27 June 1988 X-Ray Anode Surface Temperatures: The Effect Of Volume Heating
Stephen Whitaker
Author Affiliations +
Abstract
An analytical solution for the focal spot temperature is developed for the case of an exponential decay of the electron flux in the anode. The influence of volume heating on the surface temperature is dominated by the parameter (w/δ)2 where w is the focal spot width and δ is a mean depth of penetration. For typical operating conditions, the influence of volume heating becomes important when the focal spot width is less than 500 microns.
© (1988) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Stephen Whitaker "X-Ray Anode Surface Temperatures: The Effect Of Volume Heating", Proc. SPIE 0914, Medical Imaging II, (27 June 1988); https://doi.org/10.1117/12.968682