15 March 2016 Photoacoustic imaging of clinical metal needle by a LED light source integrated transducer
Toshitaka Agano, Naoto Sato, Hitoshi Nakatsuka, Kazuo Kitagawa, Takamitsu Hanaoka, Koji Morisono, Yusuke Shigeta, Chizuyo Tanaka
Author Affiliations +
Abstract
We have achieved penetration depth of 30mm by photoacoustic imaging system using LED light source integrated transducer to image a clinical metal needle inserted into a tissue mimicking phantom. We developed the transducer that integrated near-infrared LED array light source, which was connected to a photoacoustic imaging system which drove LED array light source and controlled photoacoustic data acquisition process. Conventionally solid-state laser has been used as the light source for photoacoustic imaging system. Because LED is diffused light source, laser safety glasses is not necessary, also inflexible fibers are not used to guide light close to a transducer, and we integrated LED light source inside the transducer, which became compact and practical size for conventional ultrasound equipment users. We made LED light source unit as detachable to the transducer easily, so wave-length of light can be selectable by changing the LED light source unit.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Toshitaka Agano, Naoto Sato, Hitoshi Nakatsuka, Kazuo Kitagawa, Takamitsu Hanaoka, Koji Morisono, Yusuke Shigeta, and Chizuyo Tanaka "Photoacoustic imaging of clinical metal needle by a LED light source integrated transducer", Proc. SPIE 9708, Photons Plus Ultrasound: Imaging and Sensing 2016, 97083U (15 March 2016); https://doi.org/10.1117/12.2212685
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Light sources

Transducers

Light emitting diodes

Imaging systems

Photoacoustic imaging

Ultrasonography

Data acquisition

Back to Top