Paper
22 March 2010 The dependence of image quality on the number of high and low kVp projections in dual energy CT using the prior image constrained compressed sensing (PICCS) algorithm
Timothy P. Szczykutowicz, Jiang Hsieh, Guang-Hong Chen
Author Affiliations +
Abstract
Dual energy CT using a fast kVp switching technique and the standard filtered back projection (FBP) image reconstruction method has recently been studied. With conventional FBP methods, high slew rates are required for acceptable image reconstruction with high image quality. However, high slew rates also require hardware changes to enable data acquisition. In this work, we aim at studying the necessary slew rate for dual energy CT imaging provided that the PICCS algorithm is used for image reconstruction. The results demonstrate that a slew rate of 7.5 kV / view (assuming 2,000 views were collected over 360o with a 60 kVp energy separation) was sufficient for dual energy imaging using PICCS.
© (2010) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Timothy P. Szczykutowicz, Jiang Hsieh, and Guang-Hong Chen "The dependence of image quality on the number of high and low kVp projections in dual energy CT using the prior image constrained compressed sensing (PICCS) algorithm", Proc. SPIE 7622, Medical Imaging 2010: Physics of Medical Imaging, 762221 (22 March 2010); https://doi.org/10.1117/12.844471
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications and 3 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Reconstruction algorithms

Dual energy imaging

Sensors

Data acquisition

X-rays

Bone

Image quality

Back to Top