3 March 2007 Evaluation of accuracy and workflow between different alignment techniques for correction of CTAC and PET misalignment in cardiac PET-CT imaging
Elizabeth B. Philps, Sarah J. Aivano
Author Affiliations +
Abstract
Small errors in the alignment between CT Attenuation Correction (CTAC) images and Positron Emission Tomography (PET) acquisitions can result in significant changes in PET attenuation corrected images. Misalignment due to respiratory or cardiac motion can produce mismatch between the PET and CTAC acquisitions. This contributes to artifactual hypoperfusion defects that are interpretable as myocardial ischemia or infarct. Correction for the misalignment between the PET and CTAC images can eliminate these false positive artifacts. Two methods for correcting for this respiratory and cardiac misalignment were compared. The first was an existing procedure, the manual-shift method, using point-to-point, in-plane, two-dimensional (2D) measurements of the shifts in axial, sagittal, and coronal planes. A new PET image reconstruction using the corrected attenuation map shifted by the 2D measurements was then performed. In the second method, the Interactive ACQC method, visual alignment was performed between the left ventricle boundaries on fused images and automated calculation of necessary rigid three-dimensional (3D) alignment parameters was performed. A new PET image reconstruction was then performed with an attenuation map shifted by the prescribed alignment parameters. The two methods were compared for accuracy and workflow efficiency using five cardiac PET/CT cases, scanned on GE Discovery VCT and Discovery ST systems. Alignment measurements using the visual alignment process (the interactive ACQC method) improved productivity by over five minutes, on average. The results show that the interactive ACQC procedure yields similar results to those of the point-to-point procedure while providing improved workflow for cardiac PET attenuation correction quality control.
© (2007) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Elizabeth B. Philps and Sarah J. Aivano "Evaluation of accuracy and workflow between different alignment techniques for correction of CTAC and PET misalignment in cardiac PET-CT imaging", Proc. SPIE 6512, Medical Imaging 2007: Image Processing, 65124J (3 March 2007); https://doi.org/10.1117/12.713599
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Positron emission tomography

Signal attenuation

Visualization

Image restoration

Computed tomography

Medicine

3D image processing

Back to Top