13 October 2016 Transmission electron microscope calibration methods for critical dimension standards
Ndubuisi G. Orji, Ronald G. Dixson, Domingo I. Garcia-Gutierrez, Benjamin D. Bunday, Michael Bishop, Michael W. Cresswell, Richard A. Allen, John A. Allgair
Author Affiliations +
Abstract
One of the key challenges in critical dimension (CD) metrology is finding suitable dimensional calibration standards. The transmission electron microscope (TEM), which produces lattice-resolved images having scale traceability to the SI (International System of Units) definition of length through an atomic lattice constant, has gained wide usage in different areas of CD calibration. One such area is critical dimension atomic force microscope (CD-AFM) tip width calibration. To properly calibrate CD-AFM tip widths, errors in the calibration process must be quantified. Although the use of TEM for CD-AFM tip width calibration has been around for about a decade, there is still confusion on what should be considered in the uncertainty analysis. We characterized CD-AFM tip-width samples using high-resolution TEM and high angle annular dark field scanning TEM and two CD-AFMs that are implemented as reference measurement systems. The results are used to outline how to develop a rigorous uncertainty estimate for TEM/CD-AFM calibration, and to compare how information from the two electron microscopy modes are applied to practical CD-AFM measurements. The results also represent a separate validation of previous TEM/CD-AFM calibration. Excellent agreement was observed.
© 2016 Society of Photo-Optical Instrumentation Engineers (SPIE) 1932-5150/2016/$25.00 © 2016 SPIE
Ndubuisi G. Orji, Ronald G. Dixson, Domingo I. Garcia-Gutierrez, Benjamin D. Bunday, Michael Bishop, Michael W. Cresswell, Richard A. Allen, and John A. Allgair "Transmission electron microscope calibration methods for critical dimension standards," Journal of Micro/Nanolithography, MEMS, and MOEMS 15(4), 044002 (13 October 2016). https://doi.org/10.1117/1.JMM.15.4.044002
Published: 13 October 2016
Lens.org Logo
CITATIONS
Cited by 19 scholarly publications and 3 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Transmission electron microscopy

Electron microscopes

Error analysis

Silicon

Line width roughness

Oxides

RELATED CONTENT

Comprehensive approach to MuGFET metrology
Proceedings of SPIE (March 24 2006)
Silicon single atom steps as AFM height standards
Proceedings of SPIE (August 22 2001)
Reference metrology using a next-generation CD-AFM
Proceedings of SPIE (May 24 2004)
TEM calibration methods for critical dimension standards
Proceedings of SPIE (April 05 2007)
CD-AFM reference metrology at NIST and SEMATECH
Proceedings of SPIE (May 10 2005)

Back to Top