24 April 2002 CT perfusion: comparison of gamma-variate fit and deconvolution
Zhong Min Lin, Scott Pohlman, A. J. Cook II, Shalabh Chandra
Author Affiliations +
Abstract
The purpose of this paper is to present results of a study comparing CT brain perfusion image calculated using the maximum slope and deconvolution method on data from 32 patients. The 32 patients were organized into two groups. One group was obtained using a 4cc/sec contrast injection rate and 10 sec injection period; the other group was obtained using a 6cc/sec injection rate and 7 sec injection period. All clinical data were analyzed using both the maximum slope and deconvolution methods. Perfusion maps computed from the two methods were reviewed by radiologists. The contrast enhanced CT data were noisy, especially in the white matter area. Our results showed that, for both methods, perfusion maps from 4cc/sec injection were noisier than those using a 6cc/sec injection. However, both 4cc/sec and 6cc/sec produced useful diagnostic images. Our qualitative side-by-side studies also showed that perfusion images from the maximum slope and deconvolution are both clinically useful and substantially equivalent.
© (2002) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Zhong Min Lin, Scott Pohlman, A. J. Cook II, and Shalabh Chandra "CT perfusion: comparison of gamma-variate fit and deconvolution", Proc. SPIE 4683, Medical Imaging 2002: Physiology and Function from Multidimensional Images, (24 April 2002); https://doi.org/10.1117/12.463572
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Deconvolution

Tissues

Computed tomography

Image processing

Arteries

Blood

Blood circulation

Back to Top