Paper
1 February 1985 Flash X-Ray Apparatus With Spectrum Control Functions For Medical Use And Fuji Computed Radiography
H. Isobe, E. Sato, Y. Hayasi, M. Suzuki, H. Arima, F. Hoshino
Author Affiliations +
Proceedings Volume 0491, 16th Intl Congress on High Speed Photography and Photonics; (1985) https://doi.org/10.1117/12.967906
Event: 16th International Congress on High Speed Photography and Photonics, 1984, Strasbourg, France
Abstract
Flash radiographic bio-medical studies at sub-microsecond intervals were performed by using both a new type of flash X-ray(FX) apparatus with spectrum control functions and Fuji Computed Radiography(FCR). This single flasher tends to have a comparatively long exposure time and the electric pulse width of the FX wave form is about 0.3,usec. The maximum FX dose is about 50mR at 1m per pulse, and the effective focal spot varies according to condenser charging voltage, A-C distance, etc., ranging from 1.0 to 3.0mm in diameter, but in the low dose rate region it can be reduced to less than 1.0mm in diameter. The FX dose is determined by the condenser charging voltage and the A-C distance, while the FX spectrum is determined by the average voltage of the FX tube and filters. Various clear FX images were obtained by controlling the spectrum and dose. FCR is a new storage medium for medical radiography developed by the Fuji Photo Film Co., Ltd. and this apparatus has various image forming functions: low dose radiography, film density control, image contrast control, subtraction management and others. We have used this new apparatus in conjunction with our FX radiography and have obtained some new and interesting biomedical radiograms: the edge enhancement image, the instantaneous enlarged image, and the single exposure energy subtraction image using the FX spectrum distribution.
© (1985) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
H. Isobe, E. Sato, Y. Hayasi, M. Suzuki, H. Arima, and F. Hoshino "Flash X-Ray Apparatus With Spectrum Control Functions For Medical Use And Fuji Computed Radiography", Proc. SPIE 0491, 16th Intl Congress on High Speed Photography and Photonics, (1 February 1985); https://doi.org/10.1117/12.967906
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Radiography

Aluminum

X-rays

Virtual colonoscopy

High speed photography

Image processing

Electrodes

RELATED CONTENT

Biomedical Flash Radiography
Proceedings of SPIE (February 04 1988)
Effects of x ray spectrum on the NPS of a...
Proceedings of SPIE (April 25 2000)
High Intensity Soft Flash X Ray Generator with Two Types...
Proceedings of SPIE (December 08 1986)
A Flash Radiographic System For Biomedical Use
Proceedings of SPIE (June 07 1989)

Back to Top