18 February 2011 Calibration of ultrasonic sensors using optoacoustics
Author Affiliations +
Abstract
Ultrasonic detectors are commonly calibrated by finding their response to incident plane waves. However, in optoacoustics, the response to broadband point sources is required. To induce such sources using the optoacoustic effect, the illuminated object's dimensions must be smaller than the resolution achievable by the optoacoustic system. The main difficulty in such measurements is that the magnitude of the field emitted by such sources is proportional to their dimensions, and thus may be weak compared to parasitic sources in the setup. In this work we experimentally demonstrate two methods for calibrating acoustic detectors. In both methods, acoustic sources are optoacoustically induced in large optically absorbing slabs. Despite the large dimensions of the illuminated objects, the geometry used yields wide-band acoustic fields, which are perceived by the detectors as originating from point sources.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Amir Rosenthal, Vasilis Ntziachristos, and Daniel Razansky "Calibration of ultrasonic sensors using optoacoustics", Proc. SPIE 7899, Photons Plus Ultrasound: Imaging and Sensing 2011, 78992L (18 February 2011); https://doi.org/10.1117/12.873858