6 December 2006 Frequency characteristics of images measured on the VolumeScope scanning device
V. N. Vasiliev, M. Yu. Zaytsev
Author Affiliations +
Proceedings Volume 5943, X-ray and Neutron Capillary Optics II; 59430V (2006) https://doi.org/10.1117/12.637992
Event: X-ray and Neutron Capillary Optics II, 2004, Zvenigorod, Russian Federation
Abstract
The modulation transfer function was determined on the VolumeScope, a prototype X-ray Compton scatter tomographic scanner. The linear spread function of the device was measured independently along each coordinate axis with a 0.3 mm aluminum foil slightly angled with respect to the voxel grid. A technique of the presampled MTF generation was used to increase a sampling frequency of the data and reduce a statistical noise. The LSFs and MTFs along three coordinate axes are presented. The FWHMs of the LSF peaks are 2.50, 2.3 1 and 3.72 mm along the axes X, Y and Z respectively. The MTF cutoff frequencies are 0.45,0.50 and 0.35 cycle/mmrespectively.
© (2006) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
V. N. Vasiliev and M. Yu. Zaytsev "Frequency characteristics of images measured on the VolumeScope scanning device", Proc. SPIE 5943, X-ray and Neutron Capillary Optics II, 59430V (6 December 2006); https://doi.org/10.1117/12.637992