Paper
6 June 2000 Determination of the envelope function (maximum velocity curve) in Doppler ultrasound flow velocity diagrams
Juerg Tschirren, Ronald M. Lauer, Milan Sonka
Author Affiliations +
Abstract
This paper presents a new approach for the evaluation of Doppler flow velocity diagrams, obtained during brachial artery flow mediated dilatation (FMD) studies. The velocity diagrams are stored as image sequences on VCR tape. For this reason standard signal processing methods can not be used. A method for determination of blood velocity envelopes from image data is reported that uses Doppler-data specific heuristic to achieve high accuracy and robustness. The approach was tested in 40 Doppler blood flow images. Comparisons with manually defined independent standards demonstrated a very good correlation in determined peak velocity values (r equals 0.993) and flow envelope areas (r equals 0.996). The method is currently tested in a large volume clinical study.
© (2000) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Juerg Tschirren, Ronald M. Lauer, and Milan Sonka "Determination of the envelope function (maximum velocity curve) in Doppler ultrasound flow velocity diagrams", Proc. SPIE 3979, Medical Imaging 2000: Image Processing, (6 June 2000); https://doi.org/10.1117/12.387664
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Doppler effect

Sensors

Ultrasonography

Nonlinear filtering

Blood

Image filtering

Blood circulation