Paper
4 September 2008 Evaluation of TlBr detectors with Tl electrodes
Keitaro Hitomi, Yohei Kikuchi, Tadayoshi Shoji, Keizo Ishii
Author Affiliations +
Abstract
Long-term stability of TlBr detectors with Tl electrodes was evaluated in this study. Tl/TlBr/Tl detectors were fabricated and tested with a charge sensitive preamplifier and a spectroscopy amplifier and a gated integrator. Significant improvement of energy resolutions was realized by applying the gated integrator to the TlBr detectors. The full-energy peak positions, counts and resolutions for 662 keV gamma-rays from 137Cs were monitored during the course of measurements (10 days). By changing the polarity of the applied bias voltages every 24 hours, the Tl/TlBr/Tl detectors exhibited stable spectroscopic performance for totally over 240 hours.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Keitaro Hitomi, Yohei Kikuchi, Tadayoshi Shoji, and Keizo Ishii "Evaluation of TlBr detectors with Tl electrodes", Proc. SPIE 7079, Hard X-Ray, Gamma-Ray, and Neutron Detector Physics X, 70790J (4 September 2008); https://doi.org/10.1117/12.794381
Lens.org Logo
CITATIONS
Cited by 14 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Electrodes

Thallium

Gamma radiation

Amplifiers

Spectroscopy

Polarization

Back to Top