Paper
23 December 1976 Correction Factors For The Longitudinal And Transverse Dimensions Of X-Ray Tube Focal Spots
Satish C. Prasad
Author Affiliations +
Proceedings Volume 0096, Application of Optical Instrumentation in Medicine V; (1976) https://doi.org/10.1117/12.965397
Event: Application of Optical Instrumentation in Medicine V, 1976, Washington, D.C., United States
Abstract
Correction factors are computed for pinhole images of x-ray tube focal spots to be applied along the anode-cathode dimension and the dimension perpendicular to the anode-cathode axis. These correction factors are based upon comparison of MTF's of focal spots with uniform intensity distribution to the MTF's of focal spots with single and double peaked intensity distributions.
© (1976) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Satish C. Prasad "Correction Factors For The Longitudinal And Transverse Dimensions Of X-Ray Tube Focal Spots", Proc. SPIE 0096, Application of Optical Instrumentation in Medicine V, (23 December 1976); https://doi.org/10.1117/12.965397
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Modulation transfer functions

X-rays

Spatial frequencies

Medicine

Optical instrument design

Optical transfer functions

Radiography

RELATED CONTENT

Equivalent Blur Of Intensifying Screens
Proceedings of SPIE (December 27 1977)
Focus Measurements Using The Star Test Pattern
Proceedings of SPIE (December 23 1976)
Quantum Noise-Limited Images In Screen Film Systems
Proceedings of SPIE (December 29 1982)
Optimizing Computed Tomography (CT) Scanner Geometry
Proceedings of SPIE (July 06 1979)
Digital Radiography
Proceedings of SPIE (August 18 1980)

Back to Top