Paper
16 March 2020 Assessment of therapy applicator targeting with a mechanically assisted 3D ultrasound system for minimally invasive focal liver tumor therapy
Derek J. Gillies, Jeffrey Bax, Kevin Barker, Lori Gardi, David Tessier, Nirmal Kakani, Aaron Fenster
Author Affiliations +
Abstract
Minimally invasive focal ablation of liver cancer is an alternative technique to conventional methods for early-stage tumors. Sufficient therapy is provided when ablation applicators are placed at their intended target locations, but current practices are occasionally unable to achieve the required degree of accuracy, as observed by local cancer recurrence rates. We have developed a mechanically assisted 3D ultrasound (US) imaging and guidance system capable of providing geometrically variable images to increase intraoperative spatial information and propose a new method for placing therapeutic applicators during focal liver tumor ablations. A three-motor mechanical mover was designed to provide linear, tilt, and combined hybrid geometries for user-defined 3D US fields-of-view. This mover can manipulate any clinically available 2D US transducer mounted in a transducer-specific 3D-printed holders and is held by a counterbalanced mechanical guidance system, which contains electromagnetic brakes and encoders to track the position of the transducer. Fabrication of a transducer-specific needle guide allowed for information from 3D US images and targets to be overlaid with live 2D US to perform an image-guided workflow. End-to-end testing from 3D US acquisition to needle insertion was performed with a mock phantom procedure to assess overall needle placement accuracy and a potential clinical workflow. Mean applicator placement error was 3.8 ± 1.9 mm for all trials and demonstrated that our 3D US image-guided system may be a feasible approach for guiding ablation applicators accurately during focal liver tumor ablation procedures.
© (2020) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Derek J. Gillies, Jeffrey Bax, Kevin Barker, Lori Gardi, David Tessier, Nirmal Kakani, and Aaron Fenster "Assessment of therapy applicator targeting with a mechanically assisted 3D ultrasound system for minimally invasive focal liver tumor therapy", Proc. SPIE 11315, Medical Imaging 2020: Image-Guided Procedures, Robotic Interventions, and Modeling, 113152K (16 March 2020); https://doi.org/10.1117/12.2549841
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
3D acquisition

3D image processing

Liver

Tumors

Error analysis

Target acquisition

Scanners

Back to Top