28 May 1999 Image quality evaluation of a direct digital radiography detector operating in a UK radiology department
Stephen M. Kengyelics, Arnold R. Cowen, Andrew G. Davies
Author Affiliations +
Abstract
This paper presents results of an image quality evaluation performed on a prototype DirectRay system (Sterling Diagnostic Imaging, Direct Radiography Corp.) operating in a UK radiology department. Physical imaging characteristics measured included characteristic (sensitometric) response, square-wave response function, Wiener spectrum and detective quantum efficiency (DQE). In addition, subjective image quality was assessed using a calibrated threshold contrast detail detectability (TCDD) test object. Results produced by the DirectRay system were compared with those achieved by a modern photo-stimulable computed radiography (CR) system. DirectRay and CR systems exhibit similarly wide exposure dynamic ranges, but in practice exploit this property in differing ways. Measured values of DQE(0) for the DirectRay system were found to be higher compared to those obtained for CR, for the majority of the exposure range examined. These results were confirmed subjectively by TCDD test-object analysis. No evidence of a 'memory artifact' phenomenon was found.
© (1999) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Stephen M. Kengyelics, Arnold R. Cowen, and Andrew G. Davies "Image quality evaluation of a direct digital radiography detector operating in a UK radiology department", Proc. SPIE 3659, Medical Imaging 1999: Physics of Medical Imaging, (28 May 1999); https://doi.org/10.1117/12.349494
Lens.org Logo
CITATIONS
Cited by 8 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

Chromium

X-ray imaging

Imaging systems

Radiography

Sensors

Image quality

Back to Top