5 November 2018 Method of object MRTD-testing for thermal infrared imager
Author Affiliations +
Abstract
Summarizing main methods of MRTD-testing for thermal infrared imager briefly. For the requirements of MRTD real-time measurement in engineering, the mature objective MRTD method with less manpower and less measuring time is adopted. While ensuring measurement accuracy, optimizing the measurement steps、 improve the method of eigenvector abstracted and significantly reduce the measuring time, the improved method is more suitable for engineering measurement. Compared the subjective measurement method with the improved objective measurement method, the obtained measurement results are well consistent. and the factors of influence on accuracy of objective measurement were discussed. The given method provides an effective means for MRTD engineering measurement.
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Liwei Xu, Qiang Li, and Yingming Lu "Method of object MRTD-testing for thermal infrared imager", Proc. SPIE 10815, Optical Design and Testing VIII, 108151F (5 November 2018); https://doi.org/10.1117/12.2502131
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Minimum resolvable temperature difference

Infrared imaging

Infrared radiation

Thermography

Imaging systems

Neural networks

Temperature metrology