1.INTRODUCTIONModel observers are used to mimic human observers. However, they require the knowledge of image mean and covariance, which is difficult to achieve in reality. Analytical projection to image (Prj2Img) covariance estimation methods are proposed to estimate image covariance from a single projection for commonly used iterative CT reconstruction algorithms. Iteration-based and fixed-point methods are two ways to analytically estimate covariance from projection to image. For iteration-based methods, the covariance estimation is updated with iteration formula1. For fixed-point methods, the covariance estimation is derived from the converged point of an objective function2. Li3 et.al. study the difference and consistence of these two methods. Meanwhile, analytical Prj2Img methods for iterative reconstruction with a quadratic regularization is studied by Schmitt4 et.al., while that with a non-quadratic regularization is studied by Sánchez5. Although analytical Prj2Img methods can yield reasonable covariance estimations, they are computationally expensive for large-dimensional images (pixels > 128×128). Usually, images with larger size, e.g., 512×512, are required for practical use. Fessler6 et.al. give a fast variance estimation method for the quadratic penalized weighted least square (WLS) algorithm. Fast covariance estimation methods have not been studied yet. In fact, the covariance used for model observers is the covariance of the channel response7 that has a much smaller dimension than its image covariance matrix. Therefore, we target on estimating the covariance of low-dimensional channel response from high-dimensional projection. The proposed projection to channel response (Prj2CR) covariance estimation method is derived from the Prj2Img method. We apply the proposed method to a widely used channelized Hotelling observer (CHO) with large-dimensional image inputs for validation. 2.METHODSIn this section, we briefly introduce the analytical Prj2Img covariance estimation method for non-quadratic penalized WLS studied in our previous work8, then describe the CHO, and finally deduce the proposed analytical Prj2CR covariance estimation method in detail. 2.1Analytical Prj2Img covariance estimation methodThe cost function Φ(·) of a penalized WLS reconstruction can be expressed as: where is the linear attenuation image reconstructed from its projection p ∈ ℝM×l. H ∈ ℝM×N denotes the system matrix and W ∈ ℝM×M the noise model with Wmm = exp(−pm). R(·) is the penalty function and α the penalty parameter. The fixed-point method makes use of convergence condition: Plugging Eq. (1) into Eq. (2) and finding its covariance, we have: where A ∈ ℝN×N is a coefficient matrix approximates : Our previous work proposed the linear based method (LAM) with , thereby estimates the covariance as: here, is the expectation of . We adopt the WLS penalized with the total variance (TV-WLS) as well as qth generalized Gaussian Markov random field (qGGMRF-WLS) as two representatives in this work. 2.2Channelized Hotelling observer (CHO)A CHO is calculated as: where λ denotes the decision variable and ω the template of CHO. The image of channel response is denoted by a subscript c : with V ∈ ℝC×N being the channel matrix consists of C channel profiles and C ≪ N. Meanwhile, the and in template ω represent the mean of signal present and signal absent images respectively. The intra-class channel scatter matrix Sc = (Ksc + Kbc) / 2, where the channel covariance Ksc = VKsVT and Kbc = VKbVT with K = being the image covariance. For a given threshold t, if the decision variable satisfies λ > t, we consider the target image to contain the signal; otherwise, we consider it not. Meanwhile, signal-to-noise ratio9 (SNR) is adopted to measure the performance of a CHO: 2.3Analytical Prj2CR covariance estimation methodNote that the analytical Prj2Img method in Eq. (5) involves a matrix inverse operation, which leads to high computational cost when the dimension of the reconstructed image is high. As described in section 2.2, the CHO only requires a small-dimensional covariance of channel response, where C ≪ N. Therefore, we construct a relationship between the covariance of channel response Kc = VKVT and the covariance of corresponding projection Cov(p) : where the transition matrix X* = V(VTV)–1(VTV)–1VT. Note that matrix VTV is not full rank, and hence it is irreversible. Hypothesis there exists an invertible transition matrix X ∈ ℝC×C that satisfies Eq.(9): Both covariance of channel response Kc and transition matrix X are unknown. In theory, X only depends on channels, not systems, phantoms, and reconstruction algorithms. Thus, we can calibrate X using known Kc of low dimension. X calibration is expressed as an optimization problem: where . We split Eq. (11) into two sub-problems to make it easier to get a reasonable solution: Since X is an underdetermined matrix, the optimization problem in Eq. (12) is locally convergent. Thus, we choose starting points as: with (·)† being the Moore-Penrose generalized inverse operation. Plugging the calibrated into Eq. (10), we can finally obtain the covariance estimation under arbitrary conditions: 3.EXPERIMENTSGabor function is used for CHO in this work: here the parameter configuration of Gabor channels is similar to that used in Leng’s work10, where the channel width ωs = 56.48,28.24,14.12,7.06, the channel frequency , the orientation , and the phase β = 0, π / 2. Phantoms for calibration and validation are generated from the Grand challenge dataset of Mayo clinic respectively. For calibration, images of size 128×128 are reconstructed by qGGMRF-WLS with system defined in Table 1. Since X is underdetermined, transition matrix X of various noise levels are averaged for stability. For validation, large-dimensional images are reconstructed by both TV-WLS and qGGMRF-WLS with three noise levels as is shown in Table 1. Covariance estimated by numerous image samples is used as reference: Table 1.Reconstruction parameters for calibration and validation of Prj2CR covariance estimation method.
with K being number of image samples. To better evaluate the performance of proposed Prj2CR method, we choose SNR of CHO as the figure of merit according to Eq. (8): where SNR is calculated from the covariance estimated by Prj2CR method, while SNR* is calculated from the covariance estimated by 290 image samples. Besides, we take the ground-truth as the image mean to minimize other influence factors, i.e., and . 4.RESULTSSNR performances are displayed in Figure 1. The SNR calculated by the proposed Prj2CR method is basically within the 95% confidence interval of the SNR* calculated by image samples. For qGGMRF-WLS reconstructions in case of 3×105 incident photons, the SNR by the Prj2CR method underestimates SNR* by image samples. 5.DISCUSSION AND CONCLUSIONWe proposed an analytical Prj2CR covariance method in this work, which can estimate the covariance of channel response directly from a single projection. The proposed method solves the problem of high-computational cost of Prj2Img method, and enables the covariance estimation of high-dimensional reconstructions. In this work, we introduce an invertible transition matrix to connect covariance of low-dimensional channel response and high-dimensional projection. Meanwhile, we calibrate the transition matrix to make the proposed method works for different systems, phantoms and reconstruction algorithms. The covariance of channel response estimated by Prj2CR method is comparable to that by 290 image samples. 6.ACKNOWLEGMENTSThis work is supported by National Natural Science Foundation of China (Grant No. 62031020 and 61771279). REFERENCESQi, J.,
“A unified noise analysis for iterative image estimation,”
Phys. Med. Biol, 48
(21), 3505
–3519
(2003). https://doi.org/10.1088/0031-9155/48/21/004 Google Scholar
Fessler, J. A.,
“Mean and variance of implicitly defined biased estimators (such as penalized maximum likelihood): Applications to tomography,”
in Biomed. Imaging V - Proc. 5th IEEE EMBS Int. Summer Sch. Biomed. Imaging, SSBI,
493
–506
(2002). Google Scholar
Li, Y.,
“Noise propagation for iterative penalized-likelihood image reconstruction based on Fisher information,”
Phys. Med. Biol, 56
(4), 1083
–1103
(2011). https://doi.org/10.1088/0031-9155/56/4/013 Google Scholar
Schmitt, S. M., Goodsitt, M. M. and Fessler, J. A.,
“Fast variance prediction for iteratively reconstructed ct images with locally quadratic regularization,”
IEEE Trans. Med. Imaging, 36
(1), 17
–26
(2017). https://doi.org/10.1109/TMI.2016.2593259 Google Scholar
Sánchez, A. A.,
“Estimation of noise properties for TV-regularized image reconstruction in computed tomography,”
Phys. Med. Biol, 60
(18), 7007
–7033
(2015). https://doi.org/10.1088/0031-9155/60/18/7007 Google Scholar
Zhang-O’Connor, Y. and Fessler, J. A.,
“Fast variance predictions for 3D cone-beam CT with quadratic regularization,”
Med. Imaging 2007 Phys. Med. Imaging, 6510 65105W
(2007). https://doi.org/10.1117/12.710312 Google Scholar
He, X. and Park, S.,
“Model observers in medical imaging research,”
Theranostics, 3
(10), 774
–786
(2013). https://doi.org/10.7150/thno.5138 Google Scholar
Guo, X., Zhang, L. and Xing, Y.,
“Analytical covariance estimation for iterative CT reconstruction methods,”
Biomed. Phys. Eng. Express, 8
(2022). https://doi.org/10.1088/2057-1976/ac58bf Google Scholar
Wunderlich, A. and Noo, F.,
“Image covariance and lesion detectability in direct fan-beam x-ray computed tomography,”
Phys. Med. Biol, 53
(10), 2471
–2493
(2008). https://doi.org/10.1088/0031-9155/53/10/002 Google Scholar
Leng, S., Yu, L., Zhang, Y., Carter, R., Toledano, A. Y. and McCollough, C. H.,
“Correlation between model observer and human observer performance in CT imaging when lesion location is uncertain,”
Med. Phys, 40
(8), 1
–9
(2013). https://doi.org/10.1118/1.4812430 Google Scholar
|