Paper
4 January 2021 Modelling the point source function for collimators employed in radio guided surgery and SPECT
Author Affiliations +
Proceedings Volume 11605, Thirteenth International Conference on Machine Vision; 116050A (2021) https://doi.org/10.1117/12.2587441
Event: Thirteenth International Conference on Machine Vision, 2020, Rome, Italy
Abstract
For the multi-channel collimators employed in SPECT measurements and in gamma probes (for radio guided surgery) an approach to modeling the functioning of collimator is developed. This approach involves the calculation of point source function and the sensitivity field for the considering device. The effect of finite absorption in collimator material is taken into account. The results of modelling for confocal collimator are presented. The calculations were performed for the attenuation length corresponding to the device 3D-printed from metal and in the assumption of infinite absorption. It makes possible to estimate quantitatively the effect of gamma photons partial propagation through the matrix of collimator on the sensitivity and resolution for the specified material and the geometry of the device.
© (2021) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yana L. Shabelnikova, Sergey I. Zaitsev, and Vasiliy B. Obolenskiy "Modelling the point source function for collimators employed in radio guided surgery and SPECT", Proc. SPIE 11605, Thirteenth International Conference on Machine Vision, 116050A (4 January 2021); https://doi.org/10.1117/12.2587441
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
Back to Top