3 April 2024 Swept-source Raman spectroscopy of chemical and biological materials
Author Affiliations +
Abstract

Significance

Raman spectroscopy has been used as a powerful tool for chemical analysis, enabling the noninvasive acquisition of molecular fingerprints from various samples. Raman spectroscopy has proven to be valuable in numerous fields, including pharmaceutical, materials science, and biomedicine. Active research and development efforts are currently underway to bring this analytical instrument into the field, enabling in situ Raman measurements for a wider range of applications. Dispersive Raman spectroscopy using a fixed, narrowband source is a common method for acquiring Raman spectra. However, dispersive Raman spectroscopy requires a bulky spectrometer, which limits its field applicability. Therefore, there has been a tremendous need to develop a portable and sensitive Raman system.

Aim

We developed a compact swept-source Raman (SS-Raman) spectroscopy system and proposed a signal processing method to mitigate hardware limitations. We demonstrated the capabilities of the SS-Raman spectroscopy by acquiring Raman spectra from both chemical and biological samples. These spectra were then compared with Raman spectra obtained using a conventional dispersive Raman spectroscopy system.

Approach

The SS-Raman spectroscopy system used a wavelength-swept source laser (822 to 842 nm), a bandpass filter with a bandwidth of 1.5 nm, and a low-noise silicon photoreceiver. Raman spectra were acquired from various chemical samples, including phenylalanine, hydroxyapatite, glucose, and acetaminophen. A comparative analysis with the conventional dispersive Raman spectroscopy was conducted by calculating the correlation coefficients between the spectra from the SS-Raman spectroscopy and those from the conventional system. Furthermore, Raman mapping was obtained from cross-sections of swine tissue, demonstrating the applicability of the SS-Raman spectroscopy in biological samples.

Results

We developed a compact SS-Raman system and validated its performance by acquiring Raman spectra from both chemical and biological materials. Our straightforward signal processing method enhanced the quality of the Raman spectra without incurring high costs. Raman spectra in the range of 900 to 1200 cm1 were observed for phenylalanine, hydroxyapatite, glucose, and acetaminophen. The results were validated with correlation coefficients of 0.88, 0.84, 0.87, and 0.73, respectively, compared with those obtained from dispersive Raman spectroscopy. Furthermore, we performed scans across the cross-section of swine tissue to generate a biological tissue mapping plot, providing information about the composition of swine tissue.

Conclusions

We demonstrate the capabilities of the proposed compact SS-Raman spectroscopy system by obtaining Raman spectra of chemical and biological materials, utilizing straightforward signal processing. We anticipate that the SS-Raman spectroscopy will be utilized in various fields, including biomedical and chemical applications.

1.

Introduction

Raman spectroscopy is a powerful analytical tool, providing molecular fingerprints of samples by measuring the changes in frequency and the intensity of inelastically scattered light.1,2 The frequency shifts resulting from energy changes are indicative of molecular bonding and vibrational modes. Inelastically scattered light is categorized into two types: (1) anti-Stokes scattering with energy higher than the incident light and (2) Stokes scattering with lower energy.3 Because Stokes scattering is generally much stronger than anti-Stokes scattering, most Raman spectroscopy focuses on Stokes scattering, even though anti-Stokes scattering is almost free from background interference.

Raman spectroscopy has proven its efficacy in various fields, including pharmaceutical4,5 and materials science.6 Notably, due to its capability for noninvasive, in situ chemical analysis, Raman spectroscopy is widely used in biological and biomedical applications.7 It analyzes substances such as acetaminophen,8 glucose,9 phenylalanine,10 and hydroxyapatite11 and has been used for clinical diagnostic applications in diabetes,12,13 coronary artery disease,14 and cancer.15

Dispersive Raman spectroscopy is a commonly used method for measuring Raman spectra. It utilizes a fixed-wavelength light source and a spectrometer. Dispersive Raman spectroscopy enables the simultaneous acquisition of the entire Raman spectrum using a line charge coupled device (CCD) or rectangular CCD. Because a small amount of Raman photons is dispersed over a large spectral region, Raman spectra can be obtained by detecting Raman photons with a high throughput spectrometer. However, high throughput Raman spectroscopy systems tend to be bulky due to the larger optical components and detector.16 The substantial size of the spectroscopy system poses a challenge in developing portable Raman spectroscopy. Commercial portable Raman spectroscopy systems, such as the mini-spectrometer (C14214MA, Hamamatsu, Japan) and handheld Raman spectrometer (BRAVO, Bruker, Billerica, Massachusetts, United States) have been developed; however, they are more geared to chemical analysis and not sensitive enough for many biomedical applications. Many studies have concentrated on strategies such as increasing the excitation power, using longer acquisition times, and employing cooled detectors to enhance Raman spectra.17 Nevertheless, these approaches may result in sample damage and incur high costs. Furthermore, in a recent study, the successful implementation of compact Raman spectroscopy with a picosecond laser, time-correlated photon counting detection, and a single-mode fiber significantly decreased the weight in comparison to conventional Raman spectroscopy employing a CCD-based detector.18

The wavelength-tunable swept-source laser has emerged as a critical component in biomedical imaging. Swept-source optical coherence tomography (SS-OCT) has been applied in various biomedical applications, such as intravascular imaging.19,20 Additionally, two-photon microscopy has also been demonstrated using a swept-source laser.21 In this study, we developed a Raman spectroscopy system utilizing a swept-source laser, departing from the conventional fixed-wavelength light source. Swept-source Raman (SS-Raman) spectroscopy was initially proposed in a patent (US9662047B2, portable Raman diagnostic system) as a way to implement a compact Raman system. This patent also describes the integration of OCT and Raman using a single swept-source. Recently, Atabaki et al.22 demonstrated SS-Raman spectroscopy by combining a MEMS-tunable vertical cavity surface emitting laser with a bandpass filter having a 0.5 nm bandwidth. However, further improvements, such as increasing the optical power and sample mapping capabilities, are still required to overcome limitations and enable wider applications, especially in biomedicine.

In this study, we developed a compact SS-Raman spectroscopy system utilizing a swept-source laser, a bandpass filter, and a highly sensitive silicon photoreceiver. The integration of these components enables the implementation of a cost-effective and compact Raman spectroscopy system. Moreover, we introduced a signal processing technique to enhance the Raman spectra by eliminating artifacts and minimizing signal-broadening effects, which can be achieved without incurring high costs. To validate the SS-Raman spectroscopy, we obtained Raman spectra from diverse chemical samples, including phenylalanine, hydroxyapatite, glucose, and acetaminophen, and compared them with spectra from conventional dispersive Raman spectroscopy. Additionally, we acquired Raman spectra from swine tissue cross-sections to generate biological tissue mapping plots, demonstrating the applicability of the SS-Raman spectroscopy in biological samples.

2.

Methods and Materials

2.1.

Principle of SS-Raman Spectroscopy

The recently proposed SS-Raman uses a swept-source laser, a narrow-bandwidth bandpass filter, and a highly sensitive point photoreceiver for acquiring Raman spectra.22 These components contribute to the development of compact and cost-effective Raman spectroscopy systems. The narrow linewidth of the swept-source laser sequentially tunes the wavelength across the light source’s wavelength range. SS-Raman spectroscopy takes advantage of the independence of Raman shift from the excitation wavelength,23 resulting in a consistent nonresonant Raman spectral shape obtained during the wavelength sweep. Because the wavelength of the Raman signal changes during the sweep, the use of a narrow-bandwidth filter and a highly sensitive photoreceiver enables the acquisition of the desired Raman spectra, which are determined by the difference between the center wavelength of the bandpass filter and the excitation wavelength. The Raman shift in SS-Raman spectroscopy is expressed as follows:

Eq. (1)

Raman shift  (cm1)=1λext1λbp,
where λext is the excitation wavelength, and λbp is the center wavelength of the bandpass filter. Compact SS-Raman spectroscopy operates on time-sequential detection to obtain Raman spectra. While it cannot obtain the entire spectrum simultaneously as dispersive Raman spectroscopy can, the high sensitivity of the point detector effectively mitigates this limitation.

2.2.

SS-Raman Spectroscopy System and Components

Figure 1 illustrates the compact SS-Raman spectroscopy system developed in this study. The schematic provides an overview of the entire SS-Raman spectroscopy system, which comprises two main parts: one for sample excitation and another for collecting the scattered Raman signals from the samples. The system utilizes a swept-source laser (XPERAY, Nanobase, Seoul, Republic of Korea) that sweeps the wavelength range from 822 to 842 nm with an instantaneous linewidth of 10 pm. The output laser power, sweeping wavelength interval, and sweeping speed are controlled by the laser software. For Raman spectra acquisition, we used an excitation power of 100 mW and a wavelength sweep interval of 0.05 nm with each interval having an exposure time of 0.1 s, resulting in a total acquisition time of 40 s for 400 wavelength steps. The laser output, delivered through a multimode fiber, is collimated by a lens (AC254-100-B, Thorlabs, Newton, New Jersey, United States), filtered through an 850 nm short-pass filter (FESH0850, Thorlabs, United States) to eliminate background noise and scattered light from the fiber and then focused onto the sample using a focusing lens (AC254-150-B, Thorlabs, United States). A silver mirror directs the excitation beam to the sample stage. The Raman scattering signals from the sample were collected by a collection lens (AC254-050-B, Thorlabs, United States) with a numerical aperture (NA) of 0.25. An 850 nm long-pass filter (FELH0850, Thorlabs, United States) is utilized to obtain Raman signals without Rayleigh scattering. Additionally, a 912 nm bandpass filter with a bandwidth of 1.5 nm (912-1.5 OD4 Ultra Narrow Bandpass filter, Alluxa, Santa Rosa, California, United States), located after the long-pass filter, facilitates the acquisition of Raman spectra during a single sweep of the light source. According to Eq. (1), the range of the acquired Raman spectra spans from 900 to 1200  cm1. A converging lens (AC254-050-B, Thorlabs, United States) focuses the Raman signal onto the active area of the ultra-low noise and high-sensitive silicon detector (FWPR-20-Si, FEMTO, Berlin, Germany). Data are acquired via a data acquisition (DAQ) board and a PC, and the Raman spectra are monitored in real-time through a LabVIEW program.

Fig. 1

Schematic diagram of the SS-Raman spectroscopy system. A wavelength-swept laser, ranging from 822 to 842 nm, coupled with a multimode optical fiber, was utilized. The sample was illuminated with converging lenses and a mirror. Raman signals were collected through a collection lens and detected by a photoreceiver. The Raman spectra were acquired during a single sweep using a narrow-bandwidth bandpass filter. BPF, bandpass filter; CL, collection lens.

JBO_29_S2_S22703_f001.png

3.

Results

3.1.

Hardware Limitation and Signal Processing

The developed SS-Raman spectroscopy system faced practical hardware limitations that adversely affected the quality of the Raman spectra. Notably, the use of a bandpass filter with a relatively broad bandwidth in our system led to compromised spectral resolution and broadened Raman peaks. Additionally, the unstable output power of the swept-source laser introduced ripple noises into the Raman signal, and the low optical density of the Rayleigh rejection filters contributed to a significant background signal. Although the approach of using high-performance light sources and ultra-narrow bandpass filters could mitigate these hardware limitations, it comes with a substantial cost. In response, our study proposed a straightforward signal processing method aimed at enhancing the Raman signal without incurring significant expenses.

Figure 2 shows the signal processing process to improve the Raman spectra. Figure 2(a) shows the raw Raman signal of phenylalanine powder measured by SS-Raman spectroscopy. A sawtooth-like ripple noise is introduced due to the instability of the light source. Gaussian filtering was applied, to eliminate this high-frequency noise, including the ripple noise while preserving the Raman signal, as shown in Fig. 2(b). However, both the Gaussian filtering and the broad bandwidth of the bandpass filter contribute to the broadening of the Raman peaks. The full width at half maximum (FWHM) of the 1004  cm1 Raman peak of phenylalanine was 20  cm1. The deconvolution method can be used to improve the spectral resolution of the instrumentally broadened Raman spectra caused by the slit in dispersive Raman spectroscopy.24 Figure 2(c) shows the enhanced spectral resolution after deconvolution, considering both the bandpass filter transmission and the Gaussian filter, reducing the FWHM of the 1004  cm1 Raman peak of phenylalanine to 18.3  cm1—an improvement of the spectral resolution by 8.5%. Finally, Fig. 2(d) shows the Raman spectra after implementing polynomial background removal, which eliminates the background noise arising from the low optical density of the filters, using a second-order polynomial for this study.

Fig. 2

Signal processing in SS-Raman spectroscopy. (a) Raw Raman spectrum of phenylalanine powder shows degradation due to the unstable output of the light source and the use of filters with a large bandwidth. (b) Application of Gaussian filtering effectively eliminates high-frequency noise. (c) Deconvolution enhances the spectral resolution. (d) Polynomial background removal mitigates the background signal arising from the low optical density of the filter.

JBO_29_S2_S22703_f002.png

3.2.

Raman Spectra of Chemical Materials

Raman spectroscopy measurements were conducted on chemical powder samples, including phenylalanine, hydroxyapatite, glucose, and acetaminophen, using a 100 mW power at the sample plane with postprocessing to enhance the spectral resolution. Figure 3 shows the Raman spectra of these chemical samples within the range of 900 to 1200  cm1. Figures 3(a)3(d) show the Raman spectra of the chemical samples obtained through the SS-Raman spectroscopy with notable Raman bands marked by triangle symbols. Figure 3(a) shows the Raman spectra of phenylalanine, with the Raman bands at 1030 and 1004  cm1, assigned to the CH in-plane bending mode of the substituted benzene and the symmetric ring breathing mode, respectively.25,26 Figure 3(b) shows the Raman spectra of hydroxyapatite with Raman bands at 1045 and 960  cm1, corresponding to the PO asymmetric valence and PO symmetrical valence mode, respectively.27 Figure 3(c) shows the Raman spectra of glucose with Raman bands at 1125, 1060, and 910  cm1, associated with the COH bonds’ bending mode, CO stretching mode, and CH vibration, respectively.2830 Figure 3(d) shows the Raman spectra of acetaminophen, featuring Raman bands at 1168 and 1104  cm1, linked to the aromatic β CH bending. The Raman bands at 1016 and 968  cm1 correspond to the ρ CH3 bending and aromatic γ CH bending, respectively.31 Figures 3(e)3(h) show the Raman spectra of the same samples using dispersive Raman spectroscopy. We calculated the correlation coefficients between the Raman spectra obtained through the SS-Raman spectroscopy system and dispersive Raman spectroscopy for comparison. The correlation coefficients for phenylalanine, hydroxyapatite, glucose, and acetaminophen were 0.88, 0.84, 0.87, and 0.73, respectively, indicating a strong positive correlation.

Fig. 3

Raman spectra (900 to 1200  cm1) of the chemical samples. (a)–(d) Raman spectra of phenylalanine, hydroxyapatite, glucose, and acetaminophen, respectively, obtained through the SS-Raman spectroscopy system, utilizing a wavelength sweep range from 822 to 842 nm, a power of 100 mW, and a wavelength sweep interval of 0.05 nm. (e)–(h) Corresponding Raman spectra obtained through dispersive Raman spectroscopy.

JBO_29_S2_S22703_f003.png

3.3.

Raman Spectra of Animal Tissue

The SS-Raman spectroscopy system was used to measure the Raman spectra of biological samples. The biological samples utilized in this study was purchased from a grocery store (Shaw’s Supermarket, Lynn, Massachusetts, United States). The samples were stored in a refrigerator at 18°C to prevent deformation. Because the researcher had no control over the premortem procedures or the process of euthanasia, occupational health risks were minimal; thus, the Institutional Animal Care and Use Committee at the Massachusetts Institute of Technology did not require a protocol.

In this study, pork belly slices with a thickness of 10 mm and a length of 25 mm were used as samples without any additional treatment. Raman spectra of the biological sample were acquired within the range of 900 to 1200  cm1 using a 100 mW power at the sample plane. Each Raman spectrum was acquired with a 40 s acquisition time. Figure 4 presents the results of the biological tissue mapping plot using SS-Raman spectroscopy. The mapping was conducted on the cross-sections of the pork belly using a motorized stage. Figure 4(a) illustrates the scanning direction of the pork belly tissue, as indicated by a white arrow. The biological sample was scanned at 1 mm intervals over a length of 30 mm, resulting in the collection of 30 Raman spectra. Figure 4(b) shows the biological tissue mapping plot, providing information about the composition of the biological sample at each scanning location. Figure 4(c) shows the Raman spectra of the fat layer of the pork belly measured by SS-Raman and dispersive Raman spectroscopy, whereas Fig. 4(d) shows the Raman spectra of the muscle layer of the pork belly. The Raman spectra obtained through the SS-Raman spectroscopy are represented in red, and the Raman spectra obtained through dispersive Raman spectroscopy are depicted in blue. In general, fat has a larger Raman cross-section than that of protein. In the fat layer, a relatively intense Raman signal was observed at 1080  cm1, assigned to the CC stretch mode of fat. Conversely, no clear Raman peak was observed in the muscle layer due to the low Raman cross-section and broadening of the sharp phenylalanine peak at 1004  cm1. We calculated the correlation coefficient between the Raman spectra of the fat layer obtained through the SS-Raman and dispersive Raman spectroscopy for comparison. The correlation coefficient for the fat layer was 0.91, indicating a strong positive correlation.

Fig. 4

Raman spectra of biological samples obtained through SS-Raman spectroscopy. (a) Pork belly slice with the scanning direction indicated by the white arrow. The scale bar represents 5 mm. (b) The biological tissue mapping plot of sample composition. Raman spectra of (c) fat and (d) muscle layer, obtained through the SS-Raman spectroscopy (red) and dispersive Raman spectroscopy (blue).

JBO_29_S2_S22703_f004.png

4.

Discussion and Conclusion

In this study, we demonstrated SS-Raman spectroscopy using a narrow bandwidth bandpass filter and a swept-source laser, creating a compact system without the need for a bulky spectrometer. Additionally, the use of a highly sensitive point detector enhances the sensitivity, representing an advantage of the proposed SS-Raman spectroscopy method. We validated the efficacy of our SS-Raman spectroscopy by acquiring Raman spectra from chemical specimens and providing Raman mapping of biological tissue.

However, despite the advancements in SS-Raman spectroscopy, currently, there are practical hardware limitations, particularly associated with the large bandwidth of the bandpass filter and the noise of the swept-source laser that compromise the spectral resolution. The large bandwidth filter in the SS-Raman spectroscopy enhances the throughput, which is especially advantageous for acquiring Raman signals from biological samples. However, there is a trade-off between throughput and spectral resolution. To address these limitations, we have applied Gaussian filtering to reduce high-frequency noise and performed deconvolution to enhance the spectral resolution. Moreover, we have incorporated polynomial background removal to remove background signals due to the low optical density of the filters, improving the spectral resolution by 8.5%, as measured by the FWHM calculations. The Raman spectra of the chemical samples shown in Fig. 3, including phenylalanine, hydroxyapatite, glucose, and acetaminophen, further corroborate our findings. Additionally, we validated our method with correlation coefficients of 0.88, 0.84, 0.87, and 0.73, respectively, when comparing the SS-Raman to dispersive Raman spectroscopy for these samples. We expect that the spectral resolution will be significantly improved by eliminating the broadening effects from other instruments.

Furthermore, we successfully conducted biological tissue mapping of pork belly slices using the SS-Raman spectroscopy with a motorized stage for point scanning across the tissues’ cross-sections. Through these Raman spectra, this technique enabled the clear differentiation between fat and muscle layers within the sample. Notably, the Raman spectra obtained using SS-Raman spectroscopy exhibited a high correlation with those obtained via dispersive Raman spectroscopy, achieving a correlation coefficient of 0.91, demonstrating the reliability and precision of the SS-Raman method in biological tissue characterization.

The SS-Raman spectroscopy developed in this feasibility study can be improved in spectral resolution, acquisition time, and collection efficiency. The large bandwidth of the bandpass filter (1.5 nm) currently limits the spectral resolution of the SS-Raman spectroscopy. However, recently released ultra-narrow bandpass filters with a bandwidth from 0.1 to 0.5 nm can enhance the spectral resolution of SS-Raman spectroscopy, achieving results comparable or even superior to the spectral resolution of the most dispersive Raman spectroscopy systems. The current acquisition time of 40 s for SS-Raman spectroscopy poses limitations in non-steady-state studies. To expedite the acquisition of high-quality Raman spectra, enhancing the measurement geometry of SS-Raman spectroscopy is imperative. Incorporating the confocal geometry with a high NA objective lens (0.95), as employed in the compared dispersive Raman spectroscopy, into the SS-Raman spectroscopy system can increase the collection efficiency by 14.5 times. Additionally, by mitigating background noise through the use of a high optical density (OD) optical filter and optimizing light source parameters such as the wavelength step interval and sweeping speed, the performance of SS-Raman spectroscopy can be significantly improved. Overall, we anticipate that the enhanced SS-Raman spectroscopy will enable the measurement of biological sample in <1  s. Moreover, the presented SS-Raman spectroscopy has a spectral range constrained by the laser’s sweeping capabilities. To overcome this, we are developing a multichannel SS-Raman system with multiple detectors and bandpass filters. Multichannel SS-Raman spectroscopy with four bandpass filters and detectors is expected to cover the entire fingerprint region within the same acquisition time. This enhancement will broaden the technique’s applicability across diverse application fields.

Recent trends in the field of Raman spectroscopy include the miniaturization of systems.3234 We anticipate that SS-Raman spectroscopy could offer substantial advantages in miniaturization compared with dispersive methods. Furthermore, we are developing an OCT-Raman imaging system that simultaneously obtains morphological and chemical information. These efforts are consistent with advances in multimodal Raman spectroscopy, which integrates various imaging modalities. One example is a system that combines quantitative phase imaging and Raman functionality using a swept-source laser.35 We expect that the OCT-Raman system will be a pivotal tool in various biomedical applications, including cancer imaging and intravascular imaging.

Disclosures

The authors declare that they have no competing interests.

Code and Data Availability

The datasets generated and/or analyzed during the current study are available from the corresponding author on reasonable request.

Acknowledgments

This research was supported by the National Institutes of Health (NIH; Grant Nos, P41EB015871 and UG3CA275687), the National Research Foundation of Korea (Grant No. RS202300208888), and the MISTI MIT-Korea KAIST Seed Fund program.

References

1. 

S. P. Mulvaney and C. D. Keating, “Raman spectroscopy,” Anal. Chem., 72 (12), 145 –158 https://doi.org/10.1021/a10000155 ANCHAM 0003-2700 (2000). Google Scholar

2. 

P. Rostron, S. Gaber and D. Gaber, “Raman spectroscopy, review,” Laser, 21 50 –64 (2016). Google Scholar

3. 

S. A. Asher, “UV resonance Raman spectroscopy for analytical, physical, and biophysical chemistry. Part 2,” Anal. Chem., 65 (4), 201A –210A https://doi.org/10.1021/ac00052a001 ANCHAM 0003-2700 (1993). Google Scholar

4. 

T. Vankeirsbilck et al., “Applications of Raman spectroscopy in pharmaceutical analysis,” Trends Anal. Chem., 21 (12), 869 –877 https://doi.org/10.1016/S0165-9936(02)01208-6 (2002). Google Scholar

5. 

A. Paudel, D. Raijada and J. Rantanen, “Raman spectroscopy in pharmaceutical product design,” Adv. Drug Deliv. Rev., 89 3 –20 https://doi.org/10.1016/j.addr.2015.04.003 ADDREP 0169-409X (2015). Google Scholar

6. 

M. S. Dresselhaus et al., “Raman spectroscopy of carbon nanotubes,” Phys. Rep., 409 (2), 47 –99 https://doi.org/10.1016/j.physrep.2004.10.006 PRPLCM 0370-1573 (2005). Google Scholar

7. 

Y. Lu, L. Lin and J. Ye, “Human metabolite detection by surface-enhanced Raman spectroscopy,” Mater. Today Biol., 13 100205 https://doi.org/10.1016/j.mtbio.2022.100205 (2022). Google Scholar

8. 

C. Shende et al., “Drug stability analysis by Raman spectroscopy,” Pharmaceutics, 6 (4), 651 –662 https://doi.org/10.3390/pharmaceutics6040651 (2014). Google Scholar

9. 

A. M. Enejder et al., “Raman spectroscopy for noninvasive glucose measurements,” J. Biomed. Opt., 10 (3), 031114 https://doi.org/10.1117/1.1920212 JBOPFO 1083-3668 (2005). Google Scholar

10. 

A. Kuzuhara, “Raman spectroscopic analysis of L‐phenylalanine and hydrolyzed eggwhite protein penetration into keratin fibers,” J. Appl. Polym. Sci., 122 (4), 2680 –2689 https://doi.org/10.1002/app.34311 JAPNAB 1097-4628 (2011). Google Scholar

11. 

I. Rehman et al., “Structural evaluation of human and sheep bone and comparison with synthetic hydroxyapatite by FT‐Raman spectroscopy,” J. Biomed. Mater. Res., 29 (10), 1287 –1294 https://doi.org/10.1002/jbm.820291016 JBMRBG 0021-9304 (1995). Google Scholar

12. 

J. W. Kang et al., “Direct observation of glucose fingerprint using in vivo Raman spectroscopy,” Sci. Adv., 6 (4), eaay5206 https://doi.org/10.1126/sciadv.aay5206 STAMCV 1468-6996 (2020). Google Scholar

13. 

N. C. Dingari et al., “Investigation of the specificity of Raman spectroscopy in non-invasive blood glucose measurements,” Anal. Bioanal. Chem., 400 2871 –2880 https://doi.org/10.1007/s00216-011-5004-5 ABCNBP 1618-2642 (2011). Google Scholar

14. 

III J. F. Brennan et al., “Determination of human coronary artery composition by Raman spectroscopy,” Circulation, 96 (1), 99 –105 https://doi.org/10.1161/01.CIR.96.1.99 CIRCAZ 0009-7322 (1997). Google Scholar

15. 

G. W. Auner et al., “Applications of Raman spectroscopy in cancer diagnosis,” Cancer Metastasis Rev., 37 691 –717 https://doi.org/10.1007/s10555-018-9770-9 (2018). Google Scholar

16. 

J. Lerner and A. Thevenon, “The optics of spectroscopy,” Jobin-Yvon Optical Systems/Instrumentss SA, (1988). Google Scholar

17. 

L. Barron et al., “Backscattered Raman optical activity with a CCD detector,” J. Am. Chem. Soc., 111 (23), 8731 –8732 https://doi.org/10.1021/ja00205a031 JACSAT 0002-7863 (1989). Google Scholar

18. 

Z. Meng et al., “Lightweight Raman spectroscope using time-correlated photon-counting detection,” Proc. Natl. Acad. Sci., 112 (40), 12315 –12320 https://doi.org/10.1073/pnas.1516249112 (2015). Google Scholar

19. 

S. H. Yun et al., “Comprehensive volumetric optical microscopy in vivo,” Nat. Med., 12 (12), 1429 –1433 https://doi.org/10.1038/nm1450 1078-8956 (2006). Google Scholar

20. 

M. W. Lee et al., “Comprehensive intravascular imaging of atherosclerotic plaque in vivo using optical coherence tomography and fluorescence lifetime imaging,” Sci. Rep., 8 (1), 14561 https://doi.org/10.1038/s41598-018-32951-9 SRCEC3 2045-2322 (2018). Google Scholar

21. 

J. W. Kang et al., “Two-photon microscopy by wavelength-swept pulses delivered through single-mode fiber,” Opt. Lett., 35 (2), 181 –183 https://doi.org/10.1364/OL.35.000181 OPLEDP 0146-9592 (2010). Google Scholar

22. 

A. H. Atabaki et al., “Low-power swept-source Raman spectroscopy,” Opt. Express, 29 (16), 24723 –24734 https://doi.org/10.1364/OE.427105 OPEXFF 1094-4087 (2021). Google Scholar

23. 

J. Moon et al., “Raman spectroscopy,” Springer Handbook of Advanced Catalyst Characterization, 75 –110 Springer( (2023). Google Scholar

24. 

K. Michaelian and W. Friesen, “Deconvolution of instrumental broadening in dispersive Raman spectroscopy,” Appl. Spectrosc., 42 (8), 1538 –1543 https://doi.org/10.1366/0003702884429571 APSPA4 0003-7028 (1988). Google Scholar

25. 

B. Hernández et al., “Characteristic Raman lines of phenylalanine analyzed by a multiconformational approach,” J. Raman Spectrosc., 44 (6), 827 –833 https://doi.org/10.1002/jrs.4290 JRSPAF 0377-0486 (2013). Google Scholar

26. 

J. Zhu et al., “Surface-enhanced Raman spectroscopy investigation on human breast cancer cells,” Chem. Central J., 7 37 https://doi.org/10.1186/1752-153X-7-37 (2013). Google Scholar

27. 

P. Timchenko et al., “Experimental studies of hydroxyapatite by Raman spectroscopy,” J. Opt. Technol., 85 (3), 130 –135 https://doi.org/10.1364/JOT.85.000130 JOTEE4 1070-9762 (2018). Google Scholar

28. 

M. Mathlouthi and D. V. Luu, “Laser-Raman spectra of d-glucose and sucrose in aqueous solution,” Carbohydrate Res., 81 (2), 203 –212 https://doi.org/10.1016/S0008-6215(00)85652-9 CRBRAT 0008-6215 (1980). Google Scholar

29. 

A. Golparvar et al., “Very selective detection of low physiopathological glucose levels by spontaneous Raman spectroscopy with univariate data analysis,” BioNanoScience, 11 (3), 871 –877 https://doi.org/10.1007/s12668-021-00867-w (2021). Google Scholar

30. 

Y. Uwadaira et al., “Identification of informative bands in the short-wavelength NIR region for non-invasive blood glucose measurement,” Biomed. Opt. Express, 7 (7), 2729 –2737 https://doi.org/10.1364/BOE.7.002729 BOEICL 2156-7085 (2016). Google Scholar

31. 

F. Zapata et al., “A practical beginner’s guide to Raman microscopy,” Appl. Spectrosc. Rev., 56 (6), 439 –462 https://doi.org/10.1080/05704928.2020.1797761 APSRBB 0570-4928 (2021). Google Scholar

32. 

T. A. Anderson et al., “Raman spectroscopy differentiates each tissue from the skin to the spinal cord: a novel method for epidural needle placement?,” Anesthesiology, 125 (4), 793 –804 https://doi.org/10.1097/ALN.0000000000001249 ANESAV 0003-3022 (2016). Google Scholar

33. 

J. W. Kang et al., “Analysis of subcutaneous swine fat via deep Raman spectroscopy using a fiber-optic probe,” Analyst, 145 (13), 4421 –4426 https://doi.org/10.1039/d0an00707b ANLYAG 0365-4885 (2020). Google Scholar

34. 

Y. Zheng et al., “Noninvasive blood glucose detection using a miniature wearable Raman spectroscopy system,” Chin. Opt. Lett., 15 (8), 083001 CJOEE3 1671-7694 (2017). Google Scholar

35. 

J. W. Kang et al., “Combined confocal Raman and quantitative phase microscopy system for biomedical diagnosis,” Biomed. Opt. Express, 2 (9), 2484 –2492 https://doi.org/10.1364/BOE.2.002484 BOEICL 2156-7085 (2011). Google Scholar

Biography

Jeonggeun Song is a PhD candidate student at Korea Advanced Institute of Science and Technology (KAIST), Republic of Korea. He received his BS and MS degrees in mechanical engineering from KAIST in 2019 and 2021, respectively. His current research interests include multiphoton biological imaging systems and Raman spectroscopy. He is a member of SPIE.

Biographies of the other authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Jeonggeun Song, Peter T. C. So, Hongki Yoo, and Jeon Woong Kang "Swept-source Raman spectroscopy of chemical and biological materials," Journal of Biomedical Optics 29(S2), S22703 (3 April 2024). https://doi.org/10.1117/1.JBO.29.S2.S22703
Received: 28 November 2023; Accepted: 23 February 2024; Published: 3 April 2024