Paper
24 June 1993 Simple device for objective measurements of haze following excimer laser ablation of cornea
Russell L. McCally, Bernard F. Hochheimer, Wallace Chamon, Dimitri T. Azar M.D.
Author Affiliations +
Proceedings Volume 1877, Ophthalmic Technologies III; (1993) https://doi.org/10.1117/12.147535
Event: OE/LASE'93: Optics, Electro-Optics, and Laser Applications in Scienceand Engineering, 1993, Los Angeles, CA, United States
Abstract
We have developed a simple instrument for making objective measurements of haze that develops following excimer laser ablation of the cornea. It consists of an appropriately modified slit-lamp microscope, with a fiber optic pickup, a filter system for wavelength selection, and a photomultiplier detector. The scattered intensity at 120 degree(s) from the forward direction is determined. Preliminary tests were made by measuring the haze following a deep photorefractive ablation on a rabbit cornea under conditions which ensured that rather severe haze would develop. The VISX Model 20/20 laser system was set to produce a 6.0 mm diameter, -15 D correction, with a central depth of 236 micrometers . Measurements were made on the normal cornea prior to ablation and at various times up to 114 days post-ablation and are compared to slit-lamp photographs. Scattering peaked two weeks post-ablation at a value approximately 40X that of the normal (unablated) cornea and gradually decreased to approximately 11X the normal value at 114 days.
© (1993) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Russell L. McCally, Bernard F. Hochheimer, Wallace Chamon, and Dimitri T. Azar M.D. "Simple device for objective measurements of haze following excimer laser ablation of cornea", Proc. SPIE 1877, Ophthalmic Technologies III, (24 June 1993); https://doi.org/10.1117/12.147535
Lens.org Logo
CITATIONS
Cited by 15 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Cornea

Air contamination

Scattering

Laser ablation

Excimer lasers

Scatter measurement

Visualization

Back to Top