Paper
1 February 2012 Single-shot full complex spectrum spectrometer-based OCT with a single-line photodiode array
Author Affiliations +
Abstract
An efficient technique of simultaneous obtaining of quadrature spectral components of interference signal in spectrometer-based OCT using a single-line linear photodiode array is proposed. The components are obtained in air-spaced non-polarization interferometer by partition of reference beam onto two parts and using an achromatic phase shifter. Several setups are described and compared.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Grigory V. Gelikonov, Valentin M. Gelikonov, Aleksandr A. Moiseev, and Pavel A. Shilyagin "Single-shot full complex spectrum spectrometer-based OCT with a single-line photodiode array", Proc. SPIE 8213, Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XVI, 82133L (1 February 2012); https://doi.org/10.1117/12.909003