We present a test platform for the Athena X-IFU detection chain, shared between IRAP and CNES. This test bench, housed in a commercial two-stage ADR cryostat provided by Entropy GmbH, will serve as the first demonstration of the representative end-to-end readout chain for the X-IFU, using prototypes of the future flight electronics and currently available subsystems. The focal plane array (FPA), placed at the 50 mK cold stage of the ADR, includes a 1024-pixel array of transition-edge sensor (TES) microcalorimeter spectrometers provided by NASA/GSFC, superconducting amplifiers (SQUIDs) from VTT, as well as superconducting readout electronics for frequency domain multiplexing (FDM), provided by SRON. The detection chain then continues with the prototype room temperature electronics for the X-IFU: the Warm Front-End Electronics (WFEE, provided by APC) and the Digital Readout Electronics (DRE, provided by IRAP). The test bench yields critical feedback on current subsystem designs and electronic interfaces, and in the future will also be used for refining the X-IFU calibration plan as well as laboratory astrophysics experiments relevant to future X-IFU science. In this presentation, we describe the characterization of the cryostat, various design trades for the FPA and readout chain, and recent results from our current setup.
|