|
1.IntroductionEvaluating the optical properties of brain tissue is important for the application of light in clinical diagnosis, surgery, and therapeutic procedures for brain diseases. The optical properties of biological tissue have been used to evaluate spatial and/or temporal changes of neuronal activity and tissue viability in the brain as intrinsic optical signals (IOSs). IOSs in the brain are believed to be caused primarily by the following processes: hemodynamic-related changes in absorption and scattering properties, changes in absorption due to redox states of cytochromes in mitochondria, changes in scattering generated by cell swelling or shrinkage caused by water movement between intracellular and extracellular compartments,1 and changes in scattering and absorption caused by chromophore contents and cell deformations. Light in the visible to near-infrared spectral range is sensitive to the absorption and scattering properties of biological tissue. The absorption and scattering properties of in vitro tissue slices can be estimated from the measured diffuse reflectance and transmittance of tissue slices2 based on several light transport models, such as the Kubelka-Munk theory,3 the diffusion approximation to the transport equation,4 the Monte Carlo method,5 and the adding-doubling method.6 Numerous spectroscopic methods have been investigated for in vivo determination of the scattering and absorption properties in living tissues, including time-resolved measurements,7 a frequency-domain method,8 and spatially resolved measurements with continuous wave (CW) light.9–15 Diffuse reflectance spectroscopy (DRS) based on the measurement of CW light can be simply achieved with an incandescent white light source, inexpensive optical components, and a spectrometer. DRS is one of the most promising methods for evaluating the absorption and scattering properties of in vivo brain tissue. Several approaches using a Monte Carlo simulation-based lookup table have been investigated for determining the absorption and scattering properties of biological tissue.16–19 Several imaging methods based on DRS have been used to investigate cortical hemodynamics based on changes in the absorption properties of brain tissue.20–24 In order to achieve rapid multispectral imaging, the use of an acousto-optical tunable filter25 and the combination of a lenslet array with narrow-band filters26 have been proposed. On the other hand, the reconstruction of multispectral images from a red green blue (RGB) image acquired by a digital RGB camera is promising as a method of rapid and cost-effective multispectral imaging. Several reconstruction techniques for multispectral images, such as the pseudo-inverse method,27–30 finite-dimensional modeling,29,31 the nonlinear estimation method,32 and the Wiener estimation method (WEM),33–35 have been investigated. Among these reconstruction techniques, the WEM is one of the most promising methods for practical applications because of its simplicity, cost-effectiveness, accuracy, time efficiency, and the possibility of high-resolution image acquisition. In the present study, we investigate a simple spectral imaging of reduced scattering coefficients and the absorption coefficients of in vivo exposed brain tissues in the range of wavelengths from visible to near-infrared based on DRS using a multispectral imaging system. Multispectral diffuse reflectance images of in vivo exposed brain are estimated from an RGB image captured by a digital RGB camera. The Monte Carlo simulation-based multiple regression analysis for the estimated absorbance spectra at nine wavelengths (500, 520, 540, 560, 570, 580, 600, 730, and 760 nm) is then used to specify the concentrations of oxygenated and deoxygenated hemoglobin as the absorption parameters, and the coefficient and the exponent of the reduced scattering coefficient spectrum approximated by a power law function as the scattering parameters. The absorption coefficient spectrum and the reduced scattering coefficient spectrum are finally reconstructed from the hemoglobin concentrations and the scattering parameters, respectively. In order to confirm the feasibility of a method by which to evaluate the absorption and scattering properties of the cerebral cortex, we performed in vivo experiments using exposed rat brain while changing the fraction of inspired oxygen (). 2.Principle2.1.Estimation of Spectral Diffuse Reflectance Images by Wiener EstimationThe response of a digital color camera in spatial coordinates with the ’th (, 2, 3) color channel, or red, green, and blue can be calculated as where is the wavelength, is the transmittance spectrum of the ’th filter, is the spectrum of the illuminant, is the sensitivity of the camera, and (, ; ) is the reflectance spectrum in the spatial coordinates . For convenience, Eq. (1) is expressed in discrete vector notation as where is a vector having a three-element column and is a vector having a -element column, which corresponds to the reflectance spectrum of a pixel of an image. Moreover, is a matrix and is expressed as where . Column vector denotes the transmittance spectrum of the ’th filter, and represents the transposition of a vector. Also, and are diagonal matrices representing the spectrum of the illuminant and the sensitivity of the camera, respectively. The Wiener estimation33–35 of is given by where is the Wiener estimation matrix. The purpose of is to minimize the minimum square error between the original and estimated reflectance spectra. In this case, the minimum square error is expressed asFrom Eqs. (4) and (5), the minimum square error is rewritten as The minimization of the minimum square error requires that the partial derivative of with respect to be zero, i.e., From Eq. (7), the matrix is derived as where is an ensemble-averaging operator. The derivation of matrix requires the autocorrelation matrix . In the present study, we determine based on 720 different reflectance spectra obtained from in vivo rat brain under various physiological conditions.To test the accuracy in spectral reconstruction, the estimated spectrum at 27 different wavelengths by WEM was compared with the measured spectrum by spectrometer using a goodness-of-fit coefficient (GFC).36 The GFC is based on the inequality of Schwartz and it is described as where is the measured original spectral data at the wavelength and is the estimated spectral data at the wavelength . Hernández-Andrés et al.36 suggested that colorimetrically accurate requires a ; a “good” spectral fit requires a , and is necessary for an “excellent” spectral fit.2.2.Estimation of the Absorption Coefficient and the Reduced Scattering CoefficientFigure 1 shows a flow diagram of the method used to estimate the absorption coefficient and the reduced scattering coefficient . The absorbance spectrum is defined as where is the diffuse reflectance spectrum normalized by the incident light spectrum. Since attenuation due to light scattering can be treated as a pseudochromophore, the absorbance spectrum can be approximated as the sum of attenuations due to absorption and scattering in the brain as where is the concentration, is the mean path length, is the extinction coefficient, and indicates attenuation due to light scattering in the tissue. The subscripts HbO and HbR denote oxygenated hemoglobin and deoxygenated hemoglobin, respectively. The absorption coefficient of the cortical tissue was assumed to depend only on the concentrations of HbO and HbR (, ) asFig. 1Flow diagram of the process for estimating the concentration of oxygenated hemoglobin , the concentration of deoxygenated hemoglobin , coefficient , exponent , absorption coefficient , and reduced scattering coefficient . (a) Preparation work for determining the regression equations and (b) main process for deriving and from the reflectance spectrum estimated by the Wiener estimation method (WEM). ![]() The total hemoglobin concentration is defined as the sum of and as follows: The tissue oxygen saturation is determined as The reduced scattering coefficient of the brain tissue was assumed to take the following form as a power law function:37,38 where the coefficient and the exponent represent the scattering amplitude and the scattering power, respectively. Using as the response variable and as the predictor variables, the multiple regression analysis based on the modified Lambert-Beer law (MRA1) can be applied to Eq. (11) as where , , and are the regression coefficients. The regression coefficients and describe the degree of contribution of each extinction coefficient to and are closely related to the concentrations and , respectively. The regression coefficient is expressed as where , , and are the averages of , , and , respectively, over the wavelength range, and represents the bias component of . Thus, describes the degree of contribution of the attenuation due to light scattering in the brain to the absorbance spectrum and so is related to the coefficient and the exponent in Eq. (15). At the same time, is also affected by the absorption coefficient of the brain, since is generally a function of the tissue absorption coefficient and the reduced scattering coefficient.In order to investigate the relationship between the regression coefficients and the values of , , , and , we performed MCS for the diffuse reflectance from the rat cortical tissue through the skull at , 520, 540, 560, 570, 580, 600, 730, and 760 nm using various values of , , , and . We used the MCS source code developed by Wang et al.,5 in which the Henyey-Greenstein phase function is applied to the sampling of the scattering angle of photons. The simulation model consisted of a single layer representing cortical tissue. In a single simulation of diffuse reflectance at each wavelength, 5,000,000 photons were randomly launched. The absorption coefficients of oxygenated hemoglobin and deoxygenated hemoglobin were obtained from the values of and in the literature,39 where the hemoglobin concentration of blood with a 44% hematocrit is of hemoglobin. For the reduced scattering coefficients, the values of were 60,258, 80,344, 100,430, 120,516, and 180,774 in the simulation, which were derived by multiplying the typical value40 of by 0.5, 0.75, 1.0, 1.25, and 1.5, respectively. The values of were 1.2442, 1.31332, 1.3824, 1.45156, and 1.52068, which were derived by multiplying the typical value40 of by 0.5, 0.75, 1.0, 1.25, and 1.5, respectively. The reduced scattering coefficients of the cortical tissue were obtained from Eq. (15). The sum of the absorption coefficients of oxyhemoglobin and deoxyhemoglobin, and , represents the absorption coefficient of total hemoglobin . The values for , 23.26, and were used as input to the cortical tissue in the MCS. Tissue oxygen saturation was assumed to be for all combinations. For all simulations, the refractive index of the cortical tissue was fixed at 1.4. The thickness of the cortical tissue in each case was set to 5.0 mm. Figures 2(a) and 2(b) show the values of and versus the volume concentrations of oxyhemoglobin and deoxyhemoglobin , respectively, for various values of , as obtained from the MCS. Figures 2(c) and 2(d) show the values of and versus the volume concentrations of oxyhemoglobin and deoxyhemoglobin , respectively, for various values of , as obtained from the MCS. In both Figs. 2(a) and 2(c), the value of increases with the increase of . Moreover, the value of changes with the increase in the values of and . The same tendency can be seen for , as shown in Figs. 2(b) and 2(d). Figures 2(e) and 2(f) show the values of versus the values of and , respectively, for various values of and . The value of decreases with the increase in the value of . Moreover, the value of increases with the increases in and . On the other hand, the value of increases with the increase in the value of . Therefore, the regression coefficients , , and are related to the volume concentration of oxyhemoglobin , that of deoxyhemoglobin , the coefficient , and the exponent , respectively. However, , , , and are not determined by a unique regression coefficient when using only MRA1. Fig. 2Regression coefficients versus concentrations of oxygenated hemoglobin and deoxygenated hemoglobin , coefficient , and exponent , obtained from the Monte Carlo simulations. (a) Regression coefficient versus concentration for various values of . (b) Regression coefficient versus concentration for various values of . (c) Regression coefficient versus concentration for various values of . (d) Regression coefficient versus concentration for various values of . (e) Regression coefficient versus coefficient for various values of and . (f) Regression coefficient versus exponent for various values of and . ![]() Thus, we conducted further multiple regression analyses to estimate the values of , , , and based on the combination of regression coefficients , , and that were obtained from MRA1. In this analysis, , , and were regarded as response variables, and the three regression coefficients , , and in Eq. (16) were regarded as predictor variables to determine the regression equations for , , and . The regression equations for , , and are written as The symbol represents the transposition of a vector. We refer to this analysis as MRA2. On the other hand, in the preliminary investigation, we found that adding to the predictor variables can improve the accuracy of the estimation of . Therefore, , , , and are regarded as the predictor variables, and the given values of are regarded as the response variables for determining the regression equation for , which is written as whereWe refer to this analysis as MRA3. The coefficients , , , (, 1, 2, 3), and (, 1, 2, 3, 4) are unknown and must be determined before estimating , , , and . We used MCS as the foundation to determine reliable values of , , , and . The simulation model used here consisted of a single layer of cortical tissue, in which and are homogeneously distributed. The absorption coefficients converted from the concentrations and and the reduced scattering coefficient deduced by the coefficient and the exponent were provided as inputs to the simulation, whereas the diffuse reflectance spectrum was derived as output. The input values of , , , and and the output reflectance spectra are useful as the data set in statistically determining the values of , , , and for determining the absolute values of , , , and . The five different values of 60,258, 80,344, 100,430, 120,516, and 180,774 were calculated by multiplying the typical value40 of by 0.5, 0.75, 1.0, 1.25, and 1.5, respectively, whereas the five values of 1.2442, 1.31332, 1.3824, 1.45156, and 1.52068 were calculated by multiplying the typical value40 of by 0.5, 0.75, 1.0, 1.25, and 1.5, respectively. The reduced scattering coefficients of the cortical tissue with the 25 different values were derived using Eq. (15). The sum of the absorption coefficients of oxyhemoglobin and deoxyhemoglobin for , 23.26, and was used as input to the cortical tissue in the simulation. Tissue oxygen saturation StO2 was determined by , and values of 0%, 20%, 40%, 60%, 80%, and 100% were used for the simulation. The above values were also used for the refractive index of the cortical layer. In total, 450 diffuse-reflectance spectra at , 520, 540, 560, 570, 580, 600, 730, and 760 nm were simulated under the various combinations of , , , and . The MRA1 analysis for each simulated spectrum based on Eq. (16) generated the 450 sets of vector and concentrations and and exponent , and the 450 sets of vector and coefficient . The coefficient vectors , , and were statistically determined by performing MRA2, whereas the coefficient vector was determined statistically by performing MRA3. Once , , , and were obtained, , , , and were calculated from , , and , which were derived from MRA1 for the measured reflectance spectrum, without the MCS, as shown in Fig. 1(b). Therefore, the spectrum of the reduced scattering coefficient and that of the absorption coefficient were reconstructed by Eqs. (12) and (15), respectively, from the measured reflectance spectrum. 3.Experiments3.1.Imaging SystemFigure 3(a) shows a schematic diagram of the experimental system used in the present study. A white-light emitting diode (LED) (LA-HDF158A, Hayashi Watch Works Co., Ltd., Tokyo, Japan) illuminated the surface of the exposed cortex via a light guide and a ring-shaped illuminator with a polarizer. The light source covered a range from 400 to 780 nm. Diffusely reflected light was received by a 24-bit RGB CCD camera (DFK-31BF03.H, Imaging Source LLC, Charlotte, NC, USA) without an IR cut filter via an analyzer and a camera lens to acquire an RGB image of . The primary polarization plate (ring-shaped polarizer) and the secondary polarization plate (analyzer) were placed in a crossed Nicols alignment in order to reduce specular reflection from the sample surface. A standard white diffuser was used to regulate the white balance of the camera. In order to evaluate the accuracy of the WEM, the reflectance spectra of cortex were simultaneously measured by a fiber-coupled spectrometer (USB4000-XRS-ES, Ocean Optics Inc., Dunedin, Florida, USA) for an integration time of 130 ms as reference data. Before the measurements of RGB images and reflectance spectra, the area measured using the spectrometer was confirmed by projecting light from a halogen lamp (HL-2000, Ocean Optics Inc.) onto the surface of the cortex via one lead of a bifurcated fiber, a lens, and a beam splitter. The RGB image of the sample, including the spot of light illuminated by the halogen lamp, was stored in the PC, and the size and coordinates of the spot were then specified as the area measured by the spectrometer. After the halogen lamp was turned off, the measurements of RGB images and reflectance spectra were performed simultaneously. Using the WEM, reflectance images ranging from 500 to 760 nm at intervals of 10 nm were reconstructed from an RGB image acquired at an exposure time of 65 ms. This means that the spectral reflectance images at 27 wavelengths can be obtained with a temporal resolution of 15 fps. The field of view of the system was with . The lateral resolution of the images was estimated to be . The multispectral reflectance images with at nine wavelengths (500, 520, 540, 560, 570, 580, 600, 730, and 760 nm) were then used to estimate the images of , , , , , and according to the process described in Fig. 1. 3.2.Phantom ExperimentsTo confirm the validity of the proposed method, we performed experiments using tissue-like optical phantoms. We prepared agar solution by diluting agarose powder (Fast Gene AG01; NIPPON Genetics EUROPE GmbH, Düren, NRW, Germany) with saline at a weight ratio of 1.0%. To simulate the scattering condition, a mixture of polystyrene latex beads solution with a mean particle size (LB1-15L; Sigma-Aldrich Japan K. K. Tokyo, Japan) and that with a mean particle size (LB11-15L; Sigma-Aldrich Japan K. K.) was added to the agarose solution. The resultant solution was used as the base material. The volume concentration of the polystyrene solutions ranged from 2.75% to 11.0%. An optical phantom layer was made by adding a small amount of fully oxygenated hemoglobin extracted from horse blood to the base material. All phantoms were hardened in molds having the required thickness and size by being cooled at approximately 5.5°C for 30 min. The thickness of each phantom was 0.5 cm, while the area of each phantom was . In the preliminary experiments, we attempted to measure the phantoms with hemoglobin deoxygenated by solution. However, the measured diffuse reflectance spectra had a tendency to exhibit unexpected fluctuation and it was difficult to maintain the stable condition of the deoxygenated spectra during the experiments. To avoid the potential uncertainty of the estimation of the total hemoglobin with the condition of , the phantoms with hemoglobin deoxygenated by solution were excluded from the measurements in this study. We made 22 optical phantoms with different combinations of , , , and . As preparatory measurements, we first determined the absorption coefficient and reduced scattering coefficients of each phantom at 500, 520, 540, 560, 570, 580, 600, 730, and 760 nm, because they are necessary for the MCS to deduce the empirical formula for , , , and . The spectra of the absorption coefficient and reduced scattering coefficients are also required to calculate the given values for , , , and . For this purpose, we measured the diffuse reflectance and total transmittance spectra of each phantom individually. A 150-W halogen-lamp light source (LA-150SAE; Hayashi Watch Works Co., Ltd.) illuminated the phantom via a light guide (LGC1-5L1000; Hayashi Watch Works Co., Ltd.) and lens with a spot diameter of 0.2 cm. The diameter and focal length of the lens are 5.0 and 10 cm, respectively. The thickness of each phantom in the preparatory measurements was 0.1 cm, while the area of each phantom was . The phantom was placed between two glass slides having a thickness of 0.1 cm and fixed at the sample holder of an integrating sphere (RT-060-SF, Labsphere Incorporated). The detected area of the phantom was circular with a diameter of 2.2 cm. Light diffusely reflected from the detected area was received at the input face of an optical fiber probe having a diameter of which was placed at the detector port of the sphere. The fiber transmits the received light into a multichannel spectrometer (USB2000, Ocean Optics Inc.), which measured reflectance or transmittance spectra in the visible to near-infrared wavelength region under the control of a personal computer. To determine and from the measured diffuse reflectance and total transmittance spectra, we utilized the inverse Monte Carlo method (IMC).41 In the IMC, the MCS of the reflectance and transmittance spectra were iterated for different values of and until the difference between the simulated and measured spectral values decreased below a predetermined threshold. The values used in the last step of the iteration were adopted as the final results. This process was carried out at 500, 520, 540, 560, 570, 580, 600, 730, and 760 nm, and wavelength-dependent properties of and were obtained for each phantom. In these calculations, the refractive index was assumed to be 1.33 for all phantoms in the whole wavelength range. The concentrations of oxygenated hemoglobin and deoxygenated hemoglobin in each phantom were calculated from Eq. (12) with the estimated absorption coefficient and the known values of and . The concentration of total hemoglobin and the tissue oxygen saturation in each phantom was calculated from Eqs. (13) and (14), respectively. The coefficient and the exponent for each phantom were calculated from the estimated based on Eq. (15). Those values of , , , , , , , and obtained by IMC are used as the given values to evaluate the validity of the proposed method experimentally. We also need to have the conversion vectors for the phantoms used in this study. We generated diffuse reflectance spectra at 500, 520, 540, 560, 570, 580, 600, 730, and 760 nm using the MCS with the conditions of the phantoms. For this simulation, the values of and at 500, 520, 540, 560, 570, 580, 600, 730, and 760 nm were set to be the same as those estimated by IMC. Performing MRA1, MRA2, and MRA3 as described in Sec. 2, we derived the conversion vectors and the corresponding empirical formula for estimating the volume concentrations of oxygenated hemoglobin , that of deoxygenated hemoglobin , the coefficient , and the exponent . 3.3.Animal ExperimentsAnimal care and experimental procedures were approved by the Animal Research Committee of Tokyo University of Agriculture and Technology. Intraperitoneal anesthesia was implemented with -chloralose () and urethane () in five adult male Wister rats (134 to 426 g). Anesthesia was maintained at a depth such that the rat had no response to toe pinch. The rat head was placed in a stereotaxic frame. A longitudinal incision of in length was made along the head midline. The skull bone overlying the parietal cortex was removed with a high-speed drill to form an ellipsoidal cranial window, as shown in Fig. 3(b). In order to confirm the change in reflectance spectra of exposed rat brain, we performed the measurements during normoxia ( to 5 min), hyperoxia ( to 10 min), and anoxia ( to 30 min) by varying . Hyperoxia () was induced by gas inhalation, for which a breath mask was used under spontaneous respiration, whereas anoxia () was induced by gas inhalation. In order to identify the respiration arrest (RA), the respiration of the rat was confirmed by observing the periodical movement of the lateral region of the abdomen during the experiments. In order to evaluate the magnitude of signal induced by hyperoxia and anoxia, we calculated the change in the signal based on the time series data. The signal at the onset of measurement was selected as a control , which was subtracted from each of the subsequent signals in the series. Each subtracted value, which demonstrated the change in the signal, , over time, was normalized by dividing by . The change in the signal is expressed as . The above calculation was applied to the time series of , , , , , , and . The time of RA can differ from sample to sample. Therefore, we divided the time course of the estimated value during anoxia into two periods: anoxia 1 and anoxia 2. Anoxia 1 is the period between the onset of anoxia and RA, whereas anoxia 2 is the period between RA and the end of the measurement. The time averages of the estimated values over the periods of normoxia, hyperoxia, anoxia 1, and anoxia 2 were then calculated and averaged over all five samples. 3.4.Statistical ConsiderationsA region of interest (ROI) of was placed in an image for each resultant image. An unpaired Student’s -test was used for statistical analysis when comparing the in vivo results of the normoxic condition with those for the hyperoxic condition or with those for the anoxic conditions. The normality of the averaged value over the seven samples for each condition was tested using a Shapiro-Wilk test before the Student’s -test. A value of was considered to be statistically significant. 4.Results and DiscussionFigure 4 shows the comparisons between the estimated and given values for (a) oxygenated hemoglobin , (b) deoxygenated hemoglobin , (c) total hemoglobin , (d) tissue oxygen saturation , (e) coefficient , and (f) exponent , obtained from the phantom experiments. In Figs. 4(a), 4(c), 4(e), and 4(f), the estimated values are well correlated with the given values. Correlation coefficients between the estimated and given values are 0.91 (), 0.95 (), 0.84 () and 0.46 () for , , , and , respectively. In Figs. 4(b) and 4(d), all estimated values of and are close to and 100%, respectively, which is consistent with the fact that the hemoglobin solution used in the phantoms was fully oxygenated hemoglobin, as described above. These results indicate the validity of the proposed method for estimating the volume concentrations and , the coefficient , and the exponent . Figure 5 shows the comparisons between the estimated and given values for (a) absorption coefficient and (b) reduced scattering coefficient . Reasonable results were obtained for both and . Correlation coefficients between the estimated and given values are 0.97 () and 0.78 () for and , respectively. These results demonstrate the validity of the proposed method for estimating absorption and reduced scattering coefficients from the diffuse reflectance spectrum. Fig. 4Comparisons between the estimated and given values for (a) oxygenated hemoglobin , (b) deoxygenated hemoglobin , (c) total hemoglobin , (d) tissue oxygen saturation , (e) coefficient , and (f) exponent , obtained from the phantom experiments. ![]() Fig. 5Comparisons between the estimated and given values for (a) absorption coefficient and (b) reduced scattering coefficient , obtained from the phantom experiments. ![]() Figure 6 shows the typical spectral reflectance images at 500, 520, 540, 560, 570, 580, 600, 730, and 760 nm estimated from the RGB image of in vivo rat brain under normoxia by the WEM. Spectral reflectance images of the cerebral cortex were successfully reconstructed from the RGB image using the proposed method. The distribution of blood vessels in the cortical tissue can be clearly recognized in the estimated reflectance images at 500, 520, 540, 560, 570, and 580 nm. The images at 600, 730, and 760 nm have low contrast between the parenchyma region and the blood vessel region, which indicates the lower absorption coefficient of hemoglobin at lower wavelengths. Figure 7 shows the reflectance spectra estimated using the WEM and the reflectance spectra measured using the spectrometer for (a) normoxia, (b) hyperoxia, and (c) anoxia immediately after RA. The estimated reflectance spectra for a blood vessel and that for parenchyma in each graph of Fig. 7 is the average values over the and , respectively. In this case, both and were selected to be the same as the measured area by the spectrometer. The reflectance spectra estimated using the WEM are comparable to the spectra measured using the spectrometer for the different respiration conditions. The estimated spectra under hyperoxia and anoxia are dominated by the spectral characteristics of oxygenated hemoglobin and deoxygenated hemoglobin, respectively. The values of GFC obtained from five samples summarized in Table 1 in the revised manuscript indicate the accurate spectral reconstruction by WEM. Fig. 7Comparison of the estimated reflectance spectra averaged over the ROIs in Fig. 4 (white squares) and the reference spectra measured by spectrometer for (a) normoxia, (b) hyperoxia, and (c) anoxia. ![]() Table 1The goodness-of-fit coefficient (GFC) obtained from the five samples.
Figure 8 shows the typical estimated images of exposed rat brain under normoxia obtained using the proposed method for (a) oxygenated hemoglobin , (b) deoxygenated hemoglobin , (c) total hemoglobin , (d) tissue oxygen saturation , (e) coefficient , and (f) exponent . In Figs. 8(a)–8(c), the values of , , and , respectively, in the blood vessel region are higher than those in the parenchyma region, which indicates the difference in blood volume between the blood vessels and the parenchyma. The distribution of in the blood vessel region in Fig. 8(d) is probably due to the difference between artery and vein. The average values over the entire region were calculated to be , , , , , and , whereas those over the ROI on parenchyma (white square in each image of Fig. 8) were calculated to be , , , , , and . Fig. 8Estimated images of exposed rat brain obtained using the proposed method under normoxia for (a) oxygenated hemoglobin , (b) deoxygenated hemoglobin , (c) total hemoglobin , (d) tissue oxygen saturation , (e) coefficient , and (f) exponent . ![]() Figure 9 shows the reconstructed images of the exposed rat brain for (a) the absorption coefficient spectrum and (b) the reduced scattering coefficient spectrum under normoxia. The average values of and over the ROIs [white squares in Figs. 9(a) and 9(b)] are also shown in Fig. 10. The error bars mean the standard deviations for all pixels in the ROI. The wavelength dependence of is dominated by the spectral characteristics of hemoglobin.39 The reduced scattering coefficients have a broad scattering spectrum, exhibiting a larger magnitude at shorter wavelengths. The spectral features of correspond to the typical spectrum of brain tissue published in the literature.40 Fig. 9Typical estimated images of exposed rat brain obtained using the proposed method under normoxia for (a) absorption coefficient and (b) reduced scattering coefficient . ![]() Fig. 10Estimated spectra averaged over the ROI (white square in Fig. 9) for (a) absorption coefficient and (b) reduced scattering coefficient . The error bars mean the standard deviations for all pixels in the ROI. ![]() Figure 11 shows the comparison between the measured reflectance spectrum and the predicted reflectance spectrum for the and shown in Fig. 10. The predicted reflectance spectrum is comparable to the measured spectrum. The value of GFC in this case was 0.9974, which shows a good agreement between the measured reflectance spectrum and the predicted reflectance spectrum. Fig. 11Comparison between the measured reflectance spectrum and the predicted reflectance spectrum for the and , as shown in Fig. 10. ![]() Figure 12 shows the typical in vivo results for , , , , , and while varying . Figure 13 shows the time courses of change in , , , and averaged over the area for the ROI in the parenchyma, as shown in Fig. 12, while varying . The values of and were increased and decreased, respectively, during hyperoxia, which caused the increase in . After the onset of anoxia, the values of and decreased and increased, respectively. Consequently, the value of was dramatically decreased. The value of begins to increase before RA and reaches a maximum amplitude approximately 1 min after RA, which is indicative of an increase in blood flow for compensating hypoxia. The period between the onset of anoxia and respiratory arrest averaged over all five samples was . Immediately after RA, the values of both and decreased rapidly. The time courses of , , , and while varying are consistent with the well-known hemodynamic responses to the change in the fraction of inspired oxygen. Minor fluctuations in both and approximately 2.5 min after RA are probably caused by the changes in blood volume due to vasoconstriction and vasodilatation. Fig. 12Typical images for in vivo results while varying (from top to bottom: RGB image, , , , , , and ). ![]() Fig. 13Time courses of , , , and averaged over the ROI in the parenchyma region shown in Fig. 9 while varying . ![]() Figure 14 shows the time averages over the period of normixia, hyperoxia, anoxia 1 (before RA), and anoxia 2 (after RA) after respiratory arrest for (a) , (b) , (c) , and (d) averaged over the ROIs for all five samples. The trends of , , , and shown in Figs. 12 and 13 were apparent in all five samples. The average values of during normoxia over all five samples were , which is lower than the value of around 60% reported in the literature.38,42 This may be due to the experimental conditions of the present study, such as the type of anesthetic used, the anesthetic depth, or uncontrolled body temperature. Fig. 14Time averages over the period of normoxia, hyperoxia, anoxia 1 (before RA), and anoxia 2 (after RA) after respiratory arrest for (a) , (b) , (c) , and (d) averaged over the ROIs for all five samples. The error bars show the standard deviations (). *. ![]() Figure 15 shows the time courses of coefficient , exponent , , and averaged over the area corresponding to the ROIs in the parenchyma, as shown in Fig. 12, while varying . During the period from the onset of anoxia until RA, coefficient and exponent decrease slightly and increase slightly, respectively. The changes in the scattering parameters decrease the reduced scattering coefficient . On the other hand, the values of and increased remarkably after RA. Fig. 15Time courses of , , , and averaged over the ROI in the parenchyma region shown in Fig. 9 while varying . ![]() Figure 16 shows the time average over the period of normixia, hyperoxia, anoxia 1, and anoxia 2 for (a) coefficient , (b) exponent (c) , and (d) averaged over the ROIs for all five samples. The same tendencies of , , , and shown in Figs. 12 and 15 were applicable to the average values over all five samples. Changes in the reduced scattering coefficients after the onset of -inhalation imply morphological changes in brain tissue. Mitochondrial respiration is inhibited during the ischemia-like condition by the rapid drop in tension, which causes depletion of adenosine triphosphate (ATP). Reduction of ATP production due to the inhibition of mitochondrial respiration leads to failure of the pump.43 In such a case, extracellular , , and rush in, with water following osmotically, causing cell swelling.44,45 Thus, the slight decreases in in Figs. 12, 15, and 16 are most likely caused by cell swelling due to failure of the pump. The remarkable increases in after RA are probably caused by the dendritic beading effect, which is indicative of neuronal damage.44 The necklace-like structure of a large amount of dendritic processes is highly efficient at scattering light, such that bead formation over several minutes reduces the transmitted light intensity.46–48 Fig. 16Time averages over the period of normoxia, hyperoxia, anoxia 1 (before RA), and anoxia 2 (after RA) after respiratory arrest for (a) , (b) , (c) , and (d) averaged over the ROIs for all five samples. The error bars show the standard deviations (). *. ![]() Since the method relying on diffusing reflection integrates all information along the depth direction, the method does not have depth resolution. The depth and diameter of blood vessels usually differ among samples and may change due to the age of the rat. Their correct estimation is essential for precisely estimating the concentrations of oxygenated hemoglobin and deoxygenated hemoglobin for the blood vessel regions. In the present study, we assumed the scattering spectrum of the in vivo brain tissue as a power law function. This assumption may be applicable to the parenchyma region with a low volume concentration of blood. However, the diffuse reflectance spectrum from the blood vessel region with a higher volume concentration of blood can be strongly influenced by the scattering properties of the red blood cells (RBCs). It has been reported that RBCs behave as strong scatterers of light, and the reduced scattering coefficient spectrum has a similar wavelength dependence to the absorption spectrum of hemoglobin.49 Using the empirical formulas obtained from the MCS with an inhomogeneous layer model in which the reduced scattering coefficient spectrum is considered may enable more accurate estimations of the reduced scattering coefficients for the blood vessel regions. It has been reported that the use of the Lambert-Beer approximation in the analysis of data for ischemia condition in which the values of and are dramatically changed provides serious errors.20 The most serious problem is a large overestimation of the reduction in , leading to calculated values of of less than zero. For this problem, they used a nonlinear minimization method to improve the accuracy of the estimations for and . We also analyzed the absorbance spectrum based on the Lambert-Beer law for estimating and . In this case, the absorbance spectrum was linearly related to and . In addition, we used a linear regression model to specify the relationship between the sets of regression coefficients , , and obtained from MRA1 and the values of , , , and . Therefore, the results obtained by the proposed method may also have some errors in and due to the linearization. Although the phantom experiments showed reasonable results for , , , and , the estimated values of and have relative large deviations from the given values, as shown in Fig. 4. Those deviations account for the estimation error in shown in Fig. 5(b). The use of , , and and their higher order terms for the vector and can extend the current linear empirical formula to the nonlinear regression models. It may be useful to improve the accuracy of the estimations for , , , and . These issues should be investigated in the future. 5.ConclusionsIn summary, a method for imaging reduced scattering coefficients and the absorption coefficients of in vivo exposed brain tissues based on spectral reflectance images reconstructed from a single snap shot of an RGB image using the WEM was demonstrated in the present report. In vivo experiments using exposed rat brain while changing the fraction of inspired oxygen confirmed the feasibility of the proposed method for evaluating both cortical hemodynamics and changes in tissue morphologies due to loss of tissue viability in the brain. Since the proposed method visualizes both the hemodynamic response and the morphological changes in brain tissue, it may be useful for evaluating brain function and tissue viability in neurosurgery as well as in the diagnosis of several neurological disorders, such as neurotrauma, seizure, stroke, and ischemia. We intend to extend the proposed method in order to investigate brain function in cortical spreading depression and anoxic depolarization. AcknowledgmentsThe present study was supported in part by a Grant-in-Aid for Scientific Research from the Japanese Society for the Promotion of Science. ReferencesT. BonhoefferA. Grinvald, Brain Mapping: The Methods, Academic Press, San Diego
(1996). Google Scholar
I. NishidateK. YoshidaM. Sato,
“Changes in optical properties of rat cerebral cortical slices during oxygen glucose deprivation,”
Appl. Opt., 49
(34), 6617
–6623
(2010). http://dx.doi.org/10.1364/AO.49.006617 APOPAI 0003-6935 Google Scholar
R. R. AndersonJ. A. Parrish,
“The optics of human skin,”
J. Invest. Dermatol., 77
(1), 13
–19
(1981). http://dx.doi.org/10.1111/jid.1981.77.issue-1 JIDEAE 0022-202X Google Scholar
A. Ishimaru, Wave Propagation and Scattering in Random Media, Academic Press, New York
(1978). Google Scholar
L.-H. WangS. L. JacquesL.-Q. Zheng,
“MCML-Monte Carlo modeling of photon transport in multi-layered tissues,”
Comput. Methods Programs Biomed., 47
(2), 131
–146
(1995). http://dx.doi.org/10.1016/0169-2607(95)01640-F CMPBEK 0169-2607 Google Scholar
S. A. PrahlM. J. C. van GemertA. J. Welch,
“Determining the optical properties of turbid media by using the adding-doubling method,”
Appl. Opt., 32
(4), 559
–568
(1993). http://dx.doi.org/10.1364/AO.32.000559 APOPAI 0003-6935 Google Scholar
M. S. PattersonB. ChanceB. C. Wilson,
“Time resolved reflectance and transmittance for the noninvasive measurement of tissue optical properties,”
Appl. Opt., 28
(12), 2331
–2336
(1989). http://dx.doi.org/10.1364/AO.28.002331 APOPAI 0003-6935 Google Scholar
J. B. Fishkinet al.,
“Frequency-domain photon migration measurements of normal and malignant tissue optical properties in a human subject,”
Appl. Opt., 36
(1), 10
–20
(1997). http://dx.doi.org/10.1364/AO.36.000010 APOPAI 0003-6935 Google Scholar
T. J. FarrellM. S. PattersonB. Wilson,
“A diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivo,”
Med. Phys., 19
(4), 879
–888
(1992). http://dx.doi.org/10.1118/1.596777 MPHYA6 0094-2405 Google Scholar
S. L. Jacqueset al.,
“Video reflectometry to specify optical properties of tissue in vivo,”
Proc. SPIE, IS11 211
–226
(1993). http://dx.doi.org/10.1117/12.183203 PSISDG 0277-786X Google Scholar
L. WangS. L. Jacques,
“Use of a laser beam with an oblique angle on incidence to measure the reduced scattering coefficient of a turbid medium,”
Appl. Opt., 34
(13), 2362
–2366
(1995). http://dx.doi.org/10.1364/AO.34.002362 APOPAI 0003-6935 Google Scholar
A. Kienleet al.,
“Spatially resolved absolute diffuse reflectance measurements for noninvasive determination of the optical scattering and absorption coefficients of biological tissue,”
Appl. Opt., 35
(13), 2304
–2314
(1996). http://dx.doi.org/10.1364/AO.35.002304 APOPAI 0003-6935 Google Scholar
S.-P. Linet al.,
“Measurement of tissue optical properties by the use of oblique-incidence optical fiber reflectometry,”
Appl. Opt., 36
(1), 136
–143
(1997). http://dx.doi.org/10.1364/AO.36.000136 APOPAI 0003-6935 Google Scholar
U. UtzingerR. R. Richards-Kortum,
“Fiber optic probes for biomedical optical spectroscopy,”
J. Biomed. Opt., 8
(1), 121
–147
(2003). http://dx.doi.org/10.1117/1.1528207 JBOPFO 1083-3668 Google Scholar
P. R. Bargoet al.,
“In vivo determination of optical properties of normal and tumor tissue with white light reflectance and empirical light transport model during endoscopy,”
J. Biomed. Opt., 10
(3), 034018
(2005). http://dx.doi.org/10.1117/1.1921907 JBOPFO 1083-3668 Google Scholar
N. RajaramT. H. NguyenJ. W. Tunnell,
“Lookup table-based inverse model for determining optical properties of turbid media,”
J. Biomed. Opt., 13
(5), 050501
(2008). http://dx.doi.org/10.1117/1.2981797 JBOPFO 1083-3668 Google Scholar
B. Yuet al.,
“Cost-effective diffuse reflectance spectroscopy device for quantifying tissue absorption and scattering in vivo,”
J. Biomed. Opt., 13
(6), 060505
(2008). http://dx.doi.org/10.1117/1.3041500 JBOPFO 1083-3668 Google Scholar
B. S. NicholsN. RajaramJ. W. Tunnel,
“Performance of a lookup table-based approach for measuring tissue optical properties with diffuse optical spectroscopy,”
J. Biomed. Opt., 17
(5), 057001
(2012). http://dx.doi.org/10.1117/1.JBO.17.5.057001 JBOPFO 1083-3668 Google Scholar
R. Hennessyet al.,
“Monte Carlo lookup table-based inverse model for extracting optical properties from tissue-simulating phantoms using diffuse reflectance spectroscopy,”
J. Biomed. Opt., 18
(3), 037003
(2013). http://dx.doi.org/10.1117/1.JBO.18.3.037003 JBOPFO 1083-3668 Google Scholar
P. B. Joneset al.,
“Simultaneous multispectral reflectance imaging and laser speckle flowmetry of cerebral blood flow and oxygen metabolism in focal cerebral ischemia,”
J. Biomed. Opt., 13
(4), 044007
(2008). http://dx.doi.org/10.1117/1.2950312 JBOPFO 1083-3668 Google Scholar
C. Zhouet al.,
“Diffuse optical correlation tomography of cerebral blood flow during cortical spreading depression in rat brain,”
Opt. Express, 14
(3), 1125
–1144
(2006). http://dx.doi.org/10.1364/OE.14.001125 OPEXFF 1094-4087 Google Scholar
M. B. Bouchardet al.,
“Ultra-fast multispectral optical imaging of cortical oxygenation, blood flow, and intracellular calcium dynamics,”
Opt. Express, 17
(18), 15670
–15678
(2009). http://dx.doi.org/10.1364/OE.17.015670 OPEXFF 1094-4087 Google Scholar
S. Kawauchiet al.,
“Diffuse light reflectance signals as potential indicators of loss of viability in brain tissue due to hypoxia: charge-coupled-device-based imaging and fiber-based measurement,”
J. Biomed. Opt., 18
(1), 015003
(2013). http://dx.doi.org/10.1117/1.JBO.18.1.015003 JBOPFO 1083-3668 Google Scholar
C. Yinet al.,
“Simultaneous detection of hemodynamics, mitochondrial metabolism and light scattering changes during cortical spreading depression in rats based on multi-spectral optical imaging,”
NeuroImage, 76 70
–80
(2013). http://dx.doi.org/10.1016/j.neuroimage.2013.02.079 NEIMEF 1053-8119 Google Scholar
T. ArnoldM. De BiasioR. Leitner,
“Hyper-spectral video endoscope for intra-surgery tissue classification using auto-fluorescence and reflectance spectroscopy,”
Proc. SPIE, 8087 808711
(2011). http://dx.doi.org/10.1117/12.889940 PSISDG 0277-786X Google Scholar
A. Basiriet al.,
“Use of a multi-spectral camera in the characterization of skin wounds,”
Opt. Express, 18
(4), 3244
–3257
(2010). http://dx.doi.org/10.1364/OE.18.003244 OPEXFF 1094-4087 Google Scholar
J. Y. Hardeberg,
“Acquisition and Reproduction of Color Images: Colorimetric and Multispectral Approaches,”
Ecole Nationale Superieure des Telecommunications,
(1999). Google Scholar
J. Y. HardebergF. SchmittH. Brettel,
“Multispectral color image capture using liquid crystal tunable filter,”
Opt. Eng., 41
(10), 2532
–2548
(2002). http://dx.doi.org/10.1117/1.1503346 OPEGAR 0091-3286 Google Scholar
V. Cheunget al.,
“Characterization of trichromatic color cameras by using a new multispectral imaging technique,”
J. Opt. Soc. Am., A22
(7), 1231
–1240
(2005). http://dx.doi.org/10.1364/JOSAA.22.001231 JOSAAH 0030-3941 Google Scholar
H. L. ShenJ. H. Xin,
“Spectral characterization of a color scanner based on optimized adaptive estimation,”
J. Opt. Soc. Am. A, 23
(7), 1566
–1569
(2006). http://dx.doi.org/10.1364/JOSAA.23.001566 JOAOD6 0740-3232 Google Scholar
H. L. ShenJ. H. XinS. J. Shao,
“Improved reflectance reconstruction for multispectral imaging by combining different techniques,”
Opt. Express, 15
(9), 5531
–5536
(2007). http://dx.doi.org/10.1364/OE.15.005531 OPEXFF 1094-4087 Google Scholar
F. König,
“Reconstruction of natural spectra from color sensor using nonlinear estimation methods,”
in IS&T Annual Conference; The Society for Imaging Science and Technology,
454
–457
(1997). Google Scholar
P. StigellK. MiyataM. H. Kansari,
“Wiener estimation method in estimating of spectral reflectance from RGB image,”
Pattern Recogn. Image Anal., 17
(2), 233
–242
(2007). http://dx.doi.org/10.1134/S1054661807020101 PIANES 1054-6618 Google Scholar
Y. Murakamiet al.,
“Color reproduction from low-SNR multispectral images using spatio-spectral Wiener estimation,”
Opt. Express, 16
(6), 4106
–4120
(2008). http://dx.doi.org/10.1364/OE.16.004106 OPEXFF 1094-4087 Google Scholar
S. ChenQ. Liu,
“Modified Wiener estimation of diffuse reflectance spectra from RGB values by the synthesis of new color for tissue measurements,”
J. Biomed. Opt., 17
(3), 030501
(2012). http://dx.doi.org/10.1117/1.JBO.17.3.030501 JBOPFO 1083-3668 Google Scholar
J. Hernández-AndrésJ. RomeroR. L. Lee Jr.,
“Colorimetric and spectroradiometric characteristics of narrow-field-of-view clear skylight in Granada, Spain,”
J. Opt. Soc. Am. A, 18
(2), 412
–420
(2001). http://dx.doi.org/10.1364/JOSAA.18.000412 JOAOD6 0740-3232 Google Scholar
J. R. Mourantet al.,
“Mechanisms of light scattering from biological cells relevant to noninvasive optical-tissue diagnostics,”
Appl. Opt., 37
(16), 3586
–3593
(1998). http://dx.doi.org/10.1364/AO.37.003586 APOPAI 0003-6935 Google Scholar
D. Abookasiset al.,
“Imaging cortical absorption, scattering, and hemodynamic response during ischemic stroke using spatially modulated near-infrared illumination,”
J. Biomed. Opt., 14
(2), 024033
(2009). http://dx.doi.org/10.1117/1.3116709 JBOPFO 1083-3668 Google Scholar
S. A. Prahl, Tabulated Molar Extinction Coefficient for Hemoglobin in Water,
(2014) http://omlc.ogi.edu/spectra/hemoglobin/summary.html November ). 2014). Google Scholar
V. Tuchin, Tissue Optics: Light Scattering Methods and Instruments for Medical Diagnosis, 2nd ed.SPIE Press, Bellingham, WA
(2007). Google Scholar
I. NishidateY. AizuH. Mishina,
“Estimation of melanin and hemoglobin in skin tissue using multiple regression analysis aided by Monte Carlo simulation,”
J. Biomed. Opt., 9
(4), 700
–710
(2004). http://dx.doi.org/10.1117/1.1756918 JBOPFO 1083-3668 Google Scholar
S. Satoet al.,
“Real-time optical diagnosis of the rat brain exposed to a laser-induced shock wave: observation of spreading depolarization, vasoconstriction and hypoxemia-oligemia,”
PLoS One, 9
(1), e82891
(2014). http://dx.doi.org/10.1371/journal.pone.0082891 1932-6203 Google Scholar
P. Lipton,
“Ischemic cell death in brain neuron,”
Physiol. Rev., 79
(4), 1432
–1568
(1999). PHREA7 0031-9333 Google Scholar
C. R. JarvisT. R. AndersonR. D. Andrew,
“Anoxic depolarization mediates acute damage independent of glutamate in neocortical brain slices,”
Cereb. Cortex, 11
(3), 249
–259
(2001). http://dx.doi.org/10.1093/cercor/11.3.249 53OPAV 1047-3211 Google Scholar
I. JoshiR. D. Andrew,
“Imaging anoxic depolarization during ischemia-like conditions in the mouse hemi-brain slice,”
J. Neurophysiol., 85
(1), 414
–424
(2001). JONEA4 0022-3077 Google Scholar
T. M. PolischukC. R. JarvisR. D. Andrew,
“Intrinsic optical signaling denoting neuronal damage in response to acute excitotoxic insult in the hippocampal slice,”
Neurobiol. Dis., 4
(6), 423
–437
(1998). http://dx.doi.org/10.1006/nbdi.1998.0172 NUDIEM 0969-9961 Google Scholar
C. R. Jarviset al.,
“Interpretation of intrinsic optical signals and calcein fluorescence during acute excitotoxic insult in the hippocampal slice,”
NeuroImage, 10
(4), 357
–372
(1999). http://dx.doi.org/10.1006/nimg.1999.0473 NEIMEF 1053-8119 Google Scholar
A. ObeidatC. R. JarvisR. D. Andrew,
“Glutamate does not mediate acute neuronal damage following spreading depression induced by O2/glucose deprivation in the hippocampal slice,”
J. Cereb. Blood Flow Metab., 20
(2), 412
–422
(2000). http://dx.doi.org/10.1097/00004647-200002000-00024 JCBMDN 0271-678X Google Scholar
M. Friebelet al.,
“Influence of oxygen saturation on the optical scattering properties of human red blood cells in the spectral range 250 to 2000 nm,”
J. Biomed. Opt., 14
(3), 034001
(2009). http://dx.doi.org/10.1117/1.3127200 JBOPFO 1083-3668 Google Scholar
BiographyKeiichiro Yoshida is a PhD student in the Graduate School of Bio-Applications and Systems Engineering, Tokyo University of Agriculture and Technology, Japan, where he received his MS degree in engineering in 2012. Izumi Nishidate is an associate professor in the Graduate School of Bio-Applications and Systems Engineering, Tokyo University of Agriculture and Technology. He received his MS and PhD degrees from Muroran Institute of Technology, Japan. His research interests include diffuse reflectance spectroscopy, light transport in biological tissues, multispectral imaging, and functional imaging of skin and brain tissues. Tomohiro Ishizuka is a MS student in the Graduate School of Bio-Applications and Systems Engineering, Tokyo University of Agriculture and Technology, Japan. Satoko Kawauchi is an assistant professor in the Division of Biomedical Information Sciences, National Defense Medical College Research Institute, Japan. She completed her MS degree at Keio University and received a PhD degree in medicine from Kyorin University, Japan. Her research interests encompass functional imaging of brain, photodynamic therapy, and diagnosing/imaging tissue viability using light scattering. Shunichi Sato is an associate professor in the Division of Biomedical Information Sciences, National Defense Medical College Research Institute, Japan. He received his MS and PhD degrees from Keio University, Japan. His research interests include photomechanical waves and their application to gene/drag delivery systems, photoacoustic imaging, photodynamic therapy, as well as diagnosing/imaging brain tissue viability. Manabu Sato is a professor in the Graduate School of Science and Engineering, Yamagata University, Japan. He received his MS and PhD degrees from Tohoku University, Japan. His research focus is on optical coherence tomography, spectral imaging, and their applications to imaging of intrinsic optical signals in brain, as well as development of instruments with optical fiber for biological tissues. |