20 December 1985 Dependence Of Calibration Curves Obtained From Gram-Schmidt Chromatograms From Experimental Parameters In CGC/FTIRS
Hans Malissa Jr., Kurt Martin, Karl Winsauer
Author Affiliations +
Proceedings Volume 0553, Fourier and Computerized Infrared Spectroscopy; (1985) https://doi.org/10.1117/12.970844
Event: 1985 International Conference on Fourier and Computerized Infrared Spectroscopy, 1985, Ottawa, Canada
Abstract
The calibration curve for quantitative evaluation of Gram-Schmidt chromatograms (GSC) show a non-linearity in the low concentration region. Systematic experiments revealed, that variation in the baseline height in the GSC is responsible for a shift in the calibration curve. Summarized results are: (1) small changes of baseline introduce relative large errors; (2) this non-linearity must also be considered, when small peaks are measured which are not completely separated from a major one.