13 January 2024 Analysis of polarization features of human breast cancer tissue by Mueller matrix visualization
Thi-Thu-Hien Pham, Thao-Ngan Ngoc Quach, Quoc-Hoang-Quyen Vo
Author Affiliations +
Abstract

Significance

Breast cancer ranks second in the world in terms of the number of women diagnosed. Effective methods for its early-stage detection are critical for facilitating timely intervention and lowering the mortality rate.

Aim

Polarimetry provides much useful information on the structural properties of breast cancer tissue samples and is a valuable diagnostic tool. The present study classifies human breast tissue samples as healthy or cancerous utilizing a surface-illuminated backscatter polarization imaging technique.

Approach

The viability of the proposed approach is demonstrated using 95 breast tissue samples, including 35 healthy samples, 20 benign cancer samples, 20 grade-2 malignant samples, and 20 grade-3 malignant samples.

Results

The observation results reveal that element m23 in the Mueller matrix of the healthy samples has a deeper color and greater intensity than that in the breast cancer samples. Conversely, element m32 shows a lighter color and reduced intensity. Finally, element m44 has a darker color in the healthy samples than in the cancer samples. The analysis of variance test results and frequency distribution histograms confirm that elements m23, m32, and m44 provide an effective means of detecting and classifying human breast tissue samples.

Conclusions

Overall, the results indicate that surface-illuminated backscatter polarization imaging has significant potential as an assistive tool for breast cancer diagnosis and classification.

1.

Introduction

With an estimated 2.26 million new cases of breast cancer globally and around 685,000 breast cancer-related deaths, breast cancer accounted for around 24.5% of all cancer cases around the world in 2020 and 15.5% of cancer deaths among women.1,2 Breast cancer was the most common cause of cancer mortality in women in 2020 and the fifth most common cause of cancer death.2 Furthermore, the number of breast cancer cases is expected to exceed 4.4 million by 2070.1

The human breast is made up primarily of adipose (fat) and milk-producing glandular lobules.2 The lobules connect to create lobes, with each breast having around 15 to 20 lobes in total. Within the lobules, milk glands are connected to small ducts, which join together to form larger milk ducts that emerge through the nipple. The breast is linked to a network of blood and lymph veins, which serve to regulate the local fluid balance along with filter out harmful substances. Breast cancer metastases are formed mainly in the lymph nodes near the breast. However, the lymph veins carry lymph fluid to lymph nodes throughout the body, and thus breast cancer frequently spreads to other nearby lymph nodes, such as under the arm, above the collarbone, or in the chest. The breast is physically supported by a network of connective tissue that contains collagen. The breast density in mammograms is related to the relative amount of connective tissue,3 and a higher density is generally associated with a higher cancer risk.4

Breast cancer is defined as the uncontrolled proliferation of epithelial cells in the breast and is classified according to the cellular origin and tumor stage.4 Breast cancer can develop in the ducts or the lobules. The most prevalent type of breast cancer is ductal carcinoma, which develops in the cells that coat the duct walls. Breast cancer in the early stage (stage 0) is noninvasive and is referred to as being in situ (e.g., ductal carcinoma in situ). The more advanced cancer stages (I to IV) are distinguished by invasive tumors that have spread to other breast tissue, lymph nodes, and/or organs. Stages I to III are categorized based on the tumor size and presence (or otherwise) of cancer cells in the lymph nodes. In stage IV, the cancer spreads to other organs, most commonly the bones, lungs, liver, and brain, and palliative care or end-of-life care is generally required.

Although breast cancer patients have numerous treatment options available to them, particularly in the early stage, the majority of late-stage patients must undergo tumor removal surgery.4 Depending on the tumor size, location, disease stage, menopausal state, hormone receptor status, and cancer recurrence, surgical intervention may involve breast-conserving surgery (lumpectomy), mastectomy (complete breast removal), or lymph node removal. The surgical procedure is commonly followed by radiation therapy, hormone therapy, or chemotherapy. If the breast cancer has progressed to the bones, lungs, or brain, radiation therapy is usually delivered after breast-conserving surgery or, in rare cases, after mastectomy. Hormonal treatment, by contrast, is used to treat hormone receptor-positive breast cancer, while chemotherapy is routinely used after surgery for early stage breast cancer.

Many diagnostic imaging technologies, including mammography, magnetic resonance imaging (MRI), positron emission tomography (PET), computed tomography, and ultrasound, are available for detecting breast cancer in the early stage. Based on the imaging results, histopathology may then be performed to determine the disease.5 Mammography, which uses X-ray images to identify breast cancer tissue, is the most sensitive method for detecting breast cancer but has a high incidence rate of false-positive and false-negative readings, which can lead to therapeutic complications, overtreatment, and mental distress.6 MRI can detect minute abnormalities that may be missed by mammography; however, it is expensive and has low specificity, resulting in possible overdiagnosis.7 PET is the most accurate method for visualizing the progression of cancer and its treatment response. However, it requires the injection of radioactive tracer, which can cause numerous side effects and radiation exposure.8

On some occasions, imaging techniques may yield false-positive readings, which prompt the physician to conduct a biopsy, only to obtain a benign pathologic result. This not only incurs unnecessary medical expense but also causes the patient needless pain and mental suffering. Thus, high sensitivity and accurate diagnostic imaging techniques are required to support the physician in deciding whether a biopsy procedure is required.9

Polarimetry is a well-established technique for studying matter properties using polarized light.10,11 Various levels of information can be obtained, depending on the specific configuration of samples and number of polarization measurements taken.12 Images with concentrated contrast in superficial skin layers, for example, can be obtained by taking just two measurements with a polarizer before and after the sample (one with the polarization axes aligned and one with them crossed).13 A thorough description of a sample’s polarization properties, on the other hand, necessitates numerous (at least 16) polarization-sensitive measurements. The Mueller matrix constructed in polarimetry techniques relates the input and output polarization states and contains 16 elements, which between them completely describe how a sample interacts with all types of polarized light.14,15 The polarization state of a polarized light beam changes when it interacts with the sample, and thus analyzing the state of polarization of the emergent light yields valuable information regarding the optical and structural properties of the sample.16 Mueller matrix polarimetry (MMP) technique has numerous advantages for biomedical diagnostics.15,17 For example, the Mueller matrix provides the picture contrast of the surface tissue layers by up to 85% and thus the information for early detection of precancerous alterations are figured.18,19 Furthermore, the Mueller matrix provides critical information on the architecture of abnormal tissue areas1921 and can be used in conjunction with a wide variety of optical devices, such as microscopes and endoscopes, by adding the appropriate polarization components.22,23 For instance, numerous studies have demonstrated the feasibility of detecting early stage cancer by combining MMP with endoscopy.23,24

The present study analyzes the polarization features of human breast cancer tissue utilizing a polarimetric imaging system based on a charge-coupled device (CCD) camera.2530 In the proposed approach, the backscattered light from the breast tissue is captured by the camera and processed by the Stokes–Mueller method to determine the Mueller matrix image and corresponding optical properties of the sample. The sample is then classified into one of four different classes, namely normal, cancer benign, or cancer malignant (grade 2 or 3) based on an analysis of the average intensity of the Mueller matrix elements, the analysis of variance (ANOVA) test results, and the frequency distribution histograms (FDHs) of the pixel intensity.

2.

Methodology

2.1.

Sample Preparation

As shown in Fig. 1, 95 breast tissue samples were obtained, including 35 healthy samples and 60 cancer samples. The cancer samples comprised 20 benign samples, 20 grade-2 malignant samples, and 20 grade-3 malignant samples. Each sample consisted of two to three slices of tissue attached to a microscope slide. The samples were acquired from a specialist clinic at the Ho Chi Minh City Oncology Hospital in Vietnam with the commitment that the samples would be used only for scientific research. All of the preparation-related excision, tissue processing, attachment, staining, registration, and histopathological analysis procedures were performed by qualified pathologists before transferring the samples for experiment. The tissues were sliced with a thickness of 5  μm and were stained with hematoxylin and eosin. Each slice was measured three to five times using the polarization imaging system, and the average intensity values were then calculated.

Fig. 1

Sample preparation processing.

JBO_29_5_052917_f001.png

2.2.

Experimental Setup

Figure 2 illustrates the backscattering Mueller matrix polarization imaging system constructed in the present study. As shown, the system comprised two major components: a polarization state generator (PSG) and a polarization state analyzer (PSA). The PSG block produced the desired polarization state of the incident beam, whereas the PSA block examined the polarization state of the sample-affected beam. The generator consisted of a stabilized red He-Ne laser light source (LS, HNLS008R, Thorlabs, Inc.), a quarter-wave plate polarizer (R1, QWP0-63304-4-R10, CVI Co.) controlled by a motor-driven stage (SGSP-60YAW-0B, Sigma Koki Co.), a linear polarizer (P1, GTH5M, Thorlabs Co.) controlled by a second motor-driven stage, and a beam expander (BE). The BE increased the diameter of the laser beam from 0.5 to 1.5 cm, thereby enabling more coverage the sample to improve the accuracy and quality of the results. The analyzer consisted of a linear polarizer (P2, GTH5M, Thorlabs, Inc.), a quarter-wave plate (R2, QWP0-63304-4-R10, CVI Co.), and a CCD camera (CCD, DCU224C, Thorlabs, Inc.) fitted with a 6.5× zoom lens and connected to a computer for image acquisition and analysis purposes.

Fig. 2

(a) Schematic illustration of experimental setup, and (b) photograph of experimental arrangement (LS, laser source; P1–P2, linear polarizers (operated by stepping motors: 0 deg, 45 deg, 90 deg, and 135 deg); R1–R2, quarter-wave plates (RH 45 deg, LH – 45 deg); CCD, CCD camera; BE, beam expander).

JBO_29_5_052917_f002.png

In the measurement process, the light from the laser source was passed through the BE, quarter-wave plate (R1), and linear polarizer (P1) of the generator block to produce six different polarization states: horizontal linear (H), vertical linear (V), 45-deg linear (P), 135-deg linear (M), right-hand circular (R), and left-hand circular (L). The linear states were produced by rotating the polarizer (P1) and sliding the quarter-wave plate (R1) out of the light path, while the circular polarization states were generated by just rotating R1. The polarized light was focused on the sample surface. The back-scattered light passed through the quarter-wave plate (R2) and linear polarizer (P2) and was captured by the CCD camera. The captured images were then interfaced to the computer for further processing using the Stokes–Mueller method.

In establishing the experimental setup, the linear polarizers (P1 and P2) and quarter-wave plates (R1 and R2) were individually mounted on stepping motors to minimize rotation errors. Moreover, to minimize light reflection effects from the sample surface, the angle between the generator axis and analyzer axis was set as 30 deg. For the calibration process, a standard mirror (BB1-E02, Thorlabs, Inc.) was captured and calculated the Mueller matrix images with an accuracy of 102 (see Ref. 30). After calibration, the polarimeter system was conducted the experiments.

2.3.

Mueller Matrix Imaging

Figure 3 shows the experimental procedure of the proposed study with four stages, including sample preparation, capturing 36 polarization images, image processing and calculation of Mueller matrix elements and polarization parameters, and then data analysis. In the Mueller matrix imaging technique, the input and output Stoke vectors obtained for the six polarization states are replaced by 36 images showing the corresponding states of polarization.17 The Mueller matrix image for the sample is then computed as shown in Eq. (1), where H, P, V, and M refer to linear polarized lights with orientations of 0 deg, 45 deg, 90 deg, and 135 deg, respectively. And L and R correspond to left-handed and right-handed circular polarization lights, respectively. Thus, element m11, for example, requires four measurements, HH, VV, HV, and VH since m11=HH+HV+VH+VV. In the present study, the matrix elements were constructed using a self-written program coded in Python to merge the individual polarized images as required JBO_29_5_052917_f018.png

Fig. 3

Experimental procedure.

JBO_29_5_052917_f003.png

2.4.

Mueller Matrix Transformation Parameters

Mueller matrix transformation (MMT) techniques provide a quantitative approach for characterizing the Mueller matrix elements in such a way as to evaluate specific properties of the sample, including the microstructure, density and size of the subwavelength scatterers, orientation and alignment of the fibers, presence of scatterers of various shapes and sizes, and so forth.31,32 Various MMT parameters are available,33 depending on the particular characteristics of interest. For example, He et al.34 proposed the following four parameters for characterizing the polarization properties of biological tissue samples:

Eq. (2)

A=2(m22+m33)(m22m33)2+(m23+m32)2(m22+m33)2+(m22m33)2+(m23+m32)2,

Eq. (3)

b=m22+m332,

Eq. (4)

t=(m22m33)2+(m23+m32)22,

Eq. (5)

Δ=1|m22|+|m33|+|m44|a0Δ1,
where A is the anisotropy index, b is the depolarization power factor, t is an index related to the magnitude of the anisotropy, and Δ is the depolarization power.

3.

Results

3.1.

Mueller Matrix of Healthy Human Breast Samples

A total of 210 Mueller matrix images were obtained for the 35 healthy breast samples. Figure 4(a) shows the normalized Mueller matrix image of a typical healthy human breast sample. The color bar indicates the colorimetric intensity of the matrix elements in the interval of [1,1]. As noted above, element m11 is used for normalization purposes and thus has the maximum value of 1 (red). The average intensity of each Mueller matrix element was computed and then normalized to a value in the range of 1 to 1 using the intensity of element m11 as a reference. The corresponding results for the standard deviation and average values of the matrix elements are presented in Figs. 4(b) and 4(c), respectively.

Fig. 4

(a) Normalized Mueller matrix image of typical healthy human breast sample, (b) standard deviation values, and (c) average intensity values of 210 Mueller matrix images of healthy human breast tissue samples.

JBO_29_5_052917_f004.png

3.2.

Mueller Matrix of Human Breast Cancer Samples

3.2.1.

Benign human breast cancer sample

A total of 120 Mueller matrix images were obtained of the 20 benign breast cancer samples. Figure 5(a) shows the Mueller matrix image of a typical sample in the benign breast cancer sample set, while Figs. 5(b) and 5(c) show the standard deviation and mean values of the intensities of the 16 elements in each image.

Fig. 5

(a) Normalized Mueller matrix image of typical benign human breast cancer sample, (b) standard deviation values, and (c) average intensity values of 120 Mueller matrix images of benign human breast cancer samples.

JBO_29_5_052917_f005.png

3.2.2.

Grade-2 malignant human breast cancer sample

A total of 120 Mueller matrix images were obtained of the 20 grade-2 malignant human breast cancer samples. Figure 6(a) presents the Mueller matrix image of a typical grade-2 malignant sample, while Figs. 6(b) and 6(c) show the standard deviation and mean values of the Mueller matrix element intensities for the 20 samples.

Fig. 6

(a) Normalized Mueller matrix image of typical grade-2 malignant human breast cancer sample, (b) standard deviation values, and (c) average intensity values of 120 Mueller matrix images of grade-2 malignant human breast cancer samples.

JBO_29_5_052917_f006.png

3.2.3.

Mueller matrix of grade-3 malignant human breast cancer sample

A total of 120 Mueller matrix images were similarly obtained of the 20 grade-3 malignant human breast cancer samples. Figure 7(a) shows the normalized Mueller matrix image of a typical sample, while Figs. 7(b) and 7(c) show the mean and standard deviation values of the Mueller matrix element intensities for the 20 samples.

Fig. 7

(a) Normalized Mueller matrix image of typical grade-3 malignant human breast cancer sample, (b) standard deviation values, and (c) average intensity values of 120 Mueller matrix images of grade-3 malignant human breast cancer samples.

JBO_29_5_052917_f007.png

4.

Discussion

4.1.

Mueller Matrix Analysis

4.1.1.

Comparison between healthy and cancerous human breast tissue samples

Figure 8 compares the Mueller matrix images of a healthy breast tissue sample (left) and cancerous breast tissue sample (right). The Mueller matrix image of the cancerous sample generally shows more color change than the healthy sample and thus provides a rough guideline for estimating the pathological state of the sample as either healthy or cancerous. A detailed inspection shows that the intensity difference between the two samples is particularly pronounced for elements m23,m32, and m44. For example, the color intensity of element m23 for the healthy sample is substantially brighter (more than 0.5) than that for the corresponding element in the cancerous sample (see also Table 1). By contrast, for element m32, the color intensity of the cancer sample is much higher than that of the healthy sample. Finally, for element m44, the color intensity of the healthy sample is once again darker than that of the cancer samples.

Table 1

Average intensity values of 16 elements in normalized Mueller matrix images of healthy and cancerous human breast tissue samples.

Average intensityHealthyCancerPercentage difference (%)
m111 ± 01 ± 00
m120.4845 ± 0.01410.4581 ± 0.00075.4
m130.4840 ± 0.02720.5273 ± 0.01278.2
m140.4674 ± 0.01870.5049 ± 0.00437.4
m210.4611 ± 0.01170.4737 ± 0.00302.7
m220.2275 ± 0.02730.2197 ± 0.01873.4
m230.6394 ± 0.01750.1915 ± 0.010170.1
m240.4566 ± 0.01690.4952 ± 0.00347.8
m310.4535 ± 0.01360.5737 ± 0.001821.0
m320.3201 ± 0.00870.8129 ± 0.017760.6
m330.2282 ± 0.02080.2052 ± 0.004910.1
m340.5156 ± 0.02580.4414 ± 0.002614.4
m410.5124 ± 0.01220.4833 ± 0.01005.7
m420.4803 ± 0.02860.5239 ± 0.01138.3
m430.4755 ± 0.02930.4916 ± 0.04693.3
m440.6524 ± 0.00370.3562 ± 0.002245.4

Fig. 8

Normalized Mueller matrix images of (a) healthy human breast sample and (b) cancerous sample.

JBO_29_5_052917_f008.png

Figures 9 and 10 show the average intensities of the 16 elements in the Mueller matrixes of the healthy and cancerous breast tissue samples. Clear differences are observed between the intensities of the m23,m32, and m44 elements in the two samples. In particular, the intensity of the m23 element in the cancer sample (0.1915) is substantially lower than that in the healthy sample (0.6394). Conversely, the intensity of element m32 in the cancer group (0.8129) is around three times higher than that in the healthy group (0.320). Finally, the intensity of element m44 in the cancer sample (0.3562) is around half that of the intensity in the healthy sample (0.6524). A paired t-test comparison revealed that the difference in the intensity values between the two different samples was significant for each of the three elements (p<0.00001).

Fig. 9

Average normalized intensity of each element in Mueller matrixes of healthy and cancerous human breast tissue samples.

JBO_29_5_052917_f009.png

Fig. 10

Average normalized intensities of each element in Mueller matrixes of healthy and cancerous human breast tissue samples. The star symbols represent the p-values: 0.002 (**), <0.0001 (****), as determined by a paired t-test.

JBO_29_5_052917_f010.png

As shown in Fig. 10, elements m14,m21,m31,m34, and m41 also show a substantial (but smaller) difference in intensity between the samples of different classes (p<0.0001). Thus, overall, the results confirm that the intensity of the Mueller matrix elements provides a feasible means of distinguishing between healthy and cancerous human breast tissue samples.

Figure 11 shows the correlation coefficients between each pair of elements in the Mueller matrixes of the two sample classes, demonstrating the paired relationship between 15 variables in the correlation matrix. When the correlation coefficient value is positive and has a large value (close to 1), it implies the existence of a similar and consistent relationship between the two variables. Conversely, if the correlation coefficient value is negative and has a small value (close to 1), it indicates the discrepancy between the pair of variables. As shown, the correlation coefficient has a relatively low value (<0.80) for all of the element pairs other than those between m44 and m34 and m41, respectively, and m41 and m34. In other words, the results confirm that the intensity values of elements m23, m32, and m44 enable healthy and cancerous breast tissue samples to be reliably differentiated.35

Fig. 11

Correlation matrix of Mueller matrix element intensities for healthy and cancerous human breast tissue samples.

JBO_29_5_052917_f011.png

4.1.2.

Comparison among healthy, benign, and malignant (grades 2 and 3) human breast cancer samples

Figure 12 compares the Mueller matrix images of the four sample types of a randomly selected sample set (namely, healthy, benign, grade-2 malignant, and grade-3 malignant). It is noted that, a set of 210 Mueller matrix images was obtained from the 35 healthy breast samples, each set of 120 Mueller matrix images was obtained from the 20 benign breast cancer samples, grade-2, and grade-3 malignant human breast cancer samples, respectively. For statistics of all measurement sample groups, the average intensity values and their standard deviations of four sample group were shown in Table 2 and Fig. 13. Obvious color intensity differences are observed between the four samples, particularly for elements m23,m32, and m44. The intensity of element m23 in the healthy sample is greater (i.e., darker) than that of the corresponding element in any of the cancer samples. Thus, the intensity of this element provides a quick and simple approach for screening potential cancerous samples for further histopathological analysis. Element m32 has a strong dark intensity in all three cancer samples (benign, grade-2, and grade-3). By contrast, it has a color intensity of <0.4 in the healthy sample. Consequently, the intensity of element m32 also provides a quick approach for identifying cancerous samples. Finally, the intensity of element m44 in the healthy sample has the highest value (>0.6) of all the samples. Furthermore, for the cancer samples, the intensity reduces with increasing severity. Thus, the intensity of element m44 not only provides the means to differentiate between healthy and cancerous human breast tissue samples but also to estimate the possible grade of the cancerous samples.

Fig. 12

Normalized Mueller matrix images of healthy and cancerous human breast tissue samples.

JBO_29_5_052917_f012.png

Table 2

Average intensity values of 16 elements in normalized Mueller matrix images of healthy and cancerous human breast samples.

Average intensityHealthyBenign cancerMalignant cancer (grade 2)Malignant cancer (grade 3)
m111 ± 01 ± 01 ± 01 ± 0
m120.4845 ± 0.01410.4911 ± 0.02860.4785 ± 0.01320.4581 ± 0.0007
m130.4840 ± 0.02720.4938 ± 0.00290.4825 ± 0.00560.5273 ± 0.0127
m140.4674 ± 0.01870.4528 ± 0.00250.4666 ± 0.00550.5049 ± 0.0043
m210.4611 ± 0.01170.4923 ± 0.02670.4815 ± 0.00100.4737 ± 0.0030
m220.2275 ± 0.02730.2248 ± 0.0180.2271 ± 0.02580.2197 ± 0.0187
m230.6394 ± 0.01750.2237 ± 0.00710.2432 ± 0.02670.1915 ± 0.0101
m240.4566 ± 0.01690.4669 ± 0.01630.4722 ± 0.00880.4952 ± 0.0034
m310.4535 ± 0.01360.4851± 0.00420.5244 ± 0.00620.5737 ± 0.0018
m320.3201 ± 0.00870.6840 ± 0.00270.7114 ± 0.00480.8129 ± 0.0177
m330.2282 ± 0.02080.2243 ± 0.00640.2111 ± 0.00170.2052 ± 0.0049
m340.5156 ± 0.02580.4827 ± 0.00410.4573 ± 0.01210.4414 ± 0.0026
m410.5124 ± 0.01220.4859 ± 0.00880.4831 ± 0.00520.4833 ± 0.0100
m420.4803 ± 0.02860.4886 ± 0.01210.4788 ± 0.00320.5239 ± 0.0113
m430.4755 ± 0.02930.4781 ± 0.02680.4721 ± 0.00420.4916 ± 0.0469
m440.6524 ± 0.00370.5646 ± 0.00430.4366 ± 0.00310.3562 ± 0.0022

Fig. 13

Average normalized intensity of each element in Mueller matrixes of healthy and cancerous human breast tissue samples.

JBO_29_5_052917_f013.png

Table 2 and Fig. 13 show the average intensities and standard deviations of the 16 elements in each of the four sample groups (namely, healthy, benign, malignant grade-2, and malignant grade-3). It is noted that the average intensities were calculated based on the values of pixels in the image of each Mueller matrix element and then normalized from the range of [0, 255] to [1,1] accordingly. The results confirm that the main difference between the intensities of the various classes lies in elements m23,m32, and m44. For example, element m23 in the healthy sample has a value of 0.6394 and is substantially higher than that in the benign sample (0.2237) or malignant cancer samples (grade-2: 0.2432 and grade-3: 0.1915). For element m32, the intensity in the healthy sample (0.3201) is much lower than that in any of the cancer samples (benign: 0.6840, grade-2: 0.7114, and grade-3: 0.8129). Finally, for element m44, the healthy sample shows the highest intensity value (0.6524), followed by the benign sample (0.5646), grade-2 malignant sample (0.4366), and grade-3 malignant sample (m44=0.3562), respectively. Overall, the intensity values of elements m23 and m44 are arranged in descending order as follows: healthy > benign > grade-2 malignant > grade-3 malignant. Conversely, the intensity values of element m32 are arranged as follows: grade-3 malignant > grade-2 malignant > benign > healthy.

Figures 14 and 15 show the one-way ANOVA test results for the intensity differences between the four sample classes (healthy, benign, grade-2, and grade-3) and three cancer sample classes (benign, grade-2, and grade-3), respectively. Referring to Fig. 14, the element intensities of the healthy sample differ from those of the three cancerous samples with a significance of p<0.05 for all elements other than m22 and m43. Elements m23,m31,m32,m34, m41, and m44 all show particularly significant differences in intensity between the healthy sample and the benign, grade-2, and grade-3 samples (p<0.0001). In particular, elements m23,m32, and m44 have significantly greater differences in intensities between the healthy sample and cancerous tissue group. Similarly, in Fig. 15, elements m13,m14,m23,m31,m32,m34,m42, and m44 show significant intensity differences between the three kind samples (benign, grade-2, and grade-3), with p-values ranging from <0.0001 to 0.002. It is quite surprising that there are a clear difference in intensity for each pair of elements m31, m32, and m44.

Fig. 14

One-way ANOVA results for Mueller matrix element intensities of healthy and cancerous samples. The star symbols represent the p-values: 0.033 (*), <0.001 (***), <0.0001 (****), as determined by paired t-test.

JBO_29_5_052917_f014.png

Fig. 15

One-way ANOVA results for Mueller matrix element intensities of cancer samples. The star symbols represent the p-values: 0.002 (**), <0.001 (***), <0.0001 (****), as determined by paired t-test.

JBO_29_5_052917_f015.png

4.2.

MMT Parameters

Table 3 shows the average MMT parameter values calculated from Eqs. (24) for the healthy, benign, grade-2, and grade-3 samples. The anisotropy indicator, A, has an explicit link to the anisotropy or isotropy of optical media.34 In particular, A reduces as the dispersion of the fiber orientation angle or order of alignment of the fibers decreases. In other words, a value of A closed to 1 indicates an presence of fibrous scatters and thus implies that the medium is anisotropic. The depolarization power factor, b, is a scatterer offset indicator, with a value determined by both the structure and the density of the scatterers.34 While b is associated with numerous media attributes, it is particularly sensitive to the scatterer size, particularly for scatterers with a dimension smaller than the wavelength of the incident light. Finally, parameter t is related to the magnitude of the anisotropy of the sample.34 In other words, t is sensitive to the structure of anisotropic media. In practice, t is closely related to b and varies in the interval of [0,b], where 0 and b correspond to perfectly isotropic and perfectly anisotropic materials, respectively.

Table 3

MMT parameters A,b,t, and depolarization powers for healthy and cancerous human breast tissue samples.

SampleAnisotropy indicator (A)Depolarization power factor (b)Index related to magnitude of anisotropy (t)Depolarization powers (Δ)
Healthy0.77210.22780.64180.4459
Benign cancer0.79420.22450.67500.4931
Malignant cancer (grade 2)0.78660.21910.67840.549
Malignant cancer (grade 3)0.75100.21210.71020.5944

The results presented in Table 3 show that all four sample classes have an anisotropy indicator value of A closed to 0.7. In other words, all of the samples, both healthy and cancerous, are anisotopic. Furthermore, parameter b reduces with an increasing cancer severity, whereas parameter t increases. Thus, it is inferred that the depolarization effect intensifies as the cancer severity increases.

For optical media, a lower value of the depolarization power (?) indicates a greater isotropy. Conversely, a higher depolarization power indicates a greater anisotropy.36 Table 3 presents the depolarization powers of the present samples, as calculated using Eq. (5). The healthy sample has a depolarization power of Δ=0.4459 and is thus confirmed to be more isotropic, with a simple microstructure. By contrast, the grade-3 malignant cancer sample has a depolarization power of Δ=0.5944, which indicates that the sample has a more anisotropic and complicated microstructure.

4.3.

FDHs of Mueller Matrix Element Intensities

Figure 16 shows the FDHs of the 16 elements in the Mueller matrix images of the healthy and cancerous tissue samples. As expected, the FDHs of the two classes are widely separated for the m23,m32, and m44 elements, with the distributions of the healthy samples centered at higher intensity values for the m23 and m44 elements and a lower intensity value for the m32 element. Interestingly, the results suggest that the intensity of element m31 may also provide a feasible means of differentiating between healthy and cancerous samples.

Fig. 16

FDHs of pixel intensity of each Mueller matrix element for healthy and cancerous human breast tissue samples.

JBO_29_5_052917_f016.png

Figure 17 compares the intensity FDHs of the healthy, benign, grade-2 malignant, and grade-3 malignant samples. It is seen that the FDHs of the four classes are well-spaced for the m32 and m44 elements. Specifically, at position m32, all four lines do not overlap, the blue line (healthy group) deviates completely to the left while the remaining three lines are arranged in a separate order and clearly deviate to the right. Similarly at m44, a signal is authentic when all four lines are clearly separated and located in each separate region. In other words, the intensities of these two elements are particularly suitable for distinguishing between the four sample classes.

Fig. 17

FDHs of pixel intensity of each Mueller matrix element for healthy, benign, grade-2 malignant, and grade-3 malignant samples.

JBO_29_5_052917_f017.png

It can be seen that normal cells are usually uniform in size and shape, with a well-defined nucleus, whereas cancer cells can vary greatly in size and shape, and their nuclei may be enlarged or deformation. These cause different changes in polarization characteristics including birefringence, dichroism, and depolarization, leading the distinctions of polarization images among types of samples. For example, the structure of cartilage or tendons exhibits birefringence while collagen fibers exhibit arbitrary orientation.37,38 The depolarization also depends on the different morphological parameters of tissue such as the density, its distribution, size, shape, and refractive index of the tissue scatterers.21,26 The polarization capacities of biological tissues, nanoparticles, and other materials have all been studied using the Mueller matrix approach, which has been used to identify the characteristics of various sample types. As a result, the Mueller matrix method advances light science in medicine, especially in the areas of disease and cancer diagnosis. It could lead to the development of economical and noninvasive diagnostic methods in the near future.

Overall, the results presented above indicate the potential for using MMP to differentiate between healthy and cancerous human breast tissue samples. It is noted that performing polarization measurements on thin histology slides of breast cancer tissue is the first step of the proposed study in finding differences in polarization characterizations among types of breast tissue at different levels of cancers. Once this result has been confirmed and proven, the study will be carried out in an anesthetized mouse model without the need to remove the sample from the tumor. However, it should be noted that the present trials have considered only a relatively small sample size (n=20 to 35) for each sample class. A small sample size is associated with greater variability and lower dependability since it may contribute to bias.39 As a result, future studies should repeat the trials with a larger sample size to corroborate the present findings. In addition, the measurement process used in the present study is time-consuming (from 10 to 15 min) since the polarization states of the polarizers are adjusted one by one under motor-driven rotation stages. Thus, reducing the measurement time by a software solution for automatic control of the motor-driven rotation stages is necessary.

5.

Conclusions

Optical diagnostic techniques have attracted significant attention in the literature as a means of evaluating a wide range of substances, including textiles and biological tissues, due to their nondestructive properties, speed, and affordability. The present study has used a backscattering MMP system to classify four classes of human breast tissue samples: healthy, benign, grade-2 malignant, and grade-3 malignant. The observation results have shown that the average intensities of the m23,m32, and m44 elements of the Mueller matrix image provide a reliable means of distinguishing not only between healthy and cancerous samples but also of estimating the grades of cancerous samples as benign, grade-2 malignant, or grade-3 malignant. The ANOVA test results have confirmed that the intensities of these three elements show significant differences (p<0.0001) among the four classes of samples. The FDHs of the element intensities have suggested that elements m32 and m44 are particularly effective indicators of the sample class. The tendencies of the MMT parameters derived for the four classes are consistent with the observation and statistical results. Overall, the results presented in this study suggest that backscattering MMP has significant potential as a tool for aiding pathologists in diagnostic applications for breast cancer.

Disclosures

The authors declare no conflicts of interest.

Code and Data Availability

No materials were used for the analysis. The code and data used to generate the results are available in the Code Ocean repository: https://doi.org/10.24433/CO.7469965.v1.

Acknowledgments

The authors gratefully acknowledge the financial support provided to this study by the Vietnam National University Ho Chi Minh City (VNU-HCM) (Grant No. DS2023-28-02).

References

1. 

I. Soerjomataram and F. Bray,, “Planning for tomorrow: global cancer incidence and the role of prevention 2020–2070,” Nat. Rev. Clin. Oncol., 18 (10), 663 –672 https://doi.org/10.1038/s41571-021-00514-z (2021). Google Scholar

2. 

, “Breast cancer treatment (PDQ®) - Health professional version,” https://www.cancer.gov/types/breast/hp/breast-treatment-pdq (). Google Scholar

3. 

T. Li et al., “The association of measured breast tissue characteristics with mammographic density and other risk factors for breast cancer,” Cancer Epidemiol. Biomark. Prev., 14 (2), 343 –349 https://doi.org/10.1158/1055-9965.EPI-04-0490 (2005). Google Scholar

4. 

N. F. Boyd et al., “Mammographic density and the risk and detection of breast cancer,” N. Engl. J. Med., 356 (3), 227 –236 https://doi.org/10.1056/NEJMoa062790 NEJMAG 0028-4793 (2007). Google Scholar

5. 

S. M. Shah et al., “Artificial intelligence for breast cancer analysis: trends & directions,” Comput. Biol. Med., 142 105221 https://doi.org/10.1016/j.compbiomed.2022.105221 CBMDAW 0010-4825 (2022). Google Scholar

6. 

T. Onega et al., “Facility mammography volume in relation to breast cancer screening outcomes,” J. Med. Screen., 23 (1), 31 –37 https://doi.org/10.1177/0969141315595254 (2016). Google Scholar

7. 

A. F. Flemban, “Overdiagnosis due to screening mammography for breast cancer among women aged 40 years and over: a systematic review and meta-analysis,” J. Pers. Med., 13 (3), 523 https://doi.org/10.3390/jpm13030523 (2023). Google Scholar

8. 

L. Wang, “Early diagnosis of breast cancer,” Sensors, 17 (7), 1572 https://doi.org/10.3390/s17071572 SNSRES 0746-9462 (2017). Google Scholar

9. 

C. H. Neal and M. A. Helvie, “Overdiagnosis and risks of breast cancer screening,” Radiol. Clin. North Am., 59 (1), 19 –27 https://doi.org/10.1016/j.rcl.2020.09.005 (2021). Google Scholar

10. 

R. S. Gurjar et al., “Imaging human epithelial properties with polarized light-scattering spectroscopy,” Nat. Med., 7 (11), 1245 –1248 https://doi.org/10.1038/nm1101-1245 1078-8956 (2001). Google Scholar

11. 

B. Kunnen et al., “Application of circularly polarized light for non-invasive diagnosis of cancerous tissues and turbid tissue-like scattering media,” J. Biophotonics, 8 (4), 317 –323 https://doi.org/10.1002/jbio.201400104 (2015). Google Scholar

12. 

T.-T.-H. Pham et al., “Stokes–Mueller matrix polarimetry: effective parameters of anisotropic turbid media: theory and application,” Optical Polarimetric Modalities for Biomedical Research, 47 –76 Springer International Publishing, Cham (2023). Google Scholar

13. 

S. L. Jacques et al., “Imaging superficial tissues with polarized light,” Lasers Surg. Med., 26 (2), 119 –129 https://doi.org/10.1002/(SICI)1096-9101(2000)26:2<119::AID-LSM3>3.0.CO;2-Y LSMEDI 0196-8092 (2000). Google Scholar

14. 

D. H. Goldstein, Polarized Light, 3rd ed.CRC Press( (2011). Google Scholar

15. 

D. S. Kliger et al., “Chapter 5: The Jones calculus, Mueller calculus, and Poincaré sphere revisited,” Polarized Light in Optics and Spectroscopy, 103 –152 Academic Press, Inc.( (1990). Google Scholar

16. 

C. He et al., “Polarisation optics for biomedical and clinical applications: a review,” Light: Sci. Appl., 10 (1), 194 https://doi.org/10.1038/s41377-021-00639-x (2021). Google Scholar

17. 

J. S. Baba et al., “Development and calibration of an automated Mueller matrix polarization imaging system,” J. Biomed. Opt., 7 (3), 341 –349 https://doi.org/10.1117/1.1486248 JBOPFO 1083-3668 (2002). Google Scholar

18. 

L. Qiu et al., “Multispectral scanning during endoscopy guides biopsy of dysplasia in Barrett’s esophagus,” Nat. Med., 16 (5), 603 –606 https://doi.org/10.1038/nm.2138 1078-8956 (2010). Google Scholar

19. 

V. Backman et al., “Detection of preinvasive cancer cells,” Nature, 406 (6791), 35 –36 https://doi.org/10.1038/35017638 (2000). Google Scholar

20. 

S. Alali and A. Vitkin, “Polarized light imaging in biomedicine: emerging Mueller matrix methodologies for bulk tissue assessment,” J. Biomed. Opt., 20 (6), 061104 https://doi.org/10.1117/1.JBO.20.6.061104 JBOPFO 1083-3668 (2015). Google Scholar

21. 

N. Ghosh and I. A. Vitkin, “Tissue polarimetry: concepts, challenges, applications, and outlook,” J. Biomed. Opt., 16 (11), 110801 https://doi.org/10.1117/1.3652896 JBOPFO 1083-3668 (2011). Google Scholar

22. 

O. Arteaga et al., “Mueller matrix microscope with a dual continuous rotating compensator setup and digital demodulation,” Appl. Opt., 53 (10), 2236 –2245 https://doi.org/10.1364/AO.53.002236 APOPAI 0003-6935 (2014). Google Scholar

23. 

J. Vizet et al., “In vivo imaging of uterine cervix with a Mueller polarimetric colposcope,” Sci. Rep., 7 2471 https://doi.org/10.1038/s41598-017-02645-9 SRCEC3 2045-2322 (2017). Google Scholar

24. 

J. Qi and D. Elson, “A high definition Mueller polarimetric endoscope for tissue characterisation,” Sci. Rep., 6 25953 https://doi.org/10.1038/srep25953 SRCEC3 2045-2322 (2016). Google Scholar

25. 

D. L. Le et al., “Characterization of healthy and nonmelanoma-induced mouse utilizing the Stokes–Mueller decomposition,” J. Biomed. Opt., 23 (12), 125003 https://doi.org/10.1117/1.JBO.23.12.125003 JBOPFO 1083-3668 (2018). Google Scholar

26. 

D. L. Le et al., “Characterization of healthy and cancerous human skin tissue utilizing Stokes–Mueller polarimetry technique,” Opt. Commun., 480 126460 https://doi.org/10.1016/j.optcom.2020.126460 OPCOB8 0030-4018 (2021). Google Scholar

27. 

N. T. Luu et al., “Characterization of Mueller matrix elements for classifying human skin cancer utilizing random forest algorithm,” J. Biomed. Opt., 26 (7), 075001 https://doi.org/10.1117/1.JBO.26.7.075001 JBOPFO 1083-3668 (2021). Google Scholar

28. 

H.-M. Le et al., “Mueller matrix imaging polarimetry technique for dengue fever detection,” Opt. Commun., 502 127420 https://doi.org/10.1016/j.optcom.2021.127420 OPCOB8 0030-4018 (2022). Google Scholar

29. 

L.-Y. Nguyen et al., “Feasibility of Hepatitis B virus infection detection using Mueller matrix decomposition method and Mueller matrix imaging polarimetry technique,” Opt. Commun., 514 128152 https://doi.org/10.1016/j.optcom.2022.128152 OPCOB8 0030-4018 (2022). Google Scholar

30. 

T.-T.-H. Pham et al., “Combined Mueller matrix imaging and artificial intelligence classification framework for Hepatitis B detection,” J. Biomed. Opt., 27 (7), 075002 https://doi.org/10.1117/1.JBO.27.7.075002 JBOPFO 1083-3668 (2022). Google Scholar

31. 

H. Zhai et al., “Distinguishing tissue structures via polarization staining images based on different combinations of Mueller matrix polar decomposition parameters,” Opt. Lasers Eng., 152 106955 https://doi.org/10.1016/j.optlaseng.2022.106955 OLENDN 0143-8166 (2022). Google Scholar

32. 

M. Iqbal et al., “Comparative study of Mueller matrix transformation and polar decomposition for optical characterization of turbid media,” Optik, 224 165508 https://doi.org/10.1016/j.ijleo.2020.165508 OTIKAJ 0030-4026 (2020). Google Scholar

33. 

E. Du et al., “Mueller matrix polarimetry for differentiating characteristic features of cancerous tissues,” J. Biomed. Opt., 19 (7), 076013 https://doi.org/10.1117/1.JBO.19.7.076013 JBOPFO 1083-3668 (2014). Google Scholar

34. 

H. He et al., “A possible quantitative Mueller matrix transformation technique for anisotropic scattering media/Eine mögliche quantitative Müller-Matrix-Transformations-Technik für anisotrope streuende Medien,” Photonics Lasers Med., 2 (2), 129 –137 https://doi.org/10.1515/plm-2012-0052 (2013). Google Scholar

35. 

A. G. Asuero, A. Sayago and A. G. González, “The correlation coefficient: an overview,” Crit. Rev. Anal. Chem., 36 (1), 41 –59 https://doi.org/10.1080/10408340500526766 CCACBB 1040-8347 (2006). Google Scholar

36. 

M. Sun et al., “Characterizing the microstructures of biological tissues using Mueller matrix and transformed polarization parameters,” Biomed. Opt. Express, 5 (12), 4223 –4234 https://doi.org/10.1364/BOE.5.004223 BOEICL 2156-7085 (2014). Google Scholar

37. 

M. Borovkova et al., “Role of scattering and birefringence in phase retardation revealed by locus of Stokes vector on Poincaré sphere,” J. Biomed Opt., 25 (5), 057001 https://doi.org/10.1117/1.JBO.25.5.057001 (2020). Google Scholar

38. 

A. Bajaj et al., “Detection and differentiation of normal, cancerous, and metastatic cells using nanoparticle-polymer sensor arrays,” Proc. Natl. Acad. Sci. U. S. A., 106 (27), 10912 –10916 https://doi.org/10.1073/pnas.0900975106 (2009). Google Scholar

39. 

J. Faber and L. M. Fonseca, “How sample size influences research outcomes,” Dental Press J. Orthod., 19 (4), 27 –29 https://doi.org/10.1590/2176-9451.19.4.027-029.ebo (2014). Google Scholar

Biography

Thi-Thu-Hien Pham received her BS degree in mechatronics from the Ho Chi Minh City University of Technology—Vietnam National University, Ho Chi Minh City, Vietnam, in 2003, and her MS and PhD degrees in mechanical engineering from Southern Taiwan University of Technology and National Cheng Kung University, Tainan, Taiwan, in 2007 and 2012, respectively. She is currently a head of Biomedical Photonics Lab and an associate professor at the School of Biomedical Engineering, International University—Vietnam National University HCMC, Ho Chi Minh City, Vietnam. Her research interests are in the areas of polarized light-tissue studies, polarimetry, optical techniques in precision measurement to determine the optical properties of bio-samples (glucose, collagen, and tumor) or cancer detection (skin, liver, and breast), noninvasive glucose measurement, cell/tissue characterization, laser/LED applications, and artificial intelligence applications.

Thao-Ngan Ngoc Quach received her BS degree in biomedical engineering from International University—Vietnam National University HCMC, Vietnam, in 2022. Her research interests include measurement techniques for cancer detection, imaging polarimetry, and determining the optical properties of bio-samples.

Hoang-Quyen-Quoc Vo received her BS degree in biomedical engineering from International University—Vietnam National University HCMC, Vietnam, in 2023. Her research interests include biosignal imaging throughout the human body and biosamples using electrical signal and polarimetry imaging diagnosis techniques to determine sample-specific properties.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Thi-Thu-Hien Pham, Thao-Ngan Ngoc Quach, and Quoc-Hoang-Quyen Vo "Analysis of polarization features of human breast cancer tissue by Mueller matrix visualization," Journal of Biomedical Optics 29(5), 052917 (13 January 2024). https://doi.org/10.1117/1.JBO.29.5.052917
Received: 7 September 2023; Accepted: 20 December 2023; Published: 13 January 2024