We will describe our experimental characterization of a radiation-pressure-based optical power meter. We have tested it for modulated and CW laser powers up to 92 kW in the laboratory and up to 20 kW in an experimental laser welding booth. We will describe present accuracy, temporal response, sources of measurement uncertainty, and hurdles which must be overcome to have an accurate power meter capable of routine operation as a turning mirror within a laser delivery head. |
ACCESS THE FULL ARTICLE
No SPIE Account? Create one
CITATIONS
Cited by 4 scholarly publications and 2 patents.
Mirrors
Power meters
Calibration
Laser welding
Photodiodes
High power lasers
Reflectivity