Presentation + Paper
4 April 2022 X-ray photon-counting using silicon photomultiplier-based scintillation detectors at high x-ray tube currents
Stefan van der Sar, Stefan Brunner, Dennis Schaart
Author Affiliations +
Abstract
We investigate X-ray photon-counting scintillation detectors with silicon photomultiplier (SiPM) readout. These circumvent some drawbacks of direct-conversion detectors. We measured observed count rate (OCR) versus X-ray tube current for single-pixel detectors consisting of LYSO:Ce and YAP:Ce scintillators coupled to ultrafast SiPMs. For a 30 keV threshold, the maximum OCRs equal 4.5 Mcps/pixel (LYSO:Ce) and 5.5 Mcps/pixel (YAP:Ce) for paralyzable-like counting and 10 Mcps/pixel (LYSO:Ce) and 12.5 Mcps/pixel (YAP:Ce) for nonparalyzable-like counting. We estimate that the twice as fast LaBr3:Ce scintillator yields OCRs approaching those of CdTe/CZT-based photon-counting CT detectors. We also show energy response data and discuss dose-efficient pixel miniaturization.
Conference Presentation
© (2022) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Stefan van der Sar, Stefan Brunner, and Dennis Schaart "X-ray photon-counting using silicon photomultiplier-based scintillation detectors at high x-ray tube currents", Proc. SPIE 12031, Medical Imaging 2022: Physics of Medical Imaging, 120310I (4 April 2022); https://doi.org/10.1117/12.2611365