|
1.IntroductionUrethral stricture occurs when a part of the urethral luminal diameter becomes narrowed due to ischemic fibrosis (spongiofibrosis), which manifests as scar tissue in the spongy erectile tissue (corpus spongiosum).1 Male urethral stricture accounts for about 5000 inpatient visits and 1.5 million office visits per year in the USA.2 The prevalence rate of the urethral stricture has been estimated to be (i.e., 800/100,000 cases) with 0.2% in younger men and 0.6% in older men ( years old).1 In contemporary practices, most urethral strictures encountered are idiopathic, traumatic, inflammatory, or iatrogenic.2–4 The patients with symptomatic stricture disease typically present obstructive voiding symptoms such as straining, incomplete emptying, and a weak stream; they might also have a history of recurrent urinary tract infection, prostatitis, epididymitis, hematuria, or bladder stones.2 Depending upon the characteristics of the stricture (i.e., location, etiology, size, and density of spongiofibrosis), a number of invasive treatments have been performed,1 including urethroplasty, balloon dilation, cold knife incision, and laser urethrotomy.3–8 Currently, the most popular treatment method of urethral strictures is open reconstructive surgery of the urethra (also called urethroplasty).6 Yet, in spite of a success rate of , the procedure is highly invasive, expensive, and related to considerable complications.6 The balloon dilation injects a small tube into a urethra to keep the stricture open and stretches fibrotic tissue to widen the urethral lumen.2 The cold knife incision releases scar tissue and increases the size of the urethral lumen.5 However, these minimally invasive treatment methods have presented high complications such as discomfort and pain and risk of recurrence due to incomplete ablation of the scar tissue.5–7 In the case of laser ablation, the delivery of laser light through an end-firing fiber can coagulate or ablate the target tissue for scar removal, but the forward light emission often limits the extent of treatment areas.5–8 Laser-induced interstitial thermotherapy (LITT and also called interstitial LA)9 has been investigated and used as a minimally invasive method for treating different types of tumors in liver, brain, head, neck, and other tissues in terms of thermal coagulation.10–14 LITT delivers laser light through a miniature optical fiber percutaneously to the point of treatment. Volumetric absorption of laser energy within the tissue results in the localized heating, leading to thermal damage to the target tissue in terms of irreversible thermal denaturation.10 However, laser light during LITT often results in less uniform volumetric heat deposition due to geometrical variations in the tumors.10,12,15 In addition, the eventual goal of LITT is to coagulate or even ablate the entire pathological tissue with no or minimal thermal injury to the surrounding healthy tissue.10,13 Thus, to entail more consistent light distribution as well as to real-time monitor temperature developments during laser treatment would be critical for ensuring the safety and efficacy of thermal treatment on soft tissue. Recently, temperature monitoring during LITT for pathological tissue has been performed with thermal sensors in light of contact techniques [e.g., thermocouples (TCs)16,17 and fiber Bragg gratings (FBGs)18,19] and noncontact techniques (e.g., magnetic resonance imaging20,21). On account of cost-effectiveness and easy utilization, the contact method has primarily been used for most LITT processes.10,15 In fact, TCs are moderately inexpensive, quite accurate (), and associated with a relatively short response time (i.e., shorter than 1 s).17 However, the close placement of TCs to the light source often involves direct light absorption and self-heating, leading to overestimation and rather inaccurate recording of the interstitial tissue temperature.16,17 On the other hand, due to high optical sensitivity, FBGs have recently attracted great attention as versatile optical sensing characteristics employed for various biomedical applications such as temperature recording,19,22–25 strain estimation,25,26 and pressure measurement.26,27 Unlike TCs in association with possible light absorption, significantly low light absorption () at near-infrared wavelengths may enable FBGs to overcome the limitation of TCs (i.e., temperature overestimation resulting from self-heating) and to eventually provide precise information on temporal elevations of the tissue temperature during laser treatment.23,28 Therefore, the aim of the current study is to identify the feasible application of FBG as a biomedical temperature sensor to monitor in real time the interstitial temperature in urethral tissue during diffuser-assisted LITT. Both an FBG and a TC as temperature sensing elements were quantitatively compared in terms of temperature variations at different positions away from the diffuser. The degree of tissue coagulation was also evaluated with the temperature monitored real time in tissue. 2.Materials and MethodsA customized diffusing applicator (Tecure, Busan, Republic of Korea) was fabricated to circumferentially coagulate tissue during diffuser-assisted LITT. The diffusing tip was prepared in a length of 20 mm and a diameter of 0.6 mm for delivery of high laser power (980 nm). For mechanical protection, the tip was covered with a customized glass cap (1.4 mm in outer diameter). Figure 1(a) shows a polar measurement setup to characterize spatial light distribution from the diffuser by using a customized goniometer in conjunction with a photodiode (PD-300-3W, Ophir, Jerusalem, Israel). The photodiode was initially rotated around the proximal end of the diffuser at a radius of 4 mm with an increment of 10 deg to measure polar emissions. After each round of the measurement, the diffuser was then moved toward the distal end by 1 mm, and another polar measurement continued to estimate the longitudinal emissions along the diffuser axis. As each photodiode had a spectral range of 350 to 1100 nm, the laser light (980 nm) was applied to evaluate the emission profiles. All the measured intensities were normalized by using the maximum intensity to have the minimal effect of power variations on quantitative evaluations. Figure 1(b) shows a 3-D image of the emission profile from the diffusing tip. In spite of symmetric radial diffusion, the overall longitudinal distribution was skewed to the distal end, and the maximum intensity occurred at 5 mm away from the proximal end. Figure 1(c) exhibits almost isotropic polar emissions at the maximum intensity with a deviation of less than 10% (i.e., ). The mean polar intensity was estimated to be . According to power measurements, 89% and 11% were transmitted in the radial and forward directions, respectively, (around 10% insertion loss). Numerical computations were performed to predict the spatial distribution of the interstitial temperature as well as the degree of thermal denaturation in porcine liver tissue during diffuser-assisted LITT by using COMSOL Multiphysics software (v5.0, COMSOL Inc., Burlington, Massachusetts). The geometry for the simulation used a 20-mm long diffusing tip ( core diameter) with the distal end glass-capped ( and ) as shown in Fig. 2(b). The size of the liver tissue was (tissue boundary), and 4-W 980-nm laser light was cylindrically irradiated through the diffuser for 120 s. Table 1 shows thermal properties of the materials utilized for the current computations. Table 1Thermal properties of materials.
The thermal response of the tissue during the irradiation was estimated by31 where () is the tissue density, () is the tissue specific heat, () is the tissue heat conductivity, and (°C) is the tissue temperature. The effect of blood perfusion and metabolic heat generation was de facto 0 due to the use of the ex vivo model.29,32 () is the heat source induced by laser light. In addition, the preliminary goniometric measurement confirmed that the laser intensity was delivered in two directions: radially emitted power from the diffusing part (89% of incident laser power in W) and forward emitted power from the fiber tip (11% of incident laser power in W). Thus, the spatial beam intensity in the radial direction was assumed to be cylindrical along the diffuser axis, and the corresponding heat source was expressed as where () is the absorption coefficient of the tissue, () is the scattering coefficient of the tissue, (m) is the radial distance from the diffuser surface, and (m) is the diffuser length. Based upon the beam divergence from the diffuser tip (), the beam distribution in the forward direction was assumed to be Gaussian, and the heat source was estimated as where () is the spot size of the laser beam and (m) is the axial depth in tissue. Due to application of the 980-nm wavelength, both the absorption and the scattering effects were significant.30 Thus, the effective attenuation coefficient was utilized instead of the absorption coefficient by33 The initial temperature for the entire tissue was set to be 20°C, and the external tissue surface (i.e., ) was insulated (i.e., , where is the direction of the heat flux). Moreover, Arrhenius parameters were utilized to determine the degree of thermal injury primarily due to the temperature dependence of molecular reaction rates. As a first-order rate process, the tissue thermal damage was described by using the Arrhenius damage integral as34 where is a dimensionless factor to define the irreversible thermal injury, () is the frequency factor, () is the denaturation activation energy, is the universal gas constant of 8.314 (), and (s) is the irradiation time. Thus, the onset of the irreversible thermal denaturation corresponds to the interstitial tissue temperature of 60°C to 65°C, where .35 Table 2 demonstrates the optical properties of the liver under 980 nm, which were assumed to be constant during LITT.Table 2Physical properties of liver tissue.
FBG is typically fabricated by irradiating UV light in a wavelength range of 244 to 248 nm on a Ge-doped photosensitive or hydrogen-loaded single-mode fiber with a spatial pattern, which renders the refractive index of the fiber core be periodically modulated. The periodic core index modulation creates core modes, which are reflected from or transmitted through numerous index boundaries and interfere with each other. In turn, an input beam experiences strong reflection only at the specific wavelength that is determined by a certain phase-matching condition. The reflection wavelength is referred to as the Bragg wavelength () of FBG, and the phase-matching condition, called the Bragg condition, is given by27 where and are the grating period and the effective refractive index of the fundamental core mode at the free space wavelength.37 is affected by a temperature change through the thermal expansion or contraction of the grating period and the thermo-optic effect (thermally induced change in ), which enables the use of the FBG as a temperature sensing element. A Bragg wavelength shift induced by a temperature change is represented as38 where and and are the reference and elevated temperatures applied to the FBG, respectively. is the Bragg wavelength of the FBG at , and and are the thermal expansion and thermo-optic coefficients of a single-mode fiber in which the FBG is inscribed, respectively. On the basis of Eq. (6), the thermal sensitivity of the FBG is defined as38,39 With the pre-estimated , the measurement of will thus allow the estimation of . Although the FBG actually has the cross sensitivity to temperature and strain, it is assumed here that is shifted only by temperature change as strain-induced is small enough to be ignored.23 In the current study, an FBG (SMF-28e-acrylate coating, Technica Optical Component, LLC, Beijing, China) was utilized to monitor in real time the interstitial tissue temperature during laser irradiation. The FBG with a 5-mm active length has a reflectivity of 74.5% at of 1550.03 nm, a 3-dB bandwidth of 0.38 nm, and a side-mode suppression ratio of 28 dB.Prior to application of an FBG sensor for diffuser-assisted LITT, was obtained and calibrated by using the relationship between and [based on Eq. (7)]. An experimental static calibration was performed under the controlled environmental temperature as shown in Fig. 2(a), for which the experimental setup consisted of an environmental chamber, an FBG, an amplified spontaneous emission (ASE) source, an optical spectral analyzer (OSA), and an optical circulator. Initially, the FBG was placed in the chamber within which temperature could be controlled in a range from to 150°C (TH-KE-025 Environmental Chamber 25L, Jeio Tech. Co. Inc, Daejeon, Korea). Broadband light from the ASE source (ASE-FL7701 SCL-band, FiberLabs Inc., Saitama, Japan) reaches the FBG via the circulator. Then reflected signal of the FBG enters the OSA (AQ6370C, Yokogawa, Tokyo, Japan) via the circulator. The reflected spectrum of the FBG was monitored with the OSA. The calibration testing was executed in a temperature range between 20°C and 100°C, and each temperature interval was 10°C and lasted for 5 h. Both a -type thermocouple and FBG were employed to measure temperature elevations in tissue during diffuser-assisted LITT [Fig. 2(b)]. Porcine liver tissue was used as an ex vivo sample due to easy procurement and storage for LITT. All the samples were acquired from a local slaughter house, and each specimen was prepared with a size of and stored in saline at room tempeature (20°C) prior to testing. Any discoloration in the sample during LITT indicated irreversible thermal denaturation in the tissue. Then the diffusing applicator was perpendicularly inserted into the ex vivo tissue (20-mm deep). Both the thermocouple and FBG were also inserted into the predetermined position of the tissue parallel to the diffusing device. Various distances of the positions were tested from 0 (contract) to 3.5 mm. Each tip was positioned in the middle of the diffuser in order to measure the maximum temperature increase. The -type thermocouple (TSL-101, Guageworld, Anyang, Republic of Korea) with an accuracy of was connected to a data acquisition (DAQ) module (OMB-DAQ-55, OMEGA Engineering Inc., California) to acquire in real time and record any temperature variations in the tissue. The reflectance spectrum of the FBG was collected simultaneously by using LabVIEW software and controlling OSA through a general purpose interface bus (GPIB—IEEE 4888, National Instruments, Austin, Texas). As a light source, a 980-nm laser system (60-W Laser Diode, Apollo Instrument, Irvine, California) was employed to deliver optical energy in a continuous-wavelength mode through the diffusing applicator at various power levels (3, 4, and 5 W). The corresponding average irradiances on the tissue surface were estimated to be , 3.2, and , respectively. The initial tissue temperature was maintained at 20°C (i.e., no irradiation), and the total irradiation times were varied from 10 to 120 s with an increment of 10 s. Each experiment was repeated five times (). The transmission loss of the diffusing applicator was measured by using a goniometer after every five tests, and the diffuser was replaced if the transmission dropped by more than 10% possibly due to tissue contamination or mechanical scratches on the cap. After irradiation testing, each sample was cross sectioned in longitudinal (along diffuser axis) and transverse (6 mm away from proximal end) directions. Then each cross section was imaged using a digital camera. The spatial extent of coagulation in the tested tissue was measured by using the Image J (National Institutes of Health, Bethesda, Maryland). The coagulated regions were defined as discoloration (i.e., tan color) after the laser irradiation. To investigate the feasibility of temperature measurements with the FBG sensor for urological application, porcine urethral tissue was procured from a local slaughter house and prepared in 4.5-cm length and 2-mm inner diameter. Prior to laser irradiation, a diffusing applicator was placed into the tubular urethra and positioned in the center of the lumen. Then the FBG was inserted parallel to the diffuser and situated on the inner wall of the tissue. Due to the size of the protective cap, the diffuser surface was in contact with the luminal wall. The urethral tissue was irradiated in air for 72 s at 4 W 980 nm to achieve -thick thermal coagulation, based upon the results of the ex vivo liver experiments. Five samples were tested under the same irradiation conditions. After the laser treatment, each tested tissue was stored in 10% neutral-buffered formalin (VWR International, Westchester, Pennsylvania) for three days and embedded in paraffin for histology preparation. Then the treated region was sectioned into -thick specimens along the urethra every 4 mm (i.e., total five slides per sample) and stained with hematoxylin and eosin (H&E). An optical transmission microscope was used to image the prepared slides. The degree of thermal denaturation in the urethral tissue was measured with Image J () for quantitative evaluations. For statistical analysis, Mann U test as a nonparametric method was performed and represented insignificant correlation. 3.ResultsFigure 3 shows the variation of the reflectance spectrum of the FBG, measured by the OSA at various temperatures in the environmental chamber [Fig. 2(a)] for calibration purposes. According to Fig. 3(a), the static calibration curve of the FBG demonstrated that the peak wavelength () shifted to the longer wavelengths along with temperature increase (i.e., from 1549.80 nm at 0°C to 1550.71 nm at 80°C). Figure 3(b) shows a linear relationship between the applied temperature and the wavelength shift (), indicating that the thermal sensitivity of the measuring system is constant. The thermal sensitivity (= spectral shift/temperature change) of the FBG used in the tests was estimated to be . Figure 4 shows transient responses of temperature in ex vivo bovine liver during 980-nm LITT for 90 s at various power levels (3, 4, and 5 W). For the sake of comparison, the interstitial temperature at the diffuser surface was measured by using the FBG and type thermocouple, as shown in Figs. 4(a) and 4(b), respectively. The initial temperature of the tissue was (). Regardless of the sensor type, the tissue temperature gradually increased with irradiation time as well as laser power. Given the power of 5 W for 90-s irradiation, the tissue temperature measured by using the FBG went up to (i.e., ), whereas the thermocouple measured up to (i.e., ). Figure 5 shows temperature variations as a function of irradiation time measured at various positions in the tissue (i.e., 0, 1.5, and 3.5 mm away from the diffuser surface) during 120-s irradiation at 4 W. Irrespective of the position, both the FBG and the thermocouple confirmed that the interstitial temperature increased with irradiation time. At the diffuser surface (i.e., 0 mm as direct contact), the thermocouple measured up to 41.0% higher temperature and 43.1% faster transient temperature change than the FBG did after 120-s irradiation (i.e., and for FBG versus and for thermocouple; ). However, as the distance increased, both the FBG and the thermocouple reached comparable temperature elevations (i.e., and for FBG versus and for thermocouple at 3.5 mm with 120-s irradiation; ). The temporal development of the measured temperature with the FBG was compared with that of the simulated temperature (red dashed line) at various radial positions in Fig. 6(a). Irrespective of the position, both the simulation and the experiment obtained almost similar temperature elevations except for a slight difference in the initial development. At the diffuser surface, the heating rates after 4-W irradiation for 120 s were estimated to be 0.61 and for the simulation and the measurement, respectively. Figure 6(b) exhibits the simulated temperature distribution along the diffuser axis after 4-W irradiation for 120 s. The overall distribution was symmetrically cylindrical around the axis, and the temperature decreased significantly with the radial distance. After 120-s irradiation, the temperature at the diffuser surface reached up to 93.3°C. Figure 7 shows simulation results of temperature distribution as well as the extent of thermal coagulation in liver tissue after 4-W irradiation for 72 s. The transverse images were acquired from the middle position of the diffuser. According to Fig. 7(a), the temperature developed cylindrically around the diffuser axis (black solid circles), and the maximum temperature (81.5°C) occurred at the diffuser surface. The dotted line represents the threshold temperature to initiate the irreversible thermal denaturation in the tissue. Figure 7(b) shows the corresponding distribution of the thermal coagulation in tissue. In accordance with the temperature development, the irreversible denaturation (i.e., ) was almost radially formed (black dotted line). The estimated radial coagulation thickness was . Figure 8 shows thermal responses of bovine liver tissue during 4-W diffuser assisted LITT for 120-s irradiation. According to Fig. 8(a), a series of transverse (left column) and cross-sectional (right column) images of the coagulated tissue were captured at various irradiation times ranging from 24 to 120 s. Apparently, the degree of the thermal coagulation (defined as discoloration on tissue surface) evolved with irradiation time. Due to asymmetric beam distribution in Fig. 1(b), the cross-sectional images showed coagulative responses in an oval shape. Figure 8(b) compares the temporal development of the interstitial tissue temperature (measured by the FBG was placed at the diffuser surface; black) with that of the measured radial coagulation thickness (blue). The thickness was measured 5 mm away from the proximal end of the diffuser as the position yielded the maximum intensity [Fig. 1(b)]. Overall, both the temperature and the thickness increased with irradiation time. Unlike the temperature, the extent of the tissue coagulation showed a linear increase (). In fact, the radial expansion of the tissue denaturation reached up to (; corresponding ) after 120 s. -thick coagulation was induced at the irradiation time of 72 s, corresponding to . It was noted that the radial coagulation thickness from the experiments was thicker than that from the simulation in Fig. 7(b) i.e., for tissue versus for model; ). Figure 8(c) shows the spatial distribution of light intensity (black) and coagulation thickness (blue) along the diffuser axis (i.e., longitudinal distance). Both the intensity and the thickness were normalized for qualitative comparison. According to Fig. 8(c), the overall spatial distributions were quite comparable except for two small humps observed in the measured light intensity. Table 3 summarizes the measured values of temperature and coagulation thickness as shown in Fig. 8(b). Table 3Summary of measured temperature and coagulation thickness (mean±1 SD).
To evaluate urological applications of temperature-monitored diffuser-assisted LITT, porcine urethral tissues were tested with 4 W 980 nm for 72-s irradiation and histologically examined as shown in Fig. 8. The irradiation time was selected to obtain of thermal coagulation in the urethral tissue. Figure 9(a) shows a gross image of the tested urethra along with cross-sectional (H&E stained) histology images at various positions (every 4-mm distance) along the diffuser axis. The control (c) shows no tissue deformation (i.e., discoloration), whereas the irradiated regions where the diffuser was positioned (1 to 5) are principally associated with radial tissue coagulation around the lumen. The spatial extent of the denaturation was measured to be , which was slightly thicker (16%) than the liver testing under the equivalent conditions [ in Fig. 8(a); ]. It should be noted that the treated tissue induced larger luminal area than the control did [Fig. 9(a)]. Furthermore, the middle regions of the irradiated tissue (2 and 3) created relatively 12.6% thicker coagulation than the other regions did. Figure 9(b) shows the magnified images of the control and the tested tissue. Unlike the control, the irreversible coagulation vividly involved epithelial cellular death and formation of vacuolation, resulting in dark, dense, and almost circumferential discoloration around the lumen. Layers of smooth muscle cells with no cellular deformation confirmed native urethral tissue (pinkish color) even after thermal treatment, which was observed in both the control and the treated tissues in Fig. 9(b). Figure 10(a) shows thermal responses of the urethral tissue during diffuser-assisted LITT. The FBG was located at the center of the diffuser and used to measure temperature variations. Similar to the bovine data in Fig. 8(b), the measured temperature increased with irradiation time. It was noted that the irradiation time of 72 s resulted in the tissue temperature of 55.4°C, corresponding to the temperature measured in the bovine tissue in Fig. 8(b). Figure 10(b) compares the longitudinal distribution of the normalized light intensity (black) from the diffuser and the normalized coagulation developed in a radial direction (blue). The quantitative evaluations confirmed that the radial coagulation patterns agreed well with the spatial light distribution along the axis of the urethra in the current study. 4.DiscussionThe aim of the current study was to evaluate the feasibility and effectiveness of diffuser-assisted LITT on urethra by assessing and correlating temperature and the induced thermal damage. According to Fig. 5, both the FBG and the thermocouple showed a comparable temperature response when they were positioned relatively far away from the diffuser surface (i.e., maximum measured for FBG and for thermocouple at 3.5 mm; ). However, as the distance became closer (0 and 1.5 mm), the thermocouple measured a temperature increase of up to higher than the FBG did (i.e., for FBG versus for thermocouple at 0 mm; Fig. 5). On the other hand, both the simulated and the measured temperatures achieved almost similar elevations at various radial positions [Fig. 6(a)]. All the treated tissue samples vividly experienced discoloration and a dehydrated surface due to irreversible thermal denaturation (Fig. 8), which typically occurs at the tissue temperature of less than 100°C.37 The identified temperature elevation (i.e., overestimation) could be associated with self-heating due to direct light absorption by the metal tip of the thermocouple.16,17 In fact, previous studies showed that the TCs were typically placed away from a laser-induced heat source to avoid any overestimation of the target temperature.24,40 On account of optical fibers with fairly low absorption at the irradiation wavelength (980 nm), the temperature measurements with the FBG would thereby be more position-independent as well as reliable. Furthermore, thermal coagulation threshold mainly takes place around 60°C in soft tissue.15,40 According to the current study, the coagulation of liver tissue commenced at the diffuser surface 24 s after laser irradiation (corresponding ; Fig. 8). Lower homogeneous thermal gradients from the diffuser could be responsible for the onset temperature lower than the coagulation threshold (60°C). It should be noted that the active length of the FBG (5 mm) was longer than the axial coagulation depth (i.e., at 24-s irradiation), which led to fewer precise measurements due to inconsistent thermal gradients along the FBG. However, once the coagulation depth was developed up to 6 mm 48 s after the irradiation to cover the entire length of the FBG, the measured temperature exceeded the coagulation temperature (i.e., ). Therefore, the FBG could provide more precise information on the interstitial tissue temperature even at the diffuser surface. According to Figs. 8(b) and 9, the coagulation thickness of urethral tissue was slightly thicker (16%) than that of liver tissue under the same experimental conditions (i.e., for urethra and for liver; ), possibly resulting from different optical properties of the tissues. Absorption of laser energy is primarily followed by thermal coagulation, and the degree of light absorption in tissue is dependent on the wavelength of light and the optical properties of the tissue.41,42 As both tissues had similar thermal diffusivities (i.e., versus ),43,44 the extent of the thermal coagulation could primarily be determined by the degree of light absorption, which could explain a further increase of the coagulation thickness in the urethra. However, due to unavailable data, the absorption coefficient of the urethra at 980 nm should be measured for comparative validation. Circumferential coagulation was vividly observed in both liver and urethral tissues mainly due to consistent radial emission as shown in Fig. 1(b). Unlike the liver tissue with rather homogeneous glandular tissue structure and circular coagulation [Fig. 8(a)], the urethral tissue exhibited a eccentricity of 1.1 mm in the denatured area as the inherent tissue structure was less circular with weak integrity (Fig. 9). In fact, the heterogeneous urethral tissue consisted of multilayers such as epithelium, submucosa, and subcutaneous layers.45 Furthermore, the out-of-center phenomenon could result from the blind placement of the diffusing applicator during the tests. As the lumen of the tissue was in an irregular shape, the diffusing device was difficult to be situated at the center of the urethra, leading to inconsistent tissue contact to ensure complete thermal coagulation. In spite of uniform radial emissions, longitudinal emissions [Fig. 1(b)] contributed variations in the coagulation thickness along the diffuser for both tissue samples [Figs. 8(c) and 9(b)]. According to the normalized coagulation thickness, the middle position of the diffusing applicator entailed more tissue deformation than the proximal and the distal ends, reflecting the spatial distribution of the light emission ( deviation). The two peaks in the normalized intensities could chiefly attribute to contribute to the interstitial temperature increase in the tissue, eventually maximizing the temperature between the peaks after conductive heat transfer and yielding the thickest tissue denaturation. It should be noted that the urethral lumen after LITT became up to 230% larger than the lumen of the control (Fig. 9). In fact, the eventual goal of LITT on urethral stricture is to expand the narrowed urethral pathway for easy and complete urination without development of any complications by circumscribing the thermal coagulation within the wall of the urethra. Thus, the radial dilation of the LITT-treated tissue could be a feasible treatment for tubular tissue with stenosis or stricture. Further in vivo studies will be performed in a mini pig model to validate acute and chronic responses of the urethral tissue after LITT application in terms of hyperemia, healing response, fibrosis, and stricture recurrence. The integration of an inflatable catheter with the proposed diffuser-based LITT is also under consideration to achieve reliable and concentric deployment of the device for interstitial photothermal treatment. In spite of the feasible application of FBG to estimate the interstitial tissue temperature during photothermal treatment, experimental limitations still remain in translation of the proposed sensing technique into clinical applications. Currently, the simplest method to measure the wavelength-encoded signals of the FBG is to utilize a broadband light source and a spectrum monitoring instrument such as an ASE source and OSA, respectively. However, although the FBG is relatively inexpensive, a spectrum analyzer or an interrogator is quite costly; for example, a commercially available OSA with the minimum resolution bandwidth of 0.01 nm costs around $ 40,000. As cost-effective approaches, different demodulation techniques have been suggested including ones based on a Fabry–Perot (FP) tunable filter or an edge filter. A fiber FP tunable filter can be used for FBG sensor interrogation by electronically adjusting the distance between two mirrors to make the resonance frequencies shift. A spectral shift of a Bragg resonance can be converted into a temporal shift of an electrical signal in a photodetector following the filter. The demodulation scheme based on the edge filter harnesses its linear power reflectance. That is, a wavelength shift of a Bragg resonance can be obtained by measuring a change in the output power of the edge filter.46 Although our experiments were performed by using the OSA, the above-mentioned demodulation techniques based on FP tunable filters or edge filters can also be exploited to the currently proposed sensor system for reducing the overall system cost. In addition, the minimum resolution bandwidth of the OSA used in the current study was 0.01 nm, resulting in a thermal sensitivity of for the current FBG sensor. In turn, the temperature resolution of the sensor was evaluated to be , which outperformed that of the thermocouple (). Based upon the spatial emission profile of the diffusing tip [Fig. 1(b)], a 5-mm active length of FBG with a reflectivity of 74.5% was utilized to cover the entire range of thermal gradients in the tissue. However, due to a large coupling constant in comparison with the active length, the used FBG had a large bandwidth of , which was difficult to evidently differentiate the acquired FBG peaks. Thus, the coupling constant and the active length of the FBG need to be optimized in order to achieve even a smaller bandwidth for high resolutions. On the other hand, a -phase shifted FBG, which provides a much narrower reflection bandwidth (), may be an alternative. Additionally, the current study used a single fiber to measure temperature variations at a single point in the tissue, which hardly reflected the overall temperature distribution along the diffuser. In order to evaluate thermal responses at various locations in the tissue lumen, an FBG array (i.e., ) can be incorporated into LITT.23 5.ConclusionThis study demonstrated the feasibility of a single FBG for real-time temperature monitoring during diffuser-assisted photothermal treatment on tubular tissue. Due to high thermal sensitivity and low absorption of irradiation light, the FBG sensor was able to precisely estimate the interstitial tissue temperature in comparison with a thermocouple. Further studies will be performed in vivo to optimize the FBG and its demodulation system to improve the transient response of the temperature during diffuser-assisted laser treatment. AcknowledgmentsThis research was supported by a grant from the Marine Biotechnology Program (20150220) funded by the Ministry of Oceans and Fisheries, Republic of Korea. ReferencesR. Mathur et al.,
“A retrospective analysis of urethral strictures and their management at a tertiary care center,”
J. Nephro-Urol. Mon., 3
(2), 109
–113
(2011). Google Scholar
L. A. Hampson, J. W. Mcaninch and B. N. Breyer,
“Male urethral strictures and their management,”
Nat. Rev. Urol., 11
(1), 43
–50
(2014). http://dx.doi.org/10.1038/nrurol.2013.275 Google Scholar
A. R. Mundy and D. E. Andrich,
“Urethral strictures,”
BJU Int., 107
(1), 6
–26
(2011). http://dx.doi.org/10.1111/j.1464-410X.2010.09800.x BJINFO 1464-410X Google Scholar
J. C. Buckley et al.,
“SIU/ICUD consultation on urethral strictures: dilation, internal urethrotomy, and stenting of male anterior urethral strictures,”
Urology, 83
(3), S18
–S22
(2014). http://dx.doi.org/10.1016/j.urology.2013.08.075 Google Scholar
T. Jin et al.,
“Safety and efficacy of laser and cold knife urethrotomy for urethral stricture,”
Chin. Med. J., 123
(12), 1589
–1595
(2010). http://dx.doi.org/10.3760/cma.j.issn.0366-6999.2010.12.020 Google Scholar
N. Fried and D. Ph,
“Erbium: YAG laser incision of urethral strictures for treatment of urinary incontinence after prostate cancer surgery,”
(2006). Google Scholar
G. G. Ferguson et al.,
“Minimally invasive methods for bulbar urethral strictures: a survey of members of the American Urological Association,”
Urology, 78
(3), 701
–706
(2011). http://dx.doi.org/10.1016/j.urology.2011.02.051 Google Scholar
A. G. Cavalcanti and G. Fiedler,
“Opinion: endoscopic urethrotomy,”
Int. Braz. J. Urol., 41
(4), 619
–622
(2015). http://dx.doi.org/10.1590/S1677-5538.IBJU.2015.04.03 Google Scholar
S. N. Goldberg et al.,
“Image-guided tumor ablation: standardization of terminology and reporting criteria,”
Radiol. Soc. North Am., 235 728
–739
(2012). http://dx.doi.org/10.1148/radiol.2353042205.Image-guided Google Scholar
B. Mensel, C. Weigel and N. Hosten,
“Laser-induced thermotherapy,”
Recent Results Cancer Res., 167 69
–75
(2006). http://dx.doi.org/10.1007/3-540-28137-1 Google Scholar
K. S. Lehmann et al.,
“In vivo validation of a therapy planning system for laser-induced thermotherapy (LITT) of liver malignancies,”
Int. J. Colorectal Dis., 26
(6), 799
–808
(2011). http://dx.doi.org/10.1007/s00384-011-1175-y IJCDE6 1432-1262 Google Scholar
J. W. Rolf Kiessling and O. Winqvist,
“Laser-induced interstitial thermal therapy for liver metastases,”
Health Technol. Assess., 2011-0 1
–4
(2011). Google Scholar
P. Saccomandi et al.,
“Theoretical analysis and experimental evaluation of laser-induced interstitial thermotherapy in ex vivo porcine pancreas,”
IEEE Trans. Biomed. Eng., 59
(10), 2958
–2964
(2012). http://dx.doi.org/10.1109/TBME.2012.2210895 IEBEAX 0018-9294 Google Scholar
M.-F. Marqa et al.,
“Focal laser ablation of prostate cancer: numerical simulation of temperature and damage distribution,”
Biomed. Eng. Online, 10
(1), 45
(2011). http://dx.doi.org/10.1186/1475-925X-10-45 Google Scholar
T. H. Nguyen et al.,
“Circumferential irradiation for interstitial coagulation of urethral stricture,”
Opt. Express, 23
(16), 20829
(2015). http://dx.doi.org/10.1364/OE.23.020829 OPEXFF 1094-4087 Google Scholar
F. Manns et al.,
“In situ temperature measurements with thermocouple probes during laser interstitial thermotherapy (LITT): quantification and correction of a measurement artifact,”
Lasers Surg. Med., 23
(2), 94
–103
(1998). http://dx.doi.org/10.1002/(SICI)1096-9101(1998)23:2<94::AID-LSM7>3.0.CO;2-Q LSMEDI 0196-8092 Google Scholar
S. A. van Nimwegen et al.,
“Nd:YAG surgical laser effects in canine prostate tissue: temperature and damage distribution,”
Phys. Med. Biol., 54
(1), 29
–44
(2009). http://dx.doi.org/10.1088/0031-9155/54/1/003 PHMBA7 0031-9155 Google Scholar
P. Roriz et al.,
“Review of fiber-optic pressure sensors for biomedical and biomechanical applications,”
J. Biomed. Opt., 18
(5), 050903
(2013). http://dx.doi.org/10.1117/1.JBO.18.5.050903 JBOPFO 1083-3668 Google Scholar
E. Schena et al.,
“Fiber optic sensors for temperature monitoring during thermal treatments: an overview,”
Sensors, 16
(8), 1144
(2016). http://dx.doi.org/10.3390/s16071144 SNSRES 0746-9462 Google Scholar
F. A. Jolesz, G. P. Zientara and D. Ph,
“MRI-guided laser-induced interstitial thermotherapy: basic principles,”
294
–324
(1995). Google Scholar
R. Medvid et al.,
“Current applications of MRI-guided laser interstitial thermal therapy in the treatment of brain neoplasms and epilepsy: a radiologic and neurosurgical overview,”
Am. J. Neuroradiol., 36
(11), 1998
–2006
(2015). http://dx.doi.org/10.3174/ajnr.A4362 Google Scholar
E. Schena and L. Majocchi,
“Assessment of temperature measurement error and its correction during Nd:YAG laser ablation in porcine pancreas,”
Int. J. Hyperthermia, 30
(5), 328
–334
(2014). http://dx.doi.org/10.3109/02656736.2014.928832 Google Scholar
D. Tosi, E. G. Macchi and A. Cigada,
“Fiber-optic temperature and pressure sensors applied to radiofrequency thermal ablation in liver phantom: methodology and experimental measurements,”
J. Sens., 2015 1
–22
(2015). http://dx.doi.org/10.1155/2015/909012 Google Scholar
W. Chen et al.,
“Performance assessment of FBG temperature sensors for laser ablation of tumors,”
in IEEE Int. Symp. on Medical Measurements and Applications (MeMeA ‘15),
(2015). http://dx.doi.org/10.1109/MeMeA.2015.7145221 Google Scholar
L. Dziuda et al.,
“Fiber Bragg grating-based sensor for monitoring respiration and heart activity during magnetic resonance imaging examinations,”
J. Biomed. Opt., 18
(5), 057006
(2013). http://dx.doi.org/10.1117/1.JBO.18.5.057006 Google Scholar
H. J. Kalinowski et al.,
“Application of fibre Bragg grating sensors in biomechanics,”
Trends Photonics, 661
(2), 315
–343
(2010). Google Scholar
S. Poeggel et al.,
“Optical fibre pressure sensors in medical applications,”
Sensors, 15
(7), 17115
(2015). http://dx.doi.org/10.3390/s150717115 Google Scholar
P. Saccomandi et al.,
“Temperature monitoring and lesion volume estimation during double-applicator laser-induced thermotherapy in ex vivo swine pancreas: a preliminary study,”
Lasers Med. Sci., 29
(2), 607
–614
(2014). http://dx.doi.org/10.1007/s10103-013-1360-z Google Scholar
K. J. Bergemann, R. Krasny and S. R. Forrest,
“Thermal properties of organic light-emitting diodes,”
Org. Electron., 13
(9), 1565
–1568
(2012). http://dx.doi.org/10.1016/j.orgel.2012.05.004 Google Scholar
G. J. Müller and A. Roggan, Laser-Induced Interstitial Thermotherapy, SPIE Press, Bellingham
(1995). Google Scholar
K. Hashimoto et al.,
“Permanent hair removal with a diode-pumped Nd:YAG laser: a pilot study using the direct insertion method,”
J. Am. Acad. Dermatol., 49
(6), 1071
–1080
(2003). http://dx.doi.org/10.1016/S0190-9622(03)02130-3 Google Scholar
J. Kwon et al.,
“Computational analysis of endometrial photocoagulation with diffusing optical device,”
Biomed. Opt. Express, 4
(11), 2450
–2462
(2013). http://dx.doi.org/10.1364/BOE.4.002450 Google Scholar
A. Ishimaru,
“Diffusion of light in turbid material,”
Appl. Opt., 28
(12), 2210
–2215
(1989). http://dx.doi.org/10.1364/AO.28.002210 Google Scholar
S. Thomsen, J. A. Pearce,
“Thermal damage and rate processes in biologic tissues,”
Optical-Thermal Response of Laser-Irradiated Tissue, 487
–549 Springer, Dordrecht, Netherlands
(2011). Google Scholar
B. Cox,
“Introduction to laser-tissue interactions,”
Opt. Med., 4886 1
–61
(2007). Google Scholar
J. P. Ritz et al.,
“Optical properties of native and coagulated porcine liver tissue between 400 and 2400 nm,”
Lasers Surg. Med., 29
(3), 205
–212
(2001). http://dx.doi.org/10.1002/lsm.v29:3 Google Scholar
P. Mégret and C. Caucheteur,
“Fiber Bragg gratings: fundamentals and applications,”
1
–31
(2007). Google Scholar
Y.-J. Rao,
“Fiber Bragg grating sensors: principles and applications,”
Optical Fiber Sensor Technology: Devices and Technology, 355
–379 Springer, Boston
(1998). Google Scholar
A. D. Kersey et al.,
“Fiber grating sensors,”
J. Lightwave Technol., 15
(8), 1442
–1463
(1997). http://dx.doi.org/10.1109/50.618377 Google Scholar
C. Raulin and S. Karsai, Laser and IPL Technology in Dermatology and Aesthetic Medicine, Springer Verlag, Berlin
(2011). Google Scholar
S. L. Jacques,
“Optical properties of biological tissues: a review,”
Phys. Med. Biol., 58
(11), R37
–R61
(2013). http://dx.doi.org/10.1088/0031-9155/58/11/R37 Google Scholar
R. R. Van Den Bos et al.,
“Endovenous simulated laser experiments at 940 nm and 1470 nm suggest wavelength-independent temperature profiles,”
Eur. J. Vasc. Endovasc. Surg., 44
(1), 77
–81
(2012). http://dx.doi.org/10.1016/j.ejvs.2012.04.017 Google Scholar
J. A. Balderas-López, A. Mandelis and J. A. Garcia,
“Thermal-wave resonator cavity design and measurements of the thermal diffusivity of liquids,”
Rev. Sci. Instrum., 71
(7), 2933
–2937
(2000). http://dx.doi.org/10.1063/1.1150713 Google Scholar
J. W. Valvano, J. R. Cochran and K. R. Diller,
“Thermal conductivity and diffusivity of biomaterials measured with self-heated thermistors,”
Int. J. Thermophys., 6
(3), 301
–311
(1985). http://dx.doi.org/10.1007/BF00522151 Google Scholar
H. Masumoto et al.,
“Reappraisal of intergender differences in the urethral striated sphincter explains why a completely circular arrangement is difficult in females: a histological study using human fetuses,”
Anat. Cell Biol., 45
(2), 79
–85
(2012). http://dx.doi.org/10.5115/acb.2012.45.2.79 Google Scholar
Y. Zhao and Y. Liao,
“Discrimination methods and demodulation techniques for fiber Bragg grating sensors,”
Opt. Lasers Eng., 41
(1), 1
–18
(2004). http://dx.doi.org/10.1016/S0143-8166(02)00117-3 Google Scholar
|