19 February 2010 Non-harmonic analysis for high-resolution optical coherence tomography
Changho Chong, Xu Cao, Atsushi Morosawa, Koki Totsuka, Takuya Suzuki, Shigeki Hirobayashi
Author Affiliations +
Abstract
A novel processing technique called Non-Harmonic Analysis (NHA) is proposed for high-resolution OCT imaging. Conventional Fourier-Domain OCT relies on the FFT calculation which depends on the window function and length. NHA can resolve high frequency without being influenced by window function or frame length of sampled data. The results show that NHA process realizes practical image resolution equivalent to 100nm swept range by using significantly reduced wavelength range, and also implies the potential of high resolution imaging capability without the need of a broadband source.
© (2010) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Changho Chong, Xu Cao, Atsushi Morosawa, Koki Totsuka, Takuya Suzuki, and Shigeki Hirobayashi "Non-harmonic analysis for high-resolution optical coherence tomography", Proc. SPIE 7554, Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XIV, 75541V (19 February 2010); https://doi.org/10.1117/12.842886
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical coherence tomography

Image processing

Image resolution

Image segmentation

Skin

Data acquisition

Signal analyzers

Back to Top