Paper
16 February 2011 Optoacoustic temperature determination and automatic coagulation control in rabbits
Kerstin Schlott, Stefan Koinzer, Lars Ptaszynski, Susanne Luft, Alex Baade, Marco Bever, Johann Roider M.D., Reginald Birngruber, Ralf Brinkmann
Author Affiliations +
Proceedings Volume 7885, Ophthalmic Technologies XXI; 78850T (2011) https://doi.org/10.1117/12.875104
Event: SPIE BiOS, 2011, San Francisco, California, United States
Abstract
Retinal laser photocoagulation is an established treatment method for many retinal diseases like macula edema or diabetic retinopathy. The selection of the laser parameters is so far based on post treatment evaluation of the lesion size and strength. Due to local pigment variations in the fundus and individual transmission the same laser parameters often lead to an overtreatment. Optoacoustic allows a non invasive monitoring of the retinal temperature increase during retinal laser irradiation by measuring the temperature dependent pressure amplitudes, which are induced by short probe laser pulses. A 75 ns/ 523 nm Nd:YLF was used as a probe laser at a repetition rate of 1 kHz, and a cw / 532 nm treatment laser for heating. A contact lens was modified with a ring-shaped ultrasonic transducer to detect the pressure waves at the cornea. Temperatures were collected for irradiations leading to soft or invisible lesions. Based on this data the threshold for denaturation was found. By analyzing the initial temperature increase, the further temperature development during irradiation could be predicted. An algorithm was found to calculate the irradiation time, which is needed for a soft lesion formation, from the temperature curve. By this it was possible to provide a real-time dosimetry by automatically switching off the treatment laser after the calculated irradiation time. Automatically controlled coagulations appear softer and more uniformly.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Kerstin Schlott, Stefan Koinzer, Lars Ptaszynski, Susanne Luft, Alex Baade, Marco Bever, Johann Roider M.D., Reginald Birngruber, and Ralf Brinkmann "Optoacoustic temperature determination and automatic coagulation control in rabbits", Proc. SPIE 7885, Ophthalmic Technologies XXI, 78850T (16 February 2011); https://doi.org/10.1117/12.875104
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Temperature metrology

Automatic control

Tissues

Laser beam diagnostics

Laser coagulation

Eye

In vivo imaging

Back to Top