Paper
3 April 1989 A Simple And Accurate Method Of Calibration Of Double Refracting Microinterferometers Using Object-Adapted Interferometry
Maksymilian Pluta
Author Affiliations +
Proceedings Volume 1121, Interferometry '89; (1989) https://doi.org/10.1117/12.961296
Event: Interferometry '89, 1989, Warsaw, Poland
Abstract
Determination of the relationship between the light wavelength and the interfringe spacing is a basic calibration operation in interferometry. A simple and accurate method of the calibration of double refracting microinterferometers is proposed. The method is based on the object-adapted variable wavelength interferometry (AVAWI) developed recently by the author of this contribution. Only a birefringent plate of known thickness or a glass plate with a step whose height is known are required for the calibration process.
© (1989) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Maksymilian Pluta "A Simple And Accurate Method Of Calibration Of Double Refracting Microinterferometers Using Object-Adapted Interferometry", Proc. SPIE 1121, Interferometry '89, (3 April 1989); https://doi.org/10.1117/12.961296