1 June 1992 Signal fading, erasure, and rescan in storage phosphor imaging
Chris C. Shaw, John M. Herron, David Gur
Author Affiliations +
Abstract
This paper addresses three problems in storage phosphor imaging: natural fading of latent image signals, proper erasure of the exposed plates, and re-scanning for a second readout. Signal, and signal-to-noise ratios were measured as a function of time, erasure power/time, or number of pre-scans to study these problems. The latent image signals were found to decay very rapidly during the first several minutes and stabilize after several hours. The fading effect results in a variable signal gain (signal per unit exposure) which may affect the system calibration and quantitative use of the image data. Complete erasure of the latent image signals is necessary to ensure that no residual image signals are present when the plate is exposed again. It was found that plates used in high exposure applications (GI, therapeutic imaging) may require an excessively long erasure time to prepare them for use in low exposure applications (chest imaging). Although the latent image is partially erased during the readout process, it may sometimes become necessary to re-scan the plate for a second or third readout. It was found that because a large number of energy traps are generated for each x-ray photon, a significant portion of the x-ray information remains intact for reuse after the first or second scans. Measurement of the signals and signal-to-noise ratios are presented to demonstrate and discuss the aforementioned problems or effects.
© (1992) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Chris C. Shaw, John M. Herron, and David Gur "Signal fading, erasure, and rescan in storage phosphor imaging", Proc. SPIE 1651, Medical Imaging VI: Instrumentation, (1 June 1992); https://doi.org/10.1117/12.59393
Lens.org Logo
CITATIONS
Cited by 10 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

Radiography

Image processing

Medical imaging

Signal to noise ratio

X-ray imaging

Image storage

Back to Top