Data EvaluationThe optical data were analyzed following the physical models and procedures described in Refs. 13, 36, 40, and 41 by making use of MATLAB (Release 2018b, MathWorks, Natick, Massachusetts, United States). Once the optical data were fitted by the appropriate model, , at both 687 and 830 nm were estimated by TRS, and was derived. Similarly, from DCS, a BFI was derived. For the purposes of this work, a “baseline,” i.e., a period of constant hemodynamics, was assumed, and the mean values were calculated for all the variables at each measurement position on the head. The CT scans were utilized in a more qualitative fashion to define regions of interest (ROIs) in correspondence with each probe position as detailed in Appendix A. We chose to consider an ROI as a circular area of 15-mm radius () because the interrogated region below the probe is dominated by photons that have reached approximately this depth due to the interfiber distances. As for the qualitative evaluation, the tissue composition below the probe was assessed by experienced clinicians and categorized. The following tissue types were selected and identified with an acronym: subcutaneous tissue (ST), cranial bone (CB), normal brain (NB), swollen subcutaneous tissue (SST), air (A), cerebrospinal fluid (CSF), SAH, extracerebral hematoma (EH), intracerebral hematoma (ICH), ischemic tissue (IT), and brain contusion. The combination of all tissue types going from the most external to the most internal in the head was listed per each ROI, providing a composition of different tissue types. The basic categories that were utilized are summarized in Fig. 3, and one example is provided. The skin was disregarded because it was common to all measurements. In Appendix B, we provide an example of swollen subcutaneous tissue compared with a regular one. A database was created gathering all the measured variables from optical data, CT scans, clinical information, and the different categories of tissue properties. 2.5.Dataset DescriptionThe created comprehensive dataset comprises three key components:
For the data spreadsheet, each row corresponds to a different measurement position, an ROI. Therefore, multiple measurements may be available for the same subject either for different optical probe placements within the same acquisition or for repeated acquisitions within the same measurement day and or on a different measurement day. The columns, instead, contain the measurement identification numbers, averaged values of the optical parameters (, , , and BFI), and demographic and clinical information (tissue composition, presence of injury, presence of decompressive craniectomy, sedation state, presence of hematoma, gender, presence of bone, and age). Whenever the patient repeatedly underwent data acquisition, although the ROI was different, both clinical and demographic information were repeated for all data acquisitions. The “wavelength” column distinguishes optical properties obtained from the same probed region. Regarding the MATLAB data files (namely, a structure array is a data type that groups related data using data containers called fields), they house a structured data format containing the raw and processed data, such as both optical coefficients calculated at each wavelength, the BFI, and the values at each ROI (maximum two ROIs per file). The structure also contains additional information for purposes beyond this research. Importantly, data acquired during the same measurement acquisition but with two distinct optical probe placements are stored separately within the structure: field “hemi1” contains the results for the first position and field “hemi2” for the second position. If multiple measurement acquisitions occurred on the same day or across different days, either for one or both of the optical probes, these are saved as a separate MATLAB file. In this case, all the generated data files are stored in separate folders with incremental measurement numbers. CT scan images are provided for each measurement as described in Sec. 2.2. In cases where data acquisition occurred on the same day (or closely enough in time), a single CT scan may be used for multiple ROIs. 2.6.Data AnalysisThe bulk of the results are demonstrated as case examples by investigating the measured optical signals and their detailed structure to compare them with the clinical and radiological findings. Moreover, we used the results from a healthy volunteer to provide a fairly representative scenario for comparison against the data from pathological or abnormal cases. Each case example is elucidated by the same set of information, whenever possible, as exemplified by Fig. 4. Both optical reconstructions and essential evaluated parameters are also reported as descriptors of the probed region. From the left, the corresponding slice of the CT scan of the patient is shown with the marked ROI on it. Below, the age and sex of the subject are reported. Then, DCS results are presented as the normalized electric field autocorrelation function [, with as the delay time] derived from the normalized intensity autocorrelation function.3 For each probe position, curves were averaged over a hundred consecutive as well as over the different detection channels. Therefore, only one autocorrelation curve is represented for the selected ROI, as in Fig. 4. Afterward, the fitting procedure was performed on this curve to retrieve BFI. The obtained fit is superimposed to the averaged curve in this figure to appreciate how much the data are represented by the ideal model. In the same graph, the curve obtained in the same way from the healthy volunteer is plotted in orange as a reference. In addition to , for each averaged curve, we have reported the information about the instrument parameter, count rate (intensity), and fitted BFI value in a table that can be compared with the ideal ones. The details about the typical behavior of the measurement and the range of values can be found in Appendix C. Similarly, for each probe position, a hundred consecutive measurements were averaged for TRS DTOFs. Afterward, the fitting procedure was carried out, and the physiological and optical parameters were obtained.33,42 To represent the raw data as normalized, the maximum of the DTOF is used for normalization (by simple division).43,44 Apart from a visual inspection, TRS curves were described in terms of their overall shape (qualitative), amount of broadening with respect to the instrument response function (IRF), and the dynamic range.33,45,46 Furthermore, we have quantitatively characterized the data by some descriptive parameters, such as relative full-width at half-maximum (rFWHM) of DTOF with respect to the IRF, difference of FWHM between DTOF and IRF (), SNR, , , and . The values for each case example are reported in a table divided per light wavelength, either 687 or 830 nm. The rFWHM was calculated as the ratio between the FWHM of the DTOF and the FWHM of the IRF, although more complex equations do exist for a more precise estimation, taking into account for example the laser power.47 The , instead, was calculated as the difference between the FWHM of the DTOF and the FWHM of the IRF. The SNR was computed as the ratio between the maximum of the TRS DTOF curve divided by the standard deviation over a portion of the background signal (the noise) present before the light pulse itself. Again, details about the typical curve behavior and values range can be found in Appendix C. As a side note, general backscattered light shapes and indicative optical values are valid when the diffusion approximation is valid, as well as the approximation to have greater reduced scattering coefficient over the absorption one (weakly absorbing and highly scattering media).40,48 On the contrary, other methods should be used to describe the diffused curve shape.48 Similarly, other approximations used to solve the diffusion models for DCS and TRS may not hold true (i.e., homogeneous model, slow temporal changes in the fluence rate, and random walk approximation).42 For this reason, both qualitative and quantitative analyses are helpful in the quality assessment of the results. 3.Results3.1.Population CharacteristicsOverall, data from 121 different probe positions from 36 subjects were included from June 2016 to October 2020. The distribution of the subjects among the cohorts is reported in Table 1 as well as the pertinent demographic and clinical information. Table 1Characteristics of the recruited patients are reported in this table. The parameters are expressed in proportions, such as males:females (M:F), yes:no (Y:N), and percentage over the total. Age is given as the mean [standard deviation (SD)] for the group. Subjects were recruited from three pathological categories or cohorts. TBI, traumatic brain injury; MCA, malignant infarction; SAH, subarachnoid hemorrhage.
The frequency of the different tissue composition categories (see above for the classification) is listed in Table 2. For three positions, the CT scan was not available to perform a precise assessment, but we could still obtain a proper signal and record the presence/absence of skull. Precisely, one of them was on a decompressive craniectomy area without underlying bone, whereas two were supposed to probe a healthy-like tissue, with skull and cortex below. Table 2Classification of all tissue compositions for the ROIs after the examination of the CT scans, when available, or the clinical information. Each tissue type is represented by initials and is described by going from the outer layer toward the gray matter as a sum of layers/components. Details about all acronyms are in the text. Two main categories are identified: having a regular ST as in the left column or an SST as in the right column. The tissue compositions denoted by an asterisk (*) do not include CB as a listed tissue type. The CT scans were not available in three ROIs to perform a precise assessment, which were omitted.
3.2.Case ExamplesA table summarizing the following pathologic examples is provided in Appendix E. We suggest the readers to make this table available at the bedside to facilitate their reading and the comparison of cases. 3.2.1.Healthy case H0This case illustrates the data from a healthy 25-year-old female subject (see Fig. 5) where both TRS and DCS curves possess the typical shape following their respective theoretical models. In this case, the decay for DCS is roughly exponential. Due to the normalization, the curve starts near one and approaches zero at s. The intensity rate is high, above 90 kHz, with a and very close to ideal. The BFI value is within the normal range. The TRS curves from both wavelengths show the expected behavior: the peak of the DTOF is shifted to later times with respect to the IRF peak, and the curve is broader with the FWHM of the DTOF, much larger than that of the IRF. DTOF plots are attenuated in amplitude than the IRF, with a long linearly decaying tail. The maximum signal level is acceptable, almost three decades above the noise background. The two wavelengths behaved similarly, and their shape looks alike. The obtained , , and around 67% are not far from the indicated ideal values. Additional similar examples are provided in Appendix D. 3.2.2.Case 1In this patient (male, 29 years), the probe was placed above a region that underwent decompressive craniectomy after malignant MCA infarction. The tissue composition was ST + EH + IT. The qualitative evaluation is based on Fig. 6. As seen in the figure, the results from 687 nm were not reported because there were very few detected photons. This is most likely due to the presence of highly deoxygenated blood in the infarct zone which absorbs differentially higher in this wavelength. This is coupled with the fact that the missing skull implied that the probe was resting on the brain/muscle directly, which absorbs more light than the typical scenario. Given the performance of this study’s device, the number of photons detected by the TRS for 830 nm was very minimal presumably due to the same reasons as above. The minimal DTOF curve that was observed could not be reliably analyzed with a photon diffusion model because it did not indicate a scenario where scattering dominated over absorption. Therefore, no optical properties are reported. DCS results also showed an atypical behavior that is often observed on dead tissue or on regions with minimal red blood cell movement.3 Unexpectedly, the intensity level was quite high, but the parameter was low and the decay time () was quite long lasting longer than , implying very low BFI. This may be expected because infarcted tissue is expected to have lower blood flow. On the other hand, the model did not fit the curve well and the quantitative results cannot be trusted. Note that once the fitted BFI is lower than , the fits may be unreliable because the ergodicity of the measurement may no longer be sufficient due to very low blood flow. In a case scenario like this, few considerations can be made to yield some guidelines. As for TRS, because the SNR is low, summing up more curves than a hundred could be a method to improve the DTOF or a longer acquisition time. Of course, higher injected source power and/or a larger detection area would also help. However, even with these improvements, it may not be sufficient to obtain a good fit by the diffusion model unless a device with a much narrower IRF is utilized, and the data are fitted by a more complex model (higher order approximations to the radiation transport equation or a Monte Carlo method30,31). For DCS, the current practice is to report this type of curve as being indicative of blood flow less than one percent of the normal without a specific value. As a consequence, quantitative results cannot be obtained by diffuse optical methods, but the nature of the data, especially the DCS data, reveals minimal blood flow in the region. Unfortunately, it is hard to predict how this type of tissue would reveal itself to a multi-distance CW NIRS device. 3.2.3.Case 2The ROI delineated in this case shows an area of the brain where extracerebral muscle tissue was swollen following a craniotomy, where the bone was repositioned in the head after surgery to treat a lesion (tissue composition: SST + CSF + CB/A). The patient (male, 18 years, initial GCS = 5) presented a diffuse axonal injury with epidural hematoma in the left cerebral hemisphere and cranial fractures due to a car accident and had undergone a decompressive craniectomy on the injured hemisphere. The raw data and the clinical conditions are showcased in Fig. 7. For the depicted ROI, there is a high probability that the brain was not actually reached by the diffuse optical light path. DTOF curves qualitatively appear similar for the two wavelengths: their dynamic range looks similar to the one expected from a healthy tissue, and the DTOF appears broader than the IRF and with a slower linear decay. The DCS curve, instead, had a faster decay toward 0 intensity than the given healthy example, implying a faster BFI, although the curve shape is similar to a healthy-like one and its fitting looks reliable. In quantitative terms, the DCS detected a low light intensity rate, whereas the autocorrelation function had a comparable coefficient and a higher BFI with respect to the healthy example. coefficient was high for both DTOF curves, which is expected due to the faster linear decay, whereas was quite low. On one hand, the latter may be due to an amount of direct light, which could appear during hand-held measurements, where the adherence of the probe to the skin depends on the person who holds the probe. It could be remedied with the use of common load sensors, for example. On the other hand, swollen muscle tissue is characterized by accumulated liquid, which is mostly transparent/translucent and scatters less. If these results are evaluated without care, they could be interpreted as representing true cerebral hemodynamics. However, from the CT scan, it appears evident that the probed area had a low probability to have intersected the brain cortex: the atypical increased thickness of the upper layers implies that the bulk of the photons did not reach the brain. In light of this, these data have to be interpreted as invalid for measuring the brain. Rather, they reflect the properties of the swollen muscle and, partially, the bone, which could be relevant for other reasons or studies. From the perspective of the TRS, it is possible that a better setup, with a higher laser power and tighter FWHM of the IRF could allow us to time gate the photons and retrieve the information from the deeper layers. This would also require a more complex analysis method, e.g., using layers or anatomical priors, which are known to be unstable.29–31 3.2.4.Case 3Figure 8 shows the findings from case 3, which is a common, yet very difficult case to interpret. The data were acquired above a swollen muscle with large edema and after a craniectomy procedure (tissue composition = SST + CSF + NB, male patient, 56 years). The etiology was a fall (initial GCS =13, later reduced to 8 at intubation) that produced subdural hematoma and subarachnoid hemorrhage requiring a decompressive craniectomy procedure on the left cerebral hemisphere. At the TRS level, the DTOF curves detected for both wavelengths are analogous: sufficient dynamic range in amplitude, shorter photon path lengths compared with a typical brain, and fast linear decay of the backscattered light pulse tail in line with a higher absorption value which is due to the lack of the skull. Numerically, the values found display an absorption coefficient higher than usual, comparable to the previous case 2, whereas the is intermediate between case 2 and a normal expected value. DCS shows a slower decay and comparable shape to the normal case. Quantitatively, the DCS intensity of the acquired curves was acceptable, averaged over four channels, whereas was comparable to a healthy-like case. Finally, the obtained BFI is lower than for an intact head condition. Referring to the CT, the photons have a low probability to have crossed their path with the cerebral gray matter, due to the thickness of the outer inflated tissue. Therefore, similar conclusions and suggestions to case 2 could be made here with a main difference: the absence of skull, which may account for the differences in the scattering properties and lower BFI. This is another case of data appearing “normal” while possibly not reflecting the cerebral hemodynamics. The CT scans should be considered when evaluating these results. 3.2.5.Case 4In this case example, the probe was lying upon a region that underwent a decompressive craniectomy with an accumulation of cerebrospinal fluid between a thin layer of muscle and the brain, as visible in Fig. 9. The patient (male, 53 years) had an ischemic lesion on the right capsule–thalamic region and a hemorrhagic frontoparietal contamination on the left cerebral hemisphere after a fall (GCS = 11). He was treated with a decompressive craniectomy on the right hemisphere. Despite what is shown in the figure, the skull fragment had been replaced to its original position at the time of the measurement, and therefore, the tissue composition beneath the probe was SST + CB + CSF + NB. Qualitatively, DTOF curves reveal that overall, there were fewer detected photons but with comparable light diffusion to the healthy tissue. The 687-nm backscattered light pulse is more affected than the 830 nm one, which presumably is due to highly deoxygenated hemoglobin presence (which absorbs more at 687 nm). The data could not be fitted to the model with the current set-up and acquisition conditions. DCS was unable to detect a usable intensity. Even after a long integration time, the curves shown in Fig. 9 are quite suspect, likely reflecting structures of noise (e.g., after-pulsing). CSF is mostly translucent and low scattering could lead to a lower number of photons being scattered back to the detector. Furthermore, this measurement was taken over a region where the hair was shaved but hair follicles remained in place. This complicates the data acquisition greatly. In this case, the effects are quite visible and even without the CT scans one could notice the poor quality of the measurements. This may or may not be true for multi-distance CW measurements—stressing one more time the importance of the consideration of the underlying anatomy. 3.2.6.Case 5As highlighted in Fig. 10, this case offers an example that features a partial intersection of an intracerebral hemorrhage, which is the large, oval bright area in the CT scan enclosed in the red oval. Most likely, the measurements mainly reflect the skull and gray matter but could be influenced by the hemorrhage and its surroundings to some degree. The tissue composition assigned to this probe location is ST + CB + ICH/NB. The patient (female, 28 years) suffered a TBI and SAH (diffuse lesion type III) along with a complete spinal lesion at the C3 level. A decompressive craniectomy was performed on her left cerebral hemisphere. By inspecting the DTOF curves, at a qualitative level, both wavelengths possess the same features: wide dynamic range and dispersion of the curve that is comparable to a normal tissue-like, slower linear decay which should mean lower absorption properties. DCS autocorrelation function, instead, has a damped shape, flat in the initial part, which does not appear to reflect a single exponential and has a slower decay. On the quantitative side, the TRS figures of merit feature: around 0.1 to which are typical for a healthy head, high , high SNR, and good fit to the model. However, the appears to be higher than that for a healthy brain. Conversely, the DCS curve is characterized by low intensity and high , which was probably affected by the typical afterpulsing observed in this class of photon counting detectors, i.e., in a simplistic manner, afterpulsing is the detection of a second, false photon by the detector after the initial photon in a probabilistic fashion (see details in C). This led to the observation that the fitting procedure ended up with a BFI value much lower than usual, but because the data are dominated by noise, this is not a reliable value. Under these circumstances, a blind measurement analysis could lead to the wrong conclusion: the hemorrhage is there, but the way in which it influenced the signal is unclear. In conditions of reasonably maintained blood flow, low partial pressure of tissue oxygenation () readings are reported in the literature,49 which are a symptom of low oxygen extraction by the tissues. Therefore, higher than normal (i.e., greater than 70%) tissue oxygen saturation is expected in the blood, which is confirmed by this measurement. The DCS curve is certainly affected by the hemorrhage indicating a heterogeneous result—possibly a probed volume with large volumes of low and high scatterer dynamics leading to a complex shape. Overall, if the presence of the hemorrhage is taken into account, it may be possible to select a different region to either try to obtain brain values or to get information reflecting the hemorrhagic region itself. 3.2.7.Case 6The case example presented in Fig. 11 is similar to the previous case: the probe was positioned over the skull in a region that included tissue with a subdural hemorrhage. In fact, this patient (male, 63 years) had experienced a fall from 2 m of height that led to a contusion (GCS = 14) and following lesion on the left cerebral hemisphere necessitating a craniotomy procedure to extract the hematoma on the temporal left side of the head. The probed region was assigned ST + CB + SAH + NB as tissue composition. The DTOF curves show a faster linear decay than a healthy tissue, with sufficient dynamic range and comparable dispersion with respect to their IRF for both of the wavelengths. On the DCS side, the initial part of the intensity autocorrelation curve is similar to a healthy-like case, but its final portion did not properly reach 0 after the exponential decay, rather it presents a secondary exponential decay. This is most likely indicative of a shallow region with low blood flow. Quantitatively, on the TRS side, the SNR is high, and the relative and difference of FWHM agree with the expectations for a diffused DTOF for both wavelengths. Moreover, the data agree very well with the model. Interestingly, the information found is relatable to the previous example. In this ROI, the values are still low, between 0.07 and , is quite high (80%), whereas coefficients are , lower than healthy tissue-related ones. The DCS curve deviated from the homogeneous model significantly, although the intensity rate was decent and in line with normal values. Therefore, the BFI cannot be considered reliable. It is also worth noting that the final part of the curve affects the calculated numbers due to the physics, even if the cutoff threshold for the fitting procedure of the DCS curve occurs at an autocorrelation value of 0.3. Moreover, the second exponential decay in the tail of the autocorrelation function may suggest that the method is detecting atypical tissue, as if a sudden change in the morphology creates a second layer that abruptly modifies the decay. This case is difficult the evaluate. Clinical interpretation of the CT scan reveals significant heterogeneity and abnormalities under the probe, and the optical data show deviations from the physical models for the DCS curves. A conservative approach would err toward invalidating the optical measurements, whereas it surely highlights potential issues with CW measurements, which may not reveal these problems so clearly. 3.2.8.Case 7Another possible situation is represented by the presence of accumulation of CSF between the skull bone and the brain, as shown in Fig. 12. The patient for this case was the same one as in case 4. The tissue composition for this ROI is ST + CB + CSF + NB. Observing the DTOF curves, it is immediately evident that they have an appearance that resembles that of normal tissue: the dispersion, the dynamic range, and the shape are very similar. On the DCS side, the shape is highly distorted, possibly revealing multiple decay constants. Also, at a quantitative level, the values are in line with an acquisition obtained from a healthy head for the TRS, although resulting in a low . For the DCS, the models cannot be used to fit the data. This case is particularly problematic when interpreting the data and the worst is that, from a careful evaluation of the quality of the TRS data, it is practically impossible to notice that something is hidden behind the findings. As previously mentioned, CSF is a fluid with low scattering and absorption properties. It appears that the presence of CSF is tricky and difficult to recognize, especially considering that it is only an intermediate layer in this measurement. Moreover, considering that, given the thickness of the intermediate layers, the gray matter may have not been reached, it is reasonable to obtain a low tissue saturation. In this context, the CT scan is a primary source of information. A pre-measurement view of the CT scan should not be omitted in this case. 3.2.9.Case 8Fig. 13 shows an example in which, after an injury, air got trapped between the skull and the frontal lobe of the brain. This is a quite common situation, especially after a concussion and after a surgery,50 known as pneumocephalus. Moreover, there is a natural cavity, the sinuses, which are at times more pronounced in some people and that can be present up to a couple of centimeters above the eyebrow line. In this case example, the patient (male, 29 years) had TBI (initial GCS = 3) caused by a fall that led to a therapeutical decompressive craniectomy on the left cerebral hemisphere. The indicated ROI also appears to include the brain and skull, with a tissue description as SST + CB + A + NB. The qualitative analysis of the TRS curves reveals a very strong analogy with a healthy head situation, even if the linear decay is a little less rapid. For the DCS curve, on the other hand, the situation is similar to that of cases 6 and 8, where a second decay appears in the final part of the curve. In quantitative terms, coefficients appear normal, whereas and higher than for a healthy head. As regards the curve, the intensity rate is quite reduced (14 kHz), but and BFI values seem reasonable, with a proper fit, and comparable to a healthy scenario. Consequently, it looks like diffuse optics cannot clearly distinguish between a normal head and one where air is comprised. Most likely, the presence of a region with trapped air, a non-diffusive medium, led to a complex result that could have confused traditional quality control methods. 3.3.Dataset DescriptionPertaining to the patient data spreadsheet, 121 rows for a different cerebral position (ROI) were included, with a total of 242 rows, because the optical properties were reported separately for the two wavelengths of the TRS sources. Please note that some numbers, although unusual for the normal cerebral tissue, were still kept because the criteria for rejection did not apply nor did the data fitting procedure fail. Concerning the MATLAB-structured data files, they were gathered for 91/121 ROIs. In fact, we have omitted the data for the 30 measurements that were obtained with a different device, as explained in the methods, due to the difference and complexity of the files. However, these data were used for the averages and case examples and are present in the other formats. As for the CT scan images, no CT was identified as close enough in time to be used for the proper tissue identification or was not provided for three measurement sessions with a total of six ROIs. The created database is available on CORA, Repositori de dades de Recerca repository (Repository Technology/Platform: Dataverse) at https://doi.org/10.34810/data1355, where the CT scans are available upon request to the authors. We believe this database can be valuable to other researchers. For instance, for replication or validation, by reprocessing the data with their own data processing methods, or analyzing them from a different perspective. Moreover, data integration between the anatomical data provided by the CT scans and the optical data could be performed to more accurately unveil the optical properties of the atypical tissue structures and layers as well as to simulate the light propagation through a head with a heterogeneous structure, taking advantage of these anatomical priors. 3.4.General Recommendations for Optical Measurements with Examples from LiteratureWe now speculate about practical recommendations for quality control in hybrid diffuse optical neuro-monitoring as were already partially outlined in the case examples above. It is important to note that these guidelines are not only based on the few case examples shown in this paper but also on the authors’ experience and the literature. It could be said that there are three main categories of guidelines for improving measurements on MCA, SAH, and TBI patients:
We have gathered this information in Tables 3 and 4, without the intention of compiling a complete review. Table 3Set of recommendations and guidelines: part 1. (1) Device performance limitations, (2) loss of data due to model failure due to morphology/poor quality, and (3) poor/inaccurate probe positioning.
Table 4Set of recommendations and guidelines: part 2. (1) Device performance limitations, (2) loss of data due to model failure due to morphology/poor quality, and (3) poor/inaccurate probe positioning.
In the present study, improvements in the DCS hardware were actually already achieved with respect to the currently available devices, as suggested by Carp et al.,55 where the power delivered on the head was high thanks to the double-source solution. Moreover, field programmable gate array–based correlators with faster and more reliable acquisition were used instead of the commercial ones, allowing for fast DCS acquisitions, and an increased number of detection channels, that enhanced the SNR.54,83 As a side note, some of the cited references have implemented solutions that could improve TRS and DCS issues in this context, but that are not currently implemented in these techniques. However, despite these recommendations, sometimes it is unavoidable to obtain non-optimal measurement conditions, and it is, therefore, necessary to implement data quality checks, which are less important in cohorts of healthy subjects. Blindly fitting the TRS and DCS curves is risky and may lead to the loss of the hidden information. Quality checks can be implemented on simple figures of merits, such as the ones that were used in the case examples of this paper and the ones presented in Refs. 57, 58, 63, and 77. Such figures included the minimization of the least square error between data and fitted model, minimum total number of detected photons, difference in the temporal position between IRF and DTOF, and measurement reproducibility on a calibrated phantom. 4.DiscussionSeveral reasons are listed for the lack of widespread implication of NIRS-based monitoring in TBI. Among them, there are contamination from extracranial or extracerebral tissues, difficult implementation due to scalp and facial injuries, confounding effects caused by subdural, epidural and extracranial hematomas, and dead tissue.18,18–20,84,85 However, improvements in the computational modeling of DO and its integration with subject-specific imaging do represent a clear possibility to expand its use in brain pathology assessments. In this work, we have characterized structural and optical signal abnormalities in severe brain injury, malignant MCA, and SAH patients. Overall, the results described problematic tissue types and how they affected the DO signals in a major attempt to understand if certain impaired brain regions are measurable and to promote data quality examination. In fact, the complex multi-layered structure of the adult human cranium already sets a limit to the accuracy of NIRS.86,87 We have recruited a pool of 36 patients that included 9 malignant MCA, 26 TBI, and 1 SAH. In 19% of the cases, surgery was performed on the subject, allowing for thirteen measurement acquisitions in the absence of skull. The surgical procedures included decompressive craniectomy, craniotomy, and resection. Initially, we have described eight case examples with pathologies and one case of a healthy subject qualitatively and quantitatively. The selected cases had the objective of showing typical situations that one can meet when this type of patient is recruited and to provide a fairly representative sample, albeit limited to the variety of measurements that have been obtained. In some cases, the simple view of the curves made it possible to immediately understand the difference compared with the healthy case, whereas in others, it did not. Moreover, even at the level of numerical values, it was at times impossible to discern whether it was a curve generated by healthy or damaged tissue. Later on, some guidelines for measurement practice were outlined and related to the literature to provide insights and practical examples. To the authors’ knowledge, there is no similar attempt in other studies on head-injured subjects and by hybrid DO technique. Some existing work regarding healthy subjects has been undertaken in, for instance, Refs. 88 and 77. In the latter, a strict quality control practice was also set,77 partially as we aimed to discuss in this paper. Some studies have reported oddities when CW-NIRS devices were used with hematomas, craniotomies, and craniectomies20,23,89,90 but did not inspect the cases. Other studies rather focused on the detection of hematomas, either by bi-hemispheric cerebral comparison or absorption values,67,87,91–93 by standard CW monitors. Instead, the effect of CSF has been mainly studied by means of simulation23,25,86,94 because it is difficult to isolate it in real experiments, whereas the skull effect was studied on a reduced number of subjects () and by NIRS simulations.95 Some more comprehensive studies focused on measurements of ischemic infarction patients, in which, however, averaged values for several cerebral positions were reported and limited information regarding the raw data/curves was provided.77,96 Despite the heterogeneous complexity of the morphology across the brain, the associated values partially indicate different brain conditions as well as depend on separate confounding factors. Anatomical lesions, disease progression, and therapeutic interventions (i.e., decompressive craniectomy) can all modify the regions beneath the probes. Nevertheless, TRS-DCS techniques offer some useful applications: i.e., the significant light absorption caused by hemorrhages can help diagnose injury types. In addition, the comparison of the optical properties of various areas of the head for patients where other stimulation is not possible by checking their symmetry, as previously reported, could be introduced. However, it is always preferable to refer to images of the underlying structures, if available, before performing any measurements. Therefore, we believe that this work shows how the combination of hybrid DO with other techniques could enhance multimodal neuromonitoring, offering biomarkers that represent different aspects of pathogenetic mechanisms for a holistic (and more accurate) understanding of the brain status. We provide a large, annotated dataset that could be used for further analysis and interpretation in the future. It must be noted that the characterization reported in this paper, especially at the quantitative level, is affected by the properties and intrinsic features of the device used, which determine its performance and establish its limitations. Therefore, the experimental conditions and conclusions presented in this article should be reconsidered and adapted for each device/probe and cannot be fully generalized. As previously stated, analyzing data with a model based on a homogeneous semi-infinite medium is not ideal in cases of heterogeneous probed volumes, such as for the data presented in this article. As noted in the literature reported in Secs 3.3, 3.2.3 and Table 4, a multi-layer analysis would be advantageous. However, to date, such models remain unstable due to the physics of the problem (strong dependence on specific assumed anatomical parameters) and have not been implemented as practical bedside, real-time analysis methods. Due to this, although preliminary exploratory analyses suggested potential differences in optical properties and blood flow among tissue types (e.g., skull bone and hematoma), these findings were not pursued further due to limitations in potential inaccuracies in quantitative values, as well as sample size and study design. In this article, skin was not considered in the tissue assessment, because it was present in every measurement, even though the data are affected by its presence. In Ref. 23, the authors noticed that the light intensity was halved after scalp removal. Moreover, the scalp and skull possess highly diverse optical properties that provoke a completely different effect on the light passing through and the curve’s shape.23 The skin’s is quite higher than the skull’s: the authors argued that the skull would almost not attenuate the light passing through. Therefore, taking skin in relation to its thickness into account may be important. A potential criticism is the choice to use only a few case examples from which to begin to discuss possible guidelines and improvements for using DO in these cerebral tissues. This approach might seem reductive, but the examples serve as a starting point. In a qualitative manner, we have considered various clinical and in vivo studies carried out in other populations such as ischemic infarction patients, patients with carotid stenosis, the elderly without any known brain injuries, and healthy subjects. Nonetheless, this study provides a good starting point for visualizing extreme cases of brain damage, though additional measurements are needed to reach solid conclusions. Overall, this study provides case example studies of complex, yet common clinical situations, serving as a starting point for designing new devices, probes, and algorithms and for implementing quality control assessment. The output dataset can be utilized for further quantitative analysis, such as estimating optical properties for different tissue categories and carrying out numerical simulations. We strongly recommend that even real time, bedside display of the output should be supplemented with priors from the clinical and radiological evaluations to avoid false results. Finally, we highlight that there might be situations in which typical diffuse optical neuro-monitoring is simply impractical for certain tissue volumes. 5.ConclusionsIn light of the findings and literature review, this article has successfully shown key morphological alterations and their relation to the performance of hybrid optical devices. The interpretations are truly multiple and yet necessary to confirm the practicality of this technique in the myriad of possible conditions. 6.Appendix A: Identification of the Regions of Interest on CT ScansIt was arbitrarily chosen to consider the ROI as a circular area with a radius of 15 mm and an area of . This choice was made upon the fact that the interfiber distance between the TRS source and detector tips for the device was 30 mm, and consequently, the optical sensor was probing with higher probability the density region at deep. Few steps were followed for the manual identification of each ROI. If the CT scan contained the reconstruction of the fiducial staples left on the patient’s skin (examples are provided in Fig. 2), the slice with higher contrast for this object was taken as an axial plane. The ROI was then selected “underlying” the staple contrast position. Unfortunately, there was no post-measurement CT scan with staples for all included subjects. In this eventuality, the retrieval of the ROI took advantage of the angular and radial information collected at the time of the acquisition. A mannequin head was used to relate distances and angles on a real volume and then adapt the position as much as possible to the morphology found in the CT. For few subjects, the actual position was not recorded: in this case, an estimation based on the data collection experience was carried out. The procedure consisted of reaching the axial slice with the highest contrast for the crystalline biconvex lens of the eyes and then adding to this plane a vertical shift of 15 mm (the average value between the crystalline and the eyebrows for other subjects), adding further 40 mm to reach the middle of the forehead and select this axial plane. Then, in this plane, a line was depicted following the midline of the brain up to the front, and 45-deg angles were also drawn up to the forehead surface approximately in correspondence with the middle of the eyes and symmetrically for the two cerebral hemispheres. The crossing point on the forehead line was used to position the ROI in the area below, tangent to it. 7.Appendix B: CT Scans Example of Regular Versus Swollen Subcutaneous TissueIn Fig. 14, we facilitate a visual example of regular subcutaneous tissue in comparison with swollen subcutaneous tissue to make inexperienced users aware of the appearance of the CT scans. We would like to point out that expert personnel trained in interpreting CT scans were responsible for identifying the tissue categorization. Consequently, this task may not be as straightforward for inexperienced viewers to achieve. 8.Appendix C: Expected Optical Signals Shape and Values RangeWe report the nature of typical signals measured from an uninjured, presumably healthy brain for those readers who are not familiar with TRS and DCS data. Ideally, the decay for the DCS autocorrelation curve is exponential, decreasing with the square root of the delay time.3,71,97–99 When the normalization of the curve works as expected, the curve starts near one and approaches exponentially to zero for increasing .3,71,97–99 For the source–detector separations and typical instrumentation, such as those used in this study, it reaches zero at s.3,71,97–99 Even for a specific instrument, this is influenced by the BFI and other optical parameters.3,37 In a healthy brain, good-quality data are obtained with the observed count rate well above the noise level (0.5 kHz) and well below the saturation level (200 to 300 kHz, accounting for correlator and detector effects), with an instrument parameter () near 0.5 for unpolarized light.3,13,100–102 The fitted BFI value normal range is to 26,27,37,103 which depends strongly on how the reduced scattering coefficient was calculated. It is often the case that the DCS signal is affected by detector non-idealities such as dead time, shot noise, thermal noise, dark current, and afterpulsing probability, among others. As a side note, the afterpulsing effect of a detector is a phenomenon where, following a primary detection event, spurious pulses are generated that are not due to actual incident photons but are instead a result of the detector’s internal processes.104 This effect is particularly significant in some detector types, where residual charge carriers or trapped electrons from the initial detection event can be released at later times, causing additional false signals. For instance, the false afterpulsing pulses can create artificial correlations, leading to inflated values of the at short delay times, a high parameter, and cause additional noise.105 Although some of these could be “corrected” with a sufficiently thorough characterization of the individual detector behavior, it is not advisable to do so when these become the dominant processes due to low photon counts. In an ideal scenario, the “TRS curves” from both wavelengths show an expected behavior: the peak of the DTOF is shifted to later times with respect to the IRF peak, and the curve is broader meaning that the FWHM of the DTOF is much larger than that of the IRF due to the broad distribution of the propagation times in tissue.33,62 DTOFs are generally attenuated in amplitude (often visualized and/or processed normalized to the peak) with a long decaying tail. This tail decays linearly for long path lengths where absorption occurs with greater probability and dominates over scattering.33,62 The delay of the peak comes from the finite time that light employs to go from the source to the detector.33 The broadening of the DTOF is a consequence of the different paths that photons travel inside of media because of multiple scattering events.33,62 Retrieving the proper shape of the DTOF, built by the number of detected photons yields information about the optical properties of scattering and absorption that characterize the probed medium area. To obtain a good quality signal, the maximum signal level has to be a few decades above the noise background prior to the DTOFs.33,62 Moreover, both DTOFs at the two wavelengths should behave similarly, which is expected in a healthy-like tissue where oxy- and deoxyhemoglobins affect them in a similar way. This could change dramatically in case of deeply deoxygenated, i.e., often hypoxic and/or ischemic, tissue. Overall, a dynamic range of three to four decades (or higher) for the DTOF curves is generally acceptable to ensure a proper fit of the optical parameters, especially guaranteeing to extrapolate the absorption information from the linear decay of the tail.62 The DTOF peak delay from the IRF peak of at least to 0.8 ns is also common and is generally acceptable;33 otherwise, it may entail insufficient diffusion time of the photons within the medium, hence rendering accurate the measurement of IRF even more crucial. The relative FWHM is expected to be a value greater than one when there is typical broadening due to photon diffusion in the tissue. In fact, the FWHM of the IRF defines the temporal resolution of the device. When the photon’s arrival time broadens due to the scattering events and the emerging DTOF modifies its decay given the amount of absorption events while traversing a tissue, the output DTOF retains this information: the more scattering events (and properties, the more), the broader the histogram of photon arrival times. Therefore, the FWHM of the DTOF is generally broader than the IRF one.62 This is typical for a structure such as the healthy human head and its optical properties.58 should be positive in standard diffusion/absorption conditions.40,106 The SNR level depends on many factors, such as the background noise of the detectors or how well external light is shielded. Its acceptable value has to guarantee a clear distinction between the DTOF curve and the background level,106 and generally, an SNR = 10 can be chosen as a non-strict value below which data should be considered for exclusion. Generally, a rule of thumb for good values in a healthy brain are , , and around 56% to 66%.7,33,58 Nonetheless, these are indicative values and literature examples carried out on large cohorts of patients are useful as a starting point, i.e., from Refs. 58, 77, and 88. In both cases, we have provided a qualitative guideline based on experience and the literature that could be adopted, even at the bedside as a figure of merit. For further quantitative analysis, we refer the readers to excellent reviews of the topics such as in Refs. 32, 33, 58, 69, 88, 107, and 108 and their citations. 9.Appendix D: Additional Healthy Case ExamplesIn this appendix, we provide further examples of typical TRS and DCS curves from healthy subjects to not only provide further data as an illustration but also to demonstrate that even the presumably healthy brain impresses significant heterogeneity on the measurement.32,33,57,58,109 We have produced them in the same fashion as healthy case H0 in Figs. 15Fig. 16Fig. 17–18. 10.Appendix E: Summary of the Case ExamplesThe case examples are summarized in Table 5 to facilitate the comparison among cases. Table 5Summary of the case examples presented in the results, to help the reader in their comparison. TBI, traumatic brain injury; SAH, subarachnoid hemorrhage; MCA, malignant cerebral artery stroke; SNR, signal-to-noise ratio; BFI, blood flow index; μa, absorption coefficient; μs′, reduced scattering coefficient; CSF, cerebrospinal fluid.
DisclosuresNo financial conflicts of interest were identified. No competing financial interests were present for all the other authors. Code and Data AvailabilityData are available on CORA, Repositori de dades de Recerca repository at https://doi.org/10.34810/data1355, under the title Replication Data for: How the heterogeneity of the severely injured brain affects hybrid diffuse optical signals: case examples and guidelines. The CT scans can be downloaded upon request to the authors. Author ContributionsS.T. had full access to all data in the study, took responsibility for the integrity of the data and the accuracy of the analysis, and mainly wrote the paper. S.T., M.K., and J.F. built the optical device and performed the optical measurements. S.T. and M.K. carried out the data analysis. M.F. helped during the data acquisition and provided clinical information and CT scans. A.R., J.B., and M.R. were responsible for the patient recruitment and management. M.A.P. and T.D. were responsible for the study design and contributed to the result interpretation and writing of the paper. All authors read and approved the final paper. Author DisclaimerTurgut Durduran is an inventor on relevant patents. ICFO had equity ownership in the spinoff company HemoPhotonics S.L. Potential financial conflicts of interest and objectivity of research were monitored by ICFO’s Knowledge & Technology Transfer Department. No financial conflicts of interest were identified. AcknowledgmentsWe gratefully acknowledge the collaborators and colleagues from the Department of Anesthesiology and Neurosurgery as well as the Neurotraumatology and Neurosurgery Research Unit at VHUH Barcelona for interesting and encouraging discussions and their great support throughout the whole study. In particular, the authors also thank the whole team of nurses of the Department of the Neurotrauma-ICU of the Vall d’Hebron University Hospital, who helped in the preparation and handling of patients during the measurements. This research was funded by the European Union’s Horizon 2020 research and innovation programme under grant agreements No. 675332 (BitMap), No. 101016087 (VASCOVID) and No. 101017113 (TinyBRAINS). Moreover, this work was funded by: Fundació CELLEX Barcelona, Fundació Mir-Puig, Agencia Estatal de Investigación (PHOTOMETABO, PID2019-106481RBC31/10.13039/501100011033), the “Severo Ochoa” Programme for Centres of Excellence in R&D (CEX2019-000910-S); LUX4MED and MEDLUX special programs; Generalitat de Catalunya (CERCA, AGAUR-2017-SGR-1380, RIS3CAT-001-P-001682 CECH, and AGAUR-2021SGR/00810), FEDER EC and LASERLAB-EUROPE V (EC H2020 no. 871124), KidsBrainIT (ERA-NET NEURON), la Fundació La Marató de TV3 (201724.31, 201709.31 and 202109-30), “PLEC2022-009290, SafeICP” project funded by MCIN/AEI/10.13039/501100011033 and by the “European Union NextGenerationEU/PRTR.” The study was partially financed by the National Science Centre of Poland (NCN) in the framework of project 2019/33/B/ST7/01387. ReferencesW. Cheong, S. Prahl and A. Welch,
“A review of the optical properties of tissues,”
IEEE J. Quantum Electron., 26 2166
–2185 https://doi.org/10.1109/3.64354 IEJQA7 0018-9197
(1991).
Google Scholar
T. J. Farrell, M. S. Patterson and B. A. Wilson,
“Diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivo,”
Med. Phys., 19 879 https://doi.org/10.1118/1.596777 MPHYA6 0094-2405
(1992).
Google Scholar
T. Durduran et al.,
“Diffuse optics for tissue monitoring and tomography,”
Rep. Prog. Phys., 73
(7), 076701 https://doi.org/10.1088/0034-4885/73/7/076701 RPPHAG 0034-4885
(2010).
Google Scholar
F. Martelli et al., Light Propagation Through Biological Tissue and Other Diffusive Media: Theory, Solutions, and Software, SPIE Press, Bellingham, Washington
(2009). Google Scholar
L. V. Wang and H. Wu, Biomedical Optics: Principles and Imaging, John Wiley & Sons, Hoboken
(2007). Google Scholar
V. Ntziachristos,
“Going deeper than microscopy: the optical imaging frontier in biology,”
Nat. Methods, 7 603
–614 https://doi.org/10.1038/nmeth.1483 1548-7091
(2010).
Google Scholar
L. Gagnon et al.,
“Double-layer estimation of intra- and extracerebral hemoglobin concentration with a time-resolved system,”
J. Biomed. Opt., 13
(5), 054019 https://doi.org/10.1117/1.2982524 JBOPFO 1083-3668
(2008).
Google Scholar
A. Kienle and M. S. Patterson,
“Improved solutions of the steady-state and the time-resolved diffusion equations for reflectance from a semi-infinite turbid medium,”
J. Opt. Soc. Am. A, 14
(1), 246
–254 https://doi.org/10.1364/JOSAA.14.000246
(1997).
Google Scholar
H. Dehghani et al.,
“Depth sensitivity and image reconstruction analysis of dense imaging arrays for mapping brain function with diffuse optical tomography,”
Appl. Opt., 48
(10), D137
–43 https://doi.org/10.1364/AO.48.00D137 APOPAI 0003-6935
(2009).
Google Scholar
S. L. Ferradal et al.,
“Atlas-based head modeling and spatial normalization for high-density diffuse optical tomography: in vivo validation against fMRI,”
NeuroImage, 85 117
–126 https://doi.org/10.1016/j.neuroimage.2013.03.069 NEIMEF 1053-8119
(2014).
Google Scholar
A. H. Barnett et al.,
“Robust inference of baseline optical properties of the human head with three-dimensional segmentation from magnetic resonance imaging,”
Appl. Opt., 42 3095
–3108 https://doi.org/10.1364/AO.42.003095 APOPAI 0003-6935
(2003).
Google Scholar
A. Ishimaru,
“Wave propagation and scattering in random media and rough surfaces,”
Proc. IEEE, 79
(10), 1359
–1366 https://doi.org/10.1109/5.104210 IEEPAD 0018-9219
(1991).
Google Scholar
D. Boas,
“Diffuse photon probes of structural and dynamical properties of turbid media: theory and biomedical application,”
260
(1996). Google Scholar
K. R. Ward et al.,
“Near infrared spectroscopy for evaluation of the trauma patient: a technology review,”
Resuscitation, 68 27
–44 https://doi.org/10.1016/j.resuscitation.2005.06.022 RSUSBS 0300-9572
(2006).
Google Scholar
N. Ajayan et al.,
“Limitations of near infrared spectroscopy (NIRS) in neurosurgical setting: our case experience,”
J. Clin. Monit. Comput., 33 743
–746 https://doi.org/10.1007/s10877-018-0209-1
(2019).
Google Scholar
H. Grocott and S. Davie,
“Future uncertainties in the development of clinical cerebral oximetry,”
Front. Physiol., 4 360 https://doi.org/10.3389/fphys.2013.00360 FROPBK 0301-536X
(2013).
Google Scholar
G. Greisen,
“Is near-infrared spectroscopy living up to its promises?,”
Semin. Fetal. Neonatal. Med., 11
(6), 498
–502 https://doi.org/10.1016/j.siny.2006.07.010
(2006).
Google Scholar
A. I. R. Maas and G. Citerio,
“Noninvasive monitoring of cerebral oxygenation in traumatic brain injury: a mix of doubts and hope,”
Intensive Care Med., 36
(8), 1283
–1285 https://doi.org/10.1007/s00134-010-1921-6 ICMED9 0342-4642
(2010).
Google Scholar
S. Leal-Noval and M. Jiménez Sánchez,
“La transfusión de hematíes incrementa la oxigenación tisular y mejora el resultado clínico (CON),”
Med. Int., 34 471
–475 https://doi.org/10.1016/j.medin.2010.06.001
(2010).
Google Scholar
K. Büchner et al.,
“Near-infrared spectroscopy—not useful to monitor cerebral oxygenation after severe brain injury,”
Zentralbl. Neurochir., 61
(2), 69
–73 https://doi.org/10.1055/s-2000-8262
(2000).
Google Scholar
S. B. Lewis et al.,
“Cerebral oxygenation monitoring by near-infrared spectroscopy is not clinically useful in patients with severe closed-head injury: a comparison with jugular venous bulb oximetry,”
Crit. Care Med., 24
(8), 1334
–1338 https://doi.org/10.1097/00003246-199608000-00011 CCMDC7 0090-3493
(1996).
Google Scholar
J. Meixensberger et al.,
“Multimodal hemodynamic neuromonitoring—quality and consequences for therapy of severely head injured patients,”
Acta Neurochir. Suppl., 71 260
–262 https://doi.org/10.1007/978-3-7091-6475-4_75 ANCSBM 0065-1419
(1998).
Google Scholar
A. E. R. Young et al.,
“Behaviour of near-infrared light in the adult human head: implications for clinical near-infrared spectroscopy,”
Br. J. Anaesth., 84 38
–42 https://doi.org/10.1093/oxfordjournals.bja.a013379 BJANAD 0007-0912
(2000).
Google Scholar
Y. Ogoshi et al.,
“Analysis of light propagation in adult head model by direct hybrid Monte Carlo-diffusion method,”
in IEEE EMBS Asian-Pac. Conf. Biomed. Eng. of 2003,
316
–317
(2003). https://doi.org/10.1109/APBME.2003.1302711 Google Scholar
E. Okada and D. T. Delpy,
“Near-infrared light propagation in an adult head model. II. Effect of superficial tissue thickness on the sensitivity of the near-infrared spectroscopy signal,”
Appl. Opt., 42
(16), 2915
–2922 https://doi.org/10.1364/AO.42.002915 APOPAI 0003-6935
(2003).
Google Scholar
K. Verdecchia et al.,
“Assessment of a multi-layered diffuse correlation spectroscopy method for monitoring cerebral blood flow in adults,”
Biomed. Opt. Express, 7
(9), 3659
–3674 https://doi.org/10.1364/BOE.7.003659 BOEICL 2156-7085
(2016).
Google Scholar
H. Zhao, E. Sathialingam and E. M. Buckley,
“Accuracy of diffuse correlation spectroscopy measurements of brain blood flow when using a three-layer analytical model,”
in Biophotonics Congr. of 2021,
BF1B.4
(2021). Google Scholar
D. Ancora et al.,
“Noninvasive optical estimation of CSF thickness for brain-atrophy monitoring,”
Biomed. Opt. Express, 9
(9), 4094
–4112 https://doi.org/10.1364/BOE.9.004094 BOEICL 2156-7085
(2018).
Google Scholar
A. Sudakou et al.,
“Two-layered blood-lipid phantom and method to determine absorption and oxygenation employing changes in moments of DTOFs,”
Biomed. Opt. Express, 14 3506
–3531 https://doi.org/10.1364/BOE.492168 BOEICL 2156-7085
(2023).
Google Scholar
A. Liebert et al.,
“Monte Carlo algorithm for efficient simulation of time-resolved fluorescence in layered turbid media,”
Opt. Express, 16 13188
–13202 https://doi.org/10.1364/OE.16.013188 OPEXFF 1094-4087
(2008).
Google Scholar
S. Wojtkiewicz and A. Liebert,
“Parallel, multi-purpose Monte Carlo code for simulation of light propagation in segmented tissues,”
Biocybern. Biomed. Eng., 41
(4), 1303
–1321 https://doi.org/10.1016/j.bbe.2021.03.001
(2021).
Google Scholar
A. Farina et al.,
“In-vivo multilaboratory investigation of the optical properties of the human head,”
Biomed. Opt. Express, 6
(7), 2609
–23 https://doi.org/10.1364/BOE.6.002609 BOEICL 2156-7085
(2015).
Google Scholar
A. Torricelli et al.,
“Time domain functional NIRS imaging for human brain mapping,”
NeuroImage, 85 28
–50 https://doi.org/10.1016/j.neuroimage.2013.05.106 NEIMEF 1053-8119
(2014).
Google Scholar
S. Tagliabue et al.,
“Transcranial, non-invasive evaluation of potential misery perfusion during hyperventilation therapy of traumatic brain injury patients,”
J. NeurotraumaJ. Neurotrauma, 4041
(19–2011–12), 20731469
–20861471
(2023). https://doi.org/10.1089/neu.2022.0419
(2024).
Google Scholar
C. Lindner et al.,
“Diffuse optical characterization of the healthy human thyroid tissue and two pathological case studies,”
PLoS One, 11
(1), e0147851 https://doi.org/10.1371/journal.pone.0147851 POLNCL 1932-6203
(2016).
Google Scholar
S. Tagliabue et al.,
“Comparison of cerebral metabolic rate of oxygen, blood flow and bispectral index under general anesthesia,”
Neurophotonics, 10
(1), 015006 https://doi.org/10.1117/1.NPh.10.1.015006
(2023).
Google Scholar
D. A. Boas et al.,
“Establishing the diffuse correlation spectroscopy signal relationship with blood flow,”
Neurophotonics, 3
(3), 031412 https://doi.org/10.1117/1.NPh.3.3.031412
(2016).
Google Scholar
World Medical Association,
“World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects,”
JAMA, 310
(20), 2191
–2194 https://doi.org/10.1001/jama.2013.281053
(2013).
Google Scholar
M. S. Patterson, B. Chance and B. C. Wilson,
“Time resolved reflectance and transmittance for the non-invasive measurement of tissue optical properties,”
Appl. Opt., 28
(12), 2331
–2336 https://doi.org/10.1364/AO.28.002331 APOPAI 0003-6935
(1989).
Google Scholar
J. C. Lagarias et al.,
“Convergence properties of the Nelder–Mead simplex method in low dimensions,”
SIAM J. Optim., 9 112
–147 https://doi.org/10.1137/S1052623496303470 SJOPE8 1095-7189
(1998).
Google Scholar
Y. Yamada, H. Suzuki and Y. Yamashita,
“Time-domain near-infrared spectroscopy and imaging: a review,”
Appl. Sci., 9
(6), 1127 https://doi.org/10.3390/app9061127
(2019).
Google Scholar
H. Wabnitz et al.,
“Performance assessment of time-domain optical brain imagers, part 1: basic instrumental performance protocol,”
J. Biomed. Opt., 19
(8), 086010 https://doi.org/10.1117/1.JBO.19.8.086010 JBOPFO 1083-3668
(2014).
Google Scholar
H. Wabnitz et al.,
“Performance assessment of time-domain optical brain imagers, part 2: nEUROPt protocol,”
J. Biomed. Opt., 19
(8), 086012 https://doi.org/10.1117/1.JBO.19.8.086012 JBOPFO 1083-3668
(2014).
Google Scholar
H. Wabnitz et al.,
“Depth-selective data analysis for time-domain fNIRS: moments vs. time windows,”
Biomed. Opt. Express, 11
(8), 4224
–4243 https://doi.org/10.1364/BOE.396585 BOEICL 2156-7085
(2020).
Google Scholar
M. Helton, M. Mycek and K. Vishwanath,
“Reconstruction of optical coefficients in turbid media using time-resolved reflectance and calibration-free instrument response functions,”
Biomed. Opt. Express, 13
(3), 1595
–1608 https://doi.org/10.1364/BOE.447685 BOEICL 2156-7085
(2022).
Google Scholar
H. Wabnitz et al.,
“Performance assessment of time-domain optical brain imagers, part 1: Basic instrumental performance protocol,”
J. Biomed. Opt., 19
(8), 086010 https://doi.org/10.1117/1.JBO.19.8.086010 JBOPFO 1083-3668
(2014).
Google Scholar
A. Ishimaru,
“Diffusion of a pulse in densely distributed scatterers,”
J. Opt. Soc. Am., 68 1045
–1050 https://doi.org/10.1364/JOSA.68.001045 JOSAAH 0030-3941
(1978).
Google Scholar
W. Hoffman et al.,
“Hypoxic brain tissue following subarachnoid hemorrhage,”
Anesthesiology, 92
(2), 442
–442 https://doi.org/10.1097/00000542-200002000-00026 ANESAV 0003-3022
(2000).
Google Scholar
C. B. Dabdoub et al.,
“Review of the management of pneumocephalus,”
Surg. Neurol. Int., 6 155 https://doi.org/10.4103/2152-7806.166195
(2015).
Google Scholar
I. Pirovano et al.,
“High power time domain fNIRs device,”
in Transl., Microsc., OCT, OTS, BRAIN, OSA Tech. Digest, in Biophotonics Congr.: Biomed. Opt. 2020,
JTu3A.19
(2020). Google Scholar
M. Buttafava et al.,
“A compact two-wavelength time-domain NIRS system based on SiPM and pulsed diode lasers,”
IEEE Photonics J., 9
(1), 7800114 https://doi.org/10.1109/JPHOT.2016.2632061
(2017).
Google Scholar
H. Y. Ban et al.,
“Kernel flow: a high channel count scalable time-domain functional near-infrared spectroscopy system,”
J. Biomed. Opt., 27
(7), 074710 https://doi.org/10.1117/1.JBO.27.7.074710 JBOPFO 1083-3668
(2022).
Google Scholar
L. Cortese et al.,
“The LUCA device: a multi-modal platform combining diffuse optics and ultrasound imaging for thyroid cancer screening,”
Biomed. Opt. Express, 12 3392
–3409 https://doi.org/10.1364/BOE.416561 BOEICL 2156-7085
(2021).
Google Scholar
S. A. Carp, M. B. Robinson and M. A. Franceschini,
“Diffuse correlation spectroscopy: current status and future outlook,”
Neurophotonics, 1
(10), 013509 https://doi.org/10.1117/1.NPh.10.1.013509
(2023).
Google Scholar
W. Liu et al.,
“Fast and sensitive diffuse correlation spectroscopy with highly parallelized single photon detection,”
APL Photonics, 6 026106 https://doi.org/10.1063/5.0031225
(2021).
Google Scholar
M. Giovannella et al.,
“Accuracy and precision of tissue optical properties and hemodynamic parameters estimated by the BabyLux device: a hybrid time-resolved near-infrared and diffuse correlation spectroscopy neuro-monitor,”
Biomed. Opt. Express, 10 2556
–2579 https://doi.org/10.1364/BOE.10.002556 BOEICL 2156-7085
(2019).
Google Scholar
G. Giacalone et al.,
“Cerebral time domain-NIRS: reproducibility analysis, optical properties, hemoglobin species and tissue oxygen saturation in a cohort of adult subjects,”
Biomed. Opt. Express, 8
(11), 4987
–5000 https://doi.org/10.1364/BOE.8.004987 BOEICL 2156-7085
(2017).
Google Scholar
A. Ruesch et al.,
“Estimating intracranial pressure using pulsatile cerebral blood flow measured with diffuse correlation spectroscopy,”
Biomed. Opt. Express, 11
(3), 1462
–1476 https://doi.org/10.1364/BOE.386612 BOEICL 2156-7085
(2020).
Google Scholar
A. Parthasarathy et al.,
“Clinical applications of high-speed blood flow measurements with diffuse correlation spectroscopy,”
10059
(2017). Google Scholar
A. Tosi et al.,
“Fast-gated single-photon counting technique widens dynamic range and speeds up acquisition time in time-resolved measurements,”
Opt. Express, 19
(11), 10735
–10746 https://doi.org/10.1364/OE.19.010735 OPEXFF 1094-4087
(2011).
Google Scholar
A. Dalla Mora et al.,
“Time-gated single-photon detection in time-domain diffuse optics: a review,”
Appl. Sci., 10
(3), 1101 https://doi.org/10.3390/app10031101
(2020).
Google Scholar
L. Cortese et al.,
“Recipes for diffuse correlation spectroscopy instrumentation design based on signal-to-noise ratio and precision targets,”
Biomed. Opt. Express, 12 3265
–3281 https://doi.org/10.1364/BOE.423071 BOEICL 2156-7085
(2021).
Google Scholar
V. Ntziachristos and B. Chance,
“Accuracy limits in the determination of absolute optical properties using time-resolved NIR spectroscopy,”
Med. Phys., 28
(6), 1115
–1124 https://doi.org/10.1118/1.1373674 MPHYA6 0094-2405
(2001).
Google Scholar
E. Alerstam, S. Andersson-Engels and T. Svensson,
“Improved accuracy in time-resolved diffuse reflectance spectroscopy,”
Opt. Express, 16 10440
–10448 https://doi.org/10.1364/OE.16.010440 OPEXFF 1094-4087
(2008).
Google Scholar
J. H. Choi et al.,
“Noninvasive determination of the optical properties of adult brain: near-infrared spectroscopy approach,”
J. Biomed. Opt., 9
(1), 221
–229 https://doi.org/10.1117/1.1628242 JBOPFO 1083-3668
(2004).
Google Scholar
R. Francis et al.,
“NIR light propagation in a digital head model for traumatic brain injury (TBI),”
Biomed. Opt. Express, 6 3256 https://doi.org/10.1364/BOE.6.003256 BOEICL 2156-7085
(2015).
Google Scholar
L. Gagnon et al.,
“Investigation of diffuse correlation spectroscopy in multi-layered media including the human head,”
Opt. Express, 16
(20), 15514
–15530 https://doi.org/10.1364/OE.16.015514 OPEXFF 1094-4087
(2008).
Google Scholar
J. Selb et al.,
“Sensitivity of near-infrared spectroscopy and diffuse correlation spectroscopy to brain hemodynamics: simulations and experimental findings during hypercapnia,”
Neurophotonics, 1
(1), 015005 https://doi.org/10.1117/1.NPh.1.1.015005
(2014).
Google Scholar
W. B. Baker et al.,
“Modified Beer-Lambert law for blood flow,”
Biomed. Opt. Express, 5
(11), 4053 https://doi.org/10.1364/BOE.5.004053 BOEICL 2156-7085
(2014).
Google Scholar
J. Li et al.,
“Noninvasive detection of functional brain activity with near-infrared diffusing-wave spectroscopy,”
J. Biomed. Opt., 10
(4), 044002 https://doi.org/10.1117/1.2007987 JBOPFO 1083-3668
(2005).
Google Scholar
M. M. Wu et al.,
“Improved accuracy of cerebral blood flow quantification in the presence of systemic physiology cross-talk using multi-layer Monte Carlo modeling,”
Neurophotonics, 8
(1), 015001 https://doi.org/10.1117/1.NPh.8.1.015001
(2021).
Google Scholar
M. G. Rao et al.,
“Dating of early subdural haematoma: a correlative clinico-radiological study,”
J. Clin. Diagn. Res., 10
(4), HC01
–HC05 https://doi.org/10.7860/JCDR/2016/17207.7644
(2016).
Google Scholar
P. Giacometti and S. G. Diamond,
“Compliant head probe for positioning electroencephalography electrodes and near-infrared spectroscopy optodes,”
J. Biomed. Opt., 18
(2), 027005 https://doi.org/10.1117/1.JBO.18.2.027005 JBOPFO 1083-3668
(2013).
Google Scholar
C. M. Aasted et al.,
“Anatomical guidance for functional near-infrared spectroscopy: Atlasviewer tutorial,”
Neurophotonics, 2
(2), 020801 https://doi.org/10.1117/1.NPh.2.2.020801
(2015).
Google Scholar
M. M. Wu et al.,
“Complete head cerebral sensitivity mapping for diffuse correlation spectroscopy using subject-specific magnetic resonance imaging models,”
Biomed. Opt. Express, 13
(3), 1131
–1151 https://doi.org/10.1364/BOE.449046
(2022).
Google Scholar
G. Giacalone et al.,
“Time-domain near-infrared spectroscopy in acute ischemic stroke patients,”
Neurophotonics, 6
(1), 015003 https://doi.org/10.1117/1.NPh.6.1.015003
(2019).
Google Scholar
H. Zhao et al.,
“A wide field-of-view, modular, high-density diffuse optical tomography system for minimally constrained three-dimensional functional neuroimaging,”
Biomed. Opt. Express, 11
(8), 4110
–4129 https://doi.org/10.1364/BOE.394914
(2020).
Google Scholar
M. Renna et al.,
“A contact-sensitive probe for biomedical optics,”
Sensors, 22
(6), 2361 https://doi.org/10.3390/s22062361 SNSRES 0746-9462
(2022).
Google Scholar
M. Giovannella et al.,
“BabyLux device: a diffuse optical system integrating diffuse correlation spectroscopy and time-resolved near-infrared spectroscopy for the neuromonitoring of the premature newborn brain,”
Neurophotonics, 6
(2), 025007 https://doi.org/10.1117/1.NPh.6.2.025007
(2019).
Google Scholar
M. Zanoletti et al.,
“Vascovid: an integrated platform to evaluate endothelial and microvascular impairment in severe COVID-19 patients,”
in Eur. Conf. Biomed. Opt. 2021 (ECBO),
ETh3A.4
(2021). Google Scholar
A. Rickard et al.,
“A clinical study to assess diffuse reflectance spectroscopy with an auto-calibrated, pressure-sensing optical probe in head and neck cancer,”
Curr. Oncol., 30
(3), 2751
–2760 https://doi.org/10.3390/curroncol30030208
(2023).
Google Scholar
J. B. Fischer,
“Transcranial diffuse optical measurements of pulsatility derived parameters for neuromonitoring applications,”
(2021). Google Scholar
G. Schwarz et al.,
“Cerebral oximetry in dead subjects,”
J. Neurosurg. Anesthesiol., 8
(3), 189
–193 https://doi.org/10.1097/00008506-199607000-00001 JNANEV 0898-4921
(1996).
Google Scholar
A. Caccioppola et al.,
“Ultrasound-tagged near-infrared spectroscopy does not disclose absent cerebral circulation in brain-dead adults,”
Br. J. Anaesth., 121
(3), 588
–594 https://doi.org/10.1016/j.bja.2018.04.038 BJANAD 0007-0912
(2018).
Google Scholar
P. G. Al-Rawi, P. Smielewski and P. J. Kirkpatrick,
“Evaluation of a near-infrared spectrometer (NIRO 300) for the detection of intracranial oxygenation changes in the adult head,”
Stroke, 32
(11), 2492
–2500 https://doi.org/10.1161/hs1101.098356 SJCCA7 0039-2499
(2001).
Google Scholar
D. J. Davies et al.,
“Near-infrared spectroscopy in the monitoring of adult traumatic brain injury: a review,”
J. Neurotrauma, 44 150417112024001 https://doi.org/10.1089/neu.2014.3748 JNEUE4 0897-7151
(2015).
Google Scholar
F. Scholkmann et al.,
“Absolute values of optical properties of human head tissue: dependence on head region and individual,”
Adv. Exp. Med. Biol., 1072 325
–330 AEMBAP 0065-2598
(2018).
Google Scholar
J. M. Murkin and M. Arango,
“Near-infrared spectroscopy as an index of brain and tissue oxygenation.,”
Br. J. Anaesth., 103
(Suppl.), i3
–i13 https://doi.org/10.1093/bja/aep299 BJANAD 0007-0912
(2009).
Google Scholar
V. McCredie et al.,
“The impact of red blood cell transfusion on cerebral tissue oxygen saturation in severe traumatic brain injury,”
Neurocrit. Care, 26 247
–255 https://doi.org/10.1007/s12028-016-0310-6
(2017).
Google Scholar
A. N. Sen, S. P. Gopinath and C. S. Robertson,
“Clinical application of near-infrared spectroscopy in patients with traumatic brain injury: a review of the progress of the field,”
Neurophotonics, 3
(3), 031409 https://doi.org/10.1117/1.NPh.3.3.031409
(2016).
Google Scholar
S. P. Gopinath et al.,
“Early detection of delayed traumatic intracranial hematomas using near-infrared spectroscopy,”
J. Neurosurg., 83
(3), 438
–444 https://doi.org/10.3171/jns.1995.83.3.0438 JONSAC 0022-3085
(1995).
Google Scholar
C. Liang et al.,
“Chinese military evaluation of a portable near-infrared detector of traumatic intracranial hematomas,”
Mil. Med., 183 e318
–e323 https://doi.org/10.1093/milmed/usx088 MMEDA9 0026-4075
(2018).
Google Scholar
E. Okada and D. T. Delpy,
“Near-infrared light propagation in an adult head model. I. Modeling of low-level scattering in the cerebrospinal fluid layer,”
Appl. Opt., 42
(16), 2906
–2914 https://doi.org/10.1364/AO.42.002906 APOPAI 0003-6935
(2003).
Google Scholar
P. Sawosz et al.,
“Human skull translucency: post mortem studies,”
Biomed. Opt. Express, 7 5010
–5020 https://doi.org/10.1364/BOE.7.005010 BOEICL 2156-7085
(2016).
Google Scholar
F. Moreau et al.,
“Near-infrared measurements of brain oxygenation in stroke,”
Neurophotonics, 3
(3), 031403 https://doi.org/10.1117/1.NPh.3.3.031403
(2016).
Google Scholar
G. Yu et al., Near-Infrared Diffuse Correlation Spectroscopy for Assessment of Tissue Blood Flow, CRC Press(
(2016). Google Scholar
E. M. Buckley,
“Cerebral hemodynamics in high-risk neonates probed by diffuse optical spectroscopies,”
(2011).
Google Scholar
L. Shoemaker et al.,
“Using depth-enhanced diffuse correlation spectroscopy and near-infrared spectroscopy to isolate cerebral hemodynamics during transient hypotension,”
Neurophotonics, 10
(2), 025013 https://doi.org/10.1117/1.NPh.10.2.025013
(2023).
Google Scholar
D. Wang et al.,
“Optimization of detected optical intensity for measurement of diffuse correlation spectroscopy: intralipid phantom study,”
AIP Adv., 9
(1), 015315 https://doi.org/10.1063/1.5016679 AAIDBI 2158-3226
(2019).
Google Scholar
P. A. Lemieux and D. J. Durian,
“Investigating non-gaussian scattering processes by using nth-order intensity correlation functions,”
J. Opt., 16 7
(1999).
Google Scholar
C. Zhou et al.,
“Diffuse optical correlation tomography of cerebral blood flow during cortical spreading depression in rat brain,”
Opt. Express, 14
(3), 1125
–1144 https://doi.org/10.1364/oe.14.001125
(2006).
Google Scholar
M. Giovannella et al.,
“Validation of diffuse correlation spectroscopy against -water PET for regional cerebral blood flow measurement in neonatal piglets,”
J. Cereb. Blood Flow Metab., 40
(10), 2055
–2065 https://doi.org/10.1177/0271678X19883751
(2020).
Google Scholar
R. G. W. Brown, K. D. Ridley and J. G. Rarity,
“Characterization of silicon avalanche photodiodes for photon correlation measurements. 1: passive quenching,”
Appl. Opt., 25 4122
–4126 https://doi.org/10.1364/AO.25.004122 APOPAI 0003-6935
(1986).
Google Scholar
P. R. Donoso,
“Theoretical study of the effect of detector and correlator defects on diffuse correlation spectroscopy for non-invasive blood flow measurements,”
(2015). Google Scholar
B. Chance et al.,
“Comparison of time-resolved and -unresolved measurements of deoxyhemoglobin in brain,”
Proc. Natl. Acad. Sci. U. S. A., 85 4971
–4975 https://doi.org/10.1073/pnas.85.14.4971
(1988).
Google Scholar
P. van der Zee, M. Essenpreis and D. T. Delpy,
“Optical properties of brain tissue,”
in Photonics West: Lasers and Appl. in Sci. and Eng.,
(1993). Google Scholar
D. Irwin et al.,
“Influences of tissue absorption and scattering on diffuse correlation spectroscopy blood flow measurements,”
Biomed. Opt. Express, 2
(7), 1969
–1985 https://doi.org/10.1364/BOE.2.001969 BOEICL 2156-7085
(2011).
Google Scholar
M. Diop et al.,
“Calibration of diffuse correlation spectroscopy with a time-resolved near-infrared technique to yield absolute cerebral blood flow measurements,”
Biomed. Opt. Express, 2
(7), 2068
–2081 https://doi.org/10.1364/BOE.2.002068 BOEICL 2156-7085
(2011).
Google Scholar
BiographySusanna Tagliabue obtained her PhD from the ICFO in the Medical Optics Group, Barcelona, Spain, and was a postdoc in the same group, focusing on the development of hybrid diffuse optical devices for clinical studies, particularly on the injured brain. She received her bachelor’s (2014) and master’s (2016) degrees in physics engineering from the Politecnico di Milano. She is currently an R&D engineer at NIRx Medical Technologies. Michał Kacprzak received his MSc degree in electrical engineering from the Warsaw University of Technology in 2002 and his PhD from the Institute of Biocybernetics and Biomedical Engineering in 2011, where he led the Laboratory of Biomedical Optics from 2012 to 2015. In 2015, he moved to the ICFO in Barcelona as a postdoctoral researcher for 2 years. His main area of interest is the application of diffuse optics for monitoring and diagnosis of the brain. Anna Rey-Perez obtained her bachelor of medicine and surgery from the Autonomous University of Barcelona. She is a specialist in intensive medicine, trained at the Hospital del Mar in Barcelona. She currently serves in the Intensive Care Unit (ICU) of the Hospital for Traumatology, Rehabilitation and Burns at Vall d’Hebron Hospital, where she is responsible for the neurocritical patient and safety referent while collaborating with the Neurotraumatology and Neurosurgery Unit (UNINN). Jacinto Baena is a medical specialist in intensive medicine and is affiliated with the ICU of Neurotraumatology and Major Burns of the Vall d’Hebron University Hospital in Barcelona. He collaborates with the UNINN and as a teacher in the IEM-UAB Master and the UB Master. Marilyn Riveiro is a medical specialist in intensive care medicine at the Neurotraumatology ICU of the Vall d’Hebron University Hospital, Barcelona, and a member of the UNINN. She collaborated as a teacher in the IEM-UAB master’s degree in comprehensive urgencies and emergency care (from 2013 to 2019) and in the UB master’s degree in critical illness and emergencies with the subject of head trauma. She currently performs hospital management functions as a coordinator of critical care at the Vall d’Hebrón Traumatology, Rehabilitation and Burns Hospital. Federica Maruccia received her PhD from the UNINN of the Vall d’Hebron Research Institute (VHIR), Barcelona, Spain. She received her master’s degree in rehabilitation therapy from the University of Pisa in 2013. The objective of her research was to evaluate the psychomotor development of babies and children affected by neurological diseases and to collaborate in the application of non-invasive technologies to better understand their pathological frame. Jonas B. Fischer received his PhD from ICFO, Barcelona, Spain, in 2021. He was a joint PhD student between the Medical Optics Group at ICFO and HemoPhotonics S.L. His studies focused on neuromonitoring applications in the intensive care unit and stroke unit. Particularly, he developed hardware, algorithms, and new tools related to fast, non-invasive measurements using near-infrared light of pulsatile cerebral blood flow due to the cardiac cycle. Maria A. Poca is the clinical head of neurosurgery at the Vall d’Hebron University Hospital (VHUH) and associate professor at the Universitat Autónoma de Barcelona (UAB). She is the coordinator of the UNINN of the VHIR and is responsible for research lines on hydrocephalus and craniocervical malformations. She is the coordinator of the pediatric neurosurgery unit since 2008 that has been designed as a reference center for complex pediatric neurosurgery in 2013. Turgut Durduran was trained at the University of Pennsylvania. In 2009, he moved to ICFO–The Institute of Photonic Sciences, Spain, where he leads the Medical Optics Group. His research interests revolve around the use of diffuse light to non-invasively probe tissue function. The group develops new technologies and algorithms and routinely translates them into preclinical, clinical, and industrial applications. |
Brain
Tissues
Computed tomography
Neurophotonics
Data modeling
Autocorrelation
Head