11 April 1996 Receptor modeling for use in computerized optimization of imaging systems
Michael P. Sandborg, Gudrun Alm Carlsson, David R. Dance, Jan Persliden
Author Affiliations +
Abstract
Inappropriate or poorly designed x-ray imaging systems can lead to inadequate image quality and/or excessive patient absorbed dose. To study this dilemma, we have developed a computer program, based on Monte Carlo methods, to model the imaging chain and to optimize the design of individual imaging system components. The Monte Carlo method is used to simulate the transport of x-ray photons through the patient and anti-scatter device and into the image receptor. Image quality is calculated in terms of contrast and signal-to-noise ratio (SNR) and patient risk in terms of mean absorbed dose. The model of the radiographic screen includes the statistics of x-ray to light conversion and diffusion of light photons and the spatial distribution of absorbed x rays. The optimal system was defined as that which, for a constant measure of image quality, results in the lowest mean absorbed dose in the patient. The results of the model agree with measured data. Examples are given for different measures of reducing the absorbed dose for constant SNR in a simulated lumbar spine examination. At the optimal tube potential, the dose reductions associated with using a thicker screen, a fiber-interspaced grid and an additional Cu-filtration are 53%, 42%, and 31%, respectively.
© (1996) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Michael P. Sandborg, Gudrun Alm Carlsson, David R. Dance, and Jan Persliden "Receptor modeling for use in computerized optimization of imaging systems", Proc. SPIE 2708, Medical Imaging 1996: Physics of Medical Imaging, (11 April 1996); https://doi.org/10.1117/12.237818
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Receptors

Photons

Modulation transfer functions

Signal to noise ratio

Imaging systems

X-rays

Image quality

Back to Top