Paper
8 June 1998 Measurement of pitch and width samples with the NIST calibrated atomic force microscope
Ronald G. Dixson, Rainer G.J. Koening, Theodore V. Vorburger, Joseph Fu, Vincent Wen-Chieh Tsai
Author Affiliations +
Abstract
Because atomic force microscopes (AFMs) are capable of generating three dimensional images with nanometer level resolution, these instruments are being increasingly used in many industries as tools for dimensional metrology at sub- micrometer length scales. To achieve high accuracy, the scales of an AFM must be calibrated. Presently available standards for this purpose are commonly calibrated using stylus instruments and optical techniques. We have developed the calibrated AFM (C-AFM) in order to calibrate pitch and height standards using an AFM. Our instrument has metrology traceable to the wavelength of light for all three axes. This is accomplished through the integration of a flexure x-y translation stage, heterodyne laser interferometers, and a digital-signal-processor based closed-loop feedback system to control the x-y scan motion. The z-axis translation is accomplished using a piezoelectric actuator with an integrated capacitance sensor, which is calibrated using a heterodyne laser interferometer. When fully developed, this instrument will be a calibration tool for pitch and height standards for scanning probe microscopes. We have recently completed a reevaluation of the titling motions of the C-AFM scanner. This has allowed a refinement in our estimate of the Abbe error contribution to our measurement uncertainty. Our most recent pitch measurements are consistent with this new estimate and thus support our refined uncertainty budget. We have recently completed measurements of pitch on several samples, including both grid type and linear scale patterns, for an industrial user. We are also working toward the development of linewidth standards through the comparison of C-AFM width measurements with values obtained from other methods, including an electrical resistance techniques. In this paper, we will describe the current status of the C-AFM, discuss the use of the instrument for measurements of pitch and width, and describe our plans for future measurements.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ronald G. Dixson, Rainer G.J. Koening, Theodore V. Vorburger, Joseph Fu, and Vincent Wen-Chieh Tsai "Measurement of pitch and width samples with the NIST calibrated atomic force microscope", Proc. SPIE 3332, Metrology, Inspection, and Process Control for Microlithography XII, (8 June 1998); https://doi.org/10.1117/12.308751
Lens.org Logo
CITATIONS
Cited by 10 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Error analysis

Interferometers

Metrology

Standards development

Atomic force microscopy

Atomic force microscope

RELATED CONTENT


Back to Top