|
|
1.INTRODUCTIONThe Fluorescence Imaging Spectrometer (FLORIS) instrument, embarked on ESA’s FLEX mission, is designed to monitor the photosynthetic activity of the terrestrial vegetation layer by measuring the chlorophyll sun-induced fluorescence signal.1 FLORIS is a pushbroom hyperspectral imager, flying on a sun-synchronous polar orbit, which will measure the vegetation fluorescence in the spectral range between 500 nm and 780 nm at medium spatial sampling (300 m) and over a swath of 150 km. It accommodates an imaging spectrometer with a very high spectral resolution (0.3 nm), to measure the fluorescence spectrum within two oxygen absorption bands (O2A and O2B), and a second spectrometer with lower spectral resolution to derive additional atmospheric and vegetation parameters. Both spectrometers have CCDs in their focal planes.2, 3 As reported for a number of ESA missions using CCDs either in operation or currently in development (e.g., Sentinel-4,4 Sentinel-5, HST,5 XMM, Gaia,6–8 Euclid,9–11 PLATO12), radiation effects may significantly impact their scientific performance if not properly studied and accounted for in the data processing. Indeed high-energy particles impacting the detectors result in the creation of CCD silicon lattice defects. Such defects act as traps, which may capture an electron from the conducting band or a hole from the valence band and keep them for a characteristic time which depends on the trap nature. As a result, not all the charge accumulated in a pixel can be transferred during the CCD image transfer. This is quantified by the degradation of the Charge transfer Efficiency (CTE). CTE is the fraction of the charge that is effectively transferred from pixel to pixel. Signal carriers caught by traps are eventually released and in this way distributed to other pixels. This affects the radiometric performance as well as the image quality. The amplitude of this charge redistribution is determined by the parameters describing the traps created by the damaging protons (namely the trap density, the trap capture cross section and the trap release time constant which depends also on the detector operational temperature). Because the number of traps and therefore the amount of trapped charge is fixed, the relative performance degradation is higher at low signals such as, in the case of FLEX, the sun-induced fluorescence signals. For this reason a careful characterization of the CTE impact on FLORIS performance along its lifetime is of particular importance. In this paper we present the status of an on-going study which aims at assessing the impact of in-flight degradation of CTE on the FLORIS optical performance using different and complementary approaches, namely: (i) the experimental characterization of image quality using a proton-irradiated FLEX CCD, (ii) the simulation of the CTE impact on image quality using a CCD CTE model calibrated against FLEX test data. Eventually, the simulations and on-ground characterization will be used to define a strategy for monitoring the instrument image quality over the operational life of the instrument. 2.THE FLORIS INSTRUMENTAs illustrated on Figure 1, FLORIS is composed of two separate spectrometers, sharing a common imaging telescope. The first spectrometer, referred to as Low Resolution (LR) spectrometer images the scene spectrum from 500nm to 758nm with a spectral resolution lower than 2nm. The second spectrometer, the High Resolution (HR) spectrometer, images two spectral ranges from 677nm to 697nm (centered on the O2B absorption band) and from 740nm to 780nm (centered on the O2A absorption band), both with a spectral resolution lower than 0.3nm or 0.5nm, depending on the spectral bands.2 Three identical CCDs are used in FLORIS. One in the Low Resolution spectrometer focal plane and two in the High Resolution spectrometer focal plane. These detectors are based on the CCD325 from Te2V. The main characteristics of this detector are reported in Table 1, and their architecture described in Figure 2. Table 1.Te2V CCD325 main specifications
3.MEASUREMENT TEST DATA ON AN IRRADIATED CCDA flight representative CCD was irradiated and measurements of optical quality were subsequently performed. The irradiation campaign as well as the optical measurements are described in the following sections. 3.1Irradiation campaignA proton irradiation campaign has been performed on a dedicated model of the FLEX CCD. By adequately masking the sensor surface, four different doses were deposed on the four quadrants of the CCD. Dividing the detector into four quadrants is natural considering its architecture. As depicted on Figure 2, the image section C is transferred upwards to the store section D, while the image section B is transferred downwards to the store section A. The store section D is then read-out through the output nodes H (left) and G (right) of the serial register. Similarly the store section A is read-out in the output nodes E and F of the serial register. Each quadrant is identified by the letter of the corresponding image section, store section and output (e.g. the upper left quadrant identifier is CDH). The device was exposed to a 40-MeV proton beam, with doses per quadrant given in Table 2 Table 2.Proton doses deposed on CCD per quadrant
The quadrant BAE is shielded from the proton beam and kept as reference area. These doses are to be compared to the worst dose expected in-orbit on the FLEX sensor (6.0 ×109 p+/cm2). The CDG area can therefore be considered as representative of flight conditions, with a safe margin. 3.2Extended Pixel Edge Response measurementMeasurements of the image quality were performed at ESTEC on a dedicated test bench. Two different measurements are reported here. The first type consists in illuminating the sensor in flat field by means of an integrating sphere, and modify the CCD clocking to transfer additional pixels in order to read-out the charge delayed by CTE reduction. The CTE can be inferred from the integration of the charge measured in the over-scanned region, i.e. the deferred signal. This method is referred to as the Extended Pixel Edge Response and is illustrated on Figure 3. The signal levels of the flat illumination (1 ke–, 3 ke–, 7 ke–, 10 ke–, 20 ke–, 50 ke–, 100 ke–, 500 ke–, 1100 ke–) were chosen to cover the operational signal range of FLORIS. The data of the EPER measurement is used as calibration measurements for the extraction of the lattice trap parameters in the model, as described in section 4. 3.3USAF scene projection for measurement of Contrast Transfer FunctionA direct assessment of the image quality is performed through the measurement of the Contrast Transfer Function (CTF), on the test bench shown on Figure 4. CTF is a typical figure of merit of the image quality of optical systems that is measured as the contrast of a periodic square pattern on the detector plane. Targets such as the one presented on Figure 4 contains patterns of different spatial periods, allowing to sample the CTF curve and plot it against spatial frequency. The test is performed with a clocking sequence fully representative of the operational sequence and at the operational temperature of the detector (238 K). Four different patterns are chosen in order to sample the CTF curve up to the Nyquist frequency (i.e. 11.9 lp/mm for 42 μm pixels). The patterns are projected at the two edges of each detector image quadrant (close to the center of the detector, and close to the store area) in order to cover the highest and lowest, respectively, number of transfers through the image section. These positions correspond to worst and best cases, respectively, regarding the impact of the CTE on the image quality. Additionally the pattern are slightly tilted in order to reach a sub-pixel sampling of the pattern profiles. Finally the CTF is measured at three signal levels (1 ke–, 3 ke– and 10 ke–) and two detector temperatures (238 K, 248 K). The effect of CTE on the instrument optical performances can be assessed thanks to the Modulation Transfer Function (MTF). For spatial frequencies lower than the Nyquist frequency, the CTF is directly proportional to the MTF, which allows to relate the CTF measurements to the performance model. A relation between MTF and Charge Transfer Efficiency can be found in the literature:13 where Nt is the number of transfer, f is the spatial frequency and fnyq is the Nyquist frequency. This formula is inverted to extract the CTE from the CTF measurement. The detector MTF is then derived as MTFdet(f) = MTFpixel(f) × MTFCTE with MTFpixel(f) = sinc(πfp)sinc(πfx) representing the MTF response for a pixel size p (42 μm) and cross-talk x (18 μm), fitting the detector MTF measured on-ground. The results of the measurements are presented on Figure 5 and compared to the CTF model used for instrument performance analyses. One of the key performance to monitor in order to assess the severity of the CTE effect is the spatial resolution, through the full width at half maximum (FWHM) of the point spread function. Figure 6 shows the point spread function at the end of the operational life, including the effect of CTE measured in the CDG quadrant as described above. This figure illustrates that the FWHM of the PSF remains within the range required for FLEX (100 μm in the focal plane, equivalent to roughly 360 meters on ground, i.e. 1.2 times the spatial sampling distance (SSD)). 4.CTE MODELLING AND SIMULATIONS OF FLEX-LIKE IMAGESUsing the Pyxel framework14 - a python-based detection simulation tool - we first calibrate CDM (Charge Distortion Model)15 - a commonly-used CTE model - against laboratory test data acquired under FLEX representative conditions. Once calibrated we use this model to degrade images and measure the impact of CTE on image quality. Ultimately this model can be used to generate synthetic FLEX-representative spectra including the effect of degraded CTE, which can be useful in assessing the impact of CTE in the final scientific measurements and devising an in orbit monitoring and mitigation strategy. In the following we present first the calibration of CDM - methodology and results-, and then the simulation of degraded images to measure the impact of CTE on the detector CTF. The Pyxel jupyter notebooks used for the CTE calibration and the simulation of CTE effects on the FLEX CTF are openly available online in the public repository Pyxel Data.16 4.1Calibration of the CTE modelCDM is a fast analytical model of CTE in CCDs, it computes at once the effects on a given image of the charge trapping and de-trapping of traps during parallel and/or serial transfer. CDM can simulate the effect of an arbitrary number of trap species parametrised through their trap density, capture cross section, and release time constant. CDM makes use of an extra parameter β in order to link charge packet volume and charge density for a given signal level. CDM has not been designed for accuracy and realism but for flexibility and speed; it was developed in the context of Gaia data processing with the aim of calibrating out CTE effects. It is capable of reproducing fairly well test data (from EPER trails to spectra, point sources and extended objects) and give a good idea of the type of trap species in presence and can be used to investigate the possible CTE effects on images once calibrated over a representative and large set of test data. We do so by using the so-called calibration running mode of Pyxel. This mode makes use of evolutionary algorithms implemented in the library PyGMO,17 a Python library for massively parallel optimization. The two main inputs of the calibration mode in Pyxel are: initial data to which the models are applied during the pipeline and target data which serves as a benchmark for computing the goodness of fits with given parameter sets, in PyGMO nomenclature: computing the fitness of individuals. Here, the target data is derived from the laboratory test measurement presented in Section 3: to speed up the simulation process, we do not simulate an entire image but instead simulate the transfers over a unique CCD column for this we thus average the flat field and overscan in the serial direction over a given irradiated region to obtain a representative average parallel charge profile. We used the data acquired at T = 238 K, for 6 different signal levels: 1ke−, 3ke−, 7ke−, 10ke−, 20ke− and 100ke−, and over the irradiated region CDH, the quadrant with the highest proton dose, this is 1.9 × 1010 p+/cm2 (see Table 2). The degraded CTE can be seen in the form of charge trails, as shown in figure 7. The simulation input charge profile is generated from the same test data, for which only the flat field part of the image is kept. The free parameters during calibration are: the electron cloud expansion coefficient β, the trap release times τr and the trap densities ntr, all for the parallel direction. We choose the number of 4 different trap species; this translates into a total of 9 free parameters. The exact same calibration procedure was repeated with 3 and 5 trap species; for 3 traps species the achieved fitness is poorer, while using 5 traps species does not improve the solution achieved with 4. Because of the large number of free parameters, we run the simulation on a cluster of computers, which enables us to probe large number of potential parameter sets (aka individuals), while still keeping the total time of simulation rather short (< 24 hours). We use a total number of 40 workers with 80 threads to evolve 20000 individuals for 2500 generations. The presented calibration method ultimately leads to the set of CDM parameters presented in Table 3. The associated quality of fit can be seen in Fig. 7 and Fig. 8. And the representativity of this calibrated set of parameter against other best fit sets is shown in Fig. 9 giving us confidence that the found set of parameter is close to a global minimum. For this temperature, the release time constant found, corresponds rather well to the E centre (Phosphorous Vacancy complex) for the longest release time constant and the divacancy complex for the intermediate constants. The shortest show a very high trap density which seems unrealistic, it also does not immediately seem to correspond to a trap species found in the literature. This result for the shortest release time constant could either point at one of CDM’s shortcoming - namely the fact that CDM does not simulate properly trapping at time scale shorter than a transfer period - or the fact that the test data contains a sharp edge response which is not due to CTE but some electronics effect (this could be verified looking at the trails in the non-irradiated region of the CCD). The use of this very short release time constant in simulating CTE effects on FLEX images has thus to be done with caution. Table 3.Optimal set of parameters resulting from the described calibration procedure and the possible corresponding trap species (see4 for a more complete overview and discussion of trap species).
4.2Generation of synthetic CTF measurements and comparison with test data measurementWe use the calibrated CDM model to generate synthetic CTF curves for comparison with the laboratory measurements done on USAF targets (unbinned case), described in section 3.3. We achieve this by applying the model to images of periodic square patterns and observing the change of contrast as a function of both spatial frequency and signal level. As in laboratory measurements, we do so for two different regions: BAF and CDG, and for three different signal levels: 1ke−, 3ke− and 10ke−. For this purpose the parametric mode of Pyxel can be utilized, which allows us to run pipelines and apply models for many different sets of input model parameters. We create and include in Pyxel a simple illumination model of periodic square patterns, with signal level and spatial frequency as parameters. The patterns are tilted at a small angle as in section 3.3, again to ensure sub-pixel sampling of the pattern profiles. The calibration of CDM model is done on the quadrant with the highest proton dose (CDH) and laboratory measurements are done on the quadrants with lower proton dose (CDG and BAF), thus we adjust the trap densities accordingly to match the predicted trap density in the CDG and BAF regions. Because of unreliability of the fastest trap result, explained in section 4.1, the effect of the fastest trap is omitted from the model. As is expected and seen in the laboratory measurement data, the non-irradiated detector does not show ideal contrast transfer over all spatial frequencies. We approximate the contrast degradation of a non-irradiated detector by convolving the patterns with a simple Gaussian point spread function with σ = 0.28 px. This way we match the values of simulated and measured CTF curves at the highest spatial frequency and for the case of non-irradiated detector at the value of 0.85. The combined effect of CDM and convolution with a narrow Gaussian on tilted periodic patterns can be seen in figure 10. From the patterns we compute points for the CTF curves as (Imax – Imin)/(Imax + Imin), where I is the intensity in the region of interest. The result of the simulated CTF curves versus the laboratory measurements is shown in figure 11, together with the data for the non-irradiated case. There is a clear discrepancy between CTF simulations and measurements; the laboratory measurements showing much better contrast response. Comparing the two it can be seen that in the case of simulated CTF curves there is first a quick decay of contrast even at small spatial frequencies and second a large dispersion of curves for different signal levels. Whether the mismatch is a result of a difference in CTF derivation procedure or simply a result of a parameter set that has not reached the global minimum is yet to be researched. It is already known that the CTF measurements have been taken at a temperature 6 K lower than the target data used to calibrate CDM, and that the clocking scheme used during the test campaigns may differ too greatly from the simplifications used in the simulations. Based on the findings, the model and CTF calculation will have to be refined in order to validate and match the simulations with the measurements. This way it could be used in the future for generation of synthetic FLEX-representative spectra with the effects of CTE degradation. 5.CONCLUSIONIn the previous sections we have presented the results of measurements and modeling of the effect on the FLEX CCD of Charge Transfer Inefficiency induced by proton damage in the crystalline lattice of the detector. A flight representative device has been irradiated for this purpose and optical measurements were then performed. The measurement results presented above have been used for two purposes: first to have a direct evaluation of the degradation of FLORIS optical performance through the measurement of the Contrast Transfer Function of the CCD. Second, to generate calibration input data to feed into the detector model (CDM). This calibration data is used to derive the trap species parameters, by fitting it thanks to an evolutionary algorithm. After the calibration step, the model is then used on synthetic scenes similar to the measurement scenes, and compute the CTF from these synthetic scenes. The outputs of the model are compared to the test results and show discrepancies which are so far unexplained. The next phase of this work will be dedicated to understand and solve this mismatch, aiming at eventually obtain an accurate modelling framework for assessing the optical quality of a detector affected by Charge Transfer Inefficiency. REFERENCESDrusch, M., Moreno, J., Del Bello, U., Franco, R., Goulas, Y., Huth, A., Kraft, S., Middleton, E. M., Miglietta, F., Mohammed, G., Nedbal, L., Rascher, U., Schüttemeyer, D., and Verhoef, W.,
“The fluorescence explorer mission concept—esa’s earth explorer 8,”
IEEE Transactions on Geoscience and Remote Sensing, 55
(3), 1273
–1284
(2017). https://doi.org/10.1109/TGRS.2016.2621820 Google Scholar
Coppo, P., Taiti, A., Pettinato, L., Francois, M., Taccola, M., and Drusch, M.,
“Fluorescence imaging spectrometer (floris) for esa flex mission,”
Remote Sensing, 9
(7),
(2017). https://doi.org/10.3390/rs9070649 Google Scholar
Coppo, P., Pettinato, L., Nuzzi, D., Fossati, E., Taiti, A., Gabrieli, R., Campa, A., Taccola, M., Bézy, J.-L., Francois, M., Erdmann, L. H., Triebel, P., and Lehr, D.,
“Instrument predevelopment activities for FLEX mission,”
Optical Engineering, 58
(7), 1
–25
(2019). https://doi.org/10.1117/1.OE.58.7.075102 Google Scholar
Prod’homme, T., Belloir, J. M., Weber, H., Bazalgette Courrèges-Lacoste, G., Meynart, R., Nowicki-Bringuier, Y. R., Caron, J., Levillain, Y., Woffinden, C., Lord, B., and Mackie, R.,
“Radiation-induced charge transfer inefficiency in charge-coupled devices: Sentinel-4 CCD pre-development as a case study,”
in Society of Photo-Optical Instrumentation Engineers (SPIE) Conference Series,
(2014). Google Scholar
Massey, R., Schrabback, T., Cordes, O., Marggraf, O., Israel, H., Miller, L., Hall, D., Cropper, M., Prod’homme, T., and Niemi, S.-M.,
“An improved model of charge transfer inefficiency and correction algorithm for the Hubble Space Telescope,”
439 887
–907
(2014). Google Scholar
Prod’homme, T., Holl, B., Lindegren, L., and Brown, A. G. A.,
“The impact of CCD radiation damage on Gaia astrometry - I. Image location estimation in the presence of radiation damage,”
419 2995
–3017
(2012). Google Scholar
Holl, B., Prod’homme, T., Lindegren, L., and Brown, A. G. A.,
“The impact of CCD radiation damage on Gaia astrometry - II. Effect of image location errors on the astrometric solution,”
422 2786
–2807
(2012). Google Scholar
Crowley, C., Kohley, R., Hambly, N. C., Davidson, M., Abreu, A., van Leeuwen, F., Fabricius, C., Seabroke, G., de Bruijne, J. H. J., Short, A., Lindegren, L., Brown, A. G. A., Sarri, G., Gare, P., Prusti, T., Prod’homme, T., Mora, A., Martín-Fleitas, J., Raison, F., Lammers, U., O’Mullane, W., and Jansen, F.,
“Gaia Data Release 1. On-orbit performance of the Gaia CCDs at L2,”
595
(A6),
(2016). Google Scholar
Israel, H., Massey, R., Prod’homme, T., Cropper, M., Cordes, O., Gow, J., Kohley, R., Marggraf, O., Niemi, S., Rhodes, J., Short, A., and Verhoeve, P.,
“How well can charge transfer inefficiency be corrected? A parameter sensitivity study for iterative correction,”
453 561
–580
(2015). Google Scholar
Prod’homme, T., Verhoeve, P., Oosterbroek, T., Boudin, N., Short, A., and Kohley, R.,
“Laboratory simulation of Euclid-like sky images to study the impact of CCD radiation damage on weak gravitational lensing,”
in Society of Photo-Optical Instrumentation Engineers (SPIE) Conference Series,
(2014). Google Scholar
Prod’homme, T., Liebing, P., Verhoeve, P., Menyaylov, I., Lemmel, F., Smit, H., Blommaert, S., Breeveld, D., and Shortt, B.,
“A smartphone-based arbitrary scene projector for detector testing and instrument performance evaluation,”
in Society of Photo-Optical Instrumentation Engineers (SPIE) Conference Series,
(2020). Google Scholar
Prod’homme, T., Verhoeve, P., Lemmel, F., Smit, H., Blommaert, S., van der Luijt, C., Visser, I., Beaufort, T., Levillain, Y., and Shortt, B.,
“Comparative Study of Cryogenic Versus Room-Temperature Proton Irradiation of N-Channel CCDs and Subsequent Annealing,”
IEEE Transactions on Nuclear Science, 66 134
–139
(2019). https://doi.org/10.1109/TNS.2018.2886029 Google Scholar
Janesick, J. R., Scientific Charge-Coupled Devices, SPIE Press, Bellingham, Washington
(2001). https://doi.org/10.1117/3.374903 Google Scholar
Prod’homme, T., Lemmel, F., Arko, M., Serra, B., George, E., Biancalani, E., Smit, H., and Lucsanyi, D.,
“Pyxel: the collaborative detection simulation framework,”
in Society of Photo-Optical Instrumentation Engineers (SPIE) Conference Series,
(2020). Google Scholar
Short, A., Crowley, C., de Bruijne, J. H. J., and Prod’homme, T.,
“An analytical model of radiation-induced Charge Transfer Inefficiency for CCD detectors,”
Monthly Notices of the Royal Astronomical Society, 430
(4), 3078
–3085
(2013). https://doi.org/10.1093/mnras/stt114 Google Scholar
“Pyxel Data gitlab repository,”
https://gitlab.com/esa/pyxel-data Google Scholar
Biscani, F. and Izzo, D.,
“A parallel global multiobjective framework for optimization: pagmo,”
Journal of Open Source Software, 5
(53), 2338
(2020). https://doi.org/10.21105/joss Google Scholar
|