Paper
16 March 2011 Design and development of MR-compatible SPECT systems for simultaneous SPECT-MR imaging of small animals
Benjamin M. W. Tsui, James W. Hugg, Jingyan Xu, Si Chen, Dirk Meier, William Edelstein, Abdel El-Sharkawy, Douglas J. Wagenaar, Bradley E. Patt
Author Affiliations +
Abstract
We describe a continuing design and development of MR-compatible SPECT systems for simultaneous SPECT-MR imaging of small animals. A first generation prototype SPECT system was designed and constructed to fit inside a MRI system with a gradient bore inner diameter of 12 cm. It consists of 3 angularly offset rings of 8 detectors (1"x1", 16x16 pixels MR-compatible solid-state CZT). A matching 24-pinhole collimator sleeve, made of a tungsten-compound, provides projections from a common FOV of ~25 mm. A birdcage RF coil for MRI data acquisition surrounds the collimator. The SPECT system was tested inside a clinical 3T MRI system. Minimal interference was observed on the simultaneously acquired SPECT and MR images. We developed a sparse-view image reconstruction method based on accurate modeling of the point response function (PRF) of each of the 24 pinholes to provide artifact-free SPECT images. The stationary SPECT system provides relatively low resolution of 3-5 mm but high geometric efficiency of 0.5- 1.2% for fast dynamic acquisition, demonstrated in a SPECT renal kinetics study using Tc-99m DTPA. Based on these results, a second generation prototype MR-compatible SPECT system with an outer diameter of 20 cm that fits inside a mid-sized preclinical MRI system is being developed. It consists of 5 rings of 19 CZT detectors. The larger ring diameter allows the use of optimized multi-pinhole collimator designs, such as high system resolution up to ~1 mm, high geometric efficiency, or lower system resolution without collimator rotation. The anticipated performance of the new system is supported by simulation data.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Benjamin M. W. Tsui, James W. Hugg, Jingyan Xu, Si Chen, Dirk Meier, William Edelstein, Abdel El-Sharkawy, Douglas J. Wagenaar, and Bradley E. Patt "Design and development of MR-compatible SPECT systems for simultaneous SPECT-MR imaging of small animals", Proc. SPIE 7961, Medical Imaging 2011: Physics of Medical Imaging, 79611Y (16 March 2011); https://doi.org/10.1117/12.878903
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Single photon emission computed tomography

Prototyping

Magnetic resonance imaging

Collimators

Sensors

Imaging systems

Image restoration

Back to Top