Paper
18 March 2016 Validation of MTF measurement for CBCT system using Monte Carlo simulations
Author Affiliations +
Abstract
To evaluate the spatial resolution performance of cone beam computed tomography (CBCT) system, accurate measurement of the modulation transfer function (MTF) is required. This accuracy depends on the MTF measurement method and CBCT reconstruction algorithms. In this work, the accuracy of MTF measurement of CBCT system using wire phantom is validated by Monte Carlo simulation. A Monte Carlo simulation software tool BEAMnrc/EGSnrc was employed to model X-ray radiation beams and transport. Tungsten wires were simulated with different diameters and radial distances from the axis of rotation. We adopted filtered back projection technique to reconstruct images from 360° acquisition. The MTFs for four reconstruction kernels were measured from corresponding reconstructed wire images, while the ram-lak kernel increased the MTF relative to the cosine, hamming and hann kernel. The results demonstrated that the MTF degraded radially from the axis of rotation. This study suggested that an increase in the MTF for the CBCT system is possible by optimizing scanning settings and reconstruction parameters.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ting Hao, Feng Gao, Huijuan Zhao, and Zhongxing Zhou "Validation of MTF measurement for CBCT system using Monte Carlo simulations", Proc. SPIE 9700, Design and Quality for Biomedical Technologies IX, 970017 (18 March 2016); https://doi.org/10.1117/12.2210746
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Modulation transfer functions

Monte Carlo methods

Photons

Tungsten

Interference (communication)

Spatial resolution

Electrons

Back to Top