4 February 2004 X-ray calibration of the GOES-N/O solar X-ray imagers
Paul R. Catura, Brennan L. Gantner, Bruce K. Jurcevich, Claude K. Kam, James R. Lemen, Mons D. Morrison, Lawrence Shing, Richard C. Catura
Author Affiliations +
Abstract
The X-ray calibration of the GOES Solar X-ray Imagers (SXI) was accomplished at the component level and at the instrument level. The CCD and thin film filters were characterized in the facilities at the Lockheed Martin Solar and Astrophysics Laboratory. The grazing incidence telescope mirrors and the completed instruments were calibrated at the X-ray Calibration Facility (XRCF) at NASA’s Marshall Space Flight Center. The XRCF consists of an X-ray source at one end of a 518 m long evacuated tube and a large vacuum chamber at the opposite end. The X-ray source has a variety of interchangeable anodes and filters to provide filtered characteristic K- and L-shell line emission in the range from 0.109 to 8.6 keV. The absolute Photometric calibration of the SXI telescopes is very important for analysis and interpretation of their data, and to monitor the long-term solar variations at X-ray wavelengths. This paper describes the results of these calibrations.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Paul R. Catura, Brennan L. Gantner, Bruce K. Jurcevich, Claude K. Kam, James R. Lemen, Mons D. Morrison, Lawrence Shing, and Richard C. Catura "X-ray calibration of the GOES-N/O solar X-ray imagers", Proc. SPIE 5171, Telescopes and Instrumentation for Solar Astrophysics, (4 February 2004); https://doi.org/10.1117/12.507639
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

Charge-coupled devices

Mirrors

Calibration

X-ray imaging

Beryllium

Optical filters

RELATED CONTENT

Optics for the imaging x-ray polarimetry explorer
Proceedings of SPIE (October 03 2017)
Imaging performance of the XMM-Newton x-ray telescopes
Proceedings of SPIE (July 18 2000)
Calibration of Astro D telescope with an x ray pencil...
Proceedings of SPIE (February 01 1994)
JET X instrument for the USSR Spectrum RG mission ...
Proceedings of SPIE (January 01 1992)
The solar x-ray imager for GOES
Proceedings of SPIE (February 04 2004)
Use of x ray CCDs in the calibration of the...
Proceedings of SPIE (October 31 1996)

Back to Top