Paper
24 February 2012 SinoCor: motion correction in SPECT
Debasis Mitra, Daniel Eiland, Mahmoud Abdallah, Rostyslav Bouthcko, Grant T. Gullberg, Norberto Schechtmann
Author Affiliations +
Abstract
Motion is a serious artifact in Cardiac nuclear imaging because the scanning operation takes a long time. Since reconstruction algorithms assume consistent or stationary data the quality of resulting image is affected by motion, sometimes significantly. Even after adoption of the gold standard MoCo(R) algorithm from Cedars-Sinai by most vendors, heart motion remains a significant challenge. Also, any serious study in quantitative analysis necessitates correction for motion artifacts. It is generally recognized that human eye is a very sensitive tool for detecting motion. However, two reasons prevent such manual correction: (1) it is costly in terms of specialist's time, and (2) no such tool for manual correction is available currently. Previously, at SPIE-MIC'11, we presented a simple tool (SinoCor) that allows sinograms to be corrected manually or automatically. SinoCor performs correction of sinograms containing inter-frame patient or respiratory motions using rigid-body dynamics. The software is capable of detecting the patient motion and estimating the body-motion vector using scanning geometry parameters. SinoCor applies appropriate geometrical correction to all the frames subsequent to the frame when the movement has occurred in a manual or automated mode. For respiratory motion, it is capable of automatically smoothing small oscillatory (frame-wise local) movements. Lower order image moments are used to represent a frame and the required rigid body movement compensation is computed accordingly. Our current focus is on enhancement of SinoCor with the capability to automatically detect and compensate for intra-frame motion that causes motion blur on the respective frame. Intra-frame movements are expected in both patient and respiratory motions. For a controlled study we also have developed a motion simulator. A stable version of SinoCor is available under license from Lawrence Berkeley National Laboratory.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Debasis Mitra, Daniel Eiland, Mahmoud Abdallah, Rostyslav Bouthcko, Grant T. Gullberg, and Norberto Schechtmann "SinoCor: motion correction in SPECT", Proc. SPIE 8314, Medical Imaging 2012: Image Processing, 831452 (24 February 2012); https://doi.org/10.1117/12.911101
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Head

Single photon emission computed tomography

Motion detection

Reconstruction algorithms

Sensors

Detection and tracking algorithms

Heart

Back to Top