Paper
1 June 1998 Accelerometer shock calibration using differential grating laser interferometer from 100 to 1,000,000 m/s2
Da-Hong Lian, Xiao-Mei Xu, Xin-Liang Li
Author Affiliations +
Proceedings Volume 3411, Third International Conference on Vibration Measurements by Laser Techniques: Advances and Applications; (1998) https://doi.org/10.1117/12.307706
Event: Third International Conference on Vibration Measurements by Laser Techniques: Advances and Applications, 1998, Ancona, Italy
Abstract
An accelerometer shock calibration system using grating laser interference technique is described in this paper. The shock acceleration value is directly and absolutely acquired from the base quantity and unit. The principle of the differential gating laser interferometer shock calibration device are introduced. Experimental results are given. The uncertainty of measurement is evaluated. The calibration uncertainty is in the range of 100 m/s2 to 100,000m/s2 is less than 2 percent and in the range of 100,000m/s2 to 1,000,000m/s2 less than 3 percent.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Da-Hong Lian, Xiao-Mei Xu, and Xin-Liang Li "Accelerometer shock calibration using differential grating laser interferometer from 100 to 1,000,000 m/s2", Proc. SPIE 3411, Third International Conference on Vibration Measurements by Laser Techniques: Advances and Applications, (1 June 1998); https://doi.org/10.1117/12.307706
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Doppler effect

Diffraction gratings

Interferometers

Data processing

Diffraction

Oscilloscopes

Back to Top