Abstract
In this work a combination of an ionization chamber with one-dimensional spatial resolution and a MicroCAT structure will be presented. The combination between gas gain operations and integrating front-end electronics yields a dynamic range as high as eight to nine orders of magnitude. Therefore this device is well suitable for medical imaging or applications such as small angle x-ray scattering, where the requirements on the dynamic of the detector are exceptional high. Basically the described detector is an ionization chamber adapted to fan beam geometry with an active area of 192 mm and a pitch of the anode strips of 150 micrometer. In the vertical direction beams as high as 10 mm can be accepted. Every read-out strip is connected to an analogue integrating electronics channel realized in a custom made VLSI chip. A MicroCAT structure utilized as a shielding grid enables frame rates as high as 10kHz. The high dynamic range observed stems from the fact that the MicroCAT enables active electron amplification in the gas. Thus a single photon resolution can be obtained for low photon fluxes even with the integrating electronics. The specialty of this device is that for each photon flux the gas amplification can be adjusted in such a fashion that the maximum DQE value is achieved.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ralf-Hendrik Menk and Francesco Voltolina "Imaging with high dynamic using an ionization chamber", Proc. SPIE 5030, Medical Imaging 2003: Physics of Medical Imaging, (5 June 2003); https://doi.org/10.1117/12.480239
Lens.org Logo
CITATIONS
Cited by 3 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Electronics

Ionization

Single photon

X-rays

Spatial resolution

Modulation transfer functions

RELATED CONTENT

The wide field monitor onboard the eXTP mission
Proceedings of SPIE (July 18 2018)
Ultrahigh-resolution CT and DR scanner
Proceedings of SPIE (May 28 1999)
Charge cloud asymmetry in detectors with biased MCPs
Proceedings of SPIE (January 25 2002)

Back to Top