1.IntroductionA fiber-optic probe will interrogate a specific volume within a sample that is determined by the probe geometry and the sample optical properties. For accurate optical assessment of tissue structure and function, it is imperative that this interrogation volume coincides with the location where the relevant biological processes are occurring. The importance of depth-selectivity for accurate diagnosis has been underscored experimentally and clinically, especially for the early detection of cancer, which most often originates in a few-hundred-microns-thick mucosal tissue layer. Our optical and histological investigations in human patients and animal models of colon carcinogenesis have shown diagnostic alterations in hemoglobin concentration that are present in the mucosa but attenuated at deeper submucosal depths.1–5 Studies with angle-resolved low-coherence interferometry demonstrate that elevation of nuclear diameter associated with dysplasia in Barrett’s esophagus was only detectable at a depth of 200 to 300 μm and not observable at 0 to 100 μm or 100 to 200 μm.6 Several fiber-optic probe modalities have been developed to target the mucosal layer with a wide field of view. These include polarization-gating spectroscopy (PGS),7,8 differential path length spectroscopy (DPS),9 elastic light scattering spectroscopy (ESS),10 single-fiber reflectance spectroscopy (SFS),11 and angled-illumination-collection designs.12–14 These methods have proven successful in diagnosing epithelial lesions including those in the colon, oral, and other types of mucosae.1,2,9,13,15–18 On the other hand, diffuse reflectance spectroscopy (DRS) methods typically have a sampling depth on the order of several millimeters and are consequently not very selective to the mucosal layer. However, the deeper sampling depth of DRS may be advantageous for other applications. For example, optical assessment of blood vessels under thick skin such as the palm requires light to first penetrate through the epidermal layer. This can be achieved with DRS, but not easily with the other techniques mentioned previously. It is evident from the above considerations that matching the probe sampling depth to the clinical application is one of the main driving forces behind optimal probe selection. Given the variety of spectroscopic methods that are currently available, there is a need to develop a schema for both choosing the correct technology and also selecting the optimal configuration of that technology for the application of interest. Parameters of the configuration can include size of the illumination-collection area, collection angle, fiber diameter, or inter-fiber spacing. In choosing an optical technology, there are a number of factors to consider including sampling depth, length-scale sensitivity, signal-to-noise ratio, cost, and convenience. In the framework we develop in this paper, we focus on two crucial facets of probe development: the mean sampling depth and the sensitivity of the mean depth and path length to perturbations in the optical properties of the sample. Structurally and functionally, biological tissue is multilayered with specific biological processes and diseases occurring at different depth layers. Accurate optical assessment of these processes and diseases requires light to be preferentially targeted to the layer of interest making the sampling depth a critical aspect of probe design. It is important to point out that it is not sufficient to select a probe with an application-specific depth of tissue interrogation: one needs to consider how this sampling depth may vary depending on the optical properties of tissue. The clinical rationale for considering these parameters is that optimizing both of them will maximize the effect size for a given diagnostic parameter between control and disease groups. The Cohen effect size () in statistics is defined as the difference in the means between the two groups divided by the pooled standard deviation (): Selecting an appropriate depth that targets the diagnostic layer will maximize as highlighted by the clinical studies referenced previously. The effect size can also be improved by minimizing the pooled standard deviation. Previous research has focused on reducing biomarker variability through improved data collection,19–21 calibration,22 and/or post-processing.23 Another avenue to improve variability is by minimizing the dependence of the mean sampling depth and path length on the optical properties of the medium. This ensures that a consistent depth is targeted from tissue site to tissue site and also from patient to patient. In addition, robust application of the Beer-Lambert law to tissue spectroscopy requires that the path length be insensitive to tissue optical properties.24 This fact has motivated the development of probe designs for which the path length is independent of optical properties.25,26It would be cumbersome to experimentally test side by side the individual technologies highlighted above for the optimum sampling depth and sampling depth sensitivity. To overcome this problem, we have culled the literature for mathematical expressions of the sampling depth and path length of various techniques and derived our own formulations. We have incorporated our analysis into a graphical user interface (GUI) that takes user input of the sample optical properties outputs optimum configurations of different spectroscopic techniques. The user can then select the best technique and configuration combination for their application of interest. We expect that our method and GUI tool will aid investigators who believe that optical methods would be useful in their disease screening research, but do not know which technique would be best suited for their task. Our results can also help researchers refine and optimize the geometry of their current fiber-optic probes. 2.Materials and Methods2.1.Probe Selection MetricsA variety of factors can be considered when choosing a probe for a particular application. As stated in the introduction, we focused on the probe mean sampling depth and on the sensitivity of both the sampling depth and path length to perturbations in the optical properties of the medium. The main optical properties of a sample are the scattering coefficient , the absorption coefficient , and the average cosine of the scattering angle (anisotropy factor ). We found that the depth and path length were primarily functions of the reduced scattering coefficient and . The problem of probe evaluation can be separated into two components, which need to be defined.
The probe selection criteria embodied by the mean sampling depth and the depth/path length sensitivities can automatically be calculated for a given optical technique if and are known. Expressions governing these parameters for DRS, PGS, ESS, SFS, and DPS are provided in the next section. 2.2.Depth and Path Length ExpressionsExpressions for the mean sampling depth and mean average path length of several optical spectroscopic techniques are given below. These expressions have been derived from MC simulations in which the sample was assumed to be homogeneous and in which the Henyey-Greenstein phase function was employed. In all cases, the illumination beam was assumed to have a flat intensity profile and the numerical aperture of simulated fibers was 0.22. The refractive index of the media was 1.4 to correspond to biological tissue, while the outside refractive index was 1.52 to correspond to either a lens or optical fiber.
Table 1Fitting coefficients for PGS depth models.
2.3.Probe Selection AlgorithmSelecting an optimal probe is a multicriteria decision analysis problem. One method to solve such problems is a weighted sum model (WSM). In the WSM, each criterion is given a relative weight of importance. There are two main criteria in our probe selection problem.
The sampling depth deviation as well as depth and path length sensitivities will be functions of the probe geometry for each optical technique. The geometry can be manipulated by changing , , , or . Suppose that a user has a target depth of and that probe geometry has sampling depth , depth sensitivities and , and path length sensitivities and . In our analysis, optimal probe selection will be achieved by choosing a particular to minimize the weighted sum : where are relative weights supplied by the user. These weights indicate the importance the user places on the corresponding criteria term. For example, if is set to zero, then the algorithm will not take into consideration the match between the target depth and the probe sampling depth when selecting an optimal geometry. In contrast, if is set to a high value, the algorithm will attempt to match the target depth and the probe sampling depth very accurately. To minimize Eq. (22), we used the Nelder-Mead optimization algorithm in MATLAB. In brief, the algorithm will cycle (via a simplex search) through the geometry parameters (, , , and ) until a minimum sum is reached. It will do this separately for each optical technique such that there will be an optimum geometry for DRS, PGS, SFS, DPS, and ESS.2.4.Automated Probe Selection with a MATLAB Graphical User InterfaceWe implemented a MATLAB GUI to automate probe selection using Eq. (22). A screenshot of the GUI is shown in Fig. 2(a). The first step in the GUI is for the user to specify the baseline values of, and which in turn delineate the and parameters of Eq. (5). These values should correspond with mean optical properties of the sample of interest and can be obtained from the literature or experimentally determined integrating sphere measurements. The optical properties of a wide variety of tissue types have been previously studied.44 However, care must be taken when translating ex vivo optical property determinations to in vivo measurements. The second step is for the use to enter the target depth. The target depth is the depth within the sample where the most diagnostic or scientifically relevant information is expected to be obtained. Next, the user stipulates the relative weights given in Eq. (22). These weights determine the criteria that will be the most important in the automated selection process. If these are not explicitly specified, they will default to a value of 1. Finally, the user can place upper and lower bounds on the probe geometry parameters such as , , , and . This may help the user limit the selection space to geometries that utilize commercially available components or that are more convenient in a clinical setting. After all input values have been selected and the user initiates the automated selection, the program will attempt to minimize Eq. (22) for each optical technique. After the minimization procedure is complete, the GUI will output a table of results as depicted in Fig. 2(a). Each row of the table corresponds to a different optical technique (DRS, PGS, SFS, DPS, ESS). The first column gives the weighted sum from Eq. (22) for the optimal geometry of each technique. The best technique will have the lowest value in the first column. The next five columns give the values of the optimization criteria: the depth deviation and the depth and path length sensitivities. In the remaining columns are shown the values of the geometry parameters (, , , and ) that compose the optimum geometry. The user can take these values and construct the ideal application-specific probe. The DRS technique is known to be valid across a specified range of parameters. Therefore, in our algorithm we institute two checks to make sure the user is notified as to when the diffuse reflectance assumptions may be violated. The algorithm displays a warning dialog box if the transport albedo for which the DRS result would not be accurate.45 In addition, the algorithm ensures that the output value of satisfies . A warning dialog is displayed if the upper and lower bounds on would not satisfy this property. The algorithm outputs ideal probe configurations based on single values for and that correspond to a single wavelength. Since broadband measurements are often taken, it can be useful to visualize how the depth and path length vary with wavelength. To satisfy this need, the GUI plots the depth and/or path length for each technique as a function of wavelength. This is achieved in the “depth and path length spectra” panel of the GUI, as shown in Fig. 2(a). The user selects a technique from the first drop-down menu, either the depth or path length from the second drop-down menu, and then loads a MATLAB data file. The data file consists of the wavelength in the first column followed by the corresponding and values in the second and third columns, respectively. Based on these values, the GUI will use the expressions in Sec. 2.1 to calculate the depth or path length at each wavelength for the ideal geometry specified in the GUI table. As an example, we show the SFS path length spectra for in Fig. 2(b). The value of was at 560 nm and followed a dependence. The value was at 560 nm and followed the absorption spectrum of hemoglobin with 75% oxygen saturation. The path length follows the hemoglobin spectrum with dips in the path length corresponding to the characteristic hemoglobin absorption peaks at 420, 542, and 576 nm. 2.5.Case StudiesTo illustrate the use of the GUI, we consider two biological case studies. The first is the case of detecting dyplasia in Barrett’s esophagus. Previous research with angle-resolved low coherence interferometry has shown that diagnostic increases in nuclear size occur at a depth of 200 to 300 μm beneath the esophageal surface.6 The user could set a target depth of 250 μm in the GUI. The optical properties at a wavelength of 630 nm for normal human esophagus have been supplied by Table 2 of Holmer et al.: , , and .46 The next task is assigning the relative weights from Eq. (22). It has been found that the diagnostic changes occurring with esophageal dysplasia are localized to the 200- to 300-μm layer. Therefore, it is important to have the target depth and probe sampling depth match as closely as possible and we set . Assigning weights to the sensitivity variables hinges on several considerations. The standard of deviation is high with a coefficient of variation greater than one. To maintain a consistent depth or path length given this variability, it is desirable minimize the sensitivity of the depth and path length to . The increase in nuclear size observed with low coherence interferometry might be expected to result in an increase in and consequent reduction in . To target the same region between control and dysplastic patients would necessitate minimizing the depth and path length sensitivity to . For simplicity, we will set also equal to one. The final user input is setting bounds on the geometry parameters. Choice of these bounds is primarily driven by what is commercially available or feasible to manufacture and what is clinically convenient. For example, the diameter of upper endoscope accessory channels places upper bounds on the size of a potential esophageal probe. We set a lower limit of 100 μm and upper limit of 3 mm for and while setting the lower limit of to 250 μm. We allowed to vary from 0 deg to 45 deg. Table 2Algorithm results for esophageal dysplasia detection case study.
The second case involves optically determining chromophore concentrations from target tissue using a Beer’s Law algorithm. This approach has been previously used to study microcirculatory alterations associated with dysplasia2,47 as well as to monitor chemotherapy drug concentrations in tissue.48 Insensitivity of the effective path length to tissue optical properties helps to ensure robust application of this method.24,26 For the second case study, we change the target depth to 150 μm. Hemoglobin concentration measured from this depth was diagnostic for early detection of colonic neoplasia.2 For simplicity, we will maintain the same optical properties as the first case study since the optical properties of the colon have been found to be similar.49 In general, the optical properties will need to be adjusted based on the tissue or organ being investigated. This case study will require higher weights to be placed on the path length sensitivity terms. As an example, we will consider , and . 3.Results3.1.Validation of Sampling Depth Expressions for PGSIn Fig. 3, we plot the MC simulations of the PGS sampling depth versus the models developed in Eqs. (13) to (15). Each PGS signal is shown in a different subplot of Fig. 3. with (a) co-polarized signal, (b) the cross-polarized signal, (c) total signal. In all cases, there is a clear linear correlation between the simulations and the models with Pearson correlation coefficient greater than 0.99. The mean percent differences between model and simulation were 9% for the co-polarized signal, 8% for the cross-polarized signal, 10% for the total signal. The error for the delta-polarized signal has been previously found to be ~5% in Ref. 39. This leads us to conclude that Eqs. (13) to (15) can be used as accurate condensations of the MC simulations for PGS. 3.2.Validation of Depth and Path Length Expressions for DPSWe illustrate the correspondence between our MC simulations of the DPS geometry and the models of Eqs. (18) to (20) for the depth and path length in Fig. 4(a) and 4(b). Figure 4(a) demonstrates that the simulation and model coordinates cluster around the ideal line of unity indicating good agreement. This agreement was quantified through the mean percent error, which was 3%. Figure 4(b) also demonstrates good agreement between simulation and model with an overall percent error of 3% for the path length. As previously mentioned in Sec. 2.2, our model of the DPS path length agrees well with previously published experimental data43 with a mean percent error of 8%. 3.3.Validation of Depth and Path Length Expressions for ESSIn Fig. 5(a) and 5(b) we have plotted our MC simulations of the depth and path length of the ESS geometry versus our models of the depth and path length from Eq. (21). For the depth plotted in Fig. 5(a), there is a 9% error percent difference of the depth data points about the ideal line of unity. In Fig. 5(b), the percent difference of the path length data about the unity line is 11%. Figure 5(b) also plots the path length as determined from a previously published analytical model for ESS by Reif et al.:10 . This equation is only valid for 200-μm diameter fiber with a 250-μm inter-fiber spacing. As Fig. 5(b) demonstrates, the data points from this previously validated model (shown in red) clearly overlap with the data points from model in Eq. (21) (shown in black). This underscores the validity of our MC simulations of the ESS geometry and the models we derived from them. 3.4.Sampling Depth and Sensitivity BehaviorWe now examine the behavior of the mean sampling depth for all the techniques previously mentioned. In particular, we were interested in how the depth could be tuned by varying the geometry, as this would directly affect optimal probe selection. In Fig. 6, we plot the log of the sampling depth scaled by of the various techniques versus a parameter we term the area extent. The area extent is a measure of the maximum horizontal distance a photon could travel between its entry and exit points for each technique multiplied by . For PGS the area extent is equal to , for DPS and SFS the area extent is , while for ESS and DRS the area extent is . From the expressions in Sec. 2.1 it can be observed that in the case of PGS, SFS, and DPS, for , the quantity can be represented as a function solely of the area extent. The optical properties used to generate the data in Fig. 6 are , , and From Fig. 6, it is clear that the sampling depth increases with the area extent for all the techniques though the precise behavior is not the same. For example, the delta-polarized signal saturates very quickly with area extent when compared with other methods. Next we wanted to examine the sensitivity of the depth and path length to and . In Fig. 7(a) and 7(b) we plot the sensitivity [defined in Eq. (5)] of the sampling depth to and , respectively, as a function of a geometry parameter. The geometry parameter is simply defined as for PGS, for DPS and SFS, and for DRS. The ESS technique is not shown since its sensitivity will be identical to that of SFS over the range of specified in Sec. 2.2. The optical properties used to generate the data in Fig. 7 are , , and . Examination of Fig. 7(a) shows the dependence of the depth sensitivity on the geometry parameter is dependent on the technique being used. In most cases, the sensitivity decreases when the geometry parameter is increased. Notable exceptions are the PGS delta-polarized signal for which the sensitivity increases and eventually saturates with and SFS/ESS and DRS for which the sensitivity is independent of . If the main objective was to minimize depth sensitivity, then selection of a DRS probe or analysis of the total-polarization signal from a polarization-gated probe with area extent greater than five would be appropriate. It should be noted that the validity of the DRS equations used to generate Figs. 6 and 7 are applicable when the diffusion approximation is valid, namely when . Next we examine the depth sensitivity depicted in Fig. 7(b). Here the behavior of the sensitivity as a function of the geometry parameter is more uniform across optical techniques. In nearly all cases, the sensitivity increases with the geometry parameters. The lone exception is DRS, for which the depth sensitivity is independent of . Optimal minimization of the depth sensitivity would entail selection of a DRS probe. Next, we explored the sensitivity of the path length to and in Fig. 7(c) and 7(d), respectively. The SFS/ESS, PGS (co-pol), and PGS (total-pol) methods have path length sensitivities that are independent of the geometry parameter. The PGS (delta-pol) sensitivity increases with and saturates at large . The DRS sensitivity steadily decreases with , while the PGS (cross-pol) and DPS tend to decrease with the geometry parameter. The behavior of the DPS path length sensitivity to deserves further examination. The DPS sensitivity reaches zero when is approximately equal to 2.4 or equivalently when the transport mean free path [] is equal to , which corresponds to twice the center-to-center separation between the DPS fibers. It is beyond the scope of this paper to confirm whether this phenomenon generalizes to other inter-fiber spacings, but it is useful to know where exactly the sensitivity reaches the optimal zero value. Finally, we investigated path length sensitivity in Fig. 7(d). The main pattern observed is for the sensitivity to increase with the geometry parameter for all the optical methods. The optimal techniques for path length sensitivity minimization are DPS and the PGS (delta-pol) methods. 3.5.Application of Automated Probe Selection Algorithm to Biological Case StudiesWe utilized the MATLAB GUI to implement the probe selection algorithm embodied in Eq. (22). For the case of detecting dysplasia in Barrett’s esophagus, we input the optical properties, target depth, and associated weights as described in Sec. 2.5. The automated algorithm took 0.15 s to run on a personal computer and the results were outputted in tabular format. These results are summarized in Table 2. Each row corresponds to a different technique, and each column gives the value of the criterion or geometry parameter for the optimal geometry of that technique. The optimal technique will have the lowest value in the weighted sum column. For this case study, the PGS-total, PGS-co, and SFS techniques performed similarly with weighted sum equal to ~0.7. The ideal SFS probe would have a of 0.025 cm. In general, most of the optical methods are capable of targeting the specified depth with an error less than 1%. The exceptions are ESS and DRS. This is because they tend to target deeper depths than the specified 0.025 cm. The next case study we examined was measuring chromophores concentration from a shallow depth of 0.0150 cm. Compared to the first case, the target depth was less and higher weight was placed on the path length sensitivity. The results of this case study are depicted in Table 3. For this case study, a PGS (delta-pol) probe with an of 0.005 cm and is optimal. Using this type of probe, the 0.0150-cm depth would be interrogated within 1%. For a 1% perturbation in or , the path length of the PGS (delta-pol) probe would deviate by 0.15% and 0.02%, respectively. Table 3Algorithm results for chromophores concentration measurement at a depth of 150 μm.
4.DiscussionIn this paper, we have provided a simple and flexible framework for application-specific fiber-optic probe design. The guiding principle of the algorithm was to maximize diagnostic effect size for clinical biophotonic applications as defined in Eq. (1). To accomplish this, our framework takes into account the target sampling depth, and the sensitivity of the depth and path length to fluctuations in the optical properties of the sample. We use a weighted sum algorithm incorporating the above criteria to optimize the probe geometry. The algorithm is implemented in an easy-to-use MATLAB GUI interface where the user can specify the target depth, sample optical properties, relative importance of the algorithm criteria, and bounds on elements of the probe geometry. We plan on making this software publicly available for researchers to use. Our results show that in most cast cases, both the mean sampling depth and the sensitivities of the path length and depth can be adjusted by the geometry. Notable exceptions include PGS (co-pol), PGS (total-pol), and SFS for which the path length sensitivity is independent of this value. In general, depth and path length sensitivity to increases with the geometry parameter (, , or ) for all optical techniques. The pattern is more complicated when looking at the sensitivity to . Here the sensitivity may decrease or increase with the geometry parameter depending on the optical technique. For example, the path length sensitivity increases with for the PGS (delta-pol) signal but decreases with for the DPS method. In addition, a wide range of depths. ESS and DRS tend to sample depths greater than , while the remaining methods interrogate depths on the order of 0.3-1 as shown in Fig. 6. The biological case studies we laid out gave realistic and concrete examples of probe selection scenarios. The first example was diagnosing esophageal dysplasia at a depth of 250 μm. All the techniques except for ESS and DRS had a depth percent error of less than 3%. This is not surprising, as these methods were designed largely in part to target mucosal tissue structures. If we had set a deeper target depth, then the ESS and DRS techniques would have performed better. Indeed, if the target depth is set to 650 μm, the depth percent error for ESS drops from 131% to 5%. Our next case study involved measuring chromophore concentration from a depth of 150 μm. In this example, the path length sensitivity was considered paramount for accurate application of Beer’s law. The PGS (delta-pol) was found to be ideal both because of its ability to correctly achieve the 150-μm depth and its low path length sensitivity to . At this juncture, it must be stressed that the outcome of the algorithm depends on the user-selected weights of Eq. (22). This is both an advantage and disadvantage. The advantage is that it gives the user a lot of flexibility to experiment with different weights and find what best matches the application. This, however, introduces an element of subjectivity to the problem of probe selection in the quality of the output depends on user guidance. In addition, the algorithm will be sensitive to the accuracy of the tissue optical properties input by the user. In general, the optical properties are not known precisely and must be estimated, typically from ex vivo specimens whose results do not translate exactly to the in vivo case. As an example of the algorithm sensitivity to optical properties, we increased the value of by 20% in the first biological case study described in Secs. 2.5 and 3.5. The PGS-total technique remained the optimal technique selected by the algorithm but the geometry changed from R = 0.018 cm to R = 0.02 cm. This suggests that the ultimate technique chosen by the algorithm may be robust to optical property uncertainty but that the specific geometry of that technique will be affected by optical property uncertainty. The main goal of this paper was to develop a framework for optimal probe design. Our algorithm necessitates expressions relating the sampling depth and path length to the optical properties of the medium as well as illumination-collection geometry. In the course of our main study we have used Monte Carlo simulations to develop depth and path length expressions for PGS, DPS, and ESS. These expressions have utility independent of their contributions to our algorithm. For example, they can be used to study the sampling volumes of these techniques and in particular the wavelength and system geometry dependence of these volumes. There are also several probe techniques that this paper has not considered at this time, in particular probe geometries that use tilted illumination and collection beams.10,12,14 This is due to a lack of condensed equations explaining their depth and path length behavior. However, our algorithm and GUI can be easily extended once equations for these and other techniques are known. To maintain simplicity in our algorithm, we have made use of some assumptions that need to be addressed. In both our own Monte Carlo simulations and in the simulations of other groups that we employed, it has been assumed that the optical properties are distributed homogeneously throughout the sample. In reality, biological tissue can be multilayered, have absorption localized to blood vessels, and have inhomogeneous distribution of the scattering properties. A correction factor50 has been developed for blood vessel absorption that would allow our algorithm to be fully valid as long as the corrected is input to the algorithm. However, there is currently no general solution for multilayered structures and our use of the homogeneous assumption is necessary to make the problem tractable. While the value of the one-layer assumption for studying reflectance from biological media is well established,9,10,51,52 our algorithm results must be considered as an estimate for multilayer systems and future study of the effect of multilayer structures on our algorithm is warranted. In addition, we have not explicitly considered the effect of the scattering phase function on the depth and path length. The Henyey-Greenstein phase function was used in the modeling for the all the techniques studied. It has been previously found that the details of the phase function have only a minor influence on the depth and path length properties of the techniques we investigated.10,39,53 Thus we do not expect the choice of phase function to significantly alter our results though this is an area of future study. Finally, as noted in Sec. 2.1, we used a sampling depth definition based on the expected value of the maximum depth collected photons will have reached and we applied this definition consistently across the different optical techniques. Other definitions of depth are also possible such as a weighted mean33 or the depth from which a specified percentage of photons emerge.25 Ideally, different depth metrics would be incorporated into the algorithm. This feature is limited by the availability of different depth expressions. Our algorithm and GUI could also be extended in the future to incorporate different depth definitions. This paper has considered the target depth and the depth and path length sensitivities to be the main criteria for probe selection. The chief reasons for this framework are both its relevance to increasing diagnostic effect size and that its parameters are easily computable. However, there are additional factors that can govern probe selection. Cost, ease of manufacturing, and signal to noise ratio (SNR) are crucial considerations especially for technologies that seek to be commercialized. In addition, some techniques may be more readily translated to a clinical setting. Smaller probes, for example, can fit through the various accessory channels of endoscopes. The GUI indirectly addresses this issue by allowing the user to set upper and lower bounds on the probe geometry parameters. These can be linked to the cost, size, and SNR of the final probe design. 5.ConclusionsIn this paper we have presented a framework for application-specific probe design and selection. The main outcome is a flexible and user-friendly GUI that automates probe assessment for several common optical methods. We intend to make this GUI and associated software publicly available for researchers to investigate promising probe designs for their application of interest. We expect that our algorithm will aid users in evaluating probe designs for specific applications. AcknowledgmentsNIH grants R01CA128641, R01 EB003682. The authors thank Joseph Wilson for help with software programming. ReferencesA. J. Gomeset al.,
“Rectal mucosal microvascular blood supply increase is associated with colonic neoplasia,”
Clin. Cancer Res., 15
(9), 3110
–3117
(2009). http://dx.doi.org/10.1158/1078-0432.CCR-08-2880 CCREF4 1078-0432 Google Scholar
H. K. Royet al.,
“Spectroscopic microvascular blood detection from the endoscopically normal colonic mucosa: biomarker for neoplasia risk,”
Gastroenterology, 135
(4), 1069
–1078
(2008). http://dx.doi.org/10.1053/j.gastro.2008.06.046 GASTAB 0016-5085 Google Scholar
M. P. Siegelet al.,
“Assessment of blood supply in superficial tissue by polarization-gated elastic light-scattering spectroscopy,”
Appl. Opt., 45
(2), 335
–342
(2006). http://dx.doi.org/10.1364/AO.45.000335 APOPAI 0003-6935 Google Scholar
A. K. Tiwariet al.,
“Neo-angiogenesis and the premalignant micro-circulatory augmentation of early colon carcinogenesis,”
Cancer Lett., 306
(2), 205
–213
(2011). http://dx.doi.org/10.1016/j.canlet.2011.03.008 CALEDQ 0304-3835 Google Scholar
R. K. Waliet al.,
“Increased microvascular blood content is an early event in colon carcinogenesis,”
Gut, 54
(5), 654
–660
(2005). http://dx.doi.org/10.1136/gut.2004.056010 GUTTAK 0017-5749 Google Scholar
N. G. Terryet al.,
“Detection of dysplasia in Barrett’s esophagus with in vivo depth-resolved nuclear morphology measurements,”
Gastroenterology, 140
(1), 42
–50
(2011). http://dx.doi.org/10.1053/j.gastro.2010.09.008 GASTAB 0016-5085 Google Scholar
V. M. Turzhitskyet al.,
“Measuring mucosal blood supply in vivo with a polarization-gating probe,”
Appl. Opt., 47
(32), 6046
–6057
(2008). http://dx.doi.org/10.1364/AO.47.006046 APOPAI 0003-6935 Google Scholar
A. Myakovet al.,
“Fiber optic probe for polarized reflectance spectroscopy in vivo: design and performance,”
J. Biomed. Opt., 7
(3), 388
–397
(2002). http://dx.doi.org/10.1117/1.1483314 JBOPFO 1083-3668 Google Scholar
A. Amelinket al.,
“In vivo measurement of the local optical properties of tissue by use of differential path-length spectroscopy,”
Opt. Lett., 29
(10), 1087
–1089
(2004). http://dx.doi.org/10.1364/OL.29.001087 OPLEDP 0146-9592 Google Scholar
R. ReifO. A’AmarI. J. Bigio,
“Analytical model of light reflectance for extraction of the optical properties in small volumes of turbid media,”
Appl. Opt., 46
(29), 7317
–7328
(2007). http://dx.doi.org/10.1364/AO.46.007317 APOPAI 0003-6935 Google Scholar
M. CanpolatJ. R. Mourant,
“Particle size analysis of turbid media with a single optical fiber in contact with the medium to deliver and detect white light,”
Appl. Opt., 40
(22), 3792
–3799
(2001). http://dx.doi.org/10.1364/AO.40.003792 APOPAI 0003-6935 Google Scholar
L. Niemanet al.,
“Optical sectioning using a fiber probe with an angled illumination-collection geometry: evaluation in engineered tissue phantoms,”
Appl. Opt., 43
(6), 1308
–1319
(2004). http://dx.doi.org/10.1364/AO.43.001308 APOPAI 0003-6935 Google Scholar
M. C. Skalaet al.,
“Investigation of fiber-optic probe designs for optical spectroscopic diagnosis of epithelial pre-cancers,”
Laser. Surg. Med., 34
(1), 25
–38
(2004). http://dx.doi.org/10.1002/(ISSN)1096-9101 LSMEDI 0196-8092 Google Scholar
A. M. Wanget al.,
“Depth-sensitive reflectance measurements using obliquely oriented fiber probes,”
J. Biomed. Opt., 10
(4), 44017
(2005). http://dx.doi.org/10.1117/1.1989335 JBOPFO 1083-3668 Google Scholar
A. Dharet al.,
“Elastic scattering spectroscopy for the diagnosis of colonic lesions: initial results of a novel optical biopsy technique,”
Gastrointest. Endosc., 63
(2), 257
–261
(2006). http://dx.doi.org/10.1016/j.gie.2005.07.026 GAENBQ 0016-5107 Google Scholar
J. R. Mourantet al.,
“Spectroscopic diagnosis of bladder cancer with elastic light scattering,”
Laser. Surg. Med., 17
(4), 350
–357
(1995). http://dx.doi.org/10.1002/(ISSN)1096-9101 LSMEDI 0196-8092 Google Scholar
L. T. Niemanet al.,
“Probing local tissue changes in the oral cavity for early detection of cancer using oblique polarized reflectance spectroscopy: a pilot clinical trial,”
J. Biomed. Opt., 13
(2), 024011
(2008). http://dx.doi.org/10.1117/1.2907450 JBOPFO 1083-3668 Google Scholar
A. Sharwaniet al.,
“Assessment of oral premalignancy using elastic scattering spectroscopy,”
Oral Oncol., 42
(4), 343
–349
(2006). http://dx.doi.org/10.1016/j.oraloncology.2005.08.008 EJCCER 1368-8375 Google Scholar
J. S. Leeet al.,
“Sources of variability in fluorescence spectroscopic measurements in a Phase II clinical trial of 850 patients,”
Gynecol. Oncol., 107
(Suppl. 1), S260
–S269
(2007). http://dx.doi.org/10.1016/j.ygyno.2007.07.012 GYNOA3 Google Scholar
B. M. Pikkulaet al.,
“Instrumentation as a source of variability in the application of fluorescence spectroscopic devices for detecting cervical neoplasia,”
J. Biomed. Opt., 12
(3), 034014
(2007). http://dx.doi.org/10.1117/1.2745285 JBOPFO 1083-3668 Google Scholar
S. Rudermanet al.,
“Analysis of pressure, angle and temporal effects on tissue optical properties from polarization-gated spectroscopic probe measurements,”
Biomed. Opt. Express, 1
(2), 489
–499
(2010). http://dx.doi.org/10.1364/BOE.1.000489 BOEICL 2156-7085 Google Scholar
B. YuH. L. FuN. Ramanujam,
“Instrument independent diffuse reflectance spectroscopy,”
J. Biomed. Opt., 16
(1), 011010
(2011). http://dx.doi.org/10.1117/1.3524303 JBOPFO 1083-3668 Google Scholar
Y. Zhuet al.,
“Elastic scattering spectroscopy for detection of cancer risk in Barrett’s esophagus: experimental and clinical validation of error removal by orthogonal subtraction for increasing accuracy,”
J. Biomed. Opt., 14
(4), 044022
(2009). http://dx.doi.org/10.1117/1.3194291 JBOPFO 1083-3668 Google Scholar
H. Liu,
“Unified analysis of the sensitivities of reflectance and path length to scattering variations in a diffusive medium,”
Appl. Opt., 40
(10), 1742
–1746
(2001). http://dx.doi.org/10.1364/AO.40.001742 APOPAI 0003-6935 Google Scholar
C. Fanget al.,
“Depth-selective fiber-optic probe for characterization of superficial tissue at a constant physical depth,”
Biomed. Opt. Express, 2
(4), 838
–849
(2011). http://dx.doi.org/10.1364/BOE.2.000838 BOEICL 2156-7085 Google Scholar
J. R. Mourantet al.,
“Measuring absorption coefficients in small volumes of highly scattering media: source-detector separations for which path lengths do not depend on scattering properties,”
Appl. Opt., 36
(22), 5655
–5661
(1997). http://dx.doi.org/10.1364/AO.36.005655 APOPAI 0003-6935 Google Scholar
D. Arifleret al.,
“Reflectance spectroscopy for diagnosis of epithelial precancer: model-based analysis of fiber-optic probe designs to resolve spectral information from epithelium and stroma,”
Appl. Opt., 44
(20), 4291
–4305
(2005). http://dx.doi.org/10.1364/AO.44.004291 APOPAI 0003-6935 Google Scholar
R. F. Bonneret al.,
“Model for photon migration in turbid biological media,”
J. Opt. Soc. Am. A, 4
(3), 423
–432
(1987). http://dx.doi.org/10.1364/JOSAA.4.000423 JOAOD6 0740-3232 Google Scholar
N. N. Mutyalet al.,
“A fiber optic probe design to measure depth-limited optical properties in-vivo with low-coherence enhanced backscattering (LEBS) spectroscopy,”
Opt. Express, 20
(18), 19643
–19657
(2012). http://dx.doi.org/10.1364/OE.20.019643 OPEXFF 1094-4087 Google Scholar
J. O’Dohertyet al.,
“Sub-epidermal imaging using polarized light spectroscopy for assessment of skin microcirculation,”
Skin Res. Technol., 13
(4), 472
–484
(2007). http://dx.doi.org/10.1111/srt.2007.13.issue-4 0909-752X Google Scholar
S. H. Tsenget al.,
“Determination of optical properties of superficial volumes of layered tissue phantoms,”
IEEE Trans. Biomed. Eng., 55
(1), 335
–339
(2008). http://dx.doi.org/10.1109/TBME.2007.910685 IEBEAX 0018-9294 Google Scholar
V. Turzhitskyet al.,
“Multiple scattering model for the penetration depth of low-coherence enhanced backscattering,”
J. Biomed. Opt., 16
(9), 097006
(2011). http://dx.doi.org/10.1117/1.3625402 JBOPFO 1083-3668 Google Scholar
S. C. Kanicket al.,
“Monte Carlo analysis of single fiber reflectance spectroscopy: photon path length and sampling depth,”
Phys. Med. Biol., 54
(22), 6991
–7008
(2009). http://dx.doi.org/10.1088/0031-9155/54/22/016 PHMBA7 0031-9155 Google Scholar
A. SassaroliS. Fantini,
“Comment on the modified Beer-Lambert law for scattering media,”
Phys. Med. Biol., 49
(14), N255
–N257
(2004). http://dx.doi.org/10.1088/0031-9155/49/14/N07 PHMBA7 0031-9155 Google Scholar
Y. Tsuchiya,
“Photon path distribution and optical responses of turbid media: theoretical analysis based on the microscopic Beer-Lambert law,”
Phys. Med. Biol., 46
(8), 2067
–2084
(2001). http://dx.doi.org/10.1088/0031-9155/46/8/303 PHMBA7 0031-9155 Google Scholar
S. Fantiniet al.,
“Non-invasive optical monitoring of the newborn piglet brain using continuous-wave and frequency-domain spectroscopy,”
Phys. Med. Biol., 44
(6), 1543
–1563
(1999). http://dx.doi.org/10.1088/0031-9155/44/6/308 PHMBA7 0031-9155 Google Scholar
C. Bonneryet al.,
“Changes in diffusion path length with old age in diffuse optical tomography,”
J. Biomed. Opt., 17
(5), 056002
(2012). http://dx.doi.org/10.1117/1.JBO.17.5.056002 JBOPFO 1083-3668 Google Scholar
L. C. SparlingG. H. Weiss,
“Some effects of beam thickness on photon migration in a turbid medium,”
J. Mod. Opt., 40
(5), 841
–859
(1993). http://dx.doi.org/10.1080/09500349314550861 JMOPEW 0950-0340 Google Scholar
A. J. Gomeset al.,
“Monte Carlo model of the penetration depth for polarization gating spectroscopy: influence of illumination-collection geometry and sample optical properties,”
Appl. Opt., 51
(20), 4627
–4637
(2012). http://dx.doi.org/10.1364/AO.51.004627 APOPAI 0003-6935 Google Scholar
J. D. RogersI. R. CapogluV. Backman,
“Nonscalar elastic light scattering from continuous random media in the Born approximation,”
Opt. Lett., 34
(12), 1891
–1893
(2009). http://dx.doi.org/10.1364/OL.34.001891 OPLEDP 0146-9592 Google Scholar
A. J. GomesV. Backman,
“Analytical light reflectance models for overlapping illumination and collection area geometries,”
Appl. Opt., 51
(33), 8013
–8021
(2012). http://dx.doi.org/10.1364/AO.51.008013 APOPAI 0003-6935 Google Scholar
A. AmelinkH. J. Sterenborg,
“Measurement of the local optical properties of turbid media by differential path-length spectroscopy,”
Appl. Opt., 43
(15), 3048
–3054
(2004). http://dx.doi.org/10.1364/AO.43.003048 APOPAI 0003-6935 Google Scholar
S. C. KanickH. J. SterenborgA. Amelink,
“Empirical model description of photon path length for differential path length spectroscopy: combined effect of scattering and absorption,”
J. Biomed. Opt., 13
(6), 064042
(2008). http://dx.doi.org/10.1117/1.3050424 JBOPFO 1083-3668 Google Scholar
A. N. Bashkatovet al.,
“Optical properties of human skin, subcutaneous and mucous tissues in the wavelength range from 400 to 2000 nm,”
J. Phys. D Appl. Phys., 38
(15), 2543
–2555
(2005). http://dx.doi.org/10.1088/0022-3727/38/15/004 JPAPBE 0022-3727 Google Scholar
S. T. Flocket al.,
“Monte Carlo modeling of light propagation in highly scattering tissue–I: model predictions and comparison with diffusion theory,”
IEEE Trans. Biomed. Eng., 36
(12), 1162
–1168
(1989). http://dx.doi.org/10.1109/TBME.1989.1173624 IEBEAX 0018-9294 Google Scholar
C. Holmeret al.,
“Optical properties of adenocarcinoma and squamous cell carcinoma of the gastroesophageal junction,”
J. Biomed. Opt., 12
(1), 014025
(2007). http://dx.doi.org/10.1117/1.2564793 JBOPFO 1083-3668 Google Scholar
A. Amelinket al.,
“Non-invasive measurement of the microvascular properties of non-dysplastic and dysplastic oral leukoplakias by use of optical spectroscopy,”
Oral Oncol., 47
(12), 1165
–1170
(2011). http://dx.doi.org/10.1016/j.oraloncology.2011.08.014 EJCCER 1368-8375 Google Scholar
R. Reifet al.,
“Optical method for real-time monitoring of drug concentrations facilitates the development of novel methods for drug delivery to brain tissue,”
J. Biomed. Opt., 12
(3), 034036
(2007). http://dx.doi.org/10.1117/1.2744025 JBOPFO 1083-3668 Google Scholar
R. Marchesiniet al.,
“Ex vivo optical properties of human colon tissue,”
Laser. Surg. Med., 15
(4), 351
–357
(1994). http://dx.doi.org/10.1002/(ISSN)1096-9101 LSMEDI 0196-8092 Google Scholar
R. L. van VeenW. VerkruysseH. J. Sterenborg,
“Diffuse-reflectance spectroscopy from 500 to 1060 nm by correction for inhomogeneously distributed absorbers,”
Opt. Lett., 27
(4), 246
–248
(2002). http://dx.doi.org/10.1364/OL.27.000246 OPLEDP 0146-9592 Google Scholar
G. ZoniosA. Dimou,
“Modeling diffuse reflectance from semi-infinite turbid media: application to the study of skin optical properties,”
Opt. Express, 14
(19), 8661
–8674
(2006). http://dx.doi.org/10.1364/OE.14.008661 OPEXFF 1094-4087 Google Scholar
G. Zonioset al.,
“Diffuse reflectance spectroscopy of human adenomatous colon polyps in vivo,”
Appl. Opt., 38
(31), 6628
–6637
(1999). http://dx.doi.org/10.1364/AO.38.006628 APOPAI 0003-6935 Google Scholar
S. C. KanickH. J. SterenborgA. Amelink,
“Empirical model of the photon path length for a single fiber reflectance spectroscopy device,”
Opt. Express, 17
(2), 860
–871
(2009). http://dx.doi.org/10.1364/OE.17.000860 OPEXFF 1094-4087 Google Scholar
|